Page 31
EN © Adam Equipment Company 2015 29 3016612482 Rev1-Aug15 10.4.5 NET/TOTAL & Accumulation Mode. The Net/total function allows the user to add number of items consecutively and to see the total of all the net weights, the value of which is displayed on smaller digits above the main display. The current net weight is shown on the main display. The [Tare] key functions the same way as when in normal weighing mode. Using NET/TOTAL, you can only load items up to the maximum capacity of...
Page 32
EN © Adam Equipment Company 2015 30 3016612482 Rev1-Aug15 ACCUMULATION · Press [Mode] and then [Up] or [Down] to select “ NEt T\bT NEt T\bTNEt T\bT NEt T\bT ” for Net/Total mode. Now press [Select] to enter the function. The display will show zeros in both main and upper display areas, accompanied by “ NT NTNT NT” in the upper display line to indicate that the balance is in the Net/Total mode. · Place the first item. The net weight will be displayed...
Page 33
EN © Adam Equipment Company 2015 31 3016612482 Rev1-Aug15 10.4.6 Density Determination It is possible to determine the density of solids or liquids using this mode. The user selects the type of density to be determined and then enters values to be used by the balance. The density mode allows the user to use a special Density Kit (available from your dealer), or use the below pan weighing facility to perform the necessary weighing. Press [Mode] until “ Density DensityDensity...
Page 34
EN © Adam Equipment Company 2015 32 3016612482 Rev1-Aug15 The [Print] key is lit, indicating that print output to a connected device is available. Print output will be the computed density “ XX.XXX g/cc” as shown on screen. Remove the item from the pan and press [Mode] to continue with a new sample or press [ 0/T ] to return to normal weighing. 10.4.6.2 Density of a Liquid When finding the density of a liquid, it is necessary to weigh a sample of known volume in air and...
Page 35
EN © Adam Equipment Company 2015 33 3016612482 Rev1-Aug15 11 RS-232 INTERFACE The balances have the ability to send or receive data over the serial interfaces, RS232 and USB (if fitted). Both interfaces are controlled by the parameters detailed below. If the host computer to be used does not have a serial port then a USB-RS232 convertor accessory can be used. The USB and RS232 both operate as general purpose serial data ports. Weighing data can be sent over the interface either...
Page 36
EN © Adam Equipment Company 2015 34 3016612482 Rev1-Aug15 11.2.2 STANDARD OUTPUT FORMAT The balance will print the following data as the standard form. The standard form cannot be changed. The format of the custom forms #1 and #2 will be the same as the standard form until modified by the user. Line 1 Date Line 2 Time Line 3 Blank line Line 4 ID number Line 5 Blank line Line 6 Result Line 7 Blank line Line 8 Blank line This will result in a printout that looks like:...
Page 37
EN © Adam Equipment Company 2015 35 3016612482 Rev1-Aug15 The data for each form will be preceded by a start-of-header ASCII character (01) and terminated with an end-of-transmission ASCII character (04). These characters will be ignored by a serial printer but will allow a computer program which reads the data to distinguish between this block report format and the single-line output format described above. 11.3 INPUT COMMANDS USING REMOTE KEYS The balance can be controlled...
Page 38
EN © Adam Equipment Company 2015 36 3016612482 Rev1-Aug15 If the remote display data stream format is required for development purposes then please contact the manufacturer for advice. C O P Y
Page 39
EN © Adam Equipment Company 2015 37 3016612482 Rev1-Aug15 12 ERROR CHECKING During weighing the balance is constantly checking to see if it is operating within the limited parameters. The errors likely to occur are: A/D counts below lowest allowed value A/D counts above highest allowed value A/D not operating Maximum capacity exceeded Other errors may be detected during special functions or operations. These will be described in the section that applies. Error...
Page 40
EN © Adam Equipment Company 2015 38 3016612482 Rev1-Aug15 13 SUPERVISOR MENUS Pressing the [Select] key while in normal weighing gives access to the menus. When [Select] is pressed and the Supervisor Passcode is not enabled the display will allow access to the Supervisor menus. If passcode is enabled, the balance will ask for it by displaying “ PASS CD P ASS CDPASS CD PASS CD ” shortly followed by displaying “ 0 0 0 0 ”. I f a passcode is incorrectly entered then the...