Philips Hd9 Owners Manual
Have a look at the manual Philips Hd9 Owners Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 619 Philips manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 41 on hardware options installed: Table 4.3-1 SUPPORTED PHYSICAL NETWORK INTERFACES Ethernet 10/100/1000BaseT, RJ-45, UTP, STP, Auto-Detect or manually set Speed and Duplex. 4.4 CONFIGURATION 4.4.1 AE Title/Presentation Address Mapping 4.4.1.1 Local AE Titles All local applications use the AE Titles and TCP/IP Port s configured via the Setup/DICOM Menu. All local DICOM services use the same AE Title. The system listens for Verification requests and Commitment reports on the configured Port. 4.4.1.2 Remote AE Title/Presentation Address Mapping The AE Title, host names and port numbers of remote applic ations are configured using the HD9 Setup/DICOM Menu. Use the “Test” button at any time to send a “Verification” reqest to any configured device(s). 4.4.1.2.1 Storage Use the “Add” button on the HD9 Setup/ DICOM Menu and select ‘STORAGE’ from the Services menu to set the AE Titles, Port numbers, IP Addresses and capabilities fo r the remote Image Storage SCPs. Multiple remote Image Storage SCPs can be defined. Use the “Add” button on the HD9 Setup/DI COM Menu and select ‘STORAGE SR’ from the Services menu to set the AE Titles, Port numbers, IP Addresses and capabilities for the remote Structured Report Storage SCP. Only a single remote Structured Report Storage SCP can be defined. Use the “Add” button on the HD9 Setup/DICOM Menu and select ‘SC’ from the Services menu to set the AE Titles, Port numbers, IP Addresses and capabilities for the remote Storage Commitment SCP. Only a single remote Storage Commitment SCP can be defined and only one Image Storage SCP can be assigned for Storage Commitment. 4.4.1.2.2 Workflow Use the “Add” button on the HD9 Setup/ DICOM Menu and select ‘WORKLIST’ from the Services menu to set the AE Titles, Port numbers, IP Addresses and capabilities for the remote Modality Worklist SCP. Only a single remote Modality Worklist SCP can be defined. Use the “Add” button on th e HD9 Setup/DICOM Menu and select ‘PPS’ from the Servic es menu to set the AE Titles, Port numbers, IP Addresses and capabilities for the remote MPPS SCP. Only a single remote MPPS SCP can be defined. 4.4.1.2.3 Hardcopy Use the “Add” button on the HD9 Setup/DI COM Menu and select ‘PRINT’ from the Services menu to set the AE Titles, Port numbers, IP Addresses and capabilities for the remote Print SCPs. Multiple remote Print SCPs can be defined.
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 42 4.4.2 Parameters A number of parameters related to acquisition and general operation can be configured using the Setup/DICOM Menu. The table below only shows those configuration parameters relevant to DICOM communications. See the HD9 Manual for details on general configuration capabilities. Table 4.4-1 CONFIGURATION PARAMETERS TABLE Parameter Configurable (Yes/No) Default Value Local System Parameters AE Title (Local System AE Title) Yes “Set AE Title” Station Name Yes “Set Station Name” Port No. (Local Port Number) Yes 104 Service Common Parameters Retry Interval Yes 30 Sec. Connect Timeout Yes 15 Sec. Maximum Retires Yes 1 Storage Parameters Send Cine Loops Yes Checked Send Pixel Spacing Yes Not checked DICOM Send Format Yes 2D Mode Gray Color Mode Color Display Compensation Yes Gamma 10 Brightness 0 Contrast 100 Transfer Mode Yes “Batch” Include 3D Volume Yes Not checked Include Pixel Spacing Yes Not checked Window Center (VOI LUT) Yes 128 Window Width (VOI LUT) Yes 256 Storage Commitment Parameters Associated Storage Server Yes None Worklist Modality Parameters Delay between automatic Worklist Updates Yes 5 Min. Query Worklist for specific Scheduled Station AE Title Yes Any Query Worklist for specific Sched uled Modality Value No “US” fixed Query Worklist for Specific Start Date Yes Today Print Parameters Transfer Mode Yes “Batch” Color Yes “Grayscale” Medium Type Yes “PAPER”
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 43 Format Yes 1x1 Film Size Yes 8 IN X 10 IN Orientation Yes “PORTRAIT” Destination Yes “MAGAZINE” Magnification Yes “REPLICATE” Smoothing Type Yes Blank Border Density Yes “BLACK” Empty Density Yes “BLACK” Priority Yes “HIGH” Min Density Yes Blank Max Density Yes Blank Copies Yes 1 Configuration Info Yes Blank
5 MEDIA INTERCHANGE 5.1 IMPLEMENTATION MODEL 5.1.1 Application Data Flow Removable Storage MediaBackup Offline Media Application Entity Figure 5.1-1 APPLICATION DATA FLOW DIAGRAM FOR MEDIA STORAGE - The Offline Media Application Entity exports images a nd Structured Report to a Removable Storage medium. It is associated with the local real-world activity “B ackup”, performed upon user request for selected studies. - Note: “Backup” is for exams and will create a DICOMDIR file and export images in DICOM format, “Export” is for images, and has DICOM format as a choice with BMP, JPEG and TIFF. No DICOMDIR is written with “Export” to media. 5.1.2 Functional Definition of AEs 5.1.2.1 Functional Definition of Application Entity Activation of the “Backup” menu entry w ill pass the currently selected studies to the Offline Media Application Entity. The SOP Instances associated with the selection will be collec ted into one or more export jobs. The contents of each export job will be written to a single piece of media. 5.1.3 Sequencing of Real-World Activities At least one study must exist and be selected before the Off line Media Application Entity can be invoked. The operator can insert a new media at any time before or after invocati on of the Offline Media Application Entity. If no media is available the export job can be cancelled immediately. 5.1.4 File Meta Information Options The implementation written to the File Meta Header in each file is: Table 5.1-1 DICOM IMPLEMENTATION CLASS AND VERSION FOR MEDIA STORAGE Implementation Class UID 1.3.46.670589.14.2 Implementation Version Name HD9 1.1 HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 44
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 45 5.2 AE SPECIFICATIONS 5.2.1 Offline Media Application Entity Specification The Offline Media Application Entity provides standard conformance to the Media Storage Service Class. The Application Profiles and roles are listed below: Table 5.2-1 APPLICATION PROFILES, ACTIVITIES AND ROLES FOR OFFLINE MEDIA Application Profiles Supported Real World Activity Role STD-US-SC-MF-CD-R Export To Media FSC, FSU* STD-GEN-CD Export To Media FSC, FSU STD-US-SC-MF-DVD Export To Media FSC, FSU STD-GEN-DVD Export To Media FSC, FSU STD-GEN-USB-JPEG Export To Media FSC, FSU Note: The additional Media Application Profiles are list ed to indicate support for all media types the system is capable of writing to and to include the expor t of Structured Reports to media in DICOM. * File Set Updater (FSU) is only possible on medi a which supports multiple write capability. 5.2.1.1 File Meta Information for the Application Entity The File-Set Identifier included in the File Meta Header is “MED_FSU”. The Source Application Entity Title included in the File Meta Header is configurable using the Setup/DICOM Menu. 5.2.1.2 Real-World Activities 5.2.1.2.1 Activity – Backup The Offline Media Application Entity acts as an FSC and FSU when requested to export SOP Instances from the local database to a piece of media. If the contents of the current selection do not fit on a single m edia, a separation into multiple export jobs which can be adapted by the user will be suggested. The user will be prompted to insert a ne w media for each export job. The contents of the export job will be written together with a corresponding DICOMDIR to a medi a. Writing in multi-session mode is supported. 5.2.1.2.1.1 Media Storage Application Profiles The Offline Media Application Entity supports the following Application Profiles: • STD-US-SC-MF-CD-R - For Ultrasound Images, Single and Multiframe, compressed or not • STD-GEN-CD - For supporting Structured Reports to CD • STD-US-SC-MF-DVD - For Ultrasound Images, Single and Multiframe, compressed or not • STD-GEN-DVD - For supporting Structured Reports to CD • STD-GEN-USB-JPEG - For supporting all images, compressed or not and SRs to media 5.2.1.2.1.1.1 Options The Media Application Entity supports the SOP Classe s and Transfer Syntaxes listed in the table below:
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 46 Table 5.2-2 IODS, SOP CLASSES AND TRANSFER SYNTAXES FOR OFFLINE MEDIA Information Object Definition SOP Class UID Transfer Syntax Transfer Syntax UID Media Storage Directory Storage 1.2.840.10008.1.3.10 Explicit VR Little Endian 1.2.840.10008.1.2.1 US Image Storage 1.2.840.10008.5.1.4.1.1.6.1 Explicit VR Little Endian 1.2.840.10008.1.2.1 US Multiframe Image Storage 1.2.840.10008.5.1.4.1.1.3.1 JPEG Baseline Lossy Compression 1.2.840.10008.1.2.4.50 Comprehensive Structured Report Storage 1.2.840.10008.5.1.4.1.1.88 .33 Explicit VR Little Endian 1.2.840.10008.1.2.1
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 47 6 SUPPORT OF CHARACTER SETS All HD9 DICOM applications support the following Character Sets: ISO_IR 100: Latin Alphabet No. 1 Supplementary set of ISO 8859 ISO 646 ISO_IR 144: Cyrillic Supplementary set of ISO 8859 ISO 646 ISO 2000 IR 13\ISO 2022 IR 87: Japanese JIS X 0201 Katakana JIS X 0201 Romaji JIS X 0208 Kanji JIS X 0212 Supplementary Kanji Set
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 48 7 SECURITY HD9 does not support any specific security measures. It is assumed that HD9 is used within a secured environm ent. It is assumed that a secured environment includes as minimum: a. Firewall or router protections to ensure that onl y approved external hosts have network access to HD9. b. Firewall or router protections to ensure that HD9 has only network access to approved external hosts and services. c. Any communication with external hosts and services outside the locally secured environment use appropriately secure network channels (e.g. su ch as a Virtual Private Network (VPN). Other network security procedures such as automated intrusion detection may be appropriate in some environments. Additional security features may be established by the lo cal security policy and are beyond the scope of this conformance statement.
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 49 8 ANNEXES 8.1 IOD CONTENTS 8.1.1 Created SOP Instances Table 8.1-1 specifies the attributes of Ultrasound Images transmitted by the HD9 storage applications. Table 8.1-2 specifies the attributes of Comprehensive Structur ed Reports transmitted by the HD9 storage applications. The following tables use a number of abbreviations. The abbreviations used in the “Presence of…” column are: VNAP Value Not Always Present (attribute se nds zero length if no value is present) ANAP Attribute Not Always Present ALWAYS Always Present EMPTY Attribute is sent without a value The abbreviations used in the “Source” column: MWL the attribute value source Modality Worklist USER the attribute value source is from User input AUTO the attribute value is generated automatically MPPS the attribute value is the same as t he Modality Performed Procedure Step service CONFIG the attribute value source is a configurable parameter NOTE: All dates and times are encoded in the local configured calendar and time. Date, Time and Time zones are configured using the Setup Menu.
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 50 8.1.1.1 US or US Multiframe Image IOD Table 8.1-1 IOD OF CREATED US OR US MULTIFRAME SOP INSTANCES IE Module Reference Presence of Module Patient Patient Table 8.1-3 ALWAYS Study General Study Table 8.1-4 ALWAYS Patient Study Table 8.1-5 ALWAYS Series General Series Table 8.1-6 ALWAYS Equipment General Equipment Table 8.1-7 ALWAYS Image General Image Table 8.1-8 ALWAYS Image Pixel Table 8.1-9 ALWAYS Image Plane* Table 8.1-9a ANAP Cine Table 8.1-10 Only if US Multiframe Multi-Frame Table 8.1-11 Only if US Multiframe US Region Calibration Table 8.1-12 ANAP US Image Table 8.1-13 ALWAYS VOI LUT Table 8.1-14 ALWAYS SOP Common Table 8.1-15 ALWAYS * Extended SOP Class to add Pixel Spacing depending on user setting. 8.1.1.2 Comprehensive Structured Report IOD Table 8.1-2 IOD OF CREATED COMPREHENSIVE STRUCTURED REPORT SOP INSTANCES IE Module Reference Presence of Module Patient Patient Table 8.1-3 ALWAYS Study General Study Table 8.1-4 ALWAYS Patient Study Table 8.1-5 ALWAYS Series SR Document Series Table 8.1-16 ALWAYS Equipment General Equipment Table 8.1-7 ALWAYS Document SR Document General Table 8.1-17 ALWAYS SR Document Content Table 8.1-18 ALWAYS SOP Common Table 8.1-19 ALWAYS