    OPERATING INSTRUCTIONS
    POWER & MEASURE 2-WIRE TRANSMITTERS
    PWR XMTR mA, PWR XMTR %, PWR XMTR ENG. UNIT,
    PWR XMTR, % DP FLOW
    Choose this function to simultaneously supply power to a
    2-Wire transmitter while displaying the 4-20 mA output of
    the transmitter.
    1) Disconnect one or both input wires from the 2-Wire
    Transmitter to be calibrated
    2) Connect the red 
    SOURCE lead of the calibrator to the
    plus (+) input of the device and the black SOURCE lead
    to the minus (-)
    3) Connect an appropriate sensor or calibrator to the input
    of the 2-Wire Transmitter
    The Model 934 supplies a nominal 24 Volts DC at 24 mA
    to the 2-Wire transmitter. The current passed by the
    transmitter will be accurately displayed by the 934.
    Calibrate the Transmitter in the usual manner and disconnect
    the Model 934.
    T/C
    RTD
    mA
    mV
    SENSOR OR
    CALIBRATOR
    SENSOROUTPUT
    2-WIRE
    TRANSMITTER
    ALTEK 9344.000mA
    26
    OPERATING INSTRUCTIONS
    7) The words AUTO CHEK and
    TIME appear on the LCD
    prompting you to enter the
    number of seconds between
    steps. Turn the knob until
    the correct number of
    seconds per step (from 5 to
    900) appears on the LCD.
    Press the 
    STORE push-
    button to continue.
    8) Move the QUIK-CHEK
    switch to 
    SET. Press theSTORE push-button to turn
    the AUTO-CHEK on.
    The word AUTO-CHEK will
    flash while AUTO-CHEK is
    active. The ouput pauses at
    each step for the number of
    seconds selected then
    proceeds to the next higher step.
    When the top step is reached
    (the value stored in the HI
    QUIK-CHEK) the AUTO-CHEK
    reverses direction and the
    output proceeds to the next
    lower step. This cycle repeats
    continuously until it is
    deactivated.
    USING AUTO-CHEK
    S
    E
    T
    U
    PSOURCE
    AUTO-CHEK
    TIME
    SOURCE
    AUTO-CHEK mA
    AUTO-CHEK
    10
    8.000
    20.000
    SOURCE
    AUTO-CHEK mA
    4.000
    SOURCE
    mA 
    						
    							20
    OPERATING INSTRUCTIONS
    CALIBRATE VOLTAGE INPUTS
    V, V ENG. UNIT
    mV, mV ENG. UNIT
    Choose this function to provide an output from 0.00 mV
    to 999.90 mV and from 0.000 to 20.000 VDC. Source current
    is a nominal 25 mA to provide the driving power to your
    voltage receivers.
    1) Disconnect one or both input wires from the device to be
    calibrated
    2) Connect the red 
    SOURCE lead of the calibrator to the
    plus (+) input of the device and the black SOURCElead to the minus (-)
    Output voltage is continuously adjustable with the
    QUIK-CHEK switch in the 
    SET position. Zero & Span (or any
    other values) are available by using the LO and HI
    QUIK-CHEKs.
    COMPUTER
    RECEIVER
    LOGGER
    TRANSMITTER
    ALTEK 934
    10.000V
    25
    OPERATING INSTRUCTIONS
    USING AUTO-CHEK
    5) Press the STORE push-
    button to select the function.
    AUTO-CHEK will appear
    on the display along with the
    words on or OFF.
    Rotate the knob until the
    word on appears on the
    display. If AUTO-CHEK has
    not been enabled the word
    AUTO-CHEK will not appear
    - see ENABLING RANGES
    & FUNCTIONS, page 6.
    Press the 
    STORE push-
    button to continue.
    6) The word AUTO CHEK and
    % appear on the LCD
    prompting you to enter the
    percent of span change
    between steps. Turn the knob
    until  the correct percentage
    appears on the LCD. For 2
    steps choose 100%, 3 steps
    choose 50%, 5 steps choose
    25% and for 10 steps choose
    10%. Press the 
    STOREpush-button to continue.S
    E
    T
    U
    P
    SOURCE
    mA
    onAUTO-CHEK
    S
    E
    T
    U
    PSOURCE
    25AUTO-CHEK %
    S
    E
    T
    U
    PSOURCE
    mA
    OFFAUTO-CHEK 
    						
    							21
    OPERATING INSTRUCTIONS
    CALIBRATING LOOP
    CONVERTERS & I/I ISOLATORS
    Calibrate your 4-20 milliamp and voltage loop converters
    and I/I isolators without any additional test equipment or
    power supplies.
    1) Setup the Model 934 to select both the correct SOURCE
    and READ functions and ranges to match the instrument
    being calibrated.
    2) Connect the Model 934 SOURCE and READ leads
    to the input and output connections of the instrument
    being calibrated
    Some loop isolators
    require a resistor
    placed in series with
    the output of the
    isolator of either
    250 Ohms or a value
    equivalent to the
    resistance of the loop.
    Check the manual
    for your isolator for
    specific instructions.
    mA
    ROCHESTER NY 14623ALTEK 934SOURCE20.000mA
    mA/V LOOP CALIBRATOROFF ONHI
    SET
    LOMAX
    READ
    MIN
    9VDC
    AUTO-CHEK HI
    MINREAD
    SOURCE
    STOREQUIK-CHEK ®SETUP
    READ
    RESETSETUP
    500
            MAXSOURCESOURCEREAD
    INPUT OUTPUT
    20.000mA
    24
    OPERATING INSTRUCTIONS
    USING AUTO-CHEK
    Use the AUTO-CHEK to continuously switch the Model
    934s output from Zero to Span (stored in the HI and LO QUIK-
    CHEKs) in 2, 3, 5 or 10 equal steps at an interval of between
    5 and 900 seconds per step. Set your calibrator at one point
    in the field and calibrate every device in the loop.
    1) Store the required Span &
    Zero into the HI & LO QUIK-
    CHEK positions (refer to
    STORING QUIK-CHEK
    OUTPUTS, page 16).
    2) Press & hold both push-
    buttons (
    STORE & RESET)
    for 1 second to enter the
    SETUP mode. The word
    SETUP will appear in theLCD
    and the words SOURCE and
    READ will flash (shown as
    gray).
    3) Press the 
    STORE push-
    button. The words SETUP
    and SOURCE will remain on
    the LCD and the previously
    selected function will flash.
    4) Turn the knob (digital pot) to
    scroll through the available
    functions. Each enabled
    function will flash on the LCD.
    READ
    S
    E
    T
    U
    PSOURCE
    S
    E
    T
    U
    PSOURCE
    mA
    S
    E
    T
    U
    PSOURCE
    V 
    						
    							CONTROLLER
    (TYPICAL)
    TRANSMITTER
    4-20mA
    ADDITIONAL
    4-20 mA OR 1-5V
    INSTRUMENTSRECORDER
    COMPUTER
    LOGGER
    ETC,
    1-5V DC250Ω
    ALTEK 934
    5.000V
    22
    OPERATING INSTRUCTIONS
    CHECK 1-5 VOLT INPUTS WITHOUT
    DISCONNECTING WIRES
    Most 1-5 Volt receivers in 4-20mA loops have a 250 Ohm
    resistor across the input of the receiver. This resistor may be
    mounted internally or externally. The Model 934 is connected
    directly across the input of the 1-5 Volt receiver without
    disconnecting any field wiring. This saves a great deal of time
    when a large number of voltage receivers, such as chart
    recorders or computer systems require calibration.
    Output voltage is continuously adjustable with the QUIK-
    CHEK switch in the 
    SET position. Zero & Span (or any other
    values) are available by using the LO and HI QUIK-CHEKs.
    The LO & HI “QUIK-CHEK”s are usually set to 1.000 and
    5.000 V to quickly calibrate Zero & Span.
    Make certain that changing the signal input will not
    disturb the process or cause unexpected alarms when checking
    on-line instruments. It is important to remember the Model 934
    drives 
    only the device to which it is connected. It has no
    effect on other devices in the 4 to 20 mA loop. The Model 934
    will clamp the selected value in the mV and V Ranges to
    the maximum sink current of >16 mA.
    23
    OPERATING INSTRUCTIONS
    V, V ENG. UNIT
    mV, mV ENG. UNIT
    Choose this function to calibrate any 1 to 5 Volt device in
    a 4 to 20 mA loop without breaking the loop or turning off
    the signal current.
    1) Disconnect one or both input wires from the device to be
    calibrated.
    2) Connect the red 
    SOURCE lead of the calibrator to the
    plus (+) input of the device and the black SOURCE lead
    to the minus (-). Any associated 250 Ohm resistor must
    not be disconnected. 
    						
    							CONTROLLER
    (TYPICAL)
    TRANSMITTER
    4-20mA
    ADDITIONAL
    4-20 mA OR 1-5V
    INSTRUMENTSRECORDER
    COMPUTER
    LOGGER
    ETC,
    1-5V DC250Ω
    ALTEK 934
    5.000V
    22
    OPERATING INSTRUCTIONS
    CHECK 1-5 VOLT INPUTS WITHOUT
    DISCONNECTING WIRES
    Most 1-5 Volt receivers in 4-20mA loops have a 250 Ohm
    resistor across the input of the receiver. This resistor may be
    mounted internally or externally. The Model 934 is connected
    directly across the input of the 1-5 Volt receiver without
    disconnecting any field wiring. This saves a great deal of time
    when a large number of voltage receivers, such as chart
    recorders or computer systems require calibration.
    Output voltage is continuously adjustable with the QUIK-
    CHEK switch in the 
    SET position. Zero & Span (or any other
    values) are available by using the LO and HI QUIK-CHEKs.
    The LO & HI “QUIK-CHEK”s are usually set to 1.000 and
    5.000 V to quickly calibrate Zero & Span.
    Make certain that changing the signal input will not
    disturb the process or cause unexpected alarms when checking
    on-line instruments. It is important to remember the Model 934
    drives 
    only the device to which it is connected. It has no
    effect on other devices in the 4 to 20 mA loop. The Model 934
    will clamp the selected value in the mV and V Ranges to
    the maximum sink current of >16 mA.
    23
    OPERATING INSTRUCTIONS
    V, V ENG. UNIT
    mV, mV ENG. UNIT
    Choose this function to calibrate any 1 to 5 Volt device in
    a 4 to 20 mA loop without breaking the loop or turning off
    the signal current.
    1) Disconnect one or both input wires from the device to be
    calibrated.
    2) Connect the red 
    SOURCE lead of the calibrator to the
    plus (+) input of the device and the black SOURCE lead
    to the minus (-). Any associated 250 Ohm resistor must
    not be disconnected. 
    						
    							21
    OPERATING INSTRUCTIONS
    CALIBRATING LOOP
    CONVERTERS & I/I ISOLATORS
    Calibrate your 4-20 milliamp and voltage loop converters
    and I/I isolators without any additional test equipment or
    power supplies.
    1) Setup the Model 934 to select both the correct SOURCE
    and READ functions and ranges to match the instrument
    being calibrated.
    2) Connect the Model 934 SOURCE and READ leads
    to the input and output connections of the instrument
    being calibrated
    Some loop isolators
    require a resistor
    placed in series with
    the output of the
    isolator of either
    250 Ohms or a value
    equivalent to the
    resistance of the loop.
    Check the manual
    for your isolator for
    specific instructions.
    mA
    ROCHESTER NY 14623ALTEK 934SOURCE20.000mA
    mA/V LOOP CALIBRATOROFF ONHI
    SET
    LOMAX
    READ
    MIN
    9VDC
    AUTO-CHEK HI
    MINREAD
    SOURCE
    STOREQUIK-CHEK ®SETUP
    READ
    RESETSETUP
    500
            MAXSOURCESOURCEREAD
    INPUT OUTPUT
    20.000mA
    24
    OPERATING INSTRUCTIONS
    USING AUTO-CHEK
    Use the AUTO-CHEK to continuously switch the Model
    934s output from Zero to Span (stored in the HI and LO QUIK-
    CHEKs) in 2, 3, 5 or 10 equal steps at an interval of between
    5 and 900 seconds per step. Set your calibrator at one point
    in the field and calibrate every device in the loop.
    1) Store the required Span &
    Zero into the HI & LO QUIK-
    CHEK positions (refer to
    STORING QUIK-CHEK
    OUTPUTS, page 16).
    2) Press & hold both push-
    buttons (
    STORE & RESET)
    for 1 second to enter the
    SETUP mode. The word
    SETUP will appear in theLCD
    and the words SOURCE and
    READ will flash (shown as
    gray).
    3) Press the 
    STORE push-
    button. The words SETUP
    and SOURCE will remain on
    the LCD and the previously
    selected function will flash.
    4) Turn the knob (digital pot) to
    scroll through the available
    functions. Each enabled
    function will flash on the LCD.
    READ
    S
    E
    T
    U
    PSOURCE
    S
    E
    T
    U
    PSOURCE
    mA
    S
    E
    T
    U
    PSOURCE
    V 
    						
    							20
    OPERATING INSTRUCTIONS
    CALIBRATE VOLTAGE INPUTS
    V, V ENG. UNIT
    mV, mV ENG. UNIT
    Choose this function to provide an output from 0.00 mV
    to 999.90 mV and from 0.000 to 20.000 VDC. Source current
    is a nominal 25 mA to provide the driving power to your
    voltage receivers.
    1) Disconnect one or both input wires from the device to be
    calibrated
    2) Connect the red 
    SOURCE lead of the calibrator to the
    plus (+) input of the device and the black SOURCElead to the minus (-)
    Output voltage is continuously adjustable with the
    QUIK-CHEK switch in the 
    SET position. Zero & Span (or any
    other values) are available by using the LO and HI
    QUIK-CHEKs.
    COMPUTER
    RECEIVER
    LOGGER
    TRANSMITTER
    ALTEK 934
    10.000V
    25
    OPERATING INSTRUCTIONS
    USING AUTO-CHEK
    5) Press the STORE push-
    button to select the function.
    AUTO-CHEK will appear
    on the display along with the
    words on or OFF.
    Rotate the knob until the
    word on appears on the
    display. If AUTO-CHEK has
    not been enabled the word
    AUTO-CHEK will not appear
    - see ENABLING RANGES
    & FUNCTIONS, page 6.
    Press the 
    STORE push-
    button to continue.
    6) The word AUTO CHEK and
    % appear on the LCD
    prompting you to enter the
    percent of span change
    between steps. Turn the knob
    until  the correct percentage
    appears on the LCD. For 2
    steps choose 100%, 3 steps
    choose 50%, 5 steps choose
    25% and for 10 steps choose
    10%. Press the 
    STOREpush-button to continue.S
    E
    T
    U
    P
    SOURCE
    mA
    onAUTO-CHEK
    S
    E
    T
    U
    PSOURCE
    25AUTO-CHEK %
    S
    E
    T
    U
    PSOURCE
    mA
    OFFAUTO-CHEK 
    						
    							19
    OPERATING INSTRUCTIONS
    POWER & MEASURE 2-WIRE TRANSMITTERS
    PWR XMTR mA, PWR XMTR %, PWR XMTR ENG. UNIT,
    PWR XMTR, % DP FLOW
    Choose this function to simultaneously supply power to a
    2-Wire transmitter while displaying the 4-20 mA output of
    the transmitter.
    1) Disconnect one or both input wires from the 2-Wire
    Transmitter to be calibrated
    2) Connect the red 
    SOURCE lead of the calibrator to the
    plus (+) input of the device and the black SOURCE lead
    to the minus (-)
    3) Connect an appropriate sensor or calibrator to the input
    of the 2-Wire Transmitter
    The Model 934 supplies a nominal 24 Volts DC at 24 mA
    to the 2-Wire transmitter. The current passed by the
    transmitter will be accurately displayed by the 934.
    Calibrate the Transmitter in the usual manner and disconnect
    the Model 934.
    T/C
    RTD
    mA
    mV
    SENSOR OR
    CALIBRATOR
    SENSOROUTPUT
    2-WIRE
    TRANSMITTER
    ALTEK 9344.000mA
    26
    OPERATING INSTRUCTIONS
    7) The words AUTO CHEK and
    TIME appear on the LCD
    prompting you to enter the
    number of seconds between
    steps. Turn the knob until
    the correct number of
    seconds per step (from 5 to
    900) appears on the LCD.
    Press the 
    STORE push-
    button to continue.
    8) Move the QUIK-CHEK
    switch to 
    SET. Press theSTORE push-button to turn
    the AUTO-CHEK on.
    The word AUTO-CHEK will
    flash while AUTO-CHEK is
    active. The ouput pauses at
    each step for the number of
    seconds selected then
    proceeds to the next higher step.
    When the top step is reached
    (the value stored in the HI
    QUIK-CHEK) the AUTO-CHEK
    reverses direction and the
    output proceeds to the next
    lower step. This cycle repeats
    continuously until it is
    deactivated.
    USING AUTO-CHEK
    S
    E
    T
    U
    PSOURCE
    AUTO-CHEK
    TIME
    SOURCE
    AUTO-CHEK mA
    AUTO-CHEK
    10
    8.000
    20.000
    SOURCE
    AUTO-CHEK mA
    4.000
    SOURCE
    mA 
    						
    							18
    OPERATING INSTRUCTIONS
    SIMULATE 2-WIRE TRANSMITTERS
    2-WIRE SIM mA, 2-WIRE SIM %, 2-WIRE SIM ENG. UNIT,
    2-WIRE SIM % DP FLOW
    Choose this function to simulate a 2-Wire Transmitter output
    from 3.000 to 24.000 milliamps. Operates in loops with power
    supply voltages from 2 to 45 VDC.
    1) Disconnect existing 2-Wire Transmitter from the loop
    2) Connect the red 
    SOURCE lead of the calibrator to the
    plus (+) input of the field connections and the black
    SOURCE lead to the minus (-)
    The simulated output of the 2-Wire Transmitter is continuously
    adjustable from 3.000 to 24.000 mA with the QUIK-CHEK
    switch in the 
    SET position. Zero & Span (or any other values)
    are available by using the LO and HI QUIK-CHEKs.
    RECEIVER
    (Powers External
    2-Wire Transmitters)
    POWER
    SUPPLY
    2 to
    100 VDCZEROSENSORSPAN
    2-WIRE TRANSMITTER
    (Disconnected)
    ALTEK 934
    100.00%
    27
    OPERATING INSTRUCTIONS
    USING AUTO-CHEK
    You can deactivate the
    AUTO-CHEK function at any
    time by pressing the 
    STOREpush-button  while the QUIK-
    CHEK switch is at SET or by
    moving the QUIK-CHEK switch
    to 
    HI or LO. AUTO-CHEK will
    stop flashing to indicate that it
    is deactivated. 
    SET will indicate
    the the most recent step value.
    While AUTO-CHEK is
    deactivated the QUIK-CHEK
    values stored in 
    HI and LO can
    be recalled to adjust the Span
    and Zero of the device being
    calibrated. Or store new values
    by switching to 
    HI and LO,
    turning the digital pot and
    pressing the 
    STOREpushbutton.
    To reactivate AUTO-CHEK
    press the 
    STORE push-button
    while the QUIK-CHEK switch is
    at 
    SET. When AUTO-CHEK is
    reactivated it restarts the cycle
    from the lowest output (stored
    in the 
    LO QUIK-CHEK).
    SOURCE
    AUTO-CHEK mA
    24.000
    HISOURCE
    AUTO-CHEK mA
    20.000
    HI
    SOURCE
    AUTO-CHEK mA
    4.000 
    						
    							17
    ALTEK 934
    20.000mA
    VA LV E
    OR
    RECEIVER I/P
    OPERATING INSTRUCTIONS
    SOURCE FUNCTIONS
    The SOURCE functions generate or control the desired
    signal. Multiple scales are available for each function.
    Engineering Units & % DP Flow must be enabled before they
    are available for selection.
    CALIBRATE MILLIAMP INPUTS
    mA, mA %, mA ENG. UNIT, mA % DP FLOW
    Choose this function to provide an output from 0.000 to
    24.000 milliamps. The compliance voltage is a nominal 24 VDC
    to provide the driving power to your milliamp receivers.
    1) Disconnect one or both input wires from the device to be
    calibrated
    2) Connect the red 
    SOURCE lead of the calibrator to the
    plus (+) input of the device and the black SOURCE lead
    to the minus (-)
    Output current is continuously adjustable with the QUIK-
    CHEK switch in the 
    SET position. Zero & Span (or any other
    values) are available by using the LO and HI QUIK-CHEKs.
    28
    OPERATING INSTRUCTIONS
    READ MODE
    READ
    1) Move the POWER switch to ON
    2) Press & hold both the STORE & RESET push-buttons for
    1 second to enter the SETUP mode. The word SETUP
    will appear in the LCD and the words SOURCE and READ
    will alternately flash. Release both push-buttons.
    3) Press the 
    RESET push-button. The words SETUP and
    READ will remain on the LCD and the previously selected
    function will flash.
    4) Turn the knob (Digital pot) to scroll through the available
    functions. Each enabled function will flash on the LCD
    (See ENABLING RANGES & FUNCTIONS under
    OPERATING INSTRUCTIONS, GENERAL).
    3) Press the 
    RESET push-button to select the function.  If OFF
    is selected the READ function will be deactivated and the
    bottom half of the LCD will be blank.
    4) Connect the Model 934 to the input terminals of the
    instrument or meter to be calibrated.
    5) Display the present reading, Maximum or Minimum
    Whenever READ is selected the word READ will appear on
    the LCD display.  The display will update the present reading
    twice per second. 
    						