Hitachi F7000 Instruction Manual
Here you can view all the pages of manual Hitachi F7000 Instruction Manual. The Hitachi manuals for Fluorescence Spectrophotometer are available online for free. You can easily download all the documents as PDF.
Page 331
A - 18 E.2 Advantages of Fluorometry As contrasted with fluorometry, absorptiometry for a low-concentration sample is explained in the following: A sample having 99% transmittance to blank is taken as an example. In the absorbance measurement of such a substance, inaccuracies must always be taken into consideration. Here, the inaccuracy is assumed to be 0.1%. Since it has an effect on both the blank and sample, Percent transmittance of blank 100.0 ± 0.1% Percent transmittance of...
Page 332
A - 19 Still more, since a fluorescence wavelength of a substance is different from its excitation wavelength (incident light wavelength), the fluorescence wavelength is not readily affected by the exciting radiation, thereby contributing to ensuring high sensitivity. Fig. E-2 Comparison between Absorptiometry and Fluorometry In addition to high sensitivity, the fluorometry is advantageous in that more information is attainable. An emission spectrum is also available besides an...
Page 333
A - 20 Figure E-3 shows a simplified spectral graph of measurement of a sample containing multiple components. In absorptiometry, since only the absorption spectrum can be measured, two or more component wavelengths are presented. If the absorption wavelengths are similar to each other, each component cannot be separated in measurement. In fluorometry, even if the absorption wavelengths are similar, a difference in fluorescence makes it possible to select each fluorescence wavelength...
Page 334
A - 21 E.3 Remarks on Measurement in Fluorescence Analysis For most kinds of samples, an increase of 1 °C in sample temperature causes the fluorescence intensity to decrease by 1 to 2%. It is also reported that for some kinds of biochemical samples, the fluorescence intensity decreases as much as 10% as the temperature increases by 1 °C. When analyzing a sample having a temperature- dependent property, it is advisable to use the constant-temperature cell holder (P/N 650-0150)....
Page 335
A - 22 Fig. E-4 Raman Spectrum of Water Table E-2 Raman Peak Positions at Respective Excitation Wavelengths (excitation wavelength) Water Ethanol CyclohexaneCarbon Tetrachloride Chloroform 248 271 267 267 ⎯ ⎯ 313 350 344 344 320 346 365 416 405 408 375 410 405 469 459 458 418 461 Excitation wavelength and Raman peak position (nm) 436 511 500 499 450 502 Relative intensity Excitation wavelength Raman scattering
Page 336
A - 23 In measurement of a high-concentration sample, a variety of error factors may be involved. The most significant error factor consists in that an excitation beam is absorbed at the entrance of a cell to prevent a sufficient level of excitation at the center of the cell. Figure E-5 illustrates an extreme case of this condition. Although fluorescence is emitted in the vicinity of the entrance for the excitation beam, it is not taken into the emission monochromator. Fig....
Page 337
A - 24 In any case, if there is a possibility of a measurement error due to high concentration of a sample, dilute the sample properly or carry out surface fluorescence measurement using a solid sample holder. Where the excitation and emission wavelengths are plotted near each other, care should be exercised not to mistake the Raman and Rayleigh scattering for the fluorescence spectrum as mentioned in E.3.3. Where the excitation and emission wavelengths are plotted apart from each other,...
Page 338
A - 25 Figure E-7 shows a measurement example of fluorescence spectrum. ① Scattering of exciting radiation ② Raman spectrum of solvent ③ Fluorescence of impurities, solvent, etc. ④ Fluorescence of sample ⑤ Second-order spectrum of exciting radiation Fig. E-7 Measurement Example of Fluorescence Spectrum As shown in Fig. E-7, other peaks than a fluorescence peak of sample appear in measurement of fluorescence spectrum. With reference to this example, it is necessary to...
Page 339
A - 26 APPENDIX F MEASUREMENT OF INSTRUMENTAL RESPONSE (CORRECTED SPECTRA) Spectrum correction is performed to enable measuring a true spectrum by eliminating instrumental response such as wavelength characteristics of the monochromator or detector (photomultiplier). The measurement of instrumental response is needed to perform spectrum correction. “Instrumental Response” is the function to measure and save the instrumental response. F.1 Measurement of Instrumental Response on...
Page 340
A - 27 Fig. F-1 Handling of Rhodamine B (1) Click the (analysis method) button on the Measurement toolbar. A box for setting your analysis method will appear. (2) Select the General tab. On the General tab page, specify “Wavelength scan” for the measurement mode. (3) Select the Instrument tab. (4) Set “Fluorescence” for the data mode, “400 V” for the photomultiplier voltage and “Excitation” for the scan mode. (5) Execute the Zero Adjust command from the...
All Hitachi manuals