Page 41

GE Energy
GE I-210+cElectronic Meter
Product Description and Operation, Maintenance,and Upgrade Instructions
GEH-7101-2.00-141
The default unit for the calibration pulses is watt hours. The meter may be switched to varh
calibration pulses using MeterMate™software.
4.1.1.2 LCD Display
The meter display has annunciators for quadrant, phase voltage, and energy flow indication as
shown inFigure 13. The annunciators provide valuable information during the testing process.
Figure 13 Liquid Crystal Display...

Page 42

GE I-210+cElectronic Meter
Product Description and Operation, Maintenance,and Upgrade Instructions
GE Energy
42GEH-7101-2.00-1
Load profile recording is suspended, and load profile interval status bits reflect test mode
was in effect during affected interval(s).
When test mode begins, the test annunciator is lit and the first item programmed for the test
display is shown.
4.2.2 Operation During Test Mode
The same summation quantities that are calculated during normal operation are calculated in test...

Page 43

GE Energy
GE I-210+cElectronic Meter
Product Description and Operation, Maintenance,and Upgrade Instructions
GEH-7101-2.00-143
4.2.2.2 Displays Available in Test Mode Only
The following display items are available for display only in the test mode:
Test mode maximum demands
Time remaining in test demand interval in block demand only
Time remaining in test demand subinterval in rolling demand only
Test mode summations
Test mode accumulating demand
4.2.2.3 Test Reset
A test reset is initiated by...

Page 44

GE I-210+cElectronic Meter
Product Description and Operation, Maintenance,and Upgrade Instructions
GE Energy
44GEH-7101-2.00-1
4.3 Field Accuracy Test
Test mode allows the meter to be tested for accuracy in the field without disturbing any billing
data.
4.3.1 Field Testing with Test Mode
Testing the meter in the field can be accomplished three ways in test mode by using the
following:
1. Maximum demand reading in the display
2. Disk analog display
3. Instantaneous demand feature of the I-210+c™
4.3.2...

Page 45

GE Energy
GE I-210+cElectronic Meter
Product Description and Operation, Maintenance,and Upgrade Instructions
GEH-7101-2.00-145
The meter’s magnitude indicator consists of three segments used to simulate the rotation of the
disk on an electromechanical meter. Each state change represents Kt. One complete “revolution”
represents 10 state changes, which represents Kh watt hours or varhours. The meter illuminates
the segments as follows:
If the Wh direction is delivered (or the varh direction is lagging),...

Page 46

GE I-210+cElectronic Meter
Product Description and Operation, Maintenance,and Upgrade Instructions
GE Energy
46GEH-7101-2.00-1
4.5 Shop Test
Shop testing consists of verifying the meter’s accuracy.
4.5.1 Meter Shop Equipment
The meter testing equipment must be capable of maintaining accuracy while supplying energy to
the meter’s switching power supply. Otherwise, meters may be tested in any shop that meets the
requirements outlined in the current editions of theHandbook for Electricity...

Page 47

GE Energy
GE I-210+cElectronic Meter
Product Description and Operation, Maintenance,and Upgrade Instructions
GEH-7101-2.00-147
1. Use MeterMate™software to put the meter calibration LED into var hour pulse output mode.
2. Set up the meter for testing as described above in the watt hour test procedure. The test pulse
value is now Kt var hours per pulse.
3. Begin testing according to your standard var hour test procedures. Allow 15 seconds of
settling time after applying voltage before making accuracy...

Page 48

GE I-210+cElectronic Meter
Product Description and Operation, Maintenance,and Upgrade Instructions
GE Energy
48GEH-7101-2.00-1
6. Pull the battery holder out of the slot at the top of the nameplate carrier.
7. Remove the old battery.
8. Place the new battery in the battery holder.
9. Slip the battery holder into the slot at the top of the nameplate carrier and press down until
the battery holder clicks into place.
10. Connect the battery wire to the nameplate carrier battery connector.
11. Replace the...

Page 49

GE Energy
GE I-210+cElectronic Meter
Product Description and Operation, Maintenance,and Upgrade Instructions
GEH-7101-2.00-149
4.10 Cleaning
Warning:Care must be taken during cleaning not to damage or contaminate any
gold-plated contacts of the connectors.
Caution:Do not immerse the meter in any liquid. Do not use abrasive cleaners on
the Lexan™covers. Do not use chlorinated hydrocarbon or ketone solvents on the covers.
4.11 Storage
The I-210+c™meter is a durable device; however, it should be handled...

Page 50

GE I-210+cElectronic Meter
Product Description and Operation, Maintenance,and Upgrade Instructions
GE Energy
50GEH-7101-2.00-1
Table 5 Error Code Display
Error DisplayProbable CauseRemedy
Er 000 002 Power outage occurred, and:
a. Battery disconnected
b. Battery defective
Set meter’s date and time, read meter’s load profile
data and restart load profile recording (if applicable),
and:
a. Connect battery.
b. Replace battery.
Er 000 020 Hardware failurea. Check for proper installation.
b. Check for proper...