(7)  Set the diffuser into the cell holder in the sample compartment.   
(Refer to Fig. F-12.) 
 
 
 
Fig. F-12 
 
 
(8)  Execute the Instrumental Response command from the Utility 
menu.    The Instrumental Response dialog box will open as 
shown in Fig. F-13. 
Select “EX [500 to 900 nm].”    (Refer to ①  in Fig. F-13.) 
 
 
 
Fig. F-13 
 
 Diffuse
r 
1 
2 

Page 352

A - 39 
(9) Click the Run button.  (Refer to ② in Fig. F-13.)  A message 
box “Set diffuser to cell holder.” will appear.    Click the OK button.   
(Refer to ③  in Fig. F-14.) 
 
 
 
Fig. F-14 
 
 
(10)  The EX and EM wavelengths are set at 498.6 nm and the 
instrument starts synchronous wavelength scan (from 498.6 to 
901.4 nm). 
At this stage, the following measuring conditions are automatically 
set. 
• Scan mode :  Emission 
• Data mode :  Luminescence 
•  EX slit  :  5.0 nm 
•  EM slit  :  5.0 nm...

Page 353

A - 40 
F.5    Resetting of Instrumental Response 
 
(1) Click the Reset button in the Instrumental Response dialog box.   
(Refer to ①  in Fig. F-15.) 
 
 
 
Fig. F-15 
 
 
(2)  A message box “Erase instrumental response on 
excitation/emission side. OK?” will appear.    Click the OK button.   
(Refer to ②  in Fig. F-16.) 
 
 
 
Fig. F-16 
 
 
(3)  Now, the instrumental response has been erased. 
 
1
2 

Page 354

A - 41 
APPENDIX G  PARAMETER SETTING RANGE 
FOR ANALYSIS METHOD 
 
G.1  General Tab:  Common to all measurement 
modes 
 
Parameter Name  Setting Range 
Measurement  Wavelength scan, Time scan, Photometry, 3-D scan,   
3-D Time scan 
Operator  Up to 20 characters 
Comments  Up to 255 characters 
Use sample table  ON, OFF 
 
ON :  Checked 
OFF :  Unchecked 
 
 
G.1.1  Wavelength Scan:  Instrument Tab 
 
Parameter Name  Setting Range 
Scan mode  Excitation, Emission, Synchronous 
Data mode  Fluorescence,...

Page 355

A - 42 
(cont’d) 
Parameter Name  Setting Range 
Scan speed   
(unit: nm/min) 60000 
30000 
12000 
2400 
1200 
240 
60 
30 
When Synchronous is specified for scan mode, 1200 nm/min or less 
is indicated. 
When Phosphorescence is specified for data mode, select from the 
following.   
Chopping speed is in parentheses. 
F-7000 (40 Hz)  F-7000 (20 Hz/10 Hz) 
240 
60 
30 60 
30 
  
Delay  0 to 9999 s 
EX Slit (unit: nm)  1.0 
2.5 
5.0 
10.0 
20.0 
EM Slit (unit: nm)  1.0 
2.5 
5.0 
10.0 
20.0 
PMT Voltage...

Page 356

A - 43 
G.1.2  Wavelength Scan:  Monitor Tab 
 
Parameter Name  Setting Range 
Y Axis: Max  -119999999.9 to 120000000.0 
Y Axis: Min  -120000000.0 to 119999999.9 
Open data processing window after data acquisition  ON, OFF 
Open data processing window after acquisition 
(Overlay) ON, OFF 
Print report after data acquisition  ON, OFF 
Overlay ON, OFF 
 
 
G.1.3  Wavelength Scan:  Processing Tab 
 
Parameter Name  Setting Range 
CAT ON, OFF 
Processing choices  Savitsky-Golay, Smooth, Mean Smooth,...

Page 357

A - 44 
G.1.4  Wavelength Scan:  Report Tab 
 
Parameter Name  Setting Range 
Output  Print Report, Use Microsoft(R) Excel,   
Use print generator sheet 
Orientation Portrait, Landscape 
Print items: Include date  ON, OFF 
Print items: Include method  ON, OFF 
Print items: Include graph  ON, OFF 
Print items: Include data listing  ON, OFF 
ON:    “Constant” or “Select data” is selectable. 
When “Constant” is specified: 
Parameter Setting Range 
Data interval  Minimum sampling interval to 
measuring...

Page 358

A - 45 
G.2.1  Time Scan:  Instrument Tab 
 
Parameter Name  Setting Range 
Data mode  Fluorescence, Luminescence, Phosphorescence,   
Phosphorescence Life Time,   
Phosphorescence Life Time (short)   
Chopping speed  10 Hz, 20 Hz, 40 Hz 
EX WL  Input Range:    0, 200 to 900 nm 
EM WL  Input Range:    0, 200 to 900 nm 
Time unit  s, ms 
Scan time  
Time Unit  Input Range 
s  10 to 9000 s 
ms  100 to 9999 ms 
 
When Phosphorescence Life Time (short) is specified for data 
mode: 
Chopping Speed  Scan Time...

Page 359

A - 46 
G.2.2  Time Scan:  Monitor Tab 
 
Parameter Name  Setting Range 
Y Axis: Max  -119999999.9 to 120000000.0 
Y Axis: Min  -120000000.0 to 119999999.9 
X Axis: Max (unit: s)  0.1 to 9999.0 
X Axis: Min (unit: s)  0.0 to 9998.9 
Open data processing window after 
data acquisition ON, OFF 
Open data processing window after 
acquisition (Overlay) ON, OFF 
Print report after data acquisition  ON, OFF 
Overlay ON, OFF 
 
 
G.2.3  Time Scan:  Processing Tab 
 
Parameter Name  Setting Range 
Processing...

Page 360

A - 47 
G.2.4  Time Scan:  Report Tab 
 
Parameter Name  Setting Range 
Output  Print Report, Use Microsoft(R) Excel,   
Use print generator sheet 
Orientation Portrait, Landscape 
Print items: Include date  ON, OFF 
Print items: Include method  ON, OFF 
Print items: Include graph  ON, OFF 
Print items: Include data listing  ON, OFF 
ON: “Constant” or “Select data” is selectable. 
When “Constant” is specified: 
Parameter Setting Range 
Data interval  Minimum sampling interval to 
measuring scan time...