Altek Model 934 Milliamp Voltage Loop Calibrator Manual
Have a look at the manual Altek Model 934 Milliamp Voltage Loop Calibrator Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 Altek manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
6 OPERATING INSTRUCTIONS ENABLING ADDITIONAL RANGES & FUNCTIONS The Model 934 has two additional ranges for displaying % DP Flow and Engineering Units. Additional functions are AUTO- CHEK (automatic stepping of outputs) and AUTO-OFF. These ranges & functions are enabled by configuring DIP Switches (see next page). After the functions are enabled they are available as selections during setup (see OPERATING INSTRUCTIONS, SOURCE or READ). AUTO-OFF The Model 934 can be set up to turn itself off after 30 minutes of inactivity. The internal timer is reset to 30 minutes each time the digital pot is turned or a pushbutton is pressed. % DP FLOW Check out the signal generated by DP (differential pressure) flow transmitters, square root extractors or other signals using a square law. Select %DP Flow for all milliamp functions. AUTO-CHEK The Model 934 can be set up to automatically step the output with your choice of 2, 3, 5 or 10 steps from Zero to Full Scale. ENGINEERING UNITS Scale the display on the 934 to match the display of the instrument being calibrated. For example if you have a digital indicator showing 0.0 to 1000.0 Gallons/Hour from a flowmeter with 4 - 20 mA output you can scale the output of the 934 to display 0.0 at 4 mA and 1000.0 at 20 mA. 39 CALIBRATION PROCEDURE SOURCE MODE ADJUSTMENT mA 1) Connect the Source leads in series with the 100 Ohm re- sistor. Connect the DVM parallel with the 100 Ohm resistor (See below). 2) Turn the unit on and wait for the unit to go into calibration mode (SOURCE and READ flashing). 3) Press the STORE pushbutton to select Source Mode. 4) Turn the Digital Pot (yellow knob) until mA is displayed and press the STORE pushbutton. 5) Dial the output with the READ toggle switch in the MAX position (coarse adjust) until the DVM displays approxi- mately 0.0 V. Move the READ toggle switch to READ (fine adjust) until the DVM reads 0.0000V ±0.000096V and press the STORE pushbutton. 6) Dial the output (using the same coarse and fine adjust- ments) until the DVM reads 2.4000V ±0.00024V and press the STORE pushbutton. 7) Turn the unit off.To change the output of the Model 934 during the calibration of the SOURCE functions both coarse and fine adjustments are used. Place the READ toggle switch in the MAX position for coarse adjustment and move the READ toggle switch to the READ position for fine adjustment. ALTEK 934SOURCEREAD DMM 100 OHM If calibrating with an 81/2digit multimeter directly in milliamps the calibration points are0.000 mA ±0.00096mA and 24.000 mA±0.0024mA
5 OPERATING INSTRUCTIONS GENERAL FIELD & BENCH USE The Model 934 comes with a carrying case and a built-in tilt stand/hanger. The 934 is held securely in the case by VELCRO ® for use with the carrying case open. The carrying case also has a snap-on belt loop which can also be looped around a pipe or rail. The tilt stand is easily raised by pulling the stand until it locks into place. The stand can also be reversed for use as a hanger to suspend the Model 934. RESTORE DEFAULT SETTINGS The Model 934 may be restored to the factory default setting. This will reset the HI and LO “QUIK-CHEK” memories according to the table below and the SET memory to ,idrange below HI and LO. The output and input are set to the mA range. 1) Press and hold both the STORE& RESET push-buttons while turning the Model 934 on. 2) Keep pressing the push-buttons until the display flashes once (10 seconds) then release. 3) All segments on the LCD will remain displayed until the Model 934 has been reset QUIK-CHEK DEFAULTS RANGE HI LO mA 20.000 4.000 %4-20mA100.00 0.00 %DP 100.0 0.0 mV 100.00 0.00 V 5.000 1.000 Note: Some ranges and functions are enabled by DIP switches (see pages 6 & 7 for factory settings). AUTOCAL To maintain accuracy the Model 934 periodically recalibrates its measuring circuitry against internal references. While this is occuring the word CAL will appear on the LCD display for less than 2 seconds. VELCRO® is a registered trademark of Velcro USA40 CALIBRATION PROCEDURE SOURCE MODE ADJUSTMENT 2-Wire Sim 1) Connect the Source leads in series with the current source set to 30 mA and the 100 Ohm resistor. Connect the DVM parallel with the 100 Ohm resistor (See below). 2) Turn the unit on and wait for the unit to go into calibration mode (SOURCE and READ flashing). 3) Press the STORE pushbutton to select Source Mode. 4) Turn the knob until 2-WIRE SIM is displayed and press the STORE pushbutton. 5) Disconnect the positive output lead from the 100 Ohm resistor and wait until the display reads 0.00 and press the STORE pushbutton. 6) Re-connect the positive output lead to the 100 Ohm resistor and dial the output until the DVM reads 2.4000V ±0.00024V and press the STORE pushbutton. 7) Turn the unit off. ALTEK 934 CURRENT SOURCE SET AT 30mA SOURCEREAD DMM 100 OHM
4 ROCHESTER NY 14623ALTEK 934SOURCE19.934mA mA/V LOOP CALIBRATOROFF ONHI SET LOMAX READ MIN 9VDC AUTO-CHEK HI MINREAD0.000V SOURCE STOREQUIK-CHEK ®SETUP READ RESETSETUP 500 MAXSOURCESOURCEREAD OPERATING INSTRUCTIONS GENERAL TURN-ON Each time the Model 934 is turned on, the LCD will display all segments for about 1 second. It then displays the most recently selected Source & Read Settings. 1) Move the power switch to ON. All segments on the LCD are turned on during self test. 2) The upper half of the LCD will indicate if SOURCE is enabled while the lower half will indicate if READ is enabled. SOURCE - The three QUIK-CHEK outputs will be the same as previously stored. Each time a different function is selected, the three QUIK-CHEK outputs for that function will be recalled. READ - The 934 is ready to measure the same signal as the last time it was turned on and is automatically updating the MAX & MIN readings for recall from the READ toggle switch. CONNECTIONS The Model 934 has built-in test leads with alligator clips for attachment to instruments or sensors with terminal blocks or flying leads. To prevent acci- dently overloading the instrument being tested, it is important to correctly set up the outputs before connecting the Model 934 to any instruments to be calibrated. 41 ALTEK 934SOURCEREAD DMM 100 OHM CALIBRATION PROCEDURE SOURCE MODE ADJUSTMENT mV 1) Connect the Source leads to the DVM (See below). 2) Turn the unit on and wait for the unit to go into calibration mode (SOURCE and READ flashing). 3) Press the STORE pushbutton to select Source Mode. 4) Turn the knob until mV is displayed and press the STOREpushbutton. 5) Dial the output until the DVM reads 0.000mV ±0.040mV and press the STORE pushbutton. 6) Dial the output until the DVM reads 999.900mV ±0.100mV and press the STORE pushbutton. 7) Turn the unit off. V 1) Leave the Source leads connected as in the mV calibration 2) Turn the unit on and wait for the unit to go into calibration mode (SOURCE and READ flashing). 3) Press the STORE pushbutton to select Source Mode. 4) Turn the knob unitl V is displayed and press the STORE pushbutton.5) Dial the output until the DVM reads 0.0000V ±0.0022V and press the STORE pushbutton. 6) Dial the output until the DVM reads 20.0000V ±0.002V and press the STORE pushbutton. 7) Turn the unit off. 8) Disconnect the source leads from the DVM.
3 GENERAL INFORMATION MILLIAMP CALIBRATION Use at any point in your 4 to 20 milliamp loop. Source milliamps directly into your receivers with loop loads up to 1200 Ohms. Simulate 2-Wire Transmitters for checkout of field wiring and total loop testing. Power and measure 2-Wire or 4-Wire transmitters on the bench before installation. Read Transmitter & Controller outputs up to 52.00 mA. Display directly in milliamps, percent of 4 to 20 mA, percent of DP Flow or engineering units to match the scale of your process. VOLTAGE CALIBRATION Calibrate and checkout all your DC Voltage instrumentation. Source up to 999.90 millivolts to simulate sensors and up to 20.000 Volts to check your process signal inputs. Measure voltages from -99.990 to +99.990 VDC to check your power supplies and the outputs from any process voltage equipment. Calibrate any 1 to 5 Volt device in a live 4 to 20 milliamp loop without disconnecting any wires. The Model 934 will automatically source or sink 0 to >16 milliamps to clamp the test voltage in all source ranges. Output impedance is less than 0.2 Ohm to let you calibrate voltage instruments in milliamp loops without removing the input load resistors. QUIK-CHEK & AUTO-CHEK Instantly recall three output settings in each range with the handy QUIK-CHEK switch. All output settings are remembered for each function, even with the power off. Use the AUTO-CHEK to continously switch the Model 934s output from Zero to Span in 2, 3, 5 or 10 equal steps at intervals of 5 to 900 seconds. Leave behind the panel and do your calibration from the front...or simulate your 2-Wire transmitter and calibrate every device in the loop. 42 CALIBRATION PROCEDURE READ MODE ADJUSTMENT mA 1) Connect the Read leads to a current source in series with a 100 Ohm resistor. Monitor the current by placing a DVM across the 100 Ohm resistor (See below). 2) Turn the unit on and wait for the unit to go into calibration mode (SOURCE and READ flashing). 3) Press the RESET pushbutton to select Read mode. 4) Dial the unit so mA is displayed and press the RESETpushbutton. 5) Set the current source to 0.000mA so that the DVM reads 0.0000V ±0.000096V and press the Read store pushbutton.6) Set the current source to 24.000mA so that the DVM reads 2.4000V ±0.00024V and press the RESET pushbutton. 7) Turn the unit off. ALTEK 934 CURRENT SOURCE SOURCEREAD DMM 100 OHM
2 GENERAL INFORMATION Simultaneously source and read any combination of process milliamp and voltage signals. Alteks Model 934 combines all the functionality of a milliamp and a voltage calibrator! This dual function calibrator continuously displays the output and input signals. The top half of the display indicates the output setting while the lower half of the display shows the signal being measured. Calibrate state of the art equipment such as smart transmitters and indicators to 4 1/2 or more digits using the Model 934. Full 5 digit resolution lets you check any brand of process control equipment. Your calibrations are correct with accuracy surpassing 0.02% of reading. Use the scaleable display to match the readings on the Model 934 to that of the equipment being calibrated. For example, if your 4 to 20 mA recorder scale is 0.0 to 300.0 gallons/hour you can set the 934 to display 0.0 for 4 mA and 300.0 for 20 mA. No more need to use your calculator or cheat sheet to come up with the correct milliamp values. Six AA alkaline batteries provide more than 20 hours sourcing 20 mA continuously . . . more than 3 months of typical use. From the time the low battery indicator first comes on you have hours of power remaining to get you through a shift without interruption. The 934 comes configured to automatically power down after 30 minutes of inactivity to save battery life. Calibrate loop converters and isolators by simultaneously providing the input signal and measuring the output of the converter or isolator. Internal optoisolators withstand up to 500 VDC between the source and read leads. Calibrate and checkout all your process milliamp and voltage instruments with the benchtop accurate Model 934. YOUR NEW CALIBRATOR 43 CALIBRATION PROCEDURE mV 1) Connect the Read leads to a voltage source and place the DVM across the voltage source (See below). 2) Turn the unit on and wait for the unit to go into calibration mode (SOURCE and READ flashing). 3) Press the RESET pushbutton to select Read mode. 4)Dial the unit so mV is displayed press the RESET pushbutton.5) Set the voltage source to 0.00mV so that the DVM reads 0.000mV ±0.040mV and press the RESET pushbutton. 6) Set the voltage source to 999.90mV so that the DVM reads 999.90mV ±0.100mV and press the RESET pushbutton. 7) Turn the unit off. V 1) Leave the Read leads connected as in the mV calibration2) Turn the unit on and wait for the unit to go into calibration mode (SOURCE and READ flashing). 3) Press the RESET pushbutton to select Read mode. 4) Dial the unit so V is displayed and press the RESETpushbutton to select V. 5) Set the voltage source to 0.000V so that the DVM reads 0.0000V ±0.0004V and press the RESET pushbutton. 6) Set the voltage source to 95.000V so that the DVM reads 95.0000V ±0.0019V and press the RESET pushbutton. 7) Turn the unit off. READ MODE ADJUSTMENT ALTEK 934 VOLTAGE SOURCE SOURCEREAD DMM After calibration has been completed place the unit into normal operating mode by placing the CAL Enable jumper back into the stored position (See figure page 38).
1 FEATURES • DUAL LIQUID CRYSTAL DISPLAY Simultaneous Source and Read • LABORATORY ACCURATE ±(0.012% of Reading + 0.008% of Full Scale Reading) Accurate to ±0.004 mA from 4 to 20 mA • USER SETTABLE “QUIK-CHEK®s” Set any Span & Zero • FULL FIVE DIGIT DISPLAYS Source up to 24.000 mA, 20.000 V or 999.90 mV Read up to 52.000 mA, 99.990 V or 999.90 mV • SCALEABLE DISPLAY Set in engineering units • SPEED SENSITIVE DIGIPOT Fast setting to any desired value • “AUTO-CHEK” TM OUTPUTS Automatically outputs in 2, 3, 5 or 10 steps • ISOLATED SOURCE AND READ Isolated to 500VDC Fuseless protection to 120 V AC/DC 44 THREE YEAR WARRANTY Altek products are warranted to be free from defects in material and workmanship (excluding fuses, batteries and leads) for a period of one year from the date of shipment. Warranty repairs can be obtained by returning the equipment prepaid to our factory. Products will be replaced, repaired, or adjusted at our option. Altek gives no other warranties, includ- ing any implied warranty of fitness for a particular purpose. Also, Altek shall not be liable for any special, indirect, incidental or consequential damages or losses arising from the sale or use of its products.
ORDERING INFORMATION Part No. MODEL 934 MULTIFUNCTION LOOP CALIBRATOR 934 A carrying case is included with each Model 934 Contents GENERAL INFORMATION......................................................... 2 YOUR NEW CALIBRATOR...................................................... 2 MILLIAMP CALIBRATION...................................................... 3 VOLTAGE CALIBRATION ........................................................ 3 QUIK-CHEK & AUTO-CHEK................................................... 3 GENERAL OPERATING INSTRUCTIONS.................................. 4 TURN ON................................................................................. 4 RESTORE DEFAULT SETTINGS............................................ 5 ENABLING RANGES & FUNCTIONS..................................... 6 AUTO-OFF .......................................................................... 6 % DP FLOW ........................................................................ 6 AUTO-CHEK ...................................................................... 6 ENGINEERING UNITS...................................................... 6 SETTING DIP SWITCHES...................................................... 7 RANGES.................................................................................. 8 CHANGING BATTERIES......................................................... 9 USING ENGINEERING UNITS............................................ 10 SOURCE MODE OPERATING INSTRUCTIONS.................... 14 STORING QUIK-CHEK ® OUTPUTS.................................... 16 RECALLING QUIK-CHEK OUTPUTS.................................. 16 SOURCE FUNCTIONS........................................................ 17 CALIBRATE MILLIAMP INPUTS..................................... 17 SIMULATE 2-WIRE TRANSMITTERS............................ 18 POWER & MEASURE 2-WIRE TRANSMITTERS.......... 19 CALIBRATE VOLTAGE INPUTS ...................................... 20 CALIBRATE LOOP CONVERTERS & ISOLATORS....... 21 CHECK 1-5 VOLT INPUTS WITHOUT DISCONNECTING WIRES.............................................. 22 USING AUTO-CHEK....................................................... 24 READ MODE OPERATING INSTRUCTIONS........................... 28 MIN/MAX ............................................................................... 29 READ FUNCTIONS.............................................................. 30 READ MILLIAMP OUTPUTS.......................................... 30 READ VOLTAGE OUTPUTS ............................................ 31 SPECIFICATIONS...................................................................... 32 CALIBRATION PROCEDURE................................................. 38 WARRANTY............................................................................... 44 ORDERING INFORMATION...................................................... 45
PN 1886535 Rev A April 2002 ©2002 Altek Industries, Inc. Specifications subject to change without notice. All tights reserved. Printed in U.S.A. Multifunction Loop Calibrator Model 934