    							19
    READ FUNCTIONSSelect read by pressing the DISPLAY/SOURCE/READ
    pushbutton until  the word READ appears on the LCD display. The
    READ functions measure the desired signal. Multiple scales are
    available for some functions. 
    MIN/MAXTo read the Maximum or Minimum INPUT
    since READ mode was entered, simply
    press the MAX or MIN push-button. The
    value will appear on the LCD along with the
    word MAX or MIN. The MAX/MIN values
    are automatically updated and may be
    viewed at any time without interrupting the
    other values. 
    RESTARTING MIN/MAXPressing the STORE/RESET push-button
    will cause the 830-KP to store the present
    reading into the MAX and MIN memories.
    Upon releasing the STORE/RESET push-
    button the 830-KP will resume reading the
    input and update the MAX & MIN values as
    the measured signal changes.
    OUT OF RANGE SIGNALSSignals above or below those available for
    the currently selected range will be indicat-
    ed by OVER and UNDER on the display.
    OPERATING INSTRUCTIONS
    OVER
    UNDER
    18
    CALIBRATE FREQUENCY INPUTSChoose this function to provide pulses into frequency measuring
    instruments. The 830-KP output is a zero crossing square wave
    from -1V to +5V amplitude. Available ranges are from 0.01 to
    10.00 kHz, 1 to 1000 Hz and from 1 to 1000 CPM (Counts-Per-
    Minute).
    CPM is used to simulate extremely slow frequency signals with
    greater resolution. For example, 10 Hz is equivalent to 600 CPM.
    To convert from CPM to Hz Divide by 60. To convert from Hz to
    CPM multiply by 60.
    1) Disconnect one or both input wires from the device to be
    calibrated
    2) Press the FREQUENCY/PRESSURE pushbutton until any
    frequency signal (KHz, Hz or CPM) appears on the display.
    3) Press the DISPLAY/SOURCE/READ push-button until SOURCE
    and KHz, Hz or CPM appear on the display
    4) Press the TYPE/ENG UNITS pushbuuton to select between KHz,
    Hz or CPM on the display
    5) Connect the red SOURCE lead of the calibrator to the plus (+)
    input of the device and the black SOURCE lead to the minus (-).
    Output frequency is continuously adjustable with the UP/DOWN
    push-buttons. Zero & Span (or any other values) are available by
    using the LO and HI QUIK-CHEKs
    OPERATING INSTRUCTIONS
    Frequency Receiver Input
    Flowmeter
    Controller
    Transmitter
    Computer
    Logger
    DCS 
    						
    							21
    READ DC and AC VOLTAGE OUTPUTS
    mV, V, Vhi and Vac
    Choose this function to measure from 0.00 to 110.00 millivolts (mV),
    0.00 to 10.25 DC Volts (V). Use the high voltage connection to read
    from 0.0 to 200.0 VDC (Vhi). See the next page for Volts AC.
    1) Press the MILLIAMP/V push-button until mV, V, Vhi or VAC
    appear on the display
    3) Press the DISPLAY/SOURCE/READ push-button until READ
    appears on the display
    4) Press the TYPE/ENG UNITS push-button to select mV, V, Vhi or
    VAC on the display
    5) Connect the red READ (+) lead and the black READ (-) lead of
    the calibrator across the voltage to be measured.
    Signals above or below those available for the currently selected
    range will be indicated by OVER and UNDER on the display.
    OPERATING INSTRUCTIONS
    DC VoltageOutput Signal
    Controller
    Transmitter
    Power Supply
    VoltageOutput Signal
    Power Supply
    Loop Voltage
    Connection for millivolts(mV) and Volts below 10.25 VDC (V)
    Connection for DC Volts to 200.0 (vhi)
    20
    READ MILLIAMP OUTPUTS
    mA, mA % (Percent of 4 to 20 mA), DP% (DP Flow)
    Choose this function to measure from 0.00 to +24.00 milliamps. 
    1) Open the current loop at any convenient point along the signal
    path
    2) Press the MILLIAMP/V push-button until mA or mA and %
    appear on the display
    3) Press the DISPLAY/SOURCE/READ push-button until READ
    appears on the display
    4) Press the mA/%/% DP FLOW push-button to display mA, % 4-20
    or % DP Flow.
    5) Connect the red READ (+) lead of the calibrator to the more
    positive point of the break and the black READ lead (-) to the
    more negative
    Display the present reading, Maximum or Minimum by pressing
    the READ, MAX or MIN push-buttons. If the TechChek 830-KP is
    connected in the wrong polarity, the word POL will flash on the
    display. Simply reverse the leads for correct indication.
    OPERATING INSTRUCTIONS
    Milliamp Output Signal
    Controller
    Transmitter
    P/I
    DCS 
    						
    							23 22
    MEASURE  THERMOCOUPLE SENSORSChoose this function to read a thermocouple. The input of the
    TechChek 830-KP is automatically cold junction compensated.
    1) Disconnect the thermocouple from any instrument.
    2) Press the T/C/RTD/Ωpushbutton until the word TYPE & any
    T/C appear on the display.
    3) Press the DISPLAY/SOURCE/READ push-button until READ
    and any T/C TYPE appear on the display
    4) Use the proper thermocouple wire and corresponding miniature
    T/C connector to connect TechChek 830-KP to the thermocouple.
    To Change the Thermocouple type
    1) Press the TYPE/ENG UNITS push-button. The words SELECT
    T/C TYPE and the current selection appear on the display.
    2) Press any UP/DOWN push-button until the desired T/C type
    appears.
    3) Press the STORE/RESET push-button or wait 5 seconds to
    store the selection
    OPERATING INSTRUCTIONS
    +-
    READ AC VOLTAGESChoose this function to measure from 0.0 to 250.0 V True RMS.
    1) Press the MILLIAMP/V push-button until  mV, V, Vhi or VAC
    appear on the display
    3) Press the DISPLAY/SOURCE/READ push-button until READ
    appears on the display
    4) Press the TYPE/ENG UNITS pushbuuton to select VAC on the
    display
    5) Connect the red READ (+) lead and the black READ (-) lead of
    the calibrator across the voltage to be measured.
    Signals above or below those available for the currently selected
    range will be indicated by OVER and UNDER on the display.
    CAUTION: Care should be used when measuring AC voltage.
    The included safety test probes or safety alligator clips  should be
    used. Do not exceed voltage limits shown on calibrator.
    AC Voltage Signal
    Mains
    Line Voltage
    OPERATING INSTRUCTIONS 
    						
    							25
    CHECK CONTINUITYChoose this function to check continuity. A tone will sound and a
    sound symbol will appear on the display when the resistance
    between the leads is less than 100 Ohms.
    1) Plug the leads into the TechChek 830-KP as shown below.
    2) Press the T/C/RTD/Ωpushbutton until the Ωsymbol appears
    on the display.
    3) Press the DISPLAY/SOURCE/READ push-button until READ or
    CONTINUITY appears on the display
    4) Press the TYPE/ENG UNITS pushbuuton once to switch
    between READ and CONTINUITY on the display
    OPERATING INSTRUCTIONS
    24
    READ RESISTANCE
    Ohms
    Choose this function to measure resistance from 0.0 to 1000.0
    Ohms.
    1) Disconnect one or both input wires from the device to be
    calibrated.
    2) Press the T/C/RTD/Ωpushbutton until the Ωsymbol appears
    on the display.
    3) Press the DISPLAY/SOURCE/READ push-button until READ or
    CONTINUITY appears on the display
    4) Press the TYPE/ENG UNITS push-button once to switch
    between READ and CONTINUITY on the display
    5) Connect the red READ (+) lead and the black READ (-) lead of
    the calibrator across the resistance to be measured.
    Signals above or below those available for the currently select-
    ed range will be indicated by OVER and UNDER on the display.
    OPERATING INSTRUCTIONS
    Receiver InputOhms or 2-Wire RTD
    Controller
    Transmitter
    Computer 
    						
    							27
    BLACK
    BLACK
    RED3
    2
    1RTD Element
    OPERATING INSTRUCTIONS
    MEASURE RTD SENSORS
    26
    MEASURE RTD SENSORSChoose this function to read an RTD. Three wires must be use for
    both 2 and three wire RTDs.
    1) Disconnect the RTD from the instrument being calibrated.
    2) Press the T/C/RTD/Ωpushbutton until any RTD type (Pt, Ni or
    Cu) appears on the display.
    3) Press the DISPLAY/SOURCE/READ push-button until SOURCE
    and any RTD TYPE appear on the display
    4) Connect using 3 wires as in the diagrams on the opposite
    page. Use a 3 wire connection to read a 4 wire RTD (all four
    wires must be the same gauge). Spade lugs are recommend-
    ed to minimize any contact resistance.
    To Change the RTD type
    1) Press the TYPE/ENG UNITS push-button. The words SELECT
    RTD TYPE and the current selection appear on the display.
    2) Press any UP/DOWN push-button until the required type
    appears on the display.
    3) Press the STORE/RESET push-button or wait 5 seconds to
    store the selection.
    OPERATING INSTRUCTIONS 
    						
    							29
    Frequency Output Signal
    Flowmeter
    Flow Sensor
    Variable Speed Drive
    Controller
    Transmitter
    Frequency Output Signal
    Flow Meter
    Line Frequency
    Variable Speed Drives
    Connection for signals with amplitudes below 10.25V RMS
    Connection for signals with amplitudes to 250 V RMS
    COUNT FREQUENCIES
    OPERATING INSTRUCTIONS
    28
    COUNT FREQUENCIESChoose this function to use the 830-KP as a frequency counter.
    Available ranges are from 0.01 to 15.00 kHz, 1 to 1500 Hz and
    from 1 to 1500 CPM (Counts-Per-Minute).
    To measure waveforms with amplitudes between 1 V and 10.25 V
    RMS use the low level inputs. Use the high voltage connection to
    read waveforms with amplitudes from 10.25 to 250.0 V RMS
    1) Disconnect one or both input wires from the device to be
    calibrated
    2) Press the FREQUENCY/PRESSURE pushbutton until any
    frequency signal (KHz, Hz or CPM) appears on the display.
    3) Press the DISPLAY/SOURCE/READ push-button until READ
    and KHz, Hz or CPM appear on the display
    4) Press the TYPE/ENG UNITS push-button to select between KHz,
    Hz or CPM on the display
    5) Connect the red READ lead of the calibrator to the plus (+) input
    of the device and the black READ lead to the minus (-)
    OPERATING INSTRUCTIONS 
    						
    							31
    +IN-
    REF
    Typical
    2-Wire
    Transmitter
    OPERATING INSTRUCTIONS
    CALIBRATE 2-WIRE TRANSMITTERS
    (T/C / mA Shown)
    CALIBRATE 4-WIRE TRANSMITTERS
    +IN-
    REF
    Typical
    4-Wire
    Transmitter
    Loop
    Power Supply
    24 VDC
    30
    OPERATING INSTRUCTIONS
    CALIBRATE TRANSMITTERS
    ANY SOURCE FUNCTION AND
    READ mA, READ %, READ DP%
    Choose this function to supply the input signal to the transmitter
    and displaying the 4-20 mA output of the transmitter (used with 4-
    Wire Transmitters). 
    ANY SOURCE FUNCTION AND
    P-XMTR mA, P-XMTR %, P-XMTR DP%
    Choose this function to simultaneously supply power to a 2-Wire
    transmitter while supply the input signal to the transmitter and
    display the 4-20 mA output of the transmitter. 
    1) Disconnect both the input and output wires from the 2-Wire
    Transmitter to be calibrated
    2) Choose a function of the 830-KP that matches the input of the
    transmitter (T/C, RTD, Freq, etc.). Select the proper type,
    range and temperature scale if applicable.
    3) Press the DISPLAY/SOURCE/READ push-button until READ
    or P-XMTR appears in the lower half of the display
    4) Press READ/POWER TRANSMITTER to switch between
    reading milliamps and supply voltage to power the transmitter 
    5) Connect the output of the 830-KP to the signal input of the trans-
    mitter (using the 1, 2 & 4 connectors or the T/C connector on the
    830-KP)
    6) Connect the red POWER lead of the 830-KP (Connector 5) to
    the plus (+) output of the tranmitter and the black POWER lead
    (Connector 6) to the minus (-)
    To change the output value, press the UP or DOWN key for each
    digit. To ramp the output press and hold any UP or DOWN
    pushbutton. This function operates in all three output positions (HI,
    SET & LO). Press the HI & LO Quik-Cheks to recall stored outputs.
    The TechChek 830-KP supplies a nominal 24 Volts DC at 24 mA
    to the 2-Wire transmitter. The current output of the transmitter will
    be accurately displayed by the 830-KP. Calibrate the Transmitter
    in the usual manner and disconnect the 830-KP. 
    						
    							33
    Pressure
    Source
    PROCESS CALIBRATOR
    QUIK-CHEK
    READ PRESSUREChoose this function to measure pressure with a ModPak 91
    module holder and one of the many QuikCal 90 pressure modules. 
    1) Select the proper module for the pressure range of the process
    2) Place the QuikCal 90 Module in the ModPak 91 holder
    3) Plug the ModPak cable into the mating connector on the 830-KP
    4) Turn the selector knob to PRESSURE
    5) The 830-KP will display – – – – – for up to 20 seconds while
    the module is powered up and verified
    6) The 830-KP will display the range of the module for three seconds
    7) Press the °C/°F/ZERO push-button to “Zero” the pressure
    reading
    Carefullyconnect the module to the pressure point to be
    measured.
    To Change the pressure engineering units
    1) Press the TYPE/ENG UNITS push-button. The words SELECT
    ENG. UNITS and the current selection appear on the display.
    2) Press any UP/DOWN push-button until the required type
    appears on the display.
    3) Press the STORE/RESET push-button or wait 5 seconds to
    store the selection 
    OPERATING INSTRUCTIONS
    32
    READ PRESSSURE
    WARNING! - SAFETY CONSIDERATIONS
    Sudden release of compressed or stored gas can cause personal
    injury. Always vent the system before making pressure
    connections or disconnections. Exercise standard physical
    protection practices such as eye protection, gloves, protective
    clothing, etc.
    To prevent the potentially hazardous release into the atmosphere
    of substances introduced into the pneumatic system by the user,
    no relief valves are provided in the QuikCal 90 Pressure Module.
    Consequently, if a module is overpressurized, the sensor will be
    damaged.
    FITTING INSTALLATIONAll pneumatic connections to the QuikCal 90 Module are made via
    1/8”-27 NPT threads in the pressure port. Adapter fittings may be
    used for 1/4”-18, straight 10-32 threads or any other size fittings to
    match the field connections.
    To install a fitting in a module pressure port:
    1) Apply Teflon tape or pipe dope to the fitting threads (Teflon tape
    not recommended with stainless steel fittings)
    2) Carefully install the fitting in the pressure port on top of the
    module
    3) Use a 5/8” wrench to support the pressure module port and a
    second wrench to tighten the fitting
    4) Check for leaks
    OPERATING INSTRUCTIONS 
    						
    							35
    PROCESS CALIBRATOR
    QUIK-CHEK
    34
    OPERATING INSTRUCTIONS
    CALIBRATE PRESSURE TRANSMITTERS
    READ PRESSURE AND
    P-XMTR mA, P-XMTR %, P-XMTR DP%
    Choose this function to simultaneously supply power to a 2-Wire
    pressure transmitter while supplying the input signal to the
    transmitter and displaying the 4-20 mA output of the transmitter. 
    1) Disconnect both the pressure input and output wires from the
    2-Wire Transmitter to be calibrated
    2) Select the proper module for the pressure range of the
    process
    3) Place the QuikCal 90 Module in the ModPak 91 holder
    4) Plug the ModPak cable into the mating connector on the 830-KP
    5) Turn the selector knob to PRESSURE
    6) The 830-KP will display – – – – – for up to 20 seconds while
    the module is powered up and verified
    7) The 830-KP will display the range of the module for three
    seconds
    8) Press the °C/°F/ZERO push-button to “Zero” the pressure
    reading
    9) Press the DISPLAY/SOURCE/READ push-button until READ
    or P-XMTR appears in the lower half of the display
    10) Press READ/POWER TRANSMITTER to switch between
    reading milliamps and supply voltage to power the transmitter 
    11 )
    Carefullymake all pressure connections
    12) Connect the red POWER lead of the 830-KP (Connector 5) to
    the plus (+) output of the transmitter and the black POWER lead
    (Connector 6) to the minus (-)
    Use the hand pump or regulated pressure source to adjust the
    pressure
    The TechChek 830-KP supplies a nominal 24 Volts DC at 24 mA
    to the 2-Wire transmitter. The current output of the transmitter will
    be accurately displayed by the 830-KP. Calibrate the Transmitter
    in the usual manner and disconnect the 830-KP. 
    						
    							37
    SPECIFICATIONS
    POWER & MEASURE 2-WIRE TRANSMITTERS
    RANGES & ACCURACY:Same as for MILLIAMP SOURCEOUTPUT CURRENT: up to 24.00 mATYPICAL DRIVE CAPABILITY:1200 Ohms @ 20.00 mACOMPLIANCE VOLTAGE:nominal 25 VDC @ 20 mACOMMON MODE ERROR:0.01% Full Scale/Common Mode Volt
    2-WIRE TRANSMITTER SIMULATOR
    RANGES:
    1.00 to 24.00 mA; -18.8 to 125.0% of 4 to 20 mA; % DP FlowACCURACY:Same as for MILLIAMP SOURCELOOP VOLTAGE LIMITS:Mininum, 3 VDC; Maximum 45 VDCOVERLOAD PROTECTION:Current limited to 25 mA nominalCOMMON MODE ERROR:0.01% Full Scale/Common Mode Volt
    MILLIAMP READ
    RANGES:
    0.00 to 24.00 mA; -25.0 to 125.0 % of 4 to 20 mA; % DP FlowACCURACY:Same as for MILLIAMP SOURCEOVERLOAD PROTECTION: Current limited to 25 mA nominalVOLTAGE BURDEN: 0.9V at 4 mA, 1.2V at 20 mA, 1.9V at 24 mA
    DC VOLTAGE SOURCE
    RANGES: 0.00 to 110.00 mV; 0.00 to 10.25VACCURACY:
    ±(0.05% of 110 mV+ 0.01mV) = ±0.07 mV
    ±(0.05% of 10.25 V + 0.01V) =  ±0.02V
    SOURCE CURRENT:>20 mASINK CURRENT:>20 mAOUTPUT IMPEDANCE: 
    						