HJ-150 Compact Scale
Operating Instructions
Cautions
‰ Avoid installing / leaving the scale in direct sunlight / high humidity.
‰ Place the scale on a flat and solid surface in a place free from wind.
‰ Avoid vibration, strong electric or magnetic fields and rapid changes in
temperature or humidity.
‰ Avoid using the scale near other electronic equipment.
‰ Do not mix the battery types. Replace all of the batteries at same time.
‰ If the scale is not to be used for a long period of time,...
1
Calibration switch cover
Calibration switch
HT series Compact Scale
Calibration and Function Setting Procedures
Calibration
When is Calibration Required?
Calibration may be required when the HT series compact scale was initially installed or has been moved to
another location.
Calibration Using a Calibration Weight
1. Entering the calibration mode
Remove the calibration switch cover located on the bottom of the scale.
Press the [ON:OFF] switch to turn the scale on.
Press the...
Jack de
l’adaptateur
à courant Couvercle des piles
Piles «AA»
Ruban isolant
Couvercle en
caou
tchouc
HT series Compact Scale
Operating Instructions
Cautions ‰
Avoid installing the scale in direct sunlight.
‰
Avoid over / shock load that could cause damage to the scale. ‰
Use the scale under conditions of -10°C ~ 40°C, less than 85% R.H. ‰
Do not mix the type of batteries. Replace the batteries all together. ‰
If the scale is not to be used for a long period of time, remove all...
1WMPD4002673
HT-500GD
1WMPD4002178
Jewelry Scale
SF/SG Series
Price Computing Scales
Instruction ManualWEIGHT
PRICE
UNIT PRICE
88888
8888888
888888
kg
$
O
Net Zero
Total M+
((XCEMEM
Change*
T
M+
9101112131415234567
81617
1
242526272829301819202122231234560
79
C
8
Change
125678234
11
WEIGHT kg UNIT PRICE $/kg PRICE $
TTLM+CHANGE
NETZERO 88888 8888888888888
C3456897 021M+*OOOCEMEMO
((XT1
910SF Series SG Series
$/kgSF-15KA
Max. 15kg Min 0.1kg e=0.005kg ChangeWM:PD4000340
1/2
Calibration Instructions for the
Standard Version of the SJ-H/HS series
When is to Calibration Required?
Calibration may be required when the SJ-H/HS series initially installed or if the scale is
moved a substantial distance.
Calibration using a Calibration Weight
1. Entering calibration mode.
Press the [ON/OFF] key to turn the power off. While pressing
both the [RE-ZERO] and [UNITS] keys, press the [ON/OFF] key.
Then
Cal will be displayed.
Cal can also be displayed...