Page 371
A - 58 (cont’d) Parameter Name Setting Range Response (unit: s) 0.002 0.004 0.01 0.05 0.1 0.5 2.0 4.0 8.0 Auto Corrected spectra ON, OFF Shutter control ON, OFF Replicates 2 to 99 Cycle time 2 to 3600 s ON : Checked OFF : Unchecked G.5.2 3-D Time Scan: Monitor Tab Parameter Name Setting Range Y Axis: Max -119999999.9 to 120000000.0 Y Axis: Min -120000000.0 to 119999999.9 Contour interval 0.001 to 1000 Open data processing window after data acquisition ON,...
Page 372
A - 59 G.5.4 3D Time Scan: Report Tab Parameter Name Setting Range Output Print Report, Use Microsoft(R) Excel, Use print generator sheet Orientation Portrait, Landscape Print items: Include date ON, OFF Print items: Include instrument ON, OFF Print items: Include graph ON, OFF ON: Any of the following items are selectable. • Contour • Bird’s-eye view • Contour and Bird’s-eye view • Contour and Spectrum Print items: Data listing*1 ON, OFF ON: Setting can be made for time scan...
Page 373
A - 60 APPENDIX H SAMPLING INTERVAL H.1 Wavelength Scan The data sampling interval is set as follows according to the scan speed. [Data mode = Fluorescence/Luminescence] Scan Speed (nm/min) Sampling Interval (nm) 60000 2.0 30000 1.0 12000 0.2 2400 0.2 1200 0.2 240 0.2 60 0.2 30 0.2 [Data mode = Phosphorescence, Chopping speed = 40 Hz] Scan Speed (nm/min) Sampling Interval (nm) 240 0.2 60 0.2 30 0.2 [Data mode = Phosphorescence, Chopping speed = 20/10 Hz] Scan...
Page 374
A - 61 H.2 Time Scan The data sampling interval is set as follows according to the scan time. [Data mode = Fluorescence/Luminescence/Phosphorescence/ Phosphorescence Life Time] Sampling Interval [ms] Phosphorescence Mode/Chopping Speed Scan Time [s] Fluorescence/Luminescence/ Phosphorescence Life Time Mode 40 Hz 20 Hz 10 Hz 0.1 ~ 1 1 50 100 200 ~ 3 1 50 100 200 ~ 7 2 50 100 200 ~ 18 5 50 100 200 ~ 36 10 50 100 200 ~ 72 20 50 100 200 ~ 180 50 50 100...
Page 375
A - 62 H.3 3-D Time Scan The data sampling interval is set as follows according to the scan speed. [Data mode = Fluorescence/Luminescence] Scan Speed (nm/min) Sampling Interval (nm) 60000 2.0 30000 1.0 12000 1.0 2400 1.0 1200 1.0 240 1.0 60 1.0 30 1.0 [Data mode = Phosphorescence, Chopping speed = 40 Hz] Scan Speed (nm/min) Sampling Interval (nm) 240 1.0 60 1.0 30 1.0 [Data mode = Phosphorescence, Chopping speed = 20/10 Hz] Scan Speed (nm/min) Sampling Interval (nm)...
Page 376
A - 63 APPENDIX I PARAMETER SETTING RANGE ON OPTIONS WINDOW I.1 Display Tab Parameter Name Setting Range Show trace display ON, OFF Show gridlines with increment ON, OFF ON: Gridline width is selectable from Wide, Middle and Narrow. Display items: Wavelength scan Show graph only, Show graph plus peak table Display items: Time scan Show graph only, Show graph plus kinetic data, Show graph plus peak table Display items: 3-D Scan Show contour plus EX, EM spectrum,...
Page 377
A - 64 I.2 Processing Tab Parameter Name Setting Range Integration method Rectangular, Trapezoid, Romberg Calibration formula Data = f(Conc), Conc = f(Data) Data output to peak table ON, OFF Smoothing: Processing Savitsky-Golay, Mean, Median Smoothing: Order 2 to 4 Smoothing: Number of points 5 to 101 Derivative: Smoothing order 2 to 4 Derivative: Smoothing order 5 to 101 I.3 Analysis Condition Tab Parameter Name Setting Range Integration time for current data 0.1 to...
Page 378
A - 65 APPENDIX J ERROR MESSAGE LIST Error Message Probable Cause Remedy ROM Error ROM error has occurred. Contact our local service office. RAM Error RAM error has occurred. Contact our local service office. Lamp is not lighted Lamp extinguished during initialization. Replace the lamp. Power interruption Momentary power failure detected at instrument. Check the connection of USB cable, and start up PC and spectrophotometer again. If error recurs, contact our local service...
Page 379
INDEX - 1 INDEX 1st order (linear calibration curve) ............................................................................................... 4-12, A-1 3-D scan ........................................................................................................................................ 1-2, 5-1 3-D time scan ....................................................................................................................... 1-2, 1-27, 6-1 3rd order (Cubic working curve)...
Page 380
INDEX - 2 Contour ...................................................................................................................... 1-2, 1-25, 5-19, 6-18 Contour interval .................................................................................................................................... 5-8 Corrected spectrum ............................................................................................. 1-28, 2-7, 5-7, 6-7, A-26 Cycle time...