Page 121

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 110 7.8.2 Private Templates  
LOGIQ 7 supports the following private templates for SOP Instances created by this 
product.    
 
xxvii. Private TID GEMS_US_0100 - GE Ultrasound M&A 
 
NL Relation with 
Parent Value Type Concept Name VM Req 
Type Condition Value Set Constraint 1   CONTAINER EV (43126, 
99GEMS,”Ultrasound 
M&A Document”) 1 M    > CONTAINS UIDREF EV (43140, 
99GEMS,”Series 
Instance UID”) 1 U    > CONTAINS...

Page 122

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 111 8. BASIC DIRECTORY INFORMATION OBJECT 
IMPLEMENTATION 
8.1 INTRODUCTION 
This section specifies the use of the DICOM Basic Directory IOD to represent the 
information included in directories produced by this implementation. Corresponding 
attributes are conveyed using the module construct. The contents of this section are: 
8.2 - IOD Implementation 
8.3 - IOD Entity-Relationship Model 
8.4- IOD Module Table 
8.5 - IOD...

Page 123

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 112 ILLUSTRATION 8.3-1  
BASIC DIRECTORY ENTITY RELATIONSHIP DIAGRAM 
 Root Directory Entity Patient DR Includes 1 
0 - n * Study DR * 1 - n Series DR * 1 - n Image DR  1 - n 1  
1  1  
 
8.4 IOD MODULE TABLE 
Within an entity of the Basic Directory IOD, attributes are grouped into related set of 
attributes. A set of related attributes is termed a module. A module facilitates the 
understanding of the semantics...

Page 124

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 113 The Directory Information Module is created when initializing the media. If it already 
exists, the existing information is not changed regarding patient, study, series or image 
data.  
An existing Directory Information Module may have been obtained from application 
entities using removable media. These instances are external to this conformance claim 
and the origin of the SOP instances is outside the scope of this...

Page 125

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 114 Attribute Name Tag Type Attribute Description >Offset of Referenced Lower-Level Directory 
Entity (0004,1420) 1C Is set >Directory Record Type (0004,1430) 1C The values support by FSC and 
FSU are 
 PATIENT 
 STUDY  
 SERIES  
 IMAGE 
 >Private Record UID (0004,1432) 1C Not used >Referenced File ID  (0004,1500) 1C Is set if Directory Record Type is 
IMAGE  
Contains the file path consisting 
of 5 elements: 
1....

Page 126

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 115 Key Tag Type Attribute Description >Referenced SOP Class UID (0008,1150) 1C Is filled in by FSC and FSU as in 
chapter 4, 5, and 6. >Referenced SOP Instance UID (0008,1155) 1C Is filled in by FSC and FSU as in 
chapter 4, 5, and 6. Patients Birth Time (0010,0032) 3 Is filled in by FSC and FSU as in 
chapter 4, 5, and 6. Other Patient Ids (0010,1000) 3 Is filled in by FSC and FSU as in 
chapter 4, 5, and 6. Other...

Page 127

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 116 Key Tag Type Attribute Description >Referenced SOP Instance UID (0008,1155) 1C Is filled in by FSC and FSU as in 
chapter 4, 5, and 6. Admitting Diagnoses Description (0008,1080) 3 Is filled in by FSC and FSU as in 
chapter 4, 5, and 6. Patients Age (0010,1010) 3 Is filled in by FSC and FSU as in 
chapter 4, 5, and 6. Patients Size (0010,1020) 3 Is filled in by FSC and FSU as in 
chapter 4, 5, and 6.. Patients Weight...

Page 128

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 117 Key Tag Type Attribute Description Operator’s Name (0008,1070) 3 Is filled in by FSC and FSU as in 
chapter 4, 5, and 6. Referenced Performed Procedure Step Sequence (0008,1111) 3 Is filled in by FSC and FSU as in 
chapter 4, 5, and 6. >Referenced SOP Class UID (0008,1150) 1C Is filled in by FSC and FSU as in 
chapter 4, 5, and 6.. >Referenced SOP Instance UID (0008,1155) 1C Is filled in by FSC and FSU as in 
chapter...

Page 129

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 118 Key Tag Type Attribute Description Manufacturer (0008,0070) 2 Is filled in by FSC and FSU as in 
chapter 4, 5, and 6. Institution Name (0008,0080) 3 Is filled in by FSC and FSU as in 
chapter 4, 5, and 6.. Station Name (0008,1010) 3 Is filled in by FSC and FSU as in 
chapter 4, 5, and 6.. Institutional Department Name (0008,1040) 3 Is filled in by FSC and FSU as in 
chapter 4, 5, and 6. Manufacturers Model Name...

Page 130

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 119 8.5.3.4.1 Private Directory Record Definition 
Not used.8.5.3.5 Multi-Referenced File Directory Record Definition 
Not used. 
8.6 PRIVATE DATA DICTIONARY 
If so configured, the product will send ultrasound raw data information in private data 
elements designated by the Private Creator element: 
Element Name  Tag VR VM Description Private Creator 7FE1,00xx LO 1 
GEMS_Ultrasound_MovieGroup_001  
This means that all...