Home > Adam Equipment > Balance > Adam Equipment Eclipse Precision Balances ECLIPSEEBL User Manual

Adam Equipment Eclipse Precision Balances ECLIPSEEBL User Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Adam Equipment Eclipse Precision Balances ECLIPSEEBL User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 35 Adam Equipment manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							© Adam Equipment Company 2015    
     
     
    Adam Equipment 
     
    Eclipse EBL Series (EN) 
    Série Eclipse EBL (FR)  
     
    Operating Manual 
    Notice d’utilisation 
    For internal (‘i’) and external (‘e’) calibration models 
     
    (P.N. 3.01.6.6.12482, Revision 1.00, Effective Aug 2015) 
     
    Software rev.:  V1.2155 & above (Force Motor Analytical Models) 
          V2.1827 & above (Precision Load Cell Models) 
          V6.1010 & above (Large Pan Load Cell Models) 
     
     
     
           
    C  O  P Y  
    						
    							© Adam Equipment Company 2015    
    ENGLISH:      P  1  -  54 
    FRANÇAIS:      P  55  -  111 
    EU DECLARATION: P  112
     
       
    C  O  P Y  
    						
    							EN 
    © Adam Equipment Company 2015   1             3016612482 Rev1-Aug15 
    TABLE OF CONTENTS  
    1  KNOW YOUR BALANCE .......................................................................................................................... 2 2  PRODUCT OVERVIEW ............................................................................................................................ 3 3  PRODUCT SPECIFICATIONS ................................................................................................................. 4 
    4  UNPACKING THE BALANCE ................................................................................................................. 10 
    5  LOCATING THE BALANCE .................................................................................................................... 10 
    6  SETTING UP THE BALANCE ................................................................................................................ 11  6.1  ASSEMBLING THE BALANCE ...................................................................................................... 11 6.1.1  Levelling the balance ............................................................................................................... 11 6.1.2  Warm-Up Time......................................................................................................................... 11 
    6.1.3  Weighing .................................................................................................................................. 11 
    6.2  CALIBRATION ............................................................................................................................... 12  6.2.1  Manual Calibration ................................................................................................................... 12 6.2.2  Calibration using Internal Calibration mass (if fitted) ............................................................... 12 
    6.2.3  Calibration using External Calibration mass ............................................................................ 12 6.2.4  Automatic Calibration ............................................................................................................... 12 
    6.2.5  Calibration Errors ..................................................................................................................... 13 
    7  DISPLAY ................................................................................................................................................. 14  7.1  SYMBOLS AND TEXT ................................................................................................................... 14 
    8  KEYPAD .................................................................................................................................................. 15  8.1  NUMERIC ENTRY METHOD ......................................................................................................... 16 
    9  INPUT/OUTPUT...................................................................................................................................... 17 
    10  OPERATIONS ..................................................................................................................................... 18  10.1  INITIALISATION ............................................................................................................................. 18 10.2  PASSCODES ................................................................................................................................. 18 
    10.3  WEIGHING ..................................................................................................................................... 18 
    10.4  FUNCTIONS .................................................................................................................................. 21  10.4.1  Parts Counting ..................................................................................................................... 22 10.4.2  Percentage Weighing .......................................................................................................... 23 10.4.3  Check Weighing ................................................................................................................... 24 10.4.4  Animal (Dynamic) Weighing ................................................................................................ 26 
    10.4.5  NET/TOTAL & Accumulation Mode. .................................................................................... 29 
    10.4.6  Density Determination ......................................................................................................... 31 
    11  RS-232 INTERFACE ........................................................................................................................... 33  11.1  HARDWARE .................................................................................................................................. 33 
    11.2  OUTPUT FORMATS ...................................................................................................................... 33  11.2.1  SINGLE-LINE OUTPUT FORMAT ...................................................................................... 33 
    11.2.2  STANDARD OUTPUT FORMAT ......................................................................................... 34 
    11.2.3  CUSTOM OUTPUT FORMAT ............................................................................................. 34 
    11.3  INPUT COMMANDS USING REMOTE KEYS............................................................................... 35  11.3.1  Invalid Input Command: ....................................................................................................... 35 
    12  ERROR CHECKING ............................................................................................................................ 37 
    13  SUPERVISOR MENUS ....................................................................................................................... 38  13.1  ENABLE WEIGHING UNITS .......................................................................................................... 38 
    13.2  ENABLE WEIGHING MODES ....................................................................................................... 38 
    13.3  ENABLE SERIAL INTERFACE PARAMETERS ............................................................................ 38 
    13.4  FORMAT OF CUSTOM FORMS #1 and #2 .................................................................................. 39 
    13.5  SETUP PARAMETERS .................................................................................................................. 40 
    13.6  CALIBRATION SETUP .................................................................................................................. 41 
    13.7  PASSCODES ................................................................................................................................. 41 
    14  ACCESSORIES & SPARE PARTS ..................................................................................................... 43  15  SAFETY AND MAINTENANCE ........................................................................................................... 45 
    16  TROUBLE-SHOOTING ....................................................................................................................... 46  16.1  TROUBLE-SHOOTING GUIDE. .................................................................................................... 47 
    17  SERVICE INFORMATION ................................................................................................................... 49 
    18  BALANCE MENU STRUCTURE ......................................................................................................... 50  19  LANGUAGE TABLE ............................................................................................................................ 53 
    20  WARRANTY INFORMATION .............................................................................................................. 54 
     
     
     
    C  O  P Y  
    						
    							EN 
    © Adam Equipment Company 2015   2             3016612482 Rev1-Aug15 
    1  KNOW YOUR BALANCE 
     
    Thank you for selecting the Eclipse Balance from Adam Equipment.  
     
    This Instruction Manual will familiarise you with the installation, use, general maintenance 
    etc. of the balance, and will guide you through the various applications. It also covers 
    accessories, trouble-shooting, after sales service information, and other important 
    information. 
     
    These balances are highly accurate precision instruments and contain sensitive 
    mechanisms and components. They should be transported and handled with care. When 
    in operation, be careful to place loads gently on the weighing surface and do not overload 
    or exceed recommended maximum capacity of the instrument or damage may occur. 
     
    Please read this Manual thoroughly before starting operation. If you need any 
    clarifications, feel free to contact your supplier or Adam Equipment. 
     
     
     
    C  O  P Y  
    						
    							EN 
    © Adam Equipment Company 2015   3             3016612482 Rev1-Aug15 
    2  PRODUCT OVERVIEW 
     
    The Eclipse balances are ideal for laboratory and general purpose weighing. They can 
    also be used for some advanced weighing functions.  
     
    FEATURES: 
     
    ·  External menu-driven calibration. 
    ·   Internal calibration (option) for outstanding accuracy without the need for manual calibration. 
    ·   Large easy to read 2-line LCD display with blue backlight, 24 mm tall primary display digits, 10 
    mm secondary digits, and capacity tracker bar. 
    ·   Easy to use, wipe-clean sealed keypad with pressureless capacitive-touch backlit buttons which 
    work with fingers, gloves or stylus. 
    ·   Force-restoration mechanism option for supreme accuracy, or alloy load cell technology option for 
    stable yet highly accurate weighing. 
    ·   Solid die-cast aluminium alloy construction with 304 grade stainless steel pan for durability and 
    easy cleaning. 
    ·   Standard applications include weighing, check weighing, percentage weighing, parts counting, 
    dynamic (animal) weighing, accumulation (net/total), and solid and liquid density determination. 
    ·   Bi-directional RS-232 interface and USB interface as standard. 
    ·   Can be configured to print a GLP Compliant report after each calibration to include the time, date, 
    balance number and a verification of the calibration. 
    ·   Automatic temperature compensation. 
    ·   Large range of weighing units, including customised option. 
    ·   Below balance weighing facility (accessory hook required). 
    ·   Mains powered. 
    ·   External display option. 
    ·   Display and documentation in a choice of 4 languages – English, German, French & Spanish. 
    ·   Password protection. 
    ·   Security locking point. 
     
     
      
    C  O  P Y  
    						
    							EN 
    © Adam Equipment Company 2015   4             3016612482 Rev1-Aug15 
    3  PRODUCT SPECIFICATIONS 
    Eclipse EBL Models 
    (Suffix e for external calibration models, Suffix i for internal calibration models) 
     
    Model #  EBL 104 e/i  EBL 164  e/i  EBL 214 e/i  EBL 254 e/i  EBL 314 e/i 
    Maximum 
    Capacity  100 g  160 g 
    210 g 250 g  310 g 
    Readability (d)  0.0001 g 
    Number of 
    intervals n= 1000000  1600000  2100000  2500000  3100000 
    Min. weight 
    0.01 g 
    Repeatability 
    (Std. Dev) 0.00015 g 
    0.0002 g 
    Linearity + 0.0003 g 
    Units of Measure grams, milligrams, carats, grains, Newtons, Drams,  Ounces, troy ounces, pennyweight, 
    momme, Taiwanese Taels, Hong Kong Taels, Singapore Taels, Ticals, Tola, Custom 
    Stabilization 
    Time  Typically 5 seconds 
    Operating Temp 
    15ºC  to  35ºC recommended, 40 – 60 % RH (non-condensing) 
    Power Supply  External mains power adapter - supplied as standard 
    (Input Voltage 100–240 VAC, 50/60 Hz) 
    Input Voltage  18 VDC - 830 mA 
    Weighing 
    mechanism Force Restoration Balance Motor 
    Calibration 
    Suffix i = internal calibration mechanism, e = external calibration only 
    External 
    Calibration Mass  Recommended OIML class: E2, ASTM / ANSI class: 1 
    100 g 
    Display 
    LCD with blue backlight, 7 character 24 mm high main display, with symbols. Secondary 
    information line of 10mm high alpha-numeric characters. 
    Keypad 12 capacitive-touch backlit keys – work with fingers, gloves or stylus contact. 
    Draft Shield  
    (w x d x h)  Sliding door Draft Shield  (165 x 145 x 240 mm) 
    Pan Size 
    Round, 90mm diameter 
    Overall 
    Dimensions 
    (w x d x h)  220 x 310 x 323 mm 
    8.7 x 12.2 x 12.7 in 
    Net Weight  5.2 kg / 11 lb 9 oz (external calibration model) 
    5.9 kg / 13 lb 0 oz (internal calibration model) 
     
     
       
    C  O  P Y  
    						
    							EN 
    © Adam Equipment Company 2015   5             3016612482 Rev1-Aug15 
     
    Model # EBL 223 e / i  EBL 423 e / i 
    Maximum Capacity 220 g  420 g 
    Readability (d)  0.001 g 
    Number of intervals 
    n= 220000 
    420000 
    Min. 0.02 g  0.02 g 
    Repeatability 
    (Std. Dev)  0.002 g 
    Linearity + 0.003 g 
    Units of Measure grams, milligrams, carats, grains, Newtons, Drams,  Ounces, troy ounces, pennyweight, 
    momme, Taiwanese Taels, Hong Kong Taels, Singapore Taels, Ticals, Tola, Custom 
    Stabilization Time 
    Typically 5 seconds 
    Operating Temp 15ºC  to  35ºC recommended, 40 – 60 % RH (non-condensing) 
    Power Supply  External mains power adapter - supplied as standard 
    (Input Voltage 100–240 VAC, 50/60 Hz) 
    Input Voltage  18 VDC - 830 mA 
    Weighing 
    mechanism Precision Load Cell 
    Calibration 
    Suffix i = internal calibration mechanism, e = external calibration only, 
    External Calibration 
    Mass  Recommended OIML class: E2, ASTM / ANSI class: 2 
    100 g 
    200 g 
    Display LCD with blue backlight, 7 character 24 mm high main display, with symbols. Secondary 
    information line of 10mm high alpha-numeric characters. 
    Keypad  12 capacitive-touch backlit keys – work with fingers, gloves or stylus contact. 
    Draft Shield  
    (w x d x h)  Glass Ring Draft Shield With Alloy Lid (180 mm diam. x 90 mm) 
    Pan Size 
    Round, 120 mm diameter 
    Overall Dimensions 
    (w x d x h) 220 x 310 x 90 mm without breeze ring 
    8.7 x 12.2 x 3.5 in 
    Net Weight  3.1 kg /  6 lb 12 oz (external calibration model) 
    3.7 kg / 8 lb 8 oz (internal calibration model) 
     
     
       
    C  O  P Y  
    						
    							EN 
    © Adam Equipment Company 2015   6             3016612482 Rev1-Aug15 
     
    Model # EBL 623 e / i  EBL 823 e / i  EBL 1023 e / i EBL 1623 e / i 
    Maximum Capacity  620 g  820 g  1020  1620 
    Readability (d) 0.001 g 
    Number of intervals 
    n= 620000  820000  1020000  1620000 
    Min. weight 
    0.02 g  0.02 g  0.02 g  0.02 g 
    Repeatability 
    (Std. Dev)  0.002 g 
    Linearity + 0.003 g 
    Units of Measure grams, milligrams, carats, grains, Newtons, Drams,  Ounces, troy ounces, 
    pennyweight, momme, Taiwanese Taels, Hong Kong Taels, Singapore 
    Taels, Ticals, Tola, Custom 
    Stabilization Time 
    Typically 5 seconds 
    Operating Temp 15ºC  to  35ºC recommended, 40 – 60 % RH (non-condensing) 
    Power Supply  External mains power adapter - supplied as standard 
    (Input Voltage 100–240 VAC, 50/60 Hz) 
    Input Voltage  18 VDC - 830 mA 
    Weighing 
    mechanism Force Restoration Balance Motor 
    Calibration 
    Suffix i = internal calibration mechanism, e = external calibration only 
    External Calibration 
    Mass  Recommended OIML class: E2, ASTM / ANSI class: 2 
    500 g 
    1 kg 
    Display LCD with blue backlight, 7 character 24 mm high main display, with 
    symbols. Secondary information line of 10mm high alpha-numeric  characters. 
    Keypad 12 capacitive-touch backlit keys – work with fingers, gloves or stylus 
    contact. 
    Draft Shield  
    (w x d x h) Glass Ring Draft Shield With Alloy Lid (180 mm diam. x 90 mm) 
    Pan Size 
    Round, 160 mm diameter 
    Overall Dimensions 
    (w x d x h) 220 x 310 x 90 mm without breeze ring 
    8.7 x 12.2 x 3.5 in 
    Net Weight  4.0 kg /  8 lb 13 oz (external calibration model) 
    4.8 kg / 10 lb 9 oz (internal calibration model) 
       
    C  O  P Y  
    						
    							EN 
    © Adam Equipment Company 2015   7             3016612482 Rev1-Aug15 
     
    Model # EBL 1602 e / i  EBL 2602 e / i  EBL 3602 e / i  EBL 4602 e / i  EBL 6202 e / i 
    Maximum 
    Capacity  1600 g  2600 g  3600 g 
    4600 g 6200 g 
    Readability (d)  0.01 g  0.01 g  0.01 g  0.01 g 0.01 g 
    Number of 
    intervals n=  160000  260000  360000  460000  620000 
    Min. weight 
    0.2 g 0.2 g  0.2 g  0.2 g  0.2 g 
    Repeatability 
    (Std. Dev)  0.02 g 
    Linearity + 0.03 g 
    Units of Measure grams, milligrams, carats, grains, Newtons, Drams,  Ounces, troy ounces, pennyweight, 
    momme, Taiwanese Taels, Hong Kong Taels, Singapore Taels, Ticals, Tola, Custom 
    Stabilization Time  Typically 5 seconds 
    Operating Temp 15ºC  to  35ºC recommended, 40 – 60 % RH (non-condensing) 
    Power Supply  External mains power adapter - supplied as standard 
    (Input Voltage 100–240 VAC, 50/60 Hz) 
    Input Voltage  18 VDC - 830 mA 
    Weighing 
    mechanism Precision Load Cell 
    Calibration 
    Suffix i = internal calibration mechanism, e = external calibration only 
    External 
    Calibration Mass  Recommended OIML class: F1, ASTM / ANSI class: 3 
    500 g 
    1 kg  2 kg 
    Display LCD with blue backlight, 7 character 24 mm high main display, with symbols. Secondary 
    information line of 10mm high alpha-numeric characters. 
    Keypad  12 capacitive-touch backlit keys – work with fingers, gloves or stylus contact. 
    Draft Shield  
    (w x d x h)  None 
    Pan Size 
    Round, 160 mm diameter 
    Overall 
    Dimensions 
    (w x d x h)  220 x 310 x 90 mm 
    8.7 x 12.2 x 3.5 in 
    Net Weight  3.1 kg /  6 lb 14 oz (external calibration model) 
    3.9 kg / 8 lb 10 oz (internal calibration model) 
     
       
    C  O  P Y  
    						
    							EN 
    © Adam Equipment Company 2015   8             3016612482 Rev1-Aug15 
     
    Model # EBL 4201e  EBL 6201e EBL 8201e 
    Maximum Capacity 4200 g  6200 g 8200 g 
    Readability (d)  0.1 g 0.1 g 0.1 g 
    Number of 
    intervals n=  42000 
    62000 82000 
    Min. weight  2 g 2 g 2 g 
    Repeatability 
    (Std. Dev)  0.2 g 
    Linearity + 0.3 g 
    Units of Measure grams, milligrams, carats, grains, Newtons, Drams,  Ounces, troy ounces, pennyweight, 
    momme, Taiwanese Taels, Hong Kong Taels, Singapore Taels, Ticals, Tola, Custom 
    Stabilization Time  Typically 5 seconds 
    Operating Temp 15ºC  to  35ºC recommended, 40 – 60 % RH (non-condensing) 
    Power Supply  External mains power adapter - supplied as standard 
    (Input Voltage 100–240 VAC, 50/60 Hz) 
    Input Voltage  18 VDC - 830 mA 
    Weighing 
    mechanism Precision Load Cell 
    Calibration 
    External calibration only 
    External 
    Calibration Mass Recommended OIML class: F2, ASTM / ANSI class: 4 
    2 kg 
    Display 
    LCD with blue backlight, 7 character 24 mm high main display, with symbols. Secondary 
    information line of 10mm high alpha-numeric characters. 
    Keypad  12 capacitive-touch backlit keys – work with fingers, gloves or stylus contact. 
    Draft Shield  
    (w x d x h)  None 
    Pan Size 
    Round, 160 mm diameter 
    Overall 
    Dimensions 
    (w x d x h)  220 x 310 x 90 mm 
    8.7 x 12.2 x 3.5 in 
    Net Weight  3.1 kg /  6 lb 14 oz 
     
       
    C  O  P Y  
    						
    All Adam Equipment manuals Comments (0)