Page 51
49 © Adam Equipment Company 2013 1.16. PERCENT WEIGHING AND ANIMAL WEIGHING See section 10.7 and 10.8 for details of these spe cial weighing modes. Parameter Description Options Default setting F4 PCt This parameter allows the user to enter the Percent weighing Function. See Section 10.7. None Enabled always F4 AnL Enter the Animal Weighing mode of operation, See section 10.8 Set the filter value. Enabled Always
Page 52
50 © Adam Equipment Company 2013 ERROR MESSAGES During the initial power-on testing or during opera tion, the indicator may show an error message. The meaning of the error messages i s described below. If an error message is shown, repeat the step that caused the message. If the error message is still shown then contact your dealer for support. ERROR CODE DESCRIPTION POSSIBLE CAUSES Err 1 Time input Error Invalid time entry such...
Page 53
51 © Adam Equipment Company 2013 Err 9 Low limit input error High limit is set first, then the low limit is set higher than the high limit and low limit not equal to zero. FAIL H or FAIL L Calibration error Improper calibration (should be within +10% of the factory calibration). The old calibration data will be retained until the calibration process is complete.
Page 54
52 © Adam Equipment Company 2013 SERVICE PARAMETERS 1.17. ACCESS TO PARAMETERS APPROVED INDICATORS Access to the indicator parameters and calibration is controlled in all approved indicators either by limiting access to be after th e Calibration Jumper is put on the PCB, location J1, pins 1 & 2. In this case the dis play will show the passcode request screen, “ P - - - - “ . To continue enter a passcode as described below. Or if the...
Page 55
53 © Adam Equipment Company 2013 “SPAn” “PASS” If Calibration is complete, “SPAn PASS” will be displayed. Remove the calibration weight. Or, “SPAn” “FAiLEd ” This means calibration has failed. Remove the calibration weight and repeat the process. “JP On” Remove the jumper or shorting of the pins whichever is used. The indicator will return to normal weighing. 1.18. USING “2006” TO ENTER THE SERVICE PARAMETERS Non-Approved...
Page 56
54 © Adam Equipment Company 2013 Apply power to the indicator. If the jumper has be en used the display will ask for a code number, “Pn “ on the Weight Display immediately. Or press the [Tare] key during the time the calibration counters are being displayed. Enter the number 2006 when “Pn “ is displayed and then press [Tare]. The displays will show the first parameters, called “F1” “CAL”. To select another parameter press the...
Page 57
55 © Adam Equipment Company 2013 1.18.1 F1 -CALIBRATION To enter the calibration parameter, press the [Tare] key when “F1” is displayed. The indicator will be calibrated using 2 masses of appr oximately 1/3Maximum and Maximum. If the indicator has been calibrated once the values will be stored. If this is the first time the indictor is calibrated the us er must enter the values for the calibration masses. It is necessary to set...
Page 58
56 © Adam Equipment Company 2013 Switch off the indicator, and switch it on again to continue with the other Service parameters. 1.18.2 F2–DECIMAL POINT POSITION To set the value for the decimal point location. The options are 0, 0.0, 0.00, 0.000, 0.0000 To enter this parameter, press the [Tare] key when “F2 dEC” is shown. The display will show the current setting. Press the [Func] key to change the value....
Page 59
57 © Adam Equipment Company 2013 The display will show the current increment value a s used with the current decimal point position. Press the [Func] key to change the value. Select from 1,2,5,10, 20 or 50 Not all increments may be available for the capacit y you selected. For EC Approved versions the indicator will determi ne the increment that maintains the number of divisions to be 3000 or less. Press [Tare] to accept the displayed value. Press [Zero] to...
Page 60
58 © Adam Equipment Company 2013 Press [Tare] to accept the value. Press [Zero] to return to weighing. 1.18.6 F6 -SUCCESSIVE TARE To enter this parameter, press the [Tare] key when “ F6 SCS” is shown. The display will show if the successive tare is on or off. Press the [Func] key to change the value. Press [Tare] to accept the displayed value. Press [Zero] to return to weighing. 1.18.7 F7 –ADC COUNTS To enter this parameter, press the...