    							HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 51   
       
    8.1.1.3 
    Common Modules 
    Table 8.1-3 
    PATIENT MODULE OF CREATED SOP INSTANCES 
    Attribute Name  Tag VR  Value  Presence 
    of Value  Source
    Patients Name 
    0010,0010 PN From MWL or User Input. Values 
    supplied via Modality Worklist will 
    be entered as received. Values 
    supplied via user input will contain 
    first 3 components 
    (Last^First^Middle). Maximum 64 
    characters. 
    VNAP 
    MWL/ 
    USER 
    Patient ID 
    0010,0020 LO From MWL, user input or 
    generated by device. Maximum 64 
    characters. 
    ALWAYS MWL/ 
    USER/
    AUTO 
    Patients Birth Date 
    0010,0030 DA From MWL or user input  VNAP MWL/ 
    USER 
    Patients Sex 
    0010,0040 CS From MWL or user input  VNAP MWL/ 
    USER 
     
    Table 8.1-4 
    GENERAL STUDY MODULE OF CREATED SOP INSTANCES 
    Attribute Name  Tag VR  Value  Presence 
    of Value  Source
    Study Instance UID 
    0020,000D UI From MWL or generated by device  ALWAYS MWL/ 
    AUTO 
    Study Date 
    0008,0020 DA  ALWAYS  AUTO 
    Study Time 0008,0030 TM  ALWAYS  AUTO 
    Referring Physicians 
    Name 0008,0090 
    PN From MWL or user input  VNAP MWL/ 
    USER 
    Study ID 
    0020,0010 SH System generated : Study Date + 
    Study Time 
     
    ALWAYS AUTO 
    Accession Number 
    0008,0050 SH From MWL or user input  VNAP MWL/ 
    USER 
    Study Description 
    0008,1030 LO From MWL or user input 
    In case of MWL, this system uses 
    one of the following attributes 
    according to the priority order. 
    Priority 1 : (0032,1060) Requested 
    Procedure Description 
    Priority 2 : (0040,0007) Scheduled 
    Procedure Step Description 
    Priority 3 : (0040,1002) Reason for 
    the Requested Procedure 
    Priority 4 : (0040,2001) Reason for 
    the Imaging Service Request  ANAP 
    MWL/ 
    USER  
    						
    							HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 52   
       
    Referenced Study 
    Sequence 
    0008,1110 SQ 
    From MWL  ANAP MWL 
    > Referenced SOP Class 
    UID  0008,1150 UI 
    From MWL  ANAP MWL 
    > Referenced SOP 
    Instance UID  0008,1155 UI 
    From MWL  ANAP MWL 
    Procedure Code Sequence  0008,1032 SQ From MWL  ANAP MWL 
     
     
     
     
    Table 8.1-5 
    PATIENT STUDY MODULE OF CREATED SOP INSTANCES 
    Attribute Name  Tag VR  Value  Presence 
    of Value  Source 
    Patients Size 
    0010,1020 DS From MWL or user input  ANAP MWL/ 
    USER 
    Patients Weight 
    0010,1030 DS From MWL or user input  ANAP MWL/ 
    USER 
     
    Table 8.1-6 
    GENERAL SERIES MODULE OF CREATED SOP INSTANCES 
    Attribute Name  Tag VR  Value  Presence 
    of Value  Source 
    Modality 0008,0060 
    CS US  ALWAYS AUTO 
    Series Instance UID  0020,000E UI Generated by device  ALWAYS AUTO 
    Series Number  0020,0011 IS 1  ALWAYS AUTO 
    Series Date  0008,0021 DA   ALWAYS AUTO 
    Series Time  0008,0031 TM   ALWAYS AUTO 
    Protocol Name  0018,1030 LO FreeForm  ALWAYS AUTO 
    Operators’ Name  0008,1070 PN From user input  ANAP USER 
    Referenced Performed 
    Procedure Step Sequence  0008,1111 SQ Identifies the MPPS SOP Instance 
    to which this image is related 
    ALWAYS MPPS 
    > Referenced SOP Class 
    UID 0008,1150 UI MPPS SOP Class UID 
    1.2.840.10008.3.1.2.3.3 
    ALWAYS MPPS 
    > Referenced SOP Instance 
    UID 0008,1155 
    UI MPPS SOP Instance UID  ALWAYS MPPS 
    Request Attributes 
    Sequence  0040,0275 
    SQ Zero or 1 item will be present  ANAP AUTO 
    > Requested Procedure ID  0040,1001 SH From MWL  ANAP MWL 
    > Scheduled Procedure 
    Step ID  0040,0009 SH 
    From MWL  ANAP MWL  
    						
    							HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 53   
       
    > Scheduled Procedure 
    Step Description 
    0040,0007 LO 
    From MWL  ANAP MWL 
    > Scheduled Protocol Code 
    Sequence  0040,0008 SQ 
    From MWL  ANAP MWL 
    Performed Procedure Step 
    ID  0040,0253 
    SH Same as MPPS  ALWAYS MPPS 
    Performed Procedure Step 
    Start Date  0040,0244 
    DA Same as St udy Date  ALWAYS AUTO 
    Performed Procedure Step 
    Start Time  0040,0245 
    TM Same as St udy Time  ALWAYS AUTO 
    Performed Procedure Step 
    Description  0040,0254 
    LO Same as Study Description  ANAP MWL/ 
    USER 
     
    Table 8.1-7 
    GENERAL EQUIPMENT MODULE OF CREATED SOP INSTANCES 
    Attribute Name  Tag VR  Value  Presence 
    of Value  Source 
    Manufacturer 0008,0070 
    LO Philips ALWAYS  AUTO 
    Institution Name 0008,0080 LO From user input  ANAP CONFIG 
    Station Name  0008,1010 SH From user input  ANAP CONFIG 
    Manufacturers Model Name  0008,1090 LO  HD9  ALWAYS AUTO 
    Device Serial Number  0018,1000 LO  Generated by device  ALWAYS AUTO 
    Software Versions  0018,1020 LO Generated by device  ALWAYS AUTO 
     
    8.1.1.4 US or US Multiframe Image Module 
    Table 8.1-8 
    GENERAL IMAGE MODULE OF CREATED US OR US MULTIFRAME SOP INSTANCES 
    Attribute Name  Tag VR  Value  Presence 
    of Value  Source 
    Instance Number 
    0020,0013 IS Generated by device, increments 
    from 1 in each series 
    ALWAYS AUTO 
    Patient Orientation 
    0020,0020 CS NULL  ALWAYS AUTO 
    Content Date  0008,0023 DA   ALWAYS AUTO 
    Content Time  0008,0033 TM   ALWAYS AUTO 
    Image Type  0008,0008 CS ORIGINAL and PRIMARY   ALWAYS AUTO 
    Lossy Image Compression  0028,2110 CS US = 00 (uncompressed), US-
    MF = 01 (lossy compressed)  ALWAYS AUTO 
    Lossy Image Compression 
    Ratio 
    0028,2112 DS Used if (0028, 2110) = 01, 
    Calculated by device 
    ANAP AUTO 
    Lossy Image Compression 
    Method 0028,2114 CS ISO_10918_1, used if 
    (0028,2001) = 01  ANAP AUTO 
     
    Table 8.1-9  
    						
    							HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 54   
       
    IMAGE PIXEL MODULE OF CREATED US OR US MULTIFRAME SOP INSTANCES 
    Attribute Name 
    Tag VR  Value  Presence 
    of Value  Source 
    Samples per Pixel 
    0028,0002 US 3 for RGB or YBR_FULL_422 
    “1” for MONOCHROME2  ALWAYS AUTO 
    Photometric Interpretation 
    0028,0004 CS US =RGB or “MONOCHROME2” 
    USMF = “YBR_FULL_422” or 
    “MONOCHROME2” 
    ALWAYS AUTO 
    Rows 0028,0010 
    US  US = 768,    US-MF = 480  ALWAYS AUTO 
    Columns  0028,0011 US US = 102 4, US-MF = 640  ALWAYS AUTO 
    Bits Allocated  0028,0100 US 8  ALWAYS AUTO 
    Bits Stored  0028,0101 US 8  ALWAYS AUTO 
    High Bit  0028,0102 US 7  ALWAYS AUTO 
    Pixel Representation  0028,0103 US  0  ALWAYS AUTO 
    Pixel Data  7FE0,0010 OW 
    or 
    OB 
    Generated by device 
    ALWAYS AUTO 
    Planar Configuration  0028,0006 US 0, Present only if (0028,0002) = 3  ANAP AUTO 
    Private Creator  7FE1,0010 LO “MEDISON_US”  ANAP AUTO 
    3D Volume  7FE1,1002 OB 3D Volume Data  ANAP AUTO 
     
     
    Table 8.1-10a 
    IMAGE PLANE MODULE OF CREATED US  OR US MULTIFRAME SOP INSTANCES 
    Attribute Name Tag VR  Value  Presence 
    of Value  Source 
    Pixel Spacing 
    0028,0030 DS Physical distance between the 
    centers of pixels in mm.    Only 
    present when user configures its 
    use and images have only one 
    region or if dual, the same depth. 
    ANAP USER 
     
     
    Table 8.1-11 
    CINE MODULE OF CREATED US MULTIFRAME SOP INSTANCES 
    Attribute Name  Tag VR  Value  Presence 
    of Value  Source 
    Frame Time 
    0018,1063 DS Milliseconds ALWAYS*  AUTO 
    Cine Rate 0018,0040 IS Frames  per second  ALWAYS* AUTO 
     
    * This module is only present in US Multiframe Images, and is r equired in US Multiframe Images.    Therefore the Presence of Valu e is set 
    to ALWAYS because the Module is only used in US Multiframe, not US Image. 
     
     
     
      
    						
    							HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 55   
       
    Table 8.1-12 
    MULTI-FRAME MODULE OF CREATED  US MULTIFRAME SOP INSTANCES 
    Attribute Name Tag VR  Value  Presence 
    of Value  Source 
    Number of Frames 
    0028,0008 IS  Numbers of Frames ALWAYS*  AUTO 
    Frame Increment Pointer 0028,0009 AT Frame Time : (0018, 1063)  ALWAYS* AUTO 
     
    * This module is only present in US Multiframe Images, and is r equired in US Multiframe Images.    Therefore the Presence of Valu e is set 
    to ALWAYS because the Module is only used in US Multiframe, not US Image. 
     
    Table 8.1-13 
    US REGION CALIBRATION MODULE OF CREATED US OR US MULTIFRAME SOP INSTANCES 
    Attribute Name  Tag VR  Value  Presence 
    of Value  Source 
    Sequence of Ultrasound 
    Regions  0018,6011 SQ Generated by device. A sequence 
    is present for each region in the 
    system display. 
    ANAP AUTO 
    > Region Location Min x0 
    0018,6018 UL Left position of region  ALWAYS AUTO 
    > Region Location Min y0  0018,601A UL   Top position of region  ALWAYS AUTO 
    > Region Location Max x1  0018,601C UL Right position of region  ALWAYS AUTO 
    > Region Location Max y1  0018,601E UL   Bottom position of region  ALWAYS AUTO 
    > Physical Units X Direction  0018,6024 US 2D Image : 0003H = cm 
    M-Mode : 0004H = seconds 
    Doppler : 0004H = seconds 
    ALWAYS AUTO 
    > Physical Units Y Direction 
    0018,6026 US 2D Image : 0003H = cm 
    M-Mode :    0003H = cm 
    Doppler :    0005H = hertz or 
               0007H = cm/sec 
    ALWAYS AUTO 
    > Physical Delta X 
    0018,602C FD The physical value per pixel 
    increment  ALWAYS AUTO 
    > Physical Delta Y 
    0018,602E FD The physical value per pixel 
    increment 
    ALWAYS AUTO 
    > Region Spatial Format 
    0018,6012 US 2D Tissue : 0001H 
    M-Mode Tissue or flow : 0002H 
    Spectral (CW or PW Doppler) : 
    0003H 
    ALWAYS AUTO 
    > Region Data Type 
    0018,6014 US Tissue : 0001H 
    Color Flow : 0002H 
    PW Spectral Doppler : 0003H 
    CW Spectral Doppler : 0004H 
    ALWAYS AUTO 
    > Region Flags 
    0018,6016 UL  See DICOM PS 3.3    C.8.5.5.1.3  ALWAYS AUTO 
     
     
     
     
     
      
    						
    							HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 56   
       
    Table 8.1-14 
    US IMAGE MODULE OF CREATED US OR US MULTIFRAME SOP INSTANCES 
    Attribute Name  Tag VR  Value  Presence 
    of Value  Source 
    Samples Per Pixel 
    0028,0002 US 3 for RGB or YBR_FULL_422 
    “1” for MONOCHROME2 
    ALWAYS AUTO 
    Photometric Interpretation   
    0028,0004 CS US =RGB or “MONOCHROME2” 
    USMF = “YBR_FULL_422” or 
    “MONOCHROME2” 
    ALWAYS AUTO 
    Bits Allocated 
    0028,0100 US 8  ALWAYS AUTO 
    Bits Stored  0028,0101 US 8  ALWAYS AUTO 
    High Bit  0028,0102 US 7  ALWAYS AUTO 
    Planar Configuration  0028,0006 US 0, Present only if  (0028,0002) = 3 ALWAYS AUTO 
    Pixel Representation  0028,0103 US  0  ALWAYS AUTO 
    Image Type  0008,0008 CS ORIGINAL and PRIMARY   ALWAYS AUTO 
    Lossy Image Compression  0028,2110 CS US = 00 (uncompressed), US-MF 
    = 01 (lossy compressed) 
    ALWAYS AUTO 
     
     
    Table 8.1-15 
    VOI LUT MODULE OF CREATED US OR US MULTIFRAME SOP INSTANCES 
    Attribute Name  Tag VR  Value  Presence 
    of Value  Source 
    Window Center 
    0028,1050 DS  default : 128  ALWAYS CONFIG 
    Window Width  0028,1051 DS default : 256  ALWAYS CONFIG 
     
     
    Table 8.1-16 
    SOP COMMON MODULE OF CREATED US  OR US MULTIFRAME SOP INSTANCES 
    Attribute Name Tag VR  Value  Presence 
    of Value  Source 
    SOP Class UID 
    0008,0016 UI US  = 
    1.2.840.10008.5.1.4.1.1.6.1 
    or 
    US-MF = 
    1.2.840.10008.5.1.4.1.1.3.1  ALWAYS AUTO 
    SOP Instance UID 
    0008,0018 UI Generated by device  ALWAYS AUTO 
    Specific Character Set  0008,0005 CS Only when a replacement 
    character set is used. 
    ANAP AUTO 
     
      
    						
    							HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 57   
       
     
    8.1.1.5 
    Comprehensive Structured Report Modules 
     
    Table 8.1-17 
    SR DOCUMENT SERIES MODULE OF CREA TED COMPREHENSIVE SR SOP INSTANCES 
    Attribute Name Tag VR  Value  Presence 
    of Value  Source 
    Modality 0008,0060 
    CS SR  ALWAYS AUTO 
    Series Instance UID  0020,000E UI Generated by device  ALWAYS AUTO 
    Series Number  0020,0011 IS 2  ALWAYS AUTO 
    Referenced Performed 
    Procedure Step Sequence  0008,1111 SQ Identifies the MPPS SOP Instance 
    to which this image is related 
    ALWAYS MPPS 
    > Referenced SOP Class 
    UID 0008,1150 UI MPPS SOP Class UID 
    1.2.840.10008.3.1.2.3.3 
    ALWAYS MPPS 
    > Referenced SOP 
    Instance UID 0008,1155 
    UI MPPS SOP Instance UID  ALWAYS MPPS 
     
     
    Table 8.1-18 
    SR DOCUMENT GENERAL MODULE OF CREATED COMPREHENSIVE SR SOP INSTANCES 
    Attribute Name  Tag VR  Value  Presence 
    of Value  Source 
    Instance Number 
    0020,0013 IS Generated by device, increments 
    from 1 in each series 
    ALWAYS AUTO 
    Completion Flag 
    0040,A491 CS PARTIAL  ALWAYS AUTO 
    Verification Flag  0040,A493 CS UNVERIFIED  ALWAYS AUTO 
    Content Date  0008,0023 DA   ALWAYS AUTO 
    Content Time  0008,0033 TM   ALWAYS AUTO 
    Referenced Request 
    Sequence  0040,A370 
    SQ 1 item will be present  ALWAYS AUTO 
    > Study Instance UID  0020,000D UI From MWL or generated by device  ALWAYS MWL/ 
    AUTO 
    > Referenced Study 
    Sequence 
    0008,1110 SQ 
    From MWL  ANAP MWL 
    >> Referenced SOP Class 
    UID  0008,1150 UI 
    From MWL  ANAP MWL 
    >> Referenced SOP 
    Instance UID  0008,1155 UI 
    From MWL  ANAP MWL 
    > Accession Number  0008,0050 SH From MWL or user input  VNAP MWL/ 
    USER 
    > Placer Order 
    Number/Imaging Service 
    Request  0040,2016 LO 
    NULL  VNAP AUTO 
    > Filler Order 
    Number/Imaging Service  0040,2017 LO 
    NULL  VNAP AUTO  
    						
    							HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 58   
       
    Request 
    > Requested Procedure ID 
    0040,1001 SH From MWL  VNAP MWL 
    > Requested Procedure 
    Description  0032,1060 LO 
    From MWL  VNAP MWL 
    > Requested Procedure 
    Code Sequence  0032,1064 SQ 
    From MWL  VNAP MWL 
    Performed Procedure 
    Code Sequence  0040,A372 SQ 
    NULL  VNAP AUTO 
     
     
    Table 8.1-19 
    SR DOCUMENT CONTENT MODULE OF CREATED COMPREHENSIVE SR SOP INSTANCES 
    Attribute Name  Tag VR  Value  Presence 
    of Value  Source
    Value Type 
    0040,A040 CS CONTAINER  ALWAYS AUTO 
    Concept Name Code 
    Sequence  0040,A043 
    SQ 1 item will be present  ALWAYS AUTO 
    > Include Code Sequence Macro     
    EV (125000, DCM, OB-GYN Ultrasound Procedure 
    Report) for  OB-GYN
     
    “EV (125100, DCM, “Vascular Ultrasound Report”) 
    for  Vascular
     
    “EV (125200, DCM, “Echocardiography Procedure 
    Report”) for  Adult Echocardiography
      ALWAYS AUTO 
    Include Container Macro 
        ALWAYS AUTO 
    Content Sequence  0040,A730 SQ One or more items  may be included in this sequence  ALWAYS AUTO 
    > Relationship Type  0040,A010 CS Ref. Section  9.1 TEMPLATES used in    ALWAYS AUTO 
    > Include Document Relationship 
    Macro   Ref. Section  9.1 TEMPLATES used in    ALWAYS AUTO 
    > Include Document Content 
    Macro   Ref. Section  9.1 TEMPLATES used in    ALWAYS AUTO 
     
    Table 8.1-20 
    SOP COMMON MODULE OF CREATED COMPREHENSIVE SR SOP INSTANCES 
    Attribute Name  Tag VR  Value  Presence 
    of Value  Source 
    SOP Class UID 
    0008,0016 UI 1.2.840.10008.5.1.4.1. 1.88.33 ALWAYS  AUTO 
    SOP Instance UID 0008,0018 UI Generated by device  ALWAYS AUTO 
    Specific Character Set  0008,0005 CS Only when a replacement character 
    set is used. 
    ANAP AUTO 
     
      
    						
    							HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 59   
       
    8.1.2 
    Used Fields in received IOD by application 
    The HD9 storage application does not receive SOP Instances. Th e usage of attributes received via Modality Worklist is 
    described in section  4.2.2.3.1.3. 
    8.1.3 Attribute mapping 
    The relationships between attributes received via Modality Worklist, stored in acquired images and communicated via 
    MPPS are summarized in the Table below. The format and  conversions used in Table are the same as the 
    corresponding table in IHE Technical Framewor k, Rev. 7.0 May 15, 2006, vol. II, Appendix A. 
    Table 8.1-21 
    ATTRIBUTE MAPPING BETWEEN MODALITY WORKLIST, IMAGE AND MPPS 
    Modality Worklist  Image IOD MPPS IOD 
    Patients Name Patients Name  Patients Name 
    Patient ID  Patient ID Patient ID 
    Patients Birth Date  Patients Birth Date Patients Birth Date 
    Patients Sex  Patients Sex Patients Sex 
    Patients Size  Patients Size ____ 
    Patients Weight  Patients Weight ____ 
    Referring Physicians Name  Referring Physicians Name ____ 
       ____  Scheduled Step Attributes Sequence 
    Study Instance UID  Study Instance UID  > Study Instance UID 
    Referenced Study Sequence  Referenced Study Sequence > Referenced Study Sequence 
    Accession Number  Accession Number  > Accession Number 
       Request Attributes Sequence  ____ 
    Requested Procedure ID  > Requested Procedure ID > Requested Procedure ID 
    Requested Procedure Description  ____ > Requested Procedure Description 
    Scheduled Procedure Step ID 
    > Scheduled Procedure Step ID  > Scheduled Procedure Step ID 
    Scheduled Procedure Step 
    Description  > Scheduled Procedure Step 
    Description 
    > Scheduled Procedure Step Description 
    Scheduled Protocol Code 
    Sequence  > Scheduled Protocol Code 
    Sequence 
    > Scheduled Protocol Code Sequence 
     
           
    ____  Study ID Study ID 
    ____  Performed Procedure Step ID Performed Procedure Step ID 
    ____  Performed Procedure Step Start 
    Date 
    Performed Procedure Step Start Date 
    ____  Performed Procedure Step Start 
    Time 
    Performed Procedure Step Start Time 
    ____  Performed Procedure Step 
    Description 
    Performed Procedure Step Description 
    ____  ____ Performed Series Sequence  
    						
    							HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 60   
       
    Requested Procedure Code 
    Sequence 
    Procedure Code Sequence 
    Procedure Code Sequence 
    ____  Referenced Performed Procedure 
    Step Sequence 
    ____ 
    ____  > Referenced SOP Class UID SOP Class UID 
    ____  > Referenced SOP Instance UID SOP Instance UID 
    ____  Protocol Name Protocol Name 
     
    8.1.4 Coerced/Modified Fields 
    The Modality Worklist AE will truncate attribute values receiv ed in the response to a Modality Worklist Query if the 
    value length is longer than the maximum  length permitted by the attribute’s VR. 
    8.2 DATA DICTIONARY OF PRIVATE ATTRIBUTES 
    No Private Attributes are supported. 
    8.3 CODED TERMINOLOGY AND TEMPLATES 
    The Workflow AE is capable of supporting arbitrary codi ng schemes for Procedure and Protocol Codes. The contents 
    of Requested Procedure Code Sequence (0032,1064) and  Scheduled Protocol Code Sequence (0040,0008) supplied 
    in Worklist Items will be mapped to Image IOD and MPPS  attributes as described in Section 8.1.3.   
    8.4 GRAYSCALE IMAGE CONSISTENCY 
    When available, the high resolution display monitor atta ched to HD9 can be calibrated according to the Grayscale 
    Standard Display Function (GSDF).    This is not incorporated in HD9 1.1. 
    8.5  STANDARD EXTENDED / SPECIALI ZED / PRIVATE SOP CLASSES 
    No Specialized or Private SOP Classes are supported. 
    8.5.1 US OR US MULTIFRAME IMAGE STORAGE SOP CLASS 
    The US or US Multiframe Image Storage SOP Classes  are extended to create a Standard Extended SOP Class by 
    addition of standard and private attributes to the cr eated SOP Instances as documented in section 8.1. 
    3D Volume Data is transferred to the configured Stora ge Server, if “Send 3D Volume” option is enabled in the Setup 
    Dialog. 
    The Standard Ultrasound Image and Ultrasound Multiframe  Image Storage SOP Classes are extended by adding the 
    Pixel Spacing (0028,0030) attribute when conf igured by the user.    See table 8.1-9a. 
    8.6 PRIVATE TRANSFER SYNTAXES 
    No Private Transfer Syntaxes are supported.  
    						