HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 51 8.1.1.3 Common Modules Table 8.1-3 PATIENT MODULE OF CREATED SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Patients Name 0010,0010 PN From MWL or User Input. Values supplied via Modality Worklist will be entered as received. Values supplied via user input will contain first 3 components (Last^First^Middle). Maximum 64 characters. VNAP MWL/ USER Patient ID 0010,0020 LO From MWL, user input or generated by device. Maximum 64 characters. ALWAYS MWL/ USER/ AUTO Patients Birth Date 0010,0030 DA From MWL or user input VNAP MWL/ USER Patients Sex 0010,0040 CS From MWL or user input VNAP MWL/ USER Table 8.1-4 GENERAL STUDY MODULE OF CREATED SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Study Instance UID 0020,000D UI From MWL or generated by device ALWAYS MWL/ AUTO Study Date 0008,0020 DA ALWAYS AUTO Study Time 0008,0030 TM ALWAYS AUTO Referring Physicians Name 0008,0090 PN From MWL or user input VNAP MWL/ USER Study ID 0020,0010 SH System generated : Study Date + Study Time ALWAYS AUTO Accession Number 0008,0050 SH From MWL or user input VNAP MWL/ USER Study Description 0008,1030 LO From MWL or user input In case of MWL, this system uses one of the following attributes according to the priority order. Priority 1 : (0032,1060) Requested Procedure Description Priority 2 : (0040,0007) Scheduled Procedure Step Description Priority 3 : (0040,1002) Reason for the Requested Procedure Priority 4 : (0040,2001) Reason for the Imaging Service Request ANAP MWL/ USER
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 52 Referenced Study Sequence 0008,1110 SQ From MWL ANAP MWL > Referenced SOP Class UID 0008,1150 UI From MWL ANAP MWL > Referenced SOP Instance UID 0008,1155 UI From MWL ANAP MWL Procedure Code Sequence 0008,1032 SQ From MWL ANAP MWL Table 8.1-5 PATIENT STUDY MODULE OF CREATED SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Patients Size 0010,1020 DS From MWL or user input ANAP MWL/ USER Patients Weight 0010,1030 DS From MWL or user input ANAP MWL/ USER Table 8.1-6 GENERAL SERIES MODULE OF CREATED SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Modality 0008,0060 CS US ALWAYS AUTO Series Instance UID 0020,000E UI Generated by device ALWAYS AUTO Series Number 0020,0011 IS 1 ALWAYS AUTO Series Date 0008,0021 DA ALWAYS AUTO Series Time 0008,0031 TM ALWAYS AUTO Protocol Name 0018,1030 LO FreeForm ALWAYS AUTO Operators’ Name 0008,1070 PN From user input ANAP USER Referenced Performed Procedure Step Sequence 0008,1111 SQ Identifies the MPPS SOP Instance to which this image is related ALWAYS MPPS > Referenced SOP Class UID 0008,1150 UI MPPS SOP Class UID 1.2.840.10008.3.1.2.3.3 ALWAYS MPPS > Referenced SOP Instance UID 0008,1155 UI MPPS SOP Instance UID ALWAYS MPPS Request Attributes Sequence 0040,0275 SQ Zero or 1 item will be present ANAP AUTO > Requested Procedure ID 0040,1001 SH From MWL ANAP MWL > Scheduled Procedure Step ID 0040,0009 SH From MWL ANAP MWL
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 53 > Scheduled Procedure Step Description 0040,0007 LO From MWL ANAP MWL > Scheduled Protocol Code Sequence 0040,0008 SQ From MWL ANAP MWL Performed Procedure Step ID 0040,0253 SH Same as MPPS ALWAYS MPPS Performed Procedure Step Start Date 0040,0244 DA Same as St udy Date ALWAYS AUTO Performed Procedure Step Start Time 0040,0245 TM Same as St udy Time ALWAYS AUTO Performed Procedure Step Description 0040,0254 LO Same as Study Description ANAP MWL/ USER Table 8.1-7 GENERAL EQUIPMENT MODULE OF CREATED SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Manufacturer 0008,0070 LO Philips ALWAYS AUTO Institution Name 0008,0080 LO From user input ANAP CONFIG Station Name 0008,1010 SH From user input ANAP CONFIG Manufacturers Model Name 0008,1090 LO HD9 ALWAYS AUTO Device Serial Number 0018,1000 LO Generated by device ALWAYS AUTO Software Versions 0018,1020 LO Generated by device ALWAYS AUTO 8.1.1.4 US or US Multiframe Image Module Table 8.1-8 GENERAL IMAGE MODULE OF CREATED US OR US MULTIFRAME SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Instance Number 0020,0013 IS Generated by device, increments from 1 in each series ALWAYS AUTO Patient Orientation 0020,0020 CS NULL ALWAYS AUTO Content Date 0008,0023 DA ALWAYS AUTO Content Time 0008,0033 TM ALWAYS AUTO Image Type 0008,0008 CS ORIGINAL and PRIMARY ALWAYS AUTO Lossy Image Compression 0028,2110 CS US = 00 (uncompressed), US- MF = 01 (lossy compressed) ALWAYS AUTO Lossy Image Compression Ratio 0028,2112 DS Used if (0028, 2110) = 01, Calculated by device ANAP AUTO Lossy Image Compression Method 0028,2114 CS ISO_10918_1, used if (0028,2001) = 01 ANAP AUTO Table 8.1-9
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 54 IMAGE PIXEL MODULE OF CREATED US OR US MULTIFRAME SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Samples per Pixel 0028,0002 US 3 for RGB or YBR_FULL_422 “1” for MONOCHROME2 ALWAYS AUTO Photometric Interpretation 0028,0004 CS US =RGB or “MONOCHROME2” USMF = “YBR_FULL_422” or “MONOCHROME2” ALWAYS AUTO Rows 0028,0010 US US = 768, US-MF = 480 ALWAYS AUTO Columns 0028,0011 US US = 102 4, US-MF = 640 ALWAYS AUTO Bits Allocated 0028,0100 US 8 ALWAYS AUTO Bits Stored 0028,0101 US 8 ALWAYS AUTO High Bit 0028,0102 US 7 ALWAYS AUTO Pixel Representation 0028,0103 US 0 ALWAYS AUTO Pixel Data 7FE0,0010 OW or OB Generated by device ALWAYS AUTO Planar Configuration 0028,0006 US 0, Present only if (0028,0002) = 3 ANAP AUTO Private Creator 7FE1,0010 LO “MEDISON_US” ANAP AUTO 3D Volume 7FE1,1002 OB 3D Volume Data ANAP AUTO Table 8.1-10a IMAGE PLANE MODULE OF CREATED US OR US MULTIFRAME SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Pixel Spacing 0028,0030 DS Physical distance between the centers of pixels in mm. Only present when user configures its use and images have only one region or if dual, the same depth. ANAP USER Table 8.1-11 CINE MODULE OF CREATED US MULTIFRAME SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Frame Time 0018,1063 DS Milliseconds ALWAYS* AUTO Cine Rate 0018,0040 IS Frames per second ALWAYS* AUTO * This module is only present in US Multiframe Images, and is r equired in US Multiframe Images. Therefore the Presence of Valu e is set to ALWAYS because the Module is only used in US Multiframe, not US Image.
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 55 Table 8.1-12 MULTI-FRAME MODULE OF CREATED US MULTIFRAME SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Number of Frames 0028,0008 IS Numbers of Frames ALWAYS* AUTO Frame Increment Pointer 0028,0009 AT Frame Time : (0018, 1063) ALWAYS* AUTO * This module is only present in US Multiframe Images, and is r equired in US Multiframe Images. Therefore the Presence of Valu e is set to ALWAYS because the Module is only used in US Multiframe, not US Image. Table 8.1-13 US REGION CALIBRATION MODULE OF CREATED US OR US MULTIFRAME SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Sequence of Ultrasound Regions 0018,6011 SQ Generated by device. A sequence is present for each region in the system display. ANAP AUTO > Region Location Min x0 0018,6018 UL Left position of region ALWAYS AUTO > Region Location Min y0 0018,601A UL Top position of region ALWAYS AUTO > Region Location Max x1 0018,601C UL Right position of region ALWAYS AUTO > Region Location Max y1 0018,601E UL Bottom position of region ALWAYS AUTO > Physical Units X Direction 0018,6024 US 2D Image : 0003H = cm M-Mode : 0004H = seconds Doppler : 0004H = seconds ALWAYS AUTO > Physical Units Y Direction 0018,6026 US 2D Image : 0003H = cm M-Mode : 0003H = cm Doppler : 0005H = hertz or 0007H = cm/sec ALWAYS AUTO > Physical Delta X 0018,602C FD The physical value per pixel increment ALWAYS AUTO > Physical Delta Y 0018,602E FD The physical value per pixel increment ALWAYS AUTO > Region Spatial Format 0018,6012 US 2D Tissue : 0001H M-Mode Tissue or flow : 0002H Spectral (CW or PW Doppler) : 0003H ALWAYS AUTO > Region Data Type 0018,6014 US Tissue : 0001H Color Flow : 0002H PW Spectral Doppler : 0003H CW Spectral Doppler : 0004H ALWAYS AUTO > Region Flags 0018,6016 UL See DICOM PS 3.3 C.8.5.5.1.3 ALWAYS AUTO
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 56 Table 8.1-14 US IMAGE MODULE OF CREATED US OR US MULTIFRAME SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Samples Per Pixel 0028,0002 US 3 for RGB or YBR_FULL_422 “1” for MONOCHROME2 ALWAYS AUTO Photometric Interpretation 0028,0004 CS US =RGB or “MONOCHROME2” USMF = “YBR_FULL_422” or “MONOCHROME2” ALWAYS AUTO Bits Allocated 0028,0100 US 8 ALWAYS AUTO Bits Stored 0028,0101 US 8 ALWAYS AUTO High Bit 0028,0102 US 7 ALWAYS AUTO Planar Configuration 0028,0006 US 0, Present only if (0028,0002) = 3 ALWAYS AUTO Pixel Representation 0028,0103 US 0 ALWAYS AUTO Image Type 0008,0008 CS ORIGINAL and PRIMARY ALWAYS AUTO Lossy Image Compression 0028,2110 CS US = 00 (uncompressed), US-MF = 01 (lossy compressed) ALWAYS AUTO Table 8.1-15 VOI LUT MODULE OF CREATED US OR US MULTIFRAME SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Window Center 0028,1050 DS default : 128 ALWAYS CONFIG Window Width 0028,1051 DS default : 256 ALWAYS CONFIG Table 8.1-16 SOP COMMON MODULE OF CREATED US OR US MULTIFRAME SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source SOP Class UID 0008,0016 UI US = 1.2.840.10008.5.1.4.1.1.6.1 or US-MF = 1.2.840.10008.5.1.4.1.1.3.1 ALWAYS AUTO SOP Instance UID 0008,0018 UI Generated by device ALWAYS AUTO Specific Character Set 0008,0005 CS Only when a replacement character set is used. ANAP AUTO
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 57 8.1.1.5 Comprehensive Structured Report Modules Table 8.1-17 SR DOCUMENT SERIES MODULE OF CREA TED COMPREHENSIVE SR SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Modality 0008,0060 CS SR ALWAYS AUTO Series Instance UID 0020,000E UI Generated by device ALWAYS AUTO Series Number 0020,0011 IS 2 ALWAYS AUTO Referenced Performed Procedure Step Sequence 0008,1111 SQ Identifies the MPPS SOP Instance to which this image is related ALWAYS MPPS > Referenced SOP Class UID 0008,1150 UI MPPS SOP Class UID 1.2.840.10008.3.1.2.3.3 ALWAYS MPPS > Referenced SOP Instance UID 0008,1155 UI MPPS SOP Instance UID ALWAYS MPPS Table 8.1-18 SR DOCUMENT GENERAL MODULE OF CREATED COMPREHENSIVE SR SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Instance Number 0020,0013 IS Generated by device, increments from 1 in each series ALWAYS AUTO Completion Flag 0040,A491 CS PARTIAL ALWAYS AUTO Verification Flag 0040,A493 CS UNVERIFIED ALWAYS AUTO Content Date 0008,0023 DA ALWAYS AUTO Content Time 0008,0033 TM ALWAYS AUTO Referenced Request Sequence 0040,A370 SQ 1 item will be present ALWAYS AUTO > Study Instance UID 0020,000D UI From MWL or generated by device ALWAYS MWL/ AUTO > Referenced Study Sequence 0008,1110 SQ From MWL ANAP MWL >> Referenced SOP Class UID 0008,1150 UI From MWL ANAP MWL >> Referenced SOP Instance UID 0008,1155 UI From MWL ANAP MWL > Accession Number 0008,0050 SH From MWL or user input VNAP MWL/ USER > Placer Order Number/Imaging Service Request 0040,2016 LO NULL VNAP AUTO > Filler Order Number/Imaging Service 0040,2017 LO NULL VNAP AUTO
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 58 Request > Requested Procedure ID 0040,1001 SH From MWL VNAP MWL > Requested Procedure Description 0032,1060 LO From MWL VNAP MWL > Requested Procedure Code Sequence 0032,1064 SQ From MWL VNAP MWL Performed Procedure Code Sequence 0040,A372 SQ NULL VNAP AUTO Table 8.1-19 SR DOCUMENT CONTENT MODULE OF CREATED COMPREHENSIVE SR SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Value Type 0040,A040 CS CONTAINER ALWAYS AUTO Concept Name Code Sequence 0040,A043 SQ 1 item will be present ALWAYS AUTO > Include Code Sequence Macro EV (125000, DCM, OB-GYN Ultrasound Procedure Report) for OB-GYN “EV (125100, DCM, “Vascular Ultrasound Report”) for Vascular “EV (125200, DCM, “Echocardiography Procedure Report”) for Adult Echocardiography ALWAYS AUTO Include Container Macro ALWAYS AUTO Content Sequence 0040,A730 SQ One or more items may be included in this sequence ALWAYS AUTO > Relationship Type 0040,A010 CS Ref. Section 9.1 TEMPLATES used in ALWAYS AUTO > Include Document Relationship Macro Ref. Section 9.1 TEMPLATES used in ALWAYS AUTO > Include Document Content Macro Ref. Section 9.1 TEMPLATES used in ALWAYS AUTO Table 8.1-20 SOP COMMON MODULE OF CREATED COMPREHENSIVE SR SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source SOP Class UID 0008,0016 UI 1.2.840.10008.5.1.4.1. 1.88.33 ALWAYS AUTO SOP Instance UID 0008,0018 UI Generated by device ALWAYS AUTO Specific Character Set 0008,0005 CS Only when a replacement character set is used. ANAP AUTO
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 59 8.1.2 Used Fields in received IOD by application The HD9 storage application does not receive SOP Instances. Th e usage of attributes received via Modality Worklist is described in section 4.2.2.3.1.3. 8.1.3 Attribute mapping The relationships between attributes received via Modality Worklist, stored in acquired images and communicated via MPPS are summarized in the Table below. The format and conversions used in Table are the same as the corresponding table in IHE Technical Framewor k, Rev. 7.0 May 15, 2006, vol. II, Appendix A. Table 8.1-21 ATTRIBUTE MAPPING BETWEEN MODALITY WORKLIST, IMAGE AND MPPS Modality Worklist Image IOD MPPS IOD Patients Name Patients Name Patients Name Patient ID Patient ID Patient ID Patients Birth Date Patients Birth Date Patients Birth Date Patients Sex Patients Sex Patients Sex Patients Size Patients Size ____ Patients Weight Patients Weight ____ Referring Physicians Name Referring Physicians Name ____ ____ Scheduled Step Attributes Sequence Study Instance UID Study Instance UID > Study Instance UID Referenced Study Sequence Referenced Study Sequence > Referenced Study Sequence Accession Number Accession Number > Accession Number Request Attributes Sequence ____ Requested Procedure ID > Requested Procedure ID > Requested Procedure ID Requested Procedure Description ____ > Requested Procedure Description Scheduled Procedure Step ID > Scheduled Procedure Step ID > Scheduled Procedure Step ID Scheduled Procedure Step Description > Scheduled Procedure Step Description > Scheduled Procedure Step Description Scheduled Protocol Code Sequence > Scheduled Protocol Code Sequence > Scheduled Protocol Code Sequence ____ Study ID Study ID ____ Performed Procedure Step ID Performed Procedure Step ID ____ Performed Procedure Step Start Date Performed Procedure Step Start Date ____ Performed Procedure Step Start Time Performed Procedure Step Start Time ____ Performed Procedure Step Description Performed Procedure Step Description ____ ____ Performed Series Sequence
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 60 Requested Procedure Code Sequence Procedure Code Sequence Procedure Code Sequence ____ Referenced Performed Procedure Step Sequence ____ ____ > Referenced SOP Class UID SOP Class UID ____ > Referenced SOP Instance UID SOP Instance UID ____ Protocol Name Protocol Name 8.1.4 Coerced/Modified Fields The Modality Worklist AE will truncate attribute values receiv ed in the response to a Modality Worklist Query if the value length is longer than the maximum length permitted by the attribute’s VR. 8.2 DATA DICTIONARY OF PRIVATE ATTRIBUTES No Private Attributes are supported. 8.3 CODED TERMINOLOGY AND TEMPLATES The Workflow AE is capable of supporting arbitrary codi ng schemes for Procedure and Protocol Codes. The contents of Requested Procedure Code Sequence (0032,1064) and Scheduled Protocol Code Sequence (0040,0008) supplied in Worklist Items will be mapped to Image IOD and MPPS attributes as described in Section 8.1.3. 8.4 GRAYSCALE IMAGE CONSISTENCY When available, the high resolution display monitor atta ched to HD9 can be calibrated according to the Grayscale Standard Display Function (GSDF). This is not incorporated in HD9 1.1. 8.5 STANDARD EXTENDED / SPECIALI ZED / PRIVATE SOP CLASSES No Specialized or Private SOP Classes are supported. 8.5.1 US OR US MULTIFRAME IMAGE STORAGE SOP CLASS The US or US Multiframe Image Storage SOP Classes are extended to create a Standard Extended SOP Class by addition of standard and private attributes to the cr eated SOP Instances as documented in section 8.1. 3D Volume Data is transferred to the configured Stora ge Server, if “Send 3D Volume” option is enabled in the Setup Dialog. The Standard Ultrasound Image and Ultrasound Multiframe Image Storage SOP Classes are extended by adding the Pixel Spacing (0028,0030) attribute when conf igured by the user. See table 8.1-9a. 8.6 PRIVATE TRANSFER SYNTAXES No Private Transfer Syntaxes are supported.