9                              © Adam Equipment Company 2013 
  
Figure 1A shows the connections to a 6 wire load ce ll.  Figure 1B shows a preferred 
method  to  attach  a  4  wire  load  cell,  using  a  6  cond uctor  cable  to  go  from  the 
indicator to the platform or load cell where it con nects to the 4 wires from the load 
cells.  The Excitation and sense wires are connecte d together near the load cell. 
For less exacting applications you can connect the  excitation to the sense at the...

Page 12

 
10                              © Adam Equipment Company 2013 
 
[Low  
Limit]  
    &  
[High  
Limit] It  sets  the  limits  for  check  weighing  and  
allows  setting  of  either  the  low  limit  or  the  
high limit or both.   
None 
[



Lim]  It stores and recalls any of 10 preset limits    None 
[Func]   
This  is  used  to  select  percent  weighing,  RS- 
232  parameters,  Operation  of  the  bar  
graph, RTC settings, User ID and Scale ID.      None 
[Count]   
Enter Parts Counting...

Page 13

 
11                              © Adam Equipment Company 2013 
DISPLAY 
1.2.  SYMBOLS AND INDICATORS 
  
The LCD has unique symbols to indicate the following : 
 
0  
 The display is at Zero 
 
The scale is Stable 
Net Net weight- The scale has been tared   
kg  /  lb  Symbols shown for the units    
 
 
Capacity  Tracker-  A  bar  graph  indicating  the  propor tion  of  the 
scale capacity being used by the weight on the pan   
  
bAt  LO   or 
 
Low battery    
% The scale is in Percent weighing...

Page 14

 
12                              © Adam Equipment Company 2013 
  
Next  to  the  LCD  are  a  number  of  LED’s  that  show  whe n  the  weight  is  below,  within 
or over the limits during check weighing.  
  
 
Weight   LED  LCD 
below the low limit  Amber  LO  
Within the limits  Green  OK  
Above the high limit  Red  HI 
 
NOTE:  The  LED’s  can  be  set  by  the  user  to  off,  bar,   spot  or  segment  mode.  See “F3 
LED”  in section 13.1 
  
The  LED  can  be  set  to  display  as  a  bar,...

Page 15

 
13                              © Adam Equipment Company 2013 
The  Counters  are  incremented  any  time  the  calibrati on  section  or  the  Factory 
parameters section have been modified.     
At  power  on,  the  display  will  show  the  current  soft ware  revision  number    followed 
by  the  message  of  the  Calibration  Count      “ [AL[nt”  then  a  number  i.e.    “123”.      The 
number  from  the  counter  memory.      Then  the  Paramete r  Counter  message  of 
“ PAr[nt ”    and...

Page 16

 
14                              © Adam Equipment Company 2013 
BATTERY 
·  The  indicators  can  be  operated  from  the  rechargeabl e  battery,  if 
desired. The battery  life is determined by  the numb er and impendence 
of  the  load  cells  connected.      With  a  single  load  c ell  and  backlight 
disabled  the  life  is  approximately  70  hours  before  needing  to  be 
recharged. 
  ·   When  the  battery  needs  charging  a  symbol  on  the  dis play  will  turn  on. 
The battery should be...

Page 17

 
15                              © Adam Equipment Company 2013 
OPERATION 
1.3.  ZEROING  
 
·  You  can  press  the  [Zero] key  at  any  time  to  set  the  zero  point  from 
which  all  other  weighing  and  counting  is  measured.  This  will  usually  be 
necessary when the platform is empty. When the zero  point is obtained 
the display will show the zero indicator. 
 
 
·   The indicator has an automatic re-zeroing function  to account for minor 
drifting or accumulation of material on a...

Page 18

 
16                              © Adam Equipment Company 2013 
  NOTE:  
When  the  container  is  removed  a  negative  value  will   be  shown.  If  the 
indicator was tared just before removing the contai ner, this value is the gross 
weight  of  the  container  plus  all  products  which  wer e  removed.  The  zero 
indicator  will  also  be  on  as  the  platform  is  back  t o  the  same  condition  it  was 
when  [Zero]  was pressed last. 
If the value to be tared is very large the scale ma y...

Page 19

 
17                              © Adam Equipment Company 2013 
1.5.  WEIGHING 
 
To  determine  the  weight  of  a  sample,  first  tare  an  empty  container  if  used, 
then place the sample in the container.  The displa y will show the weight and 
the unit of weight currently in use.  
 
 
  
To  change  the  weighing  unit  press  the  [Unit]  key.    The  only  alternative 
weighing  unit  is  pounds.  This  can  be  enabled  by  the   user  in  the  parameters 
section. See section 13.3....

Page 20

 
18                              © Adam Equipment Company 2013 
 
·  Either  place  10  parts  on  the  platform  for  determini ng  the  average  piece 
weight or use a different number of parts. For exam ple, place 20 parts on 
the  platform, press  [CE] to clear the last values and then enter the value 
20 using the numeric keypad.   
 
 
 
·   Press  [Cnt]  to  weigh  the  samples  and  determine  an  average  piec e 
weight.     
·   If  the  parts  are  too  light  to  measure  accurately,  t he...