Altek Model 934 Milliamp Voltage Loop Calibrator Manual
Here you can view all the pages of manual Altek Model 934 Milliamp Voltage Loop Calibrator Manual. The Altek manuals for Calibrator are available online for free. You can easily download all the documents as PDF.
Page 31
16 OPERATING INSTRUCTIONS SOURCE MODE STORING QUIK-CHEK OUTPUTS 1) Switch to HI or LO 2) Turn the knob to desired value 3) Press the STORE push-button The LCD will flash once to show that the value was saved If a value is in the SET position and you want that value stored in HI or LO, press and hold the STORE push-button while moving the switch to HI or LO. The display will flash once to indicate the value has been stored. Then release the STOREpush-button. RECALLING QUIK-CHEK OUTPUTS Any time you...
Page 32
15 OPERATING INSTRUCTIONS SOURCE MODE Whenever SOURCE mode is selected the word SOURCE will appear on the LCD display. To change the output value, turn the speed sensitive digital pot. Turning the knob slowly will cause a gradual change in the output. A faster rate of change will occur when the knob is turned faster. This function operates in all three output positions (HI, SET & LO). ALTEK 934 20.000mA VA LV E OR RECEIVER I/P 30 OPERATING INSTRUCTIONS READ FUNCTIONS The READ functions measures the...
Page 33
14 OPERATING INSTRUCTIONS SOURCE MODE SOURCE 1) Move the POWER switch to ON 2) Press & hold both the STORE & RESET push-buttons for 1 second to enter the SETUP mode. The word SETUP will appear in the LCD and the words SOURCE and READ will alternately flash. Release both push-buttons. 3) Press the STORE push-button. The words SETUP and SOURCE will remain on the LCD and the previously selected function will repeatedly flash. 4) Turn the knob (Digital-Pot) to scroll through the available functions. Each...
Page 34
13 OPERATING INSTRUCTIONS USING ENGINEERING UNITS 8) The word SPAN and ENG. UNIT will appear in the bottom half of the LCD prompting you to enter the highest value for the range in the units to match the display of the instrument being calibrated. Turn the knob until the highest value of your chosen range flashes in the bottom half of the LCD. Press the STOREor RESET push-button to continue. 9) The Model 934 will now begin operating with the ranges and Engineering Units you have selected. If Engineering...
Page 35
12 OPERATING INSTRUCTIONS USING ENGINEERING UNITS 6) You are now prompted by flashing zeros to enter the decimal point position for the engineering units you have chosen. Turn the knob until the decimal point appears in the location to match the resolution of the display on the instrument you will be calibrating (from zero to three digits to the right of the decimal point). Press the STORE or RESET push- button to continue. 7) The word ZERO and ENG. UNIT will appear in the top half of the LCD prompting...
Page 36
11 OPERATING INSTRUCTIONS USING ENGINEERING UNITS 3) Turn the knob (Digital pot) to scroll through the available functions. Each enabled function will flash on the LCD.“ ENG. UNIT” will appear as a choice for each SOURCE or READ function. Press the STOREor RESET push-button to proceed. 4) The word ZERO will appear in the top half of the LCD prompting you to enter the lowest value for the range in mA, V or mV that will be sourced or read. Turn the knob until the lowest value of your chosen range flashes...
Page 37
10 OPERATING INSTRUCTIONS USING ENGINEERING UNITS Scale the display of the Model 934 to match the display of the instrument being checked. Any span in mA, mV or V in Source or Read can be scaled to display between a maximum of -90000 to +90000 with zero to three digits to the right of the decimal. The previously stored Engineering units can be recalled by using the following procedure without turning the knob. 1) Press & hold both theSTORE & RESET push- buttons for 1 second to enter the SETUP mode. The...
Page 38
9 OPERATING INSTRUCTIONS RANGES MILLIVOLTS - mV Choose mV to display directly in millivolts DC. Use 0 to 100 millivolts to check recorder or analyzer inputs. SOURCE from 0.00 to 999.90 mVDC. In READ the 934 autoranges between -999.90 to +999.90 millivolts and -99.990 to +99.990 VDC. ENGINEERING UNITS - ENG. UNIT Choose ENG. UNIT to match the display of the instrument being calibrated (see page 10 for instructions). Any span in mA, mV or V in Source or Read can be scaled to display a maximum of between...
Page 39
8 √ OPERATING INSTRUCTIONS RANGES The selected range is displayed on the LCD by the annunciators listed below. MILLIAMPS - mA Choose mA to display directly in milliamps. Display in SOURCE and READ functions from 0.000 to 24.000 mA. In READ you can also measure up to 52.00 mA. PERCENT OF 4 - 20 MILLIAMPS - mA % Choose mA % to display in percent of span for 4 - 20 mA loops. Use with chart recorders or current trips that display in %. 0.00% 4.000 mA 25.00%8.000 mA 50.00%12.000 mA 75.00%16.000 mA 100.00%...
Page 40
BOTTOM BOARD Calibrator Enabled (Jumper Shorted) on off CAL 7 OPERATING INSTRUCTIONS ENABLING ADDITIONAL RANGES & FUNCTIONS The DIP switches are located in the battery compartment. Turn the Model 934 off and loosen the three slotted screws and remove the battery cover. 38 CALIBRATION PROCEDURE SUGGESTED EQUIPMENT 1) Digital voltmeter with accuracy of ±0.003% of reading at 2V and ±0.005% of reading at 100V Important: The required accuracy is usually found only on meters with 71/2 or more digits 2) 100...