Page 11
EN © Adam Equipment Company 2015 9 3016612482 Rev1-Aug15 Model # EBL 12001e EBL 16001e EBL 22001e EBL 32001e Maximum Capacity 12000 g 16000 g 22000 g 22000 g Readability (d) 0.1 g 0.1 g 0.1 g 0.1 g Number of intervals n= 120000 160000 220000 320000 Min. weight 2 g 2 g 2 g 2 g Repeatability (Std. Dev) 0.2 g Linearity + 0.3 g Units of Measure grams, milligrams, carats, grains, Newtons, Drams, Ounces, troy ounces, pennyweight, momme, Taiwanese Taels, Hong Kong...
Page 12
EN © Adam Equipment Company 2015 10 3016612482 Rev1-Aug15 4 UNPACKIN\b TH\f BALANC\f Remove the balance from the packing by carefully lifting it out of the box. Inside the box you will find everything needed to start using the balance- · AC mains power adapter & cord · Stainless Steel Top Pan · Alloy sub-pan · Draught shield (for analytical models only) · Glass breeze ring and lid (0.01mg models) · User documentation Carefully follow the quick setup guide...
Page 13
EN © Adam Equipment Company 2015 11 3016612482 Rev1-Aug15 6 S\fTTIN\b UP TH\f BALANC\f 6.1 ASSEMBLING THE BALANCE Carefully follow the included quick setup guide to assemble the balance. Ensure that you locate the balance on a solid level surface, free from vibration. 6.1.1 Levelling the balance After placing the balance in a suitable location, level it by using the spirit level on the front of the balance. To level the balance turn the two adjustable feet at the rear of...
Page 14
EN © Adam Equipment Company 2015 12 3016612482 Rev1-Aug15 6.2 CALIBRATION Units with an ‘i’ suffix after the model name can be calibrated using either internal calibration mechanism or by using an external mass. Units with an ‘ e’ suffix can only be calibrated with an external mass. Internal calibration option must be enabled in the setup menu options or else external calibration mode will be used when the [Cal] key is pressed. 6.2.1 Manual Calibration Pressing the [Cal] key...
Page 15
EN © Adam Equipment Company 2015 13 3016612482 Rev1-Aug15 Conditions that will trigger an automatic calibration are: · Internal temperature change greater than a pre-set amount (typically 2ºC for Precision balances). · Time since last calibration exceeds a pre-set time (typically 4 hours, or 15 minutes after power is applied). The balance will indicate the need for calibration to be carried out by flashing the “ CAL” symbol on the display. As soon as the balance is calibrated...
Page 16
EN © Adam Equipment Company 2015 14 3016612482 Rev1-Aug15 7 DISPLAY The LCD has several areas:- A large 7 digit area to display the weight with symbols for common weighing units on its right and symbols for zero, tare (Net) and stability on the left. A small 10-character display line above the main display area. This is used to provide additional instructions and information on the current operation or function being used. A capacity indicator bar at the bottom of the...
Page 17
EN © Adam Equipment Company 2015 15 3016612482 Rev1-Aug15 8 K\fYPAD The keypad has the following keys to operate the balance. Keys Primary function [POWER] To turn the balance to ON or Standby [ 0/T ] [ 0/T ] A combined zero and tare function. Also used to escape from setup functions and modes. [Cal] [Cal] Starts the calibration function [Print] [Print] Instructs the balance to print data [Mode] [Mode] Enters the Mode Selection Menu [Unit] [Unit]...
Page 18
EN © Adam Equipment Company 2015 16 3016612482 Rev1-Aug15 [Down] To decrement or change a displayed value or scroll through options backwards [Right] To advance a flashing digit by one position to the right. To go back by one step during setup functions [Left] To advance a flashing digit by one position to the left [Up] To increase or change a displayed value or scroll through options forward 8.1 NUMERIC ENTRY METHOD To set a value when required, use the keys as...
Page 19
EN © Adam Equipment Company 2015 17 3016612482 Rev1-Aug15 9 INPUT/OUTPUT The rear panel has some or all of the following connectors depending on model: · RS-232 serial - 9 pin d-subminiature plug. · Remote display connector (USB type A socket). · I/O connector (USB type B socket). · Power input socket. (Required power input is a low-voltage external supply, 18VDC @ 830mA for all models). Accepts concentric barrel plug 11.4mm length X 5.5mm outside diameter X 2.1mm...
Page 20
EN © Adam Equipment Company 2015 18 3016612482 Rev1-Aug15 10 OP\fRATIONS 10.1 INITIALISATION When the balance is first switched on, it will display the balance serial number, the software revision, the model capacity and then all segments on the display will be shown. Overall the time taken is usually 5 -10 seconds. If Operator and Supervisor passcodes have been set, the display will show “ PASS PASSPASS PASSC CC C\b\b\b \bDDD DE EE E 0 0 0 0 ”. In t his case you...