    Calibration Instructions for the 
    Standard Version of the SJ-H/HS series 
    When is to Calibration Required? 
    Calibration may be required when the SJ-H/HS series initially installed or if the scale is 
    moved a substantial distance. 
     
    Calibration using a Calibration Weight 
    1.  Entering calibration mode. 
      Press the [ON/OFF] key to turn the power off.  While pressing 
    both the [RE-ZERO] and [UNITS] keys, press the [ON/OFF] key. 
    Then 
     Cal    will be displayed. 
      Cal    can also be displayed by removing the calibration switch 
    cover at the bottom of the device and pressing the calibration 
    switch while the device is in the Weighing Mode. 
    2. Zero calibration 
      Press the [RE-ZERO] key to calibrate zero.  Cal0  will be displayed. 
      Wait for the Stable Indicator to be displayed, then press the [RE-ZERO] key.   
     Calf   will be displayed after a few seconds.   Press the [UNITS] key to perform only a 
    Zero Calibration.    The scale will then automatically return to the Weighing Mode. 
    3. Span calibration 
     When  Calf  is displayed, place the weight at the center of the platform.  Wait for the 
    Stable Indicator to be displayed, then press the [RE-ZERO] key.    The display will show 
     end   and the scale will automatically return to the Weighing Mode. 
    See “SPECIFICATIONS” about the calibration weight. 
     
    Calibration by Gravity Compensation 
    If the acceleration of gravity at your location is not 9.798 m/s2 and you do not have 
    calibration weights, the scale can be calibrated by compensating for the acceleration of 
    gravity.    (Refer to “The Value of Gravity at Various Locations”.) 
    1.  Setting a new acceleration value. 
     When  Cal    is displayed, press the [UNITS] key and  9.
    798   will be displayed. Press 
    the [RE-ZERO] key to increment the blinking digit and press the [UNITS] key to move the 
    blinking digit. 
    2.  Storing the value in the memory. 
      While pressing the [UNITS] key, press and hold the [RE-ZERO] key and release the 
    [UNITS] key.    Then the display will show 
     end   and return to  Cal  .    Turn the scale 
    off to finish the procedure. 
     
     
    Calibration switch cover
    Calibration switch
    1WMPD4001896 
    						
    							2/2 
    SPECIFICATIONS 
    MODEL SJ-1000H/HS SJ-2000H/HS SJ-5000H/HS SJ-5001H/HS SJ-12KH/HS
    (k)g  1000 g x 0.5 g  2000 g x 1 g  5000 g x 2 g  5000 g x 1g  12 kg x 5g 
    lb  2.2 lb x 0.001 lb  4.4 lb x 0.002 lb 11 lb x 0.005 lb 11 lb x 0.005 lb  26 lb x 0.01 lb
    oz  35 oz x 0.02 oz  70 oz x 0.05 oz 176 oz x 0.1 oz 176 oz x 0.1 oz  423 oz x 0.2 oz
    Capacity 
    x 
    Resolution 
    lb-oz  2.2 lb x 0.1 oz  4.4 lb x 0.1 oz 11 lb x 0.1 oz  11 lb x 0.1 oz  26 lb x 1 oz 
    Calibration weight 1000g ± 0.1g 2000g ± 0.2g 5000g ± 0.5g 5000g ± 0.5g 10 kg ± 1g 
     
     
    The Value of Gravity at Various Locations 
    Amsterdam 9.813 m/s2 
    Athens 9.807 m/s
    2 
    Auckland NZ  9.799 m/s
    2 
    Bangkok 9.783 m/s
    2 
    Birmingham 9.813 m/s
    2 
    Brussels 9.811 m/s
    2 
    Buenos Aires  9.797  m/s
    2 
    Calcutta 9.788 m/s
    2 
    Cape Town  9.796 m/s
    2 
    Chicago 9.803 m/s
    2 
    Copenhagen 9.815 m/s
    2 
    Cyprus 9.797 m/s
    2 
    Djakarta 9.781 m/s
    2 
    Frankfurt 9.810 m/s
    2 
    Glasgow 9.816 m/s
    2 Havana 9.788 m/s
    2 
    Helsinki 9.819 m/s
    2 
    Kuwait 9.793 m/s
    2 
    Lisbon 9.801 m/s
    2 
    London (
    Greenwich) 9.812 m/s2 
    Los Angeles  9.796  m/s
    2 
    Madrid 9.800 m/s
    2 
    Manila 9.784 m/s
    2 
    Melbourne 9.800 m/s
    2 
    Mexico City  9.779 m/s
    2 
    Milan 9.806 m/s
    2 
    New York  9.802 m/s
    2 
    Oslo 9.819 m/s
    2 
    Ottawa 9.806 m/s
    2 
    Paris 9.809 m/s
    2 Rio de Janeiro  9.788 m/s
    2 
    Rome 9.803 m/s
    2 
    San Francisco  9.800 m/s
    2 
    Singapore 9.781 m/s
    2 
    Stockholm 9.818 m/s
    2 
    Sydney 9.797 m/s
    2 
    Taichung 9.789 m/s
    2 
    Vancouver, BC  9.809 m/s
    2 
    Washington DC  9.801 m/s
    2 
    Wellington NZ  9.803 m/s
    2 
    Zurich 9.807 m/s
    2 
    Taiwan 9.788 m/s
    2 
    Taipei 9.790 m/s
    2 
    Tokyo 9.798 m/s
    2 
     
     
     
     
      
    						