PN 1886535 Rev A April 2002 ©2002 Altek Industries, Inc. Specifications subject to change without notice. All tights reserved. Printed in U.S.A. Multifunction Loop Calibrator Model 934
Page 2
ORDERING INFORMATION Part No. MODEL 934 MULTIFUNCTION LOOP CALIBRATOR 934 A carrying case is included with each Model 934 Contents GENERAL INFORMATION......................................................... 2 YOUR NEW CALIBRATOR...................................................... 2 MILLIAMP CALIBRATION...................................................... 3 VOLTAGE CALIBRATION ........................................................ 3 QUIK-CHEK...
Page 3
1 FEATURES • DUAL LIQUID CRYSTAL DISPLAY Simultaneous Source and Read • LABORATORY ACCURATE ±(0.012% of Reading + 0.008% of Full Scale Reading) Accurate to ±0.004 mA from 4 to 20 mA • USER SETTABLE “QUIK-CHEK®s” Set any Span & Zero • FULL FIVE DIGIT DISPLAYS Source up to 24.000 mA, 20.000 V or 999.90 mV Read up to 52.000 mA, 99.990 V or 999.90 mV • SCALEABLE DISPLAY Set in engineering units • SPEED SENSITIVE DIGIPOT Fast setting to any desired value • “AUTO-CHEK” TM OUTPUTS Automatically outputs in 2, 3,...
Page 4
2 GENERAL INFORMATION Simultaneously source and read any combination of process milliamp and voltage signals. Alteks Model 934 combines all the functionality of a milliamp and a voltage calibrator! This dual function calibrator continuously displays the output and input signals. The top half of the display indicates the output setting while the lower half of the display shows the signal being measured. Calibrate state of the art equipment such as smart transmitters and indicators to 4 1/2 or more digits...
Page 5
3 GENERAL INFORMATION MILLIAMP CALIBRATION Use at any point in your 4 to 20 milliamp loop. Source milliamps directly into your receivers with loop loads up to 1200 Ohms. Simulate 2-Wire Transmitters for checkout of field wiring and total loop testing. Power and measure 2-Wire or 4-Wire transmitters on the bench before installation. Read Transmitter & Controller outputs up to 52.00 mA. Display directly in milliamps, percent of 4 to 20 mA, percent of DP Flow or engineering units to match the scale of your...
Page 6
4 ROCHESTER NY 14623ALTEK 934SOURCE19.934mA mA/V LOOP CALIBRATOROFF ONHI SET LOMAX READ MIN 9VDC AUTO-CHEK HI MINREAD0.000V SOURCE STOREQUIK-CHEK ®SETUP READ RESETSETUP 500 MAXSOURCESOURCEREAD OPERATING INSTRUCTIONS GENERAL TURN-ON Each time the Model 934 is turned on, the LCD will display all segments for about 1 second. It then displays the most recently selected Source & Read Settings. 1) Move the power switch to ON. All segments on the LCD are turned on during self test. 2) The upper half of...
Page 7
5 OPERATING INSTRUCTIONS GENERAL FIELD & BENCH USE The Model 934 comes with a carrying case and a built-in tilt stand/hanger. The 934 is held securely in the case by VELCRO ® for use with the carrying case open. The carrying case also has a snap-on belt loop which can also be looped around a pipe or rail. The tilt stand is easily raised by pulling the stand until it locks into place. The stand can also be reversed for use as a hanger to suspend the Model 934. RESTORE DEFAULT SETTINGS The Model 934...
Page 8
6 OPERATING INSTRUCTIONS ENABLING ADDITIONAL RANGES & FUNCTIONS The Model 934 has two additional ranges for displaying % DP Flow and Engineering Units. Additional functions are AUTO- CHEK (automatic stepping of outputs) and AUTO-OFF. These ranges & functions are enabled by configuring DIP Switches (see next page). After the functions are enabled they are available as selections during setup (see OPERATING INSTRUCTIONS, SOURCE or READ). AUTO-OFF The Model 934 can be set up to turn itself off after 30...
Page 9
BOTTOM BOARD Calibrator Enabled (Jumper Shorted) on off CAL 7 OPERATING INSTRUCTIONS ENABLING ADDITIONAL RANGES & FUNCTIONS The DIP switches are located in the battery compartment. Turn the Model 934 off and loosen the three slotted screws and remove the battery cover. 38 CALIBRATION PROCEDURE SUGGESTED EQUIPMENT 1) Digital voltmeter with accuracy of ±0.003% of reading at 2V and ±0.005% of reading at 100V Important: The required accuracy is usually found only on meters with 71/2 or more digits 2) 100...
Page 10
8 √ OPERATING INSTRUCTIONS RANGES The selected range is displayed on the LCD by the annunciators listed below. MILLIAMPS - mA Choose mA to display directly in milliamps. Display in SOURCE and READ functions from 0.000 to 24.000 mA. In READ you can also measure up to 52.00 mA. PERCENT OF 4 - 20 MILLIAMPS - mA % Choose mA % to display in percent of span for 4 - 20 mA loops. Use with chart recorders or current trips that display in %. 0.00% 4.000 mA 25.00%8.000 mA 50.00%12.000 mA 75.00%16.000 mA 100.00%...