Miele G 7881 User Manual
Have a look at the manual Miele G 7881 User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 171 Miele manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
Please refer to the installation diagram supplied with the appliance. ,In order to reduce the risk of steam damage, the area around the machine should be limited to furniture and fittings that are designed for use in commercial wash environments. The washer-disinfector must be in a stable and level horizontal position. Use the front screw feet to compensate for unevenness in the floor or to adjust the height of the appliance. Installation options: –Free-standing – Slide-in: Next to other machines or furniture or in a suitable space. The space must be at least 60 cm wide and 60 cm deep. – Undercounter: The washer-disinfector can be installed under a continuous countertop. The recess must be at least 60 cm wide, 60 cm deep and 85 cm high. Undercounter installation To install the machine under a continuous countertop, the machine lid can be removed, as follows: ^Open the door. ^Unscrew the left and right fastening screws. ^Pull the lid forward approx. 5 mm, then lift it upwards and off. Washer-disinfectors with a steam condensor: To prevent steam damage to the countertop, the adhesive protective foil (25 x 58 cm) delivered with the machine needs to be attached underneath the counter near the steam condensor. Installation 61
How to position and secure the machine To ensure stability the washer-disinfector must first be aligned and then secured to the countertop. ^Open the door. ^Screw the washer-disinfector to the countertop through the left and right holes in the front trim. ^Do NOT use silicone sealant to seal the gaps between the washer-disinfector and any neighboring units as this would hinder ventilation to the circulation pump. How to protect the countertop Depending on your requirements for installation, contact Miele Customer Service for countertop protection kits. The underside of the countertop is protected from steam damage by a stainless steel plate.Decorative frame and decorative panel (Option) The front of the washer-disinfector can be fitted with a decorative panel to match the cabinetry. Decorative panels should only be exchanged in accordance with all applicable safety regulations. Decorative panel dimensions: Outer door panel H 442 / W 586 mm Mounting plate H 117 / W 586 mm / D 1.2 mm thick Toe kick service panel H 117 mm / W 586 mm / D 1.2 mm thick Installation 62
,All electrical work must be performed by a suitably qualified electrician in accordance with local and national safety regulations. –Electrical connection should be made via a suitable isolator with an on-off switch which should be easily accessible for servicing work. –The electrical connection is made through a receptacle according to national requirements. The receptacle has to be easily accessible after installation. – For increased safety, it is URGENTLY recommended to install a protective switch (30 mA) before the machine. – To replace the power cord, use only original Miele replacement parts or a suitable cord with core cable ends. – For technical data, see the data plate or attached wiring diagram. The machine must only be operated with the voltage, frequency and fusing shown on thedata plate. The machine can be converted according to the supplied conversion diagram and wiring diagram. Thedata plateis located on the rear of the machine, behind the plinth, on the plastic cover.Thewiring diagramis secured to the machine. See the supplied installation diagram. WARNING THIS APPLIANCE MUST BE GROUNDED Grounding connection The ground lead must be connected to the screw connection point (marked with the ground symbol) at the back of the machine. ,The machine must only be operated with the voltage, frequency and fusing shown on the data plate located on the rear of the machine and on the plinth (behind the service panel). Theconversion diagramand the wiring diagramare located in the document pouch. Electrical connection 63
Connecting to the water inlet ,The water in the machine is NOT suitable for drinking. –The machine must be connected to the water supply in accordance with local and national regulations. –For technical application reasons, the machine is supplied as standard for connection tocold water only. The inlet hoses (water inlet and steam condenser) must only be connected to the cold water inlets. See also "DI water connection" (hose marked H2O pure). – High iron content can leave a rust film on stainless steel items and on the machine itself. If the chloride content in the domestic water supply is above 100mg/l, the risk of corrosion damage to items is greatly increased. – Acceptable water pressure (flow rate): 10 - 147 psi – Recommended water pressure (flow rate): 25 - 60 psi – If the water pressure is not in the range listed above, the "Fill/Drain" indicator may come on and the fault code "F..E" will appear in the display. If this occurs, contact the Miele Technical Service Department for advice.–All inlet hoses are approx. 5ft (1.5 m) long, ending with a 3/4" GHT female thread. Under no circumstances should the inlet filters be removed. –Water valves with 3/4" thread are to be provided on site. They should be easily accessible because the water supply will need to be turned off when not in use. –Large surface area filters are supplied with the machine for installation between the inlet and the inlet hose (see illustration under "How to clean the water inlet filters"). The large surface area filter for distilled water is made of stainless steel and can be recognized by the matte surface. ,NEVERattempt to shorten or otherwise compromise the inlet hose (see illustration). See the supplied installation diagram. Plumbing 64
DI water connection The machine is equipped for optional connection to a DI water supply. When the water pressure is below the recommended pressure, the water intake time is automatically extended. Acceptable water pressure (flow rate): 10 - 147 psi Recommended water pressure (flow rate): 25-60psi ^Connect the DI hose (pressure-tested and marked with "H 2O pure") with 3/4" GHT male thread to the on-site water valve for distilled water. If the DI connection will not be used, the electronics may need to be reprogrammed. The water intake hose then remains at the rear side of the machine unconnected.The DI water connection should not be hooked up (or Y connected) to any water source, it should remain unconnected, unless DI water is available. The instructions below re-program the unit to draw cold tap water for the final rinse instead of the DI water. Access programming mode by pressing and holding the "Display" and "Start" buttons while simultaneously turning the machine on. ^Press "Drying" once and E01 (programming level) will appear in the display. ^Turn the program knob to the 11 o'clock position (will most likely display 110). ^110 = Set for DI water intake for final rinse. ^111 = Always cold water for final rinse instead of DI water. ^ Press "Display" until 111 appears in the display. ^ Press "Start" twice to save. Plumbing 65
Connect the water drain –The drainage system is fitted with a non-return valve which prevents wastewater from flowing back into the machine through the drain hose. –It is best to connect the appliance to aseparatedrainage system on site. The on site drain connection point must be located between 0.3 m (1ft) and 1 m (3ft)above the lower edge of the appliance. If it is lower than 0.3 m, lay the hose in a curve at a height of at least 0.3 m. The drainage system must be able to handle a minimum drainage flow of 16 l/min. – The drain hose is approx. 1.4 m long and flexible (internal diameter 22 mm). It must not be shortened. Use the supplied hose clamps to secure it into position. – Longer drain hoses (up to 4 m) can be ordered through Miele. – The drainage system must not exceed 4 m. See the supplied installation diagram. Plumbing 66
Height including lid Height without lid85 cm (33 7/16") 82 cm (32 5/16") Width 60 cm (23 5/8") Depth Depth with door open60 cm (23 5/8") 120 cm (47 1/4") Weight (net) 70 kg (154 lbs) Voltage / Fuse see data plate Power cord approx. 1.8 m (5' 11") Water temperature: Cold water DI water connectionmax. 20°C max. 70°C Static water pressure max. 145 psi (10 bar) Min. flow pressure: Cold water DI water connection10 - 147 psi (7 - 10 bar) 10 - 147 psi (7 - 10 bar) Recommended flow pressure: Cold water DI water connection025 - 60 psi (1.72 - 4.13 bar) 025 - 60 psi (1.72 - 4.13 bar) Drain height min. 0.3 m (1 ft), max. 1 m (3.3 ft) Room temperature 5°C (40°F) to 40°C (104°F) Relative humidity: maximum linear decreasing to80% for temperatures up to 31°C 50% for temperatures up to 40°C Altitude above sea level up to 1500 m (1640 yds) Contamination level P2 (as per IEC/EN 61010-1) Ingress protection rating (as per IEC 60529)IP20 (ingress of dust) Noise level in dB (A), peak LpA for cleaning and drying cycles
Choose your program according to the types of instruments or utensils being cleaned and the degree of soiling. Programs (fixed)Program Description WashWash program without disinfection. Rinse (cold)Rinse only. To rinse items before a wash program, to avoid excessive foam. DrainFor draining water out of the machine, e.g. when a program has been interrupted or the "Fill/Drain" indicator illuminates. (Turn the program selector to the "Stop"fposition first). Disinfection varioRecommended program for protein soils such as blood and tissue. For cleaning and thermal disinfection at 93C with 10 minutes holding time for stainless steel instruments. Disinfection 93°C-10’For thermal disinfection and cleaning of stainless steel instruments at 93C with 10 minutes holding time. Program guide 68
Program Sequence 1. Pre-wash 2. Cleaning and/or thermal disinfection and/or cleaning3. Interim rinse I with Neutralization4. Interim rinse II5. Final rinse 1) and/or disinfection6. Drying (extra program) XX Cleaning 60°C/3’XX Rinse 65°C/1’ (DI) DOS 2(X) X XX Cleaning 55°C/5’X DOS 3XX Rinse and Disinfection 93°C/10’ (DI) DOS 2(X) X Disinfection 93°C/10’X DOS 3X Rinse 75°C/3’ (DI) DOS 2(X) X = Sections included in a program (with temperature/temperature holding time) DOS 2 = For dispensing rinse aid DOS 3 = For dispensing neutralizer DI = Deionized water. Program guide 69