GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 130 Attribute Name Tag Req. Type 
N-CREATE   Req. Type 
N-SET   >>Referenced SOP Instance UID (0008,1155) 1C, supported Not allowed Performed Procedure Step Information Performed Procedure Step ID (0040,0253) 1 Not allowed Performed Station AE Title (0040,0241) 1 Not allowed Performed Station Name (0040,0242) 2, supported Not allowed Performed Location (0040,0243) 2, supported 
(Institution 
Name, 
truncated if 
necessary...

Page 142

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 131 Attribute Name Tag Req. Type 
N-CREATE   Req. Type 
N-SET   >Series Instance UID (0020,000E) 1C 
(Required if 
Sequence Item 
is present) 1C 
(Required if Sequence Item 
is present) >Series Description (0008,103E) 2C 
(Required if 
Sequence Item 
is present) 2C 
(Required if Sequence Item 
is present) >Retrieve AE Title (0008,0054) 2C 
(Required if 
Sequence Item 
is present) 2C 
(Required if Sequence Item 
is...

Page 143

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 132 LOGIQ 7 sends N-SET after the exam is ended. The N-SET will include all acquired 
images’ UIDs and the status of COMPLETED or DISCONTINUED. 
10.2.2.2 Status Codes 
No Service Class specific status values are defined for the N-ACTION Service.  See PS 
3.7 for general response status codes.  

Page 144

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 133 11. STORAGE COMMITMENT PUSH MODEL SOP CLASS 
DEFINITION 
11.1 INTRODUCTION 
This section of the DICOM Conformance Statement specifies the Storage Commitment 
Push Model SOP Class, the optional attributes and service elements supported, the valid 
range of values for mandatory and optional attributes, and the status code behavior.  
11.2 STORAGE COMMITMENT PUSH MODEL SOP CLASS DEFINITION 
11.2.1 IOD Description...

Page 145

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 134 11.2.3 Operations 
11.2.3.1 Action Information 
Covered under IOD Description in Table 11.2.1 
 
11.2.3.2 Service Class User Behavior 
LOGIQ 7 sends the N-ACTION primitive (Storage Commitment Request) after 
successful exam save to a DICOM Storage SCP.  
LOGIQ 7 may request storage commitment for all generated SOP Class UIDs: 
Ultrasound Multi-frame Image Storage 1.2.840.10008.5.1.4.1.1.3.1 Ultrasound Multi-frame...

Page 146

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 135 11.2.4.1 Event Information 
TABLE 11.2-2 
STORAGE COMMITMENT RESULT - EVENT INFORMATION 
Event Type 
Name Event 
Type 
ID Attribute Tag Requirement Type SCU/SCP Storage 
Commitment 
Request 
Successful 1 Transaction UID (0008,1195) -/1   Retrieve AE Title (0008,0054) Not used   Storage Media File-Set 
ID (0088,0130) Not used   Storage Media File-Set 
UID (0088,0140) Not used   Referenced SOP 
Sequence (0008,1199) -/1...

Page 147

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 136   Failed SOP Sequence (0008,1198) -/1   >Referenced SOP Class 
UID (0008,1150) -/1   >Referenced SOP 
Instance UID (0008,1155) -/1   >Failure Reason (0008,1197) -/1 11.2.4.2 Service Class User Behavior 
If a successful answer is received, the request will be removed without warning the user. 
If a non-successful answer is received, the request will be left in the holding queue.  
If no answer is received, the request...

Page 148

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 137 12. PRINT MANAGEMENT SOP CLASS DEFINITION 
12.1 INTRODUCTION 
This section of the DICOM Conformance Statement specifies the supported Print 
Management SOP and Meta SOP Classes, the optional attributes and service elements 
supported, the valid range of values for mandatory and optional attributes, and the status 
code behavior.  
12.2 - Basic Print Management Meta SOP Classes 
12.3 - Print Management SOP Class...

Page 149

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 138 TABLE 12.2.2-1 BASIC COLOR PRINT MANAGEMENT META SOP CLASS 
SOP Class Name Usage 
SCU Reference Basic Film Session SOP Class      M see 12.3.1 Basic Film Box SOP Class      M see  12.3.2 Basic Color Image Box SOP Class      M see 12.3.3.2 Printer SOP Class      M see 12.3.4 12.3 PRINT MANAGEMENT SOP CLASS DEFINITIONS 
12.3.1 Basic Film Session SOP Class 
The Basic Film Session IOD describes the presentation...

Page 150

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 139 TABLE 12.3.2-1 DIMSE SERVICE GROUP 
DIMSE Service Element Usage 
SCU Reference N-CREATE      M see 12.3.2.1.1 N-ACTION      M see 12.3.2.1.2 N-DELETE      U see 12.3.2.1.3 N-SET      U see 12.3.2.1.4 12.3.2.1 DIMSE Service Group 
12.3.2.1.1 N-CREATE 
The N-CREATE DIMSE Service is used by LOGIQ 7 to request that the SCP create a 
Film Box SOP Instance.  Table 12.4.2-1 defines the Basic Film Box Presentation Module...