GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 10 - Care & Maintenance 10 - 13 10-7-2GEHC Leakage Current Limits The following limits are summarized for NFPA 99/UL2601-1/UL60601-1 (for USA) and IEC 60601-1 Medical Equipment Safety Standards. These limits are GEHC standards and in some cases are lower than the above standards listed. NOTE: *Mains Applied refers to the sink leakage test where mains (supply) voltage is applied to the part to determine the am ount...
Page 832
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 10 - 14 Section 10-6 - Using a Phantom 10-7-3 Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated. Any problems found should be reported to the hospital immediately and the receptacle should not be used. NOTE: No outlet tester can detect the condition where the...
Page 833
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 10 - Care & Maintenance 10 - 15 10-7-4-1 Meter Procedure Follow these steps to test the ground wire resistance. 1.) Turn the Vivid 7 unit OFF. 2.) Plug the unit into the meter, and the meter into the tested AC wall outlet. 3.) Plug the black chassis cable in to the meters “CHASSIS” connecto r and attach the black chassis cable clamp to an exposed metal part of the Vivid 7 unit. 4.) Set the meters “FUNCTION” swit ch to the...
Page 834
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 10 - 16 Section 10-6 - Using a Phantom 10-7-5 Chassis Leakage Current Test 10-7-5-1 Definition This test measures the current t hat would flow in a grounded person who touched accessible metal parts of the bedside station if the ground wire shoul d break. The test verifies the isolation of the power line from the chassis. The meter is connected from accessible metal parts of the case to ground. Measurements should be made wi th...
Page 835
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 10 - Care & Maintenance 10 - 17 10-7-5-2 Generic Procedure (cont’d) . 8.) Keep a record of the results with other hard copies of PM data kept on site. 10-7-5-3 Data Sheet for Chassis Source Leakage Current The test passes when all readings measure less than the value shown in Table 10-12. Record all data on the PM Inspection Certificate. 10-7-6Isolated Patient Lead (Source) Leakage–Lead to Ground 10-7-6-1 Definition...
Page 836
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 10 - 18 Section 10-6 - Using a Phantom 10-7-6-2 Generic Procedure Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with the unit Off and On. For each combination, the operating controls such as the lead switch should be operated to find the worst case condition. 10-7-7 Isolated Patient Lead (Source) Leakage–Lead to Lead Reference the procedure in the IEC 60601-1....
Page 837
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 10 - Care & Maintenance 10 - 19 10-7-8 Isolated Patient Lead (S ink) Leakage-Isolation Test Reference the procedure in the IEC 60601-1. When using the Dale 600, switch the meter’s function selector to the LEAD-ISO. Select the ALL position on the lead selector. Depress the rocker switch to ISO TEST to test lead isolation. NOTE: It is not necessary to test each lead individua lly or power condition combinations as required...
Page 838
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 10 - 20 Section 10-6 - Using a Phantom 10-7-8-1 Data Sheet for ECG Leakage Current The test passes when all readings measure less than the value shown in the table below. Record all data on the PM Inspection Certificate. Table 10-18 Maximum Allowance Limit for ECG Leakage Current AC Power Source Maximum Allowance Limit GROUND OPEN GROUND CLOSED Patient Lead to Ground Leakage Current Test and Patient Lead to Lead Leakage Current...
Page 839
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 10 - Care & Maintenance 10 - 21 10-7-9 Probe Leakage Current Test 10-7-9-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface. 10-7-9-2 Generic Procedure Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with the unit Off...
Page 840
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 10 - 22 Section 10-6 - Using a Phantom 10-7-9-3 No Meter Probe Adapter Procedure Follow these steps to test each transducer for leakage current. 1.) Turn the Vivid 7 unit OFF. 2.) Plug the unit into the test meter, and the meter into the tested AC wall outlet. 3.) Plug the external probe into the meters “EXTERNAL” connector. 4.) Set the meters “FUNCTION” switch to EXTERNAL position. 5.) Connect the probe for test with the...