Altek TechChek Model 830KP Manual
Here you can view all the pages of manual Altek TechChek Model 830KP Manual. The Altek manuals for Calibrator are available online for free. You can easily download all the documents as PDF.
Page 11
19 READ FUNCTIONSSelect read by pressing the DISPLAY/SOURCE/READ pushbutton until the word READ appears on the LCD display. The READ functions measure the desired signal. Multiple scales are available for some functions. MIN/MAXTo read the Maximum or Minimum INPUT since READ mode was entered, simply press the MAX or MIN push-button. The value will appear on the LCD along with the word MAX or MIN. The MAX/MIN values are automatically updated and may be viewed at any time without interrupting the other...
Page 12
21 READ DC and AC VOLTAGE OUTPUTS mV, V, Vhi and Vac Choose this function to measure from 0.00 to 110.00 millivolts (mV), 0.00 to 10.25 DC Volts (V). Use the high voltage connection to read from 0.0 to 200.0 VDC (Vhi). See the next page for Volts AC. 1) Press the MILLIAMP/V push-button until mV, V, Vhi or VAC appear on the display 3) Press the DISPLAY/SOURCE/READ push-button until READ appears on the display 4) Press the TYPE/ENG UNITS push-button to select mV, V, Vhi or VAC on the display 5) Connect the...
Page 13
23 22 MEASURE THERMOCOUPLE SENSORSChoose this function to read a thermocouple. The input of the TechChek 830-KP is automatically cold junction compensated. 1) Disconnect the thermocouple from any instrument. 2) Press the T/C/RTD/Ωpushbutton until the word TYPE & any T/C appear on the display. 3) Press the DISPLAY/SOURCE/READ push-button until READ and any T/C TYPE appear on the display 4) Use the proper thermocouple wire and corresponding miniature T/C connector to connect TechChek 830-KP to the...
Page 14
25 CHECK CONTINUITYChoose this function to check continuity. A tone will sound and a sound symbol will appear on the display when the resistance between the leads is less than 100 Ohms. 1) Plug the leads into the TechChek 830-KP as shown below. 2) Press the T/C/RTD/Ωpushbutton until the Ωsymbol appears on the display. 3) Press the DISPLAY/SOURCE/READ push-button until READ or CONTINUITY appears on the display 4) Press the TYPE/ENG UNITS pushbuuton once to switch between READ and CONTINUITY on the...
Page 15
27 BLACK BLACK RED3 2 1RTD Element OPERATING INSTRUCTIONS MEASURE RTD SENSORS 26 MEASURE RTD SENSORSChoose this function to read an RTD. Three wires must be use for both 2 and three wire RTDs. 1) Disconnect the RTD from the instrument being calibrated. 2) Press the T/C/RTD/Ωpushbutton until any RTD type (Pt, Ni or Cu) appears on the display. 3) Press the DISPLAY/SOURCE/READ push-button until SOURCE and any RTD TYPE appear on the display 4) Connect using 3 wires as in the diagrams on the opposite page....
Page 16
29 Frequency Output Signal Flowmeter Flow Sensor Variable Speed Drive Controller Transmitter Frequency Output Signal Flow Meter Line Frequency Variable Speed Drives Connection for signals with amplitudes below 10.25V RMS Connection for signals with amplitudes to 250 V RMS COUNT FREQUENCIES OPERATING INSTRUCTIONS 28 COUNT FREQUENCIESChoose this function to use the 830-KP as a frequency counter. Available ranges are from 0.01 to 15.00 kHz, 1 to 1500 Hz and from 1 to 1500 CPM (Counts-Per-Minute). To measure...
Page 17
31 +IN- REF Typical 2-Wire Transmitter OPERATING INSTRUCTIONS CALIBRATE 2-WIRE TRANSMITTERS (T/C / mA Shown) CALIBRATE 4-WIRE TRANSMITTERS +IN- REF Typical 4-Wire Transmitter Loop Power Supply 24 VDC 30 OPERATING INSTRUCTIONS CALIBRATE TRANSMITTERS ANY SOURCE FUNCTION AND READ mA, READ %, READ DP% Choose this function to supply the input signal to the transmitter and displaying the 4-20 mA output of the transmitter (used with 4- Wire Transmitters). ANY SOURCE FUNCTION AND P-XMTR mA, P-XMTR %, P-XMTR...
Page 18
33 Pressure Source PROCESS CALIBRATOR QUIK-CHEK READ PRESSUREChoose this function to measure pressure with a ModPak 91 module holder and one of the many QuikCal 90 pressure modules. 1) Select the proper module for the pressure range of the process 2) Place the QuikCal 90 Module in the ModPak 91 holder 3) Plug the ModPak cable into the mating connector on the 830-KP 4) Turn the selector knob to PRESSURE 5) The 830-KP will display – – – – – for up to 20 seconds while the module is powered up and verified...
Page 19
35 PROCESS CALIBRATOR QUIK-CHEK 34 OPERATING INSTRUCTIONS CALIBRATE PRESSURE TRANSMITTERS READ PRESSURE AND P-XMTR mA, P-XMTR %, P-XMTR DP% Choose this function to simultaneously supply power to a 2-Wire pressure transmitter while supplying the input signal to the transmitter and displaying the 4-20 mA output of the transmitter. 1) Disconnect both the pressure input and output wires from the 2-Wire Transmitter to be calibrated 2) Select the proper module for the pressure range of the process 3) Place...
Page 20
37 SPECIFICATIONS POWER & MEASURE 2-WIRE TRANSMITTERS RANGES & ACCURACY:Same as for MILLIAMP SOURCEOUTPUT CURRENT: up to 24.00 mATYPICAL DRIVE CAPABILITY:1200 Ohms @ 20.00 mACOMPLIANCE VOLTAGE:nominal 25 VDC @ 20 mACOMMON MODE ERROR:0.01% Full Scale/Common Mode Volt 2-WIRE TRANSMITTER SIMULATOR RANGES: 1.00 to 24.00 mA; -18.8 to 125.0% of 4 to 20 mA; % DP FlowACCURACY:Same as for MILLIAMP SOURCELOOP VOLTAGE LIMITS:Mininum, 3 VDC; Maximum 45 VDCOVERLOAD PROTECTION:Current limited to 25 mA nominalCOMMON...