    							Process Calibrator
    TechChek™ 830-KP 
    						
    							1
    Your New TechChek 830-KP
    Lighten your load…take the TechChek 830-KP to every
    site. It’s like bringing a cartload of test equipment from the
    shop to the control room or the field. The TechChek 830-KP
    sources and reads DC like a milliamp or voltage calibrator,
    simulates and measures T/Cs & RTDs like a temperature
    calibrator, generates and counts frequency and Counts-Per-
    Minute like a frequency calibrator and displays pressure like
    a precision test gauge. Troubleshooting? It checks continuity
    with a beeper and measures AC line voltage like a
    multimeter!
    Calibrate Milliamp Inputs
    Calibrate controllers, recorders and other devices in 4
    to 20 or 0 to 20 mA loops.  Source and read 0.00 to 24.00
    mA, or Simulate a 2-Wire Transmitter. Use the optional AC
    adaptor for continuous operation. Display to 0.01 mA , 0.1
    % of 4 to 20 and 0.1 % DP Flow.
    Calibrate 2-Wire Transmitters
    Easily calibrate 2-Wire Pressure and Electronic
    Transmitters by connecting the TechChek 830-KP to both
    the input and output of the transmitter. The TechChek 830-
    KP will simultaneously indicate the input and output of the
    transmitter on the graphical display.
    Calibrate Pressure Systems
    Read pressure with extreme accuracy using an external
    QuikCal pressure module in a companion ModPak module
    holder. Attach the module directly to the pressure
    connection for the best accuracy or with optional tubing for
    tight spots. Display to five digits within 0.025% of reading
    in up to 20 engineering units including psi, pa, Kpa, Mpa,
    GENERAL INFORMATIONCONTENTSGENERAL OPERATING INSTRUCTIONS...........................3
    TURN ON ...................................................................3
    CONNECTIONS .........................................................3
    FIELD & BENCH USE (TILT STAND) ........................4
    CHANGING BATTERIES............................................4
    CONFIGURING TEMPERATURE SCALES ...............5
    ENABLING AUTO-OFF ..............................................5
    RESTORING DEFAULT SETTINGS ..........................5
    DISPLAY CONTRAST & BACKLIGHTING .................6
    SELECTING FUNCTIONS & RANGES......................7
    SOURCE MODE OPERATING INSTRUCTIONS.................8
    STORING & RECALLING QUIK-CHEK®OUTPUTS .8SOURCE MODE CONNECTION INSTRUCTIONSCALIBRATE MILLIAMP INPUTS ................................9
    SIMULATE 2-WIRE TRANSMITTERS .....................10
    CALIBRATE VOLTAGE INPUTS...............................11
    CHECK 1-5 VOLT INPUTS WITHOUT
    DISCONNECTING WIRES .......................................12
    CALIBRATE THERMOCOUPLE INPUTS ................14
    CALIBRATE RESISTANCE INPUTS ........................15
    CALIBRATE RTD INPUTS .......................................16
    CALIBRATE FREQUENCY INPUTS ........................18
    READ MODE OPERATING INSTRUCTIONS....................19
    MIN/MAX ..................................................................19
    OUT OF RANGE SIGNALS......................................19
    READ MODE CONNECTION INSTRUCTIONSREAD MILLIAMP OUTPUTS ....................................20
    READ VOLTAGE OUTPUTS ....................................21
    READ AC  VOLTAGES ..............................................22
    MEASURE THERMOCOUPLE SENSORS ..............23
    READ RESISTANCE ................................................24
    CHECK CONTINUITY ..............................................25
    MEASURE RTD SENSORS .....................................26
    COUNT FREQUENCIES ..........................................28
    CALIBRATE 2 & 4 WIRE TRANSMITTERS .............30
    PRESSURE (WITH COMPANION PRESSURE MODULES)READ PRESSURE ...................................................32
    CALIBRATE PRESSURE TRANSMITTERS ............34
    SPECIFICATIONS...............................................................36WARRANTY........................................................................44 
    						
    							3
    GENERAL
    TURN-ON
    Each time you turn on the TechChek 830-KP it runs through a self
    check then returns to the most recently selected settings.
    SOURCE - The three QUIK-CHEK outputs will be the same as
    previously stored. Each time a different function is selected, the
    associated three QUIK-CHEK outputs will be recalled. 
    READ - The 830-KP is ready to measure the same signal as the
    last time it was used and automatically updates the MAX & MIN
    readings for recall at any time.
    CONNECTIONSTechChek 830-KP has protected banana jacks compatible with
    standard and safety banana plugs.
    Included with your TechChek are:a pair of safety test leads with
    test probes, safety alligator clips, standard alligator clips and
    spade lugs for attachment to a wide variety of instruments. An
    additional test lead and spade lug are also included for 3-Wire RTD
    connections.
    A second pair of test leads with right angled safety banana plugs
    and alligator clips for mA Read and Power Transmitter functions.
    Thermocouple connections are made through a miniature
    thermocouple socket. 
    Optional KIT-1 includes T/C wires for Type J, T, E & K with miniature
    T/C plugs (not included).
    CAUTION! To prevent accidently overloading the instrument
    being tested, correctly set up the outputs before connecting
    the TechChek 830-KP to any instruments to be calibrated.
    !
    !
    !!
    2
    BAR, mBar, Atm, Kgf plus torr, inches and mm of mercury
    or water at a variety of temperatures. Characterized
    modules for laboratory accuracy in the shop, control room or
    field.
    Voltage Calibration
    Calibrate all your DC millivolt and voltage
    instrumentation. Source from 0.00 to 110.00 mV and 0.00
    to 10.25 V. Read up to 110.00 mV, 11.00 V and 200.0 VDC.
    Temperature Calibration
    Source and Read directly in °C and °F for T/C types J,
    K, T, E, R, S & N and four Pt 100 Ohm, Ni 120 Ohm and Cu
    10 Ohm RTDs. Resolution to 0.1°. Cold junction
    compensation tracks ambient temperature changes.
    Frequency Calibration
    Generate zero crossing square waves from 1 to 1000 Hz,
    0.01 to 10.00 kHz and from 1 to 1000 CPM (Counts-Per-
    Minute). Built-in frequency counter measures Hz, kHz & CPM.
    Measure AC Voltage
    Check line voltage or mains from 0.0 to 250.0 volts AC.
    Great for troubleshooting power problems.
    Check Continuity
    Locate pairs of wires, open connections and shorts with
    Rear Label - Condensed Operation Guide 
    						
    							5
    GENERAL
    CONFIGURING TEMPERATURE SCALESThe thermocouple and RTD ranges may be configured for full time
    use of °C, full time use of °F or selectable °C and °F operation.
    This configuration is part of the DEFAULT SETTINGS below.
    AUTO-OFFTechChek 830-KP can be set up to turn itself off after 30 minutes
    of inactivity. The internal timer is reset to 30 minutes each time a
    pushbutton is pressed.
    DEFAULT SETTINGS
    TechChek 830-KP may be restored to the factory default setting.
    This will reset the HI and LO “QUIK-CHEK” memories according
    to the table below and the SET memory to midrange between HI and LO.
    OPERATING INSTRUCTIONS
    1) Press and hold the STORE/RESET push-button while turning
    the 830-KP on.
    2) Keep pressing the push-button until SETTING UP DEFAULT
    appears on the display then release the push-button.
    3)°F/°C SETUP appears on the display.
    4) Press an UP/DOWN key to select °F & °C, °F Only or °C Only.
    °C & °F is the default if no selection is made.
    5) Press STORE/RESET push-button to store your choice or wait
    five seconds for the 830-KP to automatically store your choice
    and  AUTO-OFF SETUP appears on the display.
    6) Press an UP/DOWN key to select AUTO-OFF ENABLE.
    AUTO-OFF causes the 830-KP to turn itself off after 30
    minutes of inactivity to preserve the batteries. Select AUTO-
    OFF DISABLE for continuous
    operation.
    7) Press STORE/RESET push-button
    to store your choice or wait five
    seconds for the 830-KP to
    automatically store your choice. 
    8) The firmware revision will be
    displayed, then the unit will begin
    QUIK-CHEK DEFAULTSRANGE LO       SET HI
    mA 4.00    12.00   20.00
    mV 1.00     5.00   10.00
    V 1.00     5.00   10.00
    T/C J,T,E,K,N  All points 0°C/32°F
    T/C R & S  All points 538°C/1000°F
    Ohms      100.0   200.0  400.0
    RTD All points 0°C/32°F
    kHz 1.00    5.00   10.00
    Hz 100 500    1000
    CPM 100 500    1000
    4
    OPERATING INSTRUCTIONS
    GENERAL
    FIELD & BENCH USETechChek 830-KP comes with a carrying
    case and a built-in tilt stand/hanger. The
    830-KP is held securely in the case by
    VELCRO®for use with the carrying case
    open. The carrying case also has a snap-
    on belt loop which can be looped around a
    pipe or rail.
    The tilt stand is easily raised by pulling the
    stand until it locks into place. The stand can
    also be reversed for use as a hanger to
    suspend the 830-KP.
    CHANGING BATTERIES Low battery is indicated by a battery symbol
    on the display. Approximately four hours of
    operation remain before the LCD blanks
    and TechChek 830-KP shuts itself down.
    Turn the 830-KP off, loosen the  captive
    screw securing the battery compartment
    and lift off the  cover from the bottom of the
    case. The six “AA” batteries are easily
    removed and replaced (alkaline supplied
    and recommended). Replace the battery
    compartment cover by inserting the tabs
    and tightening the screw. 
    						
    							GENERAL
    SELECTING FUNCTIONS
    Press a function key to choose among mA,
    V, T/C, Ω, RTD, FREQ and PRESSURE.
    SELECTING RANGESPress the RANGE/TYPE pushbutton to
    select the desired range and scale. 
    SOURCE RANGESMilliamp:
    Source: mA, %mA (% 4-20), DP% (% Differential Pressure) 
    2-Wire Simulator: mA, %mA (% 4-20), DP% (% Differential Pressure)
    VDC: mV, V
    T/C: Types J, T, E, K, N, R & S in °C &°F
    Ohms: Ohms
    RTD: Four Pt 100Ω, Ni 120Ω& Cu 10Ωin °C &°F
    Frequency: KHz, Hz, CPM
    READ RANGESMilliamp: mA, %mA (% 4-20), DP% (% Differential Pressure) 
    VDC: mV, 10V, 200V
    T/C: Types J, T, E, K, N, R & S in °C &°F
    Ohms: Ohms
    RTD: Four Pt 100Ω, Ni 120Ω& Cu 10Ωin °C &°F
    Frequency: KHz, Hz, CPM
    AC Volts: VAC
    TRANSMITTER RANGESPwr Xmtr: Supplies nominal 24 VDC to power the transmitter while 
    measuring the transmitter milliamp signal . Can be simul-
    taneously  displayed with any Source Range or Pressure 
    Measurement
    Read mA: Measures transmitter output
    7
    TYPE
    ENG UNITS
    OPERATING INSTRUCTIONS
    6
    OPERATING INSTRUCTIONS
    GENERAL
    DISPLAY CONTRAST
    The contrast of the Liquid Crystal Display may adjusted for best
    readability.
    1) Press and hold the DISPLAY/SOURCE/READ push-button
    while turning the 830-KP on to adjust the contrast of the
    display.
    2) Press any UP or DOWN key until the display is most legible.
    3) Press the STORE/RESET push-button or wait five seconds for
    your selection to be stored.
    DISPLAY BACKLIGHTINGThe LCD may have too little contrast where the lighting is dim or
    the 830-KP is in shadow. Turn on the LCD backlighting to make
    the display easy to read.
    1) If the TechChek 830-KP is off, turn it on and wait for the display
    to come up in normal mode.
    2) Press and hold the DISPLAY/SOURCE/READ push-button for
    three seconds and the backlight will turn on.
    To extend the life of the batteries, it is recommended that the
    backlighting be turned off when the LCD is in normal light. Press
    and hold the DISPLAY/SOURCE/READ push-button for  three
    seconds and the backlight will turn off.
    TURN OFFPress the POWER push-button to turn the
    830-KP off. If AUTO-OFF is enabled, the
    830-KP will turn itself off after 30 minutes of
    inactivity.POWER 
    						
    							9
    CALIBRATE MILLIAMP INPUTS
    mA, mA % (Percent of 4 to 20 mA), DP% (DP Flow)
    Choose this function to provide an output from 0.00 to 24.00
    milliamps. The compliance voltage is a nominal 24 VDC to provide
    the driving power to your milliamp receivers.
    1) Disconnect one or both input wires from the device to be
    calibrated.
    2) Press the MILLIAMP/V push-button until mA or mA and %
    appear on the display
    3) Press the DISPLAY/SOURCE/READ push-button until SOURCE
    or 2-WIRE appear on the display
    4) If 2-WIRE is on the display, press the TYPE/ENG UNITS push-
    button once to indicate SOURCE on the display
    5) Press the mA/%/% DP FLOW push-button to display mA, % 4-20
    or % DP Flow.
    6) Connect the red SOURCE lead of the calibrator to the plus (+)
    input of the device and the black SOURCE lead to the minus (-).
    Output current is continuously adjustable with the UP/DOWN
    push-buttons. Zero & Span (or any other values) are available by
    using the LO and HI QUIK-CHEKs.
    OPERATING INSTRUCTIONS
    Milliamp Receiver Input
    Controller
    Transmitter
    Computer
    Logger
    I/P
    DCS
    SOURCE MODESelect source by pressing the DISPLAY/SOURCE/READ
    pushbutton until  the word SOURCE appears on the LCD display.
    To change the output value, press the UP or DOWN key for each
    digit. To ramp the output press and hold any UP or DOWN
    pushbutton. The display will continue to change in increments
    corresponding to the digit being changed and will automatically
    carry up and down until the limits of the range are reached. This
    function operates in all three output positions (HI, SET & LO). 
    STORING QUIK-CHEK OUTPUTS
    1) Press HI or LO 
    2) Press the UP/DOWN keys to desired
    value
    3) Press the STORE push-button
    The LCD will flash once to show that the
    value was saved
    If a value is in the SET position and you want
    that value stored in HI or LO, press and hold
    the STORE push-button, then press the HI
    or LO push-button. The display will flash
    once to indicate the value has been stored.
    Then release both push-buttons.
    RECALLING QUIK-CHEK OUTPUTSWhen you need a stored value just press a
    Quik-Chek push-button. Any value for the
    selected range may be stored in HI & LO.
    The TechChek 830-KP remembers the HI,
    LO and SET values for each function with the
    power on or off. Each time a different function
    is selected, the last three QUIK-CHEK values
    for that function will be recalled.
    OPERATING INSTRUCTIONS
    8 
    						
    							11
    CALIBRATE VOLTAGE INPUTS
    V,  m V
    Choose this function to provide an output from 0.00 mV to 110.00
    mV and from 0.00 to 10.25 VDC. Current compliance up to 20 mA
    to provide the driving power to your voltage receivers.
    1) Disconnect one or both input wires from the device to be
    calibrated
    2) Press the MILLIAMP/V push-button until V or mV appear on
    the display
    3) Press the DISPLAY/SOURCE/READ push-button until SOURCE
    and V or mV appear on the display
    4) Press the TYPE/ENG UNITS push-button once to switch
    between V and mV on the display
    5) Connect the red SOURCE lead of the calibrator to the plus (+)
    input of the device and the black SOURCE lead to the minus (-).
    Output voltage is continuously adjustable with the UP/DOWN
    push-buttons. Zero & Span (or any other values) are available by
    using the LO and HI QUIK-CHEKs.
    OPERATING INSTRUCTIONS
    Voltage Receiver Input
    Controller
    Transmitter
    Computer
    Logger
    DCS
    10
    OPERATING INSTRUCTIONS
    SIMULATE 2-WIRE TRANSMITTERS
    2-WIRE mA, 2-WIRE % (Percent of 4 to 20 mA) 2-WIRE DP %
    Choose this function to simulate a 2-Wire Transmitter output from
    1.00 to 24.00 milliamps. Operates in loops with power supply
    voltages from 3 to 45 VDC.
    1) Disconnect existing 2-Wire Transmitter from the loop
    2) Press the MILLIAMP/V push-button until mA or mA and %
    appear on the display
    3) Press the DISPLAY/SOURCE/READ push-button until SOURCE
    or 2-WIRE appear on the display
    4) If SOURCE is on the display, press the TYPE/ENG UNITS push-
    button once to indicate 2-WIRE on the display
    5) Press the mA/%/% DP FLOW push-button to display mA, % 4-20
    or % DP Flow.
    6) Connect the red SOURCE lead of the calibrator to the plus (+)
    input of the device and the black SOURCE lead to the minus (-).
    The simulated output of the 2-Wire Transmitter is continuously
    adjustable from 1.00 to 24.00 mA with the UP/DOWN push-
    buttons. Zero & Span (or any other values) are available by using
    the LO and HI QUIK-CHEKs.
    +IN-
    REF +OUT-
    Power Supply
    2 to 45 VDC
    Receiver
    (Powers external
    2-Wire Transmitter)
    To
    Sensor
    Typical
    2-Wire
    Transmitter
    (Disconnected) 
    						
    							13
    CONTROLLER
    (TYPICAL)
    ADDITIONAL
    4-20 mA OR 1-5V
    INSTRUMENTS RECORDER
    COMPUTER
    LOGGER
    ETC,
    1-5V DC 250Ω
    Typical
    2-Wire
    Transmitter
    +IN-
    REF +OUT-
    CHECK 1-5 VOLT INPUTSWITHOUT DISCONNECTING WIRES
    OPERATING INSTRUCTIONS
    12
    CHECK 1-5 VOLT INPUTSWITHOUT DISCONNECTING WIRESMost 1-5 Volt receivers in 4-20mA loops have a 250 Ohm
    resistor across the input of the receiver. This resistor may be
    mounted internally or externally. TechChek 830-KP is connected
    directly across the input of the 1-5 Volt receiver without
    disconnecting any field wiring. This saves a great deal of time
    when a large number of voltage receivers, such as chart recorders
    or computer systems, require calibration.
    Make certain that changing the signal input will not disturb the
    process or cause unexpected alarms when checking on-line
    instruments. 
    It is important to remember the 830-KP drives only
    the device to which it is connected.It has no effect on other
    devices in the 4 to 20 mA loop. TechChek 830-KP will clamp the
    selected value in the mV and V Ranges to the maximum source
    or sink current of >16 mA.
    1) Press the MILLIAMP/V push-button until V or mV appear on
    the display
    2) Press the DISPLAY/SOURCE/READ push-button until SOURCE
    and V or mV appear on the display
    3) Press the TYPE/ENG UNITS push-button once to switch
    between V and mV on the display
    4) Connect the red SOURCE lead of the calibrator to the plus (+)
    input of the device and the black SOURCE lead to the minus (-).
    Any associated 250 Ohm resistor must not be disconnected.
    OPERATING INSTRUCTIONS 
    						
    							15
    CALIBRATE RESISTANCE INPUTSChoose this function to simulate a resistance into a variety of
    instruments. 
    1) Disconnect one or both input wires from the device to be
    calibrated
    2) Press the T/C/RTD/Ωpushbutton until the Ωsymbol appears
    on the display.
    3) Press the DISPLAY/SOURCE/READ push-button until SOURCE
    appears on the display
    4) Connect the red SOURCE lead of the calibrator to the plus (+)
    input of the device and the black SOURCE lead to the minus (-). 
    Output resistance is continuously adjustable with the UP/DOWN
    push-buttons. Zero & Span (or any other values) are available by
    using the LO and HI QUIK-CHEKs”.
    OPERATING INSTRUCTIONS
    Receiver InputOhms or 2-Wire RTD
    Controller
    Transmitter
    Computer
    14
    CALIBRATE THERMOCOUPLE INPUTSChoose this function to simulate a thermocouple signal into any
    instrument requiring a thermocouple input. The output of the 830-
    KP is automatically cold junction compensated.
    1) Disconnect the thermocouple from the instrument being
    calibrated.
    2) Press the T/C/RTD/Ωpushbutton until the word TYPE & any
    T/C appear on the display.
    3) Press the DISPLAY/SOURCE/READ push-button until SOURCE
    and any T/C TYPE appear on the display
    4) Use the proper thermocouple wire and corresponding miniature
    T/C connector to connect TechChek 830-KP to the thermocouple.
    Output temperature is continuously adjustable with the UP/DOWN
    push-buttons. Zero & Span (or any other values) are available by
    using the LO and HI QUIK-CHEKs”.
    To Change the Thermocouple type
    1) Press the TYPE/ENG UNITS push-button. The words SELECT
    T/C TYPE and the current selection appear on the display.
    2) Press any UP/DOWN push-button until the desired T/C type
    appears.
    3) Press the STORE/RESET push-button or wait 5 seconds to
    store the selection
    OPERATING INSTRUCTIONS
    +-
    Instrument withThermocouple Input
    Controller
    Temperature Indicator
    Temperature Trip or Alarm
    Temperature Transmitter 
    						
    							17
    OPERATING INSTRUCTIONS
    Instrument withRTD Input
    Controller
    Temperature Transmitter
    Temperature Indicator
    Temperature Trip or Alarm
    Receiver InputOhms or 2-Wire RTD
    Controller
    Transmitter
    Computer
    Three Wire RTD Connection
    Two Wire RTD Connection
    CALIBRATE RTD INPUTS
    Four Wire RTD Connection
    SENSOR}
    + COMP
    - COMP
    Jumper
    Jumper
    16
    CALIBRATE RTD INPUTSChoose this function to simulate a temperature signal into any
    instrument requiring an RTD input. 
    1) Disconnect the RTD from the instrument being calibrated.
    2) Press the T/C/RTD/Ωpushbutton until any RTD type (Pt, Ni or
    Cu) appears on the display.
    3) Press the DISPLAY/SOURCE/READ push-button until SOURCE
    and any RTD TYPE appear on the display
    4) Connect using 2 or 3 wires as in the diagrams on the opposite
    page. Spade lugs are recommended to minimize any contact
    resistance.
    Output temperature is continuously adjustable with the
    UP/DOWN push-buttons. Zero & Span (or any other values)
    are available by using the LO and HI QUIK-CHEKs”.
    To Change the RTD type
    1) Press the TYPE/ENG UNITS push-button. The words SELECT
    RTD TYPE and the current selection appear on the display.
    2) Press any UP/DOWN push-button until the required type
    appears on the display.
    3) Press the STORE/RESET push-button or wait 5 seconds to
    store the selection 
    Note on 4-Wire Connections:
    The TechChek 830-KP may be used to calibrate instruments
    requiring a 4-Wire RTD connection. This connection is valid
    only when the lead wires included with the 830-KP are used to
    connect the 830-KP directly to the instrument. Use of longer
    wires can cause errors in the output setting.
    For long 4-Wire cable runs use the 3-Wire connection and add
    a stacking banana jack (not available from Transmation) to
    connect a second wire to jack #1. Connect this extra wire to the
    fourth wire field connection.
    OPERATING INSTRUCTIONS 
    						