HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 51 8.1.1.3 Common Modules Table 8.1-3 PATIENT MODULE OF CREATED SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Patients Name 0010,0010 PN From MWL or User Input. Values supplied via Modality Worklist will be entered as received. Values supplied via user input will contain first 3 components (Last^First^Middle). Maximum 64 characters. VNAP MWL/ USER Patient ID 0010,0020 LO From MWL, user input...
Page 52
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 52 Referenced Study Sequence 0008,1110 SQ From MWL ANAP MWL > Referenced SOP Class UID 0008,1150 UI From MWL ANAP MWL > Referenced SOP Instance UID 0008,1155 UI From MWL ANAP MWL Procedure Code Sequence 0008,1032 SQ From MWL ANAP MWL Table 8.1-5 PATIENT STUDY MODULE OF CREATED SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Patients Size 0010,1020 DS From MWL or user input ANAP...
Page 53
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 53 > Scheduled Procedure Step Description 0040,0007 LO From MWL ANAP MWL > Scheduled Protocol Code Sequence 0040,0008 SQ From MWL ANAP MWL Performed Procedure Step ID 0040,0253 SH Same as MPPS ALWAYS MPPS Performed Procedure Step Start Date 0040,0244 DA Same as St udy Date ALWAYS AUTO Performed Procedure Step Start Time 0040,0245 TM Same as St udy Time ALWAYS AUTO Performed Procedure Step Description...
Page 54
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 54 IMAGE PIXEL MODULE OF CREATED US OR US MULTIFRAME SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Samples per Pixel 0028,0002 US 3 for RGB or YBR_FULL_422 “1” for MONOCHROME2 ALWAYS AUTO Photometric Interpretation 0028,0004 CS US =RGB or “MONOCHROME2” USMF = “YBR_FULL_422” or “MONOCHROME2” ALWAYS AUTO Rows 0028,0010 US US = 768, US-MF = 480 ALWAYS AUTO Columns 0028,0011 US US = 102 4,...
Page 55
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 55 Table 8.1-12 MULTI-FRAME MODULE OF CREATED US MULTIFRAME SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Number of Frames 0028,0008 IS Numbers of Frames ALWAYS* AUTO Frame Increment Pointer 0028,0009 AT Frame Time : (0018, 1063) ALWAYS* AUTO * This module is only present in US Multiframe Images, and is r equired in US Multiframe Images. Therefore the Presence of Valu e is set to ALWAYS...
Page 56
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 56 Table 8.1-14 US IMAGE MODULE OF CREATED US OR US MULTIFRAME SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Samples Per Pixel 0028,0002 US 3 for RGB or YBR_FULL_422 “1” for MONOCHROME2 ALWAYS AUTO Photometric Interpretation 0028,0004 CS US =RGB or “MONOCHROME2” USMF = “YBR_FULL_422” or “MONOCHROME2” ALWAYS AUTO Bits Allocated 0028,0100 US 8 ALWAYS AUTO Bits Stored 0028,0101 US 8 ALWAYS...
Page 57
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 57 8.1.1.5 Comprehensive Structured Report Modules Table 8.1-17 SR DOCUMENT SERIES MODULE OF CREA TED COMPREHENSIVE SR SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Modality 0008,0060 CS SR ALWAYS AUTO Series Instance UID 0020,000E UI Generated by device ALWAYS AUTO Series Number 0020,0011 IS 2 ALWAYS AUTO Referenced Performed Procedure Step Sequence 0008,1111 SQ Identifies the MPPS SOP...
Page 58
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 58 Request > Requested Procedure ID 0040,1001 SH From MWL VNAP MWL > Requested Procedure Description 0032,1060 LO From MWL VNAP MWL > Requested Procedure Code Sequence 0032,1064 SQ From MWL VNAP MWL Performed Procedure Code Sequence 0040,A372 SQ NULL VNAP AUTO Table 8.1-19 SR DOCUMENT CONTENT MODULE OF CREATED COMPREHENSIVE SR SOP INSTANCES Attribute Name Tag VR Value Presence of Value Source Value Type...
Page 59
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 59 8.1.2 Used Fields in received IOD by application The HD9 storage application does not receive SOP Instances. Th e usage of attributes received via Modality Worklist is described in section 4.2.2.3.1.3. 8.1.3 Attribute mapping The relationships between attributes received via Modality Worklist, stored in acquired images and communicated via MPPS are summarized in the Table below. The format and conversions used in Table...
Page 60
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 60 Requested Procedure Code Sequence Procedure Code Sequence Procedure Code Sequence ____ Referenced Performed Procedure Step Sequence ____ ____ > Referenced SOP Class UID SOP Class UID ____ > Referenced SOP Instance UID SOP Instance UID ____ Protocol Name Protocol Name 8.1.4 Coerced/Modified Fields The Modality Worklist AE will truncate attribute values receiv ed in the response to a Modality Worklist Query if the...