Page 821
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 10 - Care & Maintenance 10 - 3 Section 10-3 Maintenance Task Schedule 10-3-1 How often should maintenance tasks be performed? The Customer Care Schedule (provided on page 10-3) specifies how often your Vivid 7 should be serviced and outlines items requiring special attention. NOTE: It is the customer’s responsi bility to ensure the Vivid 7 care & maintenance is performed as scheduled in order to retain its high level...
Page 822
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 10 - 4 Section 10-3 - Maintenance Task Schedule Endocavity Probe Leakage Current Checks• also after corrective maintenance or as required by your facilities QA program Transesphongeal Probe Leakage Current Checks • also after corrective maintenance or as required by your facilities QA program Surgical Probe Leakage Current Checks • also after corrective maintenance or as required by your facilities QA program Functional...
Page 823
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 10 - Care & Maintenance 10 - 5 Section 10-4 Tools Required 10-4-1 Special Tools, S upplies and Equipment 10-4-1-1 Specific Requirements for Periodic Maintenance Table 10-3 Overview of Requirements for Periodic Maintenance Tool Part NumberComments Digital Volt Meter (DVM) Anti Static Kit 46–194427P231 46–194427P279 46–194427P369 46–194427P373 46–194427P370Kit includes anti–static mat, wrist strap and cables for 200 to...
Page 824
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 10 - 6 Section 10-5 - System Maintenance Section 10-5 System Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary. Table 10-4 System Preliminary Checks StepItemDescription 1 Ask & Listen Ask the customer if they have any problems or questions about the equipment. 2 Paperwork Fill in the top of the Ultrasound Inspection...
Page 825
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 10 - Care & Maintenance 10 - 7 10-5-2 Functional Checks (S ee Also Chapter 4) The functional checks take about 60 minutes to pe rform. Refer to the system user documentation whenever necessary. 10-5-2-1 System Checks NOTE: * Some software may be considered standa rd depending upon system model configuration. Table 10-5 System Functional Checks CheckStepDescription 2D-Mode (B-Mode) Verify basic B-Mode (2D) operation....
Page 826
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 10 - 8 Section 10-5 - System Maintenance 10-5-2-2 Peripheral/Option Checks If any peripherals or options are not part of t he system configuration, the check can be omitted. 10-5-3 Input Power 10-5-3-1 Mains Cable Inspection Table 10-6 GE Approved Peripheral/Hardware Option Functional Checks StepItemDescription 1DVR/VCR Verify record/playback capabilities of the uni t: Clean heads (VCR ONLY) and covers if necessary. 2 B/W...
Page 827
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 10 - Care & Maintenance 10 - 9 10-5-4 Cleaning 10-5-4-1 General Cleaning Table 10-8 General Cleaning StepItemDescription 1 Console Use a fluid detergent in warm water on a soft, damp cl oth to carefully wipe the entire system. Be careful not to get the cloth too wet so that moisture does not enter the console. 2 Probe Holder Clean probe holders. (they may need to be soaked to remove excess gel). 3 Op Panel Only use...
Page 828
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 10 - 10 Section 10-5 - System Maintenance 10-5-4-2 Air Filter Cleaning NOTE: For your convenience or if the air filter is to o dirty, replacement filters are available. Refer to Chapter 9 for the air filter replacement part number. 10-5-5 Physical Inspection Table 10-9 Air Filter Cleaning - frequency varies with your environment StepItemDescription 1Remove Filter Cover Refer to Section 8-5 Air Filter Replacement Procedure...
Page 829
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 10 - Care & Maintenance 10 - 11 10-5-6 Optional Diagnostic Checks Optionally you can access the diagnostic software as described in Chapters 5 or 7. View the error logs and run desired diagnostics. 10-5-6-1 View the Logs 1.) Review the system error log for any problems. 2.) Check the temperature log to see if there are any trends that could cause problems in the future. 10-5-7 Probe Maintenance 10-5-7-1 Probe Related...
Page 830
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 10 - 12 Section 10-6 - Using a Phantom Section 10-6 Using a Phantom See section 4 for information on using a phantom ( 4-3-5 Performance Tests on page 4-31 ) and quality assurance tests. Section 10-7 Electrical Safety Tests 10-7-1 Safety Test Overview The electrical safety tests in this section are ba sed on and conform to NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards. They ar e intended for the...