Page 51

 
49                              © Adam Equipment Company 2013 
 
1.16.  PERCENT WEIGHING AND ANIMAL WEIGHING  
 
See section 10.7 and 10.8  for details of these spe cial weighing modes.   
Parameter  Description  Options  Default  
setting 
F4   PCt  This  parameter  allows  the  
user  to  enter  the  Percent 
weighing  Function.  See 
Section 10.7.     None 
Enabled  
always 
F4 AnL  Enter  the  Animal  Weighing  
mode  of  operation,  See  
section 10.8  Set the filter value. 
Enabled  
Always 
  

Page 52

 
50                              © Adam Equipment Company 2013 
ERROR MESSAGES 
During  the  initial  power-on  testing  or  during  opera tion,  the  indicator  may  show  an 
error message.  The meaning of the error messages i s described below. 
If  an  error  message  is  shown,  repeat  the  step  that  caused  the  message.  If  the  error 
message is still shown then contact your dealer for  support. 
ERROR  
CODE   DESCRIPTION  POSSIBLE CAUSES 
 
Err  1 Time input Error 
Invalid time entry such...

Page 53

 
51                              © Adam Equipment Company 2013 
Err 9 Low limit input error 
High limit is set first, then the low  
limit is set higher than the high limit  
and low limit not equal to zero. 
FAIL H 
or  
FAIL L  Calibration error 
Improper calibration (should be  
within +10% of the factory  
calibration). The old calibration data  
will be retained until the calibration  
process is complete. 
  

Page 54

 
52                              © Adam Equipment Company 2013 
SERVICE PARAMETERS 
1.17.  ACCESS TO PARAMETERS 
APPROVED INDICATORS  
Access  to  the  indicator  parameters  and  calibration  is  controlled  in  all  approved 
indicators  either  by  limiting  access  to  be  after  th e  Calibration  Jumper  is  put  on  the 
PCB, location J1, pins 1 & 2.  In this case the dis play will show the passcode request 
screen,  “ P - - - - “  .  To continue enter a passcode as described below.   
Or if the...

Page 55

 
53                              © Adam Equipment Company 2013 
“SPAn”  “PASS” If  Calibration  is  complete, “SPAn  PASS” will  be 
displayed. Remove the calibration weight. 
Or, “SPAn”  “FAiLEd ” This  means  calibration  has  failed.  Remove  the  
calibration weight and repeat the process. 
“JP On” Remove  the  jumper  or  shorting  of  the  pins  
whichever  is  used.  The  indicator  will  return  to  
normal weighing. 
  
1.18.   USING “2006” TO ENTER THE SERVICE PARAMETERS 
Non-Approved...

Page 56

 
54                              © Adam Equipment Company 2013 
Apply power to the indicator.  If the jumper has be en used the display will ask for a 
code  number,   “Pn    “ on  the  Weight  Display  immediately.        Or  press  the  [Tare]  key 
during the time the calibration counters are being  displayed. 
Enter the number 2006 when  “Pn  “ is displayed and  then press  [Tare]. 
The displays will show the first parameters, called  “F1”  “CAL”.   
To  select  another  parameter  press  the...

Page 57

 
55                              © Adam Equipment Company 2013 
  
1.18.1 F1 -CALIBRATION 
 
To enter the calibration parameter, press the  [Tare] key when  “F1” is displayed.  The 
indicator  will  be  calibrated  using  2  masses  of  appr oximately  1/3Maximum  and 
Maximum.  If the indicator has been calibrated once  the values will be stored.  If this 
is  the  first  time  the  indictor  is  calibrated  the  us er  must  enter  the  values  for  the 
calibration masses.    
It is necessary to set...

Page 58

 
56                              © Adam Equipment Company 2013 
Switch  off  the  indicator,  and  switch  it  on  again  to   continue  with  the  other  Service 
parameters.       
1.18.2  F2–DECIMAL POINT POSITION 
 
To  set  the  value  for  the  decimal  point  location.        The  options  are    0,  0.0,    0.00,  
0.000,  0.0000    
To enter this parameter, press the  [Tare] key when  “F2 dEC” is shown. 
The display will show the current setting.  
Press the  [Func] key to change the value....

Page 59

 
57                              © Adam Equipment Company 2013 
The display will show the current increment value a s used with the current decimal 
point position.  
Press the  [Func] key to change the value. Select from 1,2,5,10, 20  or 50 
Not all increments may be available for the capacit y you selected. 
For EC Approved versions the indicator will determi ne the increment that maintains 
the number of divisions to be 3000 or less. 
Press  [Tare]  to accept the displayed value. 
Press  [Zero]  to...

Page 60

 
58                              © Adam Equipment Company 2013 
Press [Tare]  to accept the value. 
Press  [Zero]  to return to weighing. 
  
1.18.6  F6 -SUCCESSIVE TARE  
 
To enter this parameter, press the  [Tare] key when “ F6 SCS” is shown. 
The display will show if the successive tare is on  or off. 
Press the  [Func] key to change the value. 
Press  [Tare]  to accept the displayed value. 
Press  [Zero]  to return to weighing. 
  
1.18.7  F7 –ADC COUNTS  
 
To enter this parameter, press the...