Page 841
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 10 - Care & Maintenance 10 - 23 10-7-9-4 Data Sheet for Transducer Source Leakage Current The test passes when all readings measure less than the values shown in Table 10-13 and Table 10-14 . Record all data on the PM Inspection Certificate. . CAUTIONEquipment damage possibility. Never switch the Polarity and the status of Neutral when the unit is powered ON. Be sure to turn the unit power OFF before switching them...
Page 842
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 10 - 24 Section 10-8 - When Theres Too Much Leakage Current... Section 10-8 When Theres Too Much Leakage Current... CHASSIS FAILS Check the ground on the power cord and plug for cont inuity. Ensure the ground is not broken, frayed, or intermittent. Replac e any defective part. Tighten all grounds. Ensure star washers are under all ground studs. Inspect wiring for bad crimps, poor connections, or damage. Test the wall outlet;...
Page 843
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 10 - Care & Maintenance 10 - 25 ULTRASOUND INSPECTION CERTIFICATE * Scan Format: Phased Array, Linear Array, Curved Array, Mechanical Array or Other FUNCTIONAL CHECKS PHYSICAL IN SPECTION AND CLEANING COMMENTS: ________________________________ ______________________________________________ ________________________________ ______________________________________________ ________________________________...
Page 844
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 10 - 26 Section 10-8 - When Theres Too Much Leakage Current... ELECTRICAL SAFETY Final Check. All system covers are in plac e. System scans with all probes as expected. Accepted by: ______________________________________________________________________ Electrical Test Performed Max Value Allowed Value Measured OK? Comments Outlet (correct ground &wiring config.) System Ground Continuity Chassis Source Leakage Current - Probe...
Page 845
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 10 - Care & Maintenance 10 - 27 Section 10-9 Site Log Table 10-22 Site Log DATESERVICE PERSONPROBLEMCOMMENTS
Page 846
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 10 - 28 Section 10-9 - Site Log Table 10-22 Site Log (cont’d) DATESERVICE PERSONPROBLEMCOMMENTS
Page 847
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 10 - Care & Maintenance 10 - 29 Table 10-22 Site Log (cont’d) DATESERVICE PERSONPROBLEMCOMMENTS
Page 848
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 10 - 30 Section 10-9 - Site Log This page was intentionally left blank.
Page 849
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Index Index - 1 Numerics 2D Mode Checks 4-32 4D Volume Rendering Test 4-55 A Abbrevations Definitions 9-3 AC Power Replacement Procedure 8-82 AC Power Parts 9-61 Accessory Boxes, Vivid 7 / Vivid 7 PRO 9-105 Acclimate Time 2-2, 3-2 Acoustic Noise Output 3-14 Add Printer 3-35 Adjust LCDs Contrast and Backlight Intensity 3-34 Advanced Settings 3-51 AE Title 3-126 Air Filter Cleaning 10-10 Application Turnover Check List 4-82...
Page 850
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Index - 2 - Connect Pulse Pressure Transducer 3-20 Connect USB Flash Card 3-19 Connection to a DICOM Server in a Network 3-58 Connections on the External IO 3-19 Connections on the Patient I/O Panel 3-20 Connectivity 5-148 Setup 3-42, 3-89 , 3-109 Connectivity Overview 3-41 Console Environmental Requirements 2-2 Mechanical Checks 4-78 Console Lock 9-22 Vivid 7 PRO 9-24 VIVID 7/VIVID 7 Dimension 9-23 Contrast adjustments on...