Philips Hd9 Owners Manual
Here you can view all the pages of manual Philips Hd9 Owners Manual. The Philips manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 11
4 NETWORKING 4.1 IMPLEMENTATION MODEL 4.1.1 Application Data Flow Acquire Images / Reports Storage Application Entity Workflow Application Entity Hardcopy Application Entity Remote Application Entity Receives Images & Reports Search Worklist Acquire Images Remote Application Entity Provides Worklist Items Remote Application Entity Receives MPPS Create/Update Remote Application Entity Confirms CommitmentsRequest Commitments Acquire ImagesRemote Application Entity Prints...
Page 12
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 12 automatically. Methods to send images depend on user configuration, “Batch”, “Send As You Go” or “Manual”. “Manual” mode is performed upon user request for each study or for specific images selected. “Batch” mode starts to send images at End Exam for each study. “Send As You Go” mode starts when the first image is acquired for each study and images are tr ansferred immediately after acquisition. Structured...
Page 13
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 13 o Past Month o Custom (specific Date, single value matching) The Workflow AE performs the creation of an MPPS Instance automatically whenever the first image is acquired for each study. The MPPS final states can only be set by “End Exam” for each study. 4.1.2.3 Functional Definition of Hardcopy Application Entity The existence of a print job will activate the Hardcopy AE. An association is established with the...
Page 14
4.1.3 Sequencing of Real-World Activities Figure 4.1-2 SEQUENCING CONTRAINTS – SEND AS YOU GO HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 14
Page 15
StorageHardcopyWorkflowDepartment SchedulerPrinterImage Manager 1. Query Worklist 2. Receive Worklist 3. Select Workitem 5. Start Acquisition (Create MPPS) 9. Store Acquired Images 8. Print Acquired Images 10. Complete Acquisition (Finalize MPPS) 6. Acquire Image 12. Commit Acquired Images & SR Documents 13. Receive Commitment Result 7. End Exam 11. Store SR Documents 4. Start Exam Figure 4.1-3 SEQUENCING CONSTRAINTS – BATCH MODE Under normal scheduled workflow conditions, the sequencing...
Page 16
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 16 4.2 AE SPECIFICATIONS 4.2.1 Storage Application Entity Specification 4.2.1.1 SOP Classes HD9 provides Standard Conformance to the following SOP Classes: Table 4.2-1 SOP CLASSES FOR AE STORAGE SOP Classes SOP Class UID SCU SCP Ultrasound Image Storage 1.2.840.10008.5.1.4.1.1.6.1 Yes No Ultrasound Multi-frame Image Storage 1.2.840.10008.5.1.4.1.1.3.1 Yes No Comprehensive Structured Report Storage...
Page 17
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 17 4.2.1.2.4 Implementation Identifying Information The implementation information for this Application Entity is: Table 4.2-6 DICOM IMPLEMENTATION CLASS AND VERSION FOR AE STORAGE Implementation Class UID 1.3.46.670589.14.2 Implementation Version Name HD9 1.1 4.2.1.3 Association Initiation Policy 4.2.1.3.1 Activity – Send Images and Structured Reports and Request Commitment 4.2.1.3.1.1 Description and Sequencing of...
Page 18
Storage AEImage Manager 1. Open Association 2. C-STORE (Images) 3. Close Association 4. Open Association 5. C-STORE (SR Documents) 6. Close Association 7. Open Association 8. N-ACTION (Storage Commitment Request for Images & SR Documents) 9. Close Association 10. Open Association 11. N-EVENT-REPORT (Storage Commitment Response) 12. Close Association Figure 4.2-1 SEQUENCING OF ACTIVITY - SEND IMAGES A possible sequence of interactions between the Storage AE and an Image Manager (e.g. a storage or...
Page 19
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 19 4.2.1.3.1.2 Proposed Presentation Contexts HD9 is capable of proposing the Presentation Contexts shown in the following table. Table 4.2-7 PROPOSED PRESENTATION CONTEXTS FOR ACTIVITY SEND IMAGES Presentation Context Table Abstract Syntax Transfer Syntax Role Ext. Neg. Name UID Name List UID List Ultrasound Image Storage 1.2.840.10008.5. 1.4.1.1.6.1 Implicit VR Little Endian JPEG Lossy Baseline 1.2.840.10008.1.2...
Page 20
HD9 1.1.x DICOM Conformance Statement 989605375630044 Rev. B Page 20 Understand Warning Coercion of Data Elements B000 Image transmission is considered successful. Warning Data Set does not match SOP Class B007 Same as Warning above. Warning Elements Discards B006 Same as Warning above. * * Any other status code. Same as Refused above. The Behavior of Storage AE during communication failure is summarized in the Table below: Table 4.2-9 STORAGE COMMUNICATION FAILURE BEHAVIOR...