LOGIQ7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 120 9. MODALITY WORKLIST INFORMATION MODEL DEFINITION 9.1 INTRODUCTION This section specifies the use of the DICOM Modality Worklist Information Model used to organize data and against which a Modality Worklist Query will be performed. The contents of this section are: 9.2- Information Model Description 9.3- Information Model Entity-Relationship Model 9.4- Information Model Module Table 9.5- Information Model...
Page 132
LOGIQ7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 121 ILLUSTRATION 9.3-1 MODALITY WORKLIST INFORMATION MODEL E/R DIAGRAM Scheduled Procedure Step Requested Procedure contained in requested for Patient done for Imaging Service Request 1 1 1 Visit is included 0,1 Worklist Item 9.3.1 Entity Descriptions Please refer to DICOM Standard PS 3.3. (Information Object Definitions) and PS 3.4 (Service Class Specifications) for a description of each of the Entities contained...
Page 133
LOGIQ7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 122 9.3.1.3 Imaging Service Request Entity Description Image Service is implemented in a basic form to allow for the user to retrieve a subset of attributes. 9.3.1.4 Visit Entity Description Visit Entity is implemented in a basic form to allow for the user to retrieve a subset of attributes. 9.3.1.5 Patient Entity Description Patient Entity Description is implemented in a basic form to allow for the user to...
Page 134
LOGIQ7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 123 TABLE 9.4-1 MODALITY WORKLIST INFORMATION MODEL MODULES Entity Name Module Name Reference Scheduled Procedure Step SOP Common 9.5.2.1 Scheduled Procedure Step 9.5.2.2 Requested Procedure Requested Procedure 9.5.3.1 Imaging Service Request Imaging Service Request 9.5.4.1 Visit Visit Identification 9.5.5.1 Visit Status 9.5.5.2 Visit Relationship 9.5.5.3 Visit Admission Not Used Patient Patient Relationship Not...
Page 135
LOGIQ7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 124 Attribute Name Tag Expected Matching Key Type Expected Returned Key Type Mapped into Instance/MPPS Matching Specific Character Set (0008,0005) O 1C Yes/Yes Attribute is supported if the query contains matching keys in other than the default character repertoire. ISO IR 100 or ISO 2022 IR 87 is supported in responses. 9.5.2.2 Scheduled Procedure Step Module TABLE 9.5-2 SCHEDULED PROCEDURE STEP MODULE...
Page 136
LOGIQ7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 125 >Scheduled Protocol Code Sequence (0040,0008) O 1C Yes/Yes Matching is supported. 9.5.3 Requested Procedure Entity 9.5.3.1 Requested Procedure Module TABLE 9.5-3 REQUESTED PROCEDURE MODULE ATTRIBUTES Attribute Name Tag Expected Matching Key Type Expected Returned Key Type Mapped into Instance/MPPS Matching Requested Procedure ID (0040,1001) O 1 Yes/Yes (to Requested Procedure ID and Study ID)...
Page 137
LOGIQ7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 126 TABLE 9.5-4 IMAGING SERVICE REQUEST MODULE ATTRIBUTES Attribute Name Tag Expected Matching Key Type Expected Returned Key Type Mapped into Instance/MPPS Matching Accession Number (0008,0050) O 2 Yes/Yes Matching is supported. Filtering is supported. Referring Physicians Name (0008,0090) O 2 Yes/No Matching is supported. Imaging Service Request Comments (0040,2400) O 3 No/No Matching is supported....
Page 138
LOGIQ7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 127 Referenced Patient Sequence (0008,1120) O 2 Yes/Yes Matching is supported. >Referenced SOP Class UID (0008,1150) O 2 Yes/Yes Matching is supported. >Referenced SOP Instance UID (0008,1155) O 2 Yes/Yes Matching is supported. 9.5.6 Patient Entity 9.5.6.1 Patient Identification TABLE 9.5-8 PATIENT IDENTIFICATION MODULE ATTRIBUTES Attribute Name Tag Expected Matching Key Type Expected Returned Key...
Page 139
LOGIQ7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 128 9.5.6.3 Patient Medical TABLE 9.5-10 PATIENT MEDICAL MODULE ATTRIBUTES Attribute Name Tag Expected Matching Key Type Expected Returned Key Type Mapped into Instance/MPPS Matching Additional Patient History (0010,21B0) O 3 Yes/No Matching is supported. Contrast Allergies (0010,2210) O 2 No/No Matching is supported. Medical Alerts (0010,2000) O 2 No/No Matching is supported. Pregnancy Status (0010,21C0) O...
Page 140
LOGIQ7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 129 10. MODALITY PERFORMED PROCEDURE STEP SOP CLASS DEFINITION 10.1 INTRODUCTION This section of the DICOM Conformance Statement specifies the Modality Performed Procedure Step SOP Class, the optional attributes and service elements supported, the valid range of values for mandatory and optional attributes, and the status code behavior. 10.2 MODALITY PERFORMED PROCEDURE STEP SOP CLASS DEFINITION In this section,...