    							Connection to the water inlet
    ,Water in the machine is not suit
    -
    able for drinking.
    ^The machine must be connected to
    the water supply in accordance with
    local and national regulations (WRC
    the U.K.).
    ^The machine is constructed to com
    -
    ply with German gas and water regu
    -
    lations (DVGW) and may be con
    -
    nected to a suitable supply without
    an extra non-return valve if national
    regulations allow this.
    ^The water pressure (flow rate) must
    be a minimum of 0.5 bar (50 kPa).
    If the water pressure (flow rate) is be-
    low 2 bar (200 kPa) the fill time is au-
    tomatically extended.
    The max. permitted static pressure is
    10 bar (1000 kPa).
    If the water pressure is not in the
    range of 0.5 to 10 bar (50-1000 kPa)
    thep/6indicator light may come
    on and the fault code F..E show in
    the display. Contact the Miele Service
    Department for advice.
    ^
    For technical application reasons the
    machine is supplied as standard for
    connection to cold water only
    .
    The inlet hoses (water inlet and
    steam condenser) must only be con-
    nected to the cold water stopcocks
    .
    See next page for the AD water con-
    nection (hose marked H
    2O pur).^Stopcocks with 3/4 BSP male thread
    are to be provided on site. They
    should be easily accessible so that
    the water supply can be turned off
    when the machine is not in use.
    ^The DN 10 inlet hose is approx. 1.7 m
    long terminating in a 3/4 inch female
    thread. On no account must the inlet
    filters be removed.
    ^Large surface area filters are en
    -
    closed in the kit supplied with the
    machine for installing between the
    stopcock and the inlet hose
    (see illustration in Cleaning and Care
    Water inlet).
    The large surface area filter for AD
    water is made from chrome nickel
    steel and can be recognised by the
    matt top surface.
    ,The inlet hose must notbe short-
    ened or damaged in any way (see il-
    lustration).
    See also the installation diagram sup-
    plied.
    Plumbing
    57 
    						
    							AD water connection
    ö1.5 - 10 bar (150 - 1000 kPa) over
    pressure
    The machine is supplied as standard
    for connection to an AD water supply of
    1.5 - 10 bar (150 - 1000 kPa).
    ^The AD pressure tested hose
    (marked H
    2O pur) with a 3/4
    threaded union must be connected to
    the on-site AD stopcock for purified
    water.
    If the AD water connection is not go
    -
    ing to be used, the electronics will
    need to be reprogrammed by a Miele
    Service Technician. The inlet hose
    will remain on the rear of the ma-
    chine.
    Plumbing
    58 
    						
    							Drainage
    ^The drainage system is fitted with a
    non-return valve which prevents dirty
    water from flowing back into the ma
    -
    chine via the drain hose.
    ^The machine should preferably be
    connected to a separate
    drainage
    system on-site.
    If separate drainage is not available
    contact your Miele application spe
    -
    cialist for professional advice.
    The on-site drain connection point
    should be sited between 0.3 m and 1
    m above the lower edge of the ma
    -
    chine.
    If it is lower than 0.3 m lay the hose in
    a curve at a height of at least 0.3 m.
    The drainage system must be able to
    take a minimum drainage flow of 16
    l/min.
    ^The inlet hose is approx. 1.4 m long,
    is flexible and has an internal diame-
    ter of 22 mm. It must not be short-
    ened.
    Hose clips are supplied for securing
    it in position.
    ^
    A longer drain hose (up to 4 m long)
    is available to order from the Miele
    Spare Parts Dept.
    ^
    The drainage system must not ex-
    ceed 4 metres.
    See also the installation diagram sup-
    plied.
    Plumbing
    59 
    						
    							Height: 85 (82) cm
    Width: 60 cm
    Depth: 60 cm
    Voltage: see data plate
    Rated load: see data plate
    Fuse rating: see data plate
    Mains connection cable: approx. 1.8 m
    Water pressure (flow rate): 0.5 - 10 bar (50 - 1000 kPa)
    Cold water connection
    AD-water connectionü150 - 1000 kPa (1.5 - 10 bar)
    Delivery head: min. 0.3 m, max. 1 m
    Drainage length: max. 4 m
    Inlet hose: approx. 1.7 m
    Drain hose: approx. 1.4 m
    Noise level in dB (A):
    Sound level LpA Free standing: 
    						
    							61 
    						
    							63 
    						
    							M.-Nr. 05 800 640 / VAlteration rights reserved / 4902