Philips Fm2 Service Manual
Have a look at the manual Philips Fm2 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 619 Philips manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
![](/img/blank.gif)
Service Menu Chapter 4 - Power-up Default Settings and the Service Menu 29 System A/D Values For the end user or service personnel, the system A/D values require no interpretation. The A/D screen displays the current value of each analog-to-digital (A/D) channel in volts. The A/D channels read DC voltages, for example, power supply voltages and battery voltage. The A/D values displayed are static, being acquired when the screen is first displayed. None of the displayed values may be changed or reset in this screen.When in the A/D screen, the Return item is always highlighted; a press of the wheel will return the user to the Service Menu. Rotating the wheel while in the A/D screen will have no effect. The A/D Values screen appears as follows: Figure 3 System A/D Values Note—The screenshot above shows untypical system A/D values that are not representative of normal operation.
![](/img/blank.gif)
Service Menu 30Chapter 4 - Power-up Default Settings and the Service Menu
![](/img/blank.gif)
Chapter 5 - Troubleshooting31 5 Troubleshooting General This section provides a guide for qualified service personnel in locating the possible cause of an apparent malfunction of the monitor or its accessories. Use of built-in diagnostics, flow charts, and step-by-step procedures are provided as aids in the troubleshooting process. The chapter leads you to a probable field replaceable assembly or component. Items that are serviceable only at the factory are also identified, as well as instructions for their return for repair/replacement. The initial analysis determines whether the apparent malfunction is caused by a replaceable accessory, or by a replaceable assembly within the monitor case.
![](/img/blank.gif)
General 32Chapter 5 - Troubleshooting Initial Trouble Diagnosis The first step in diagnosing a possible malfunction is to duplicate, if possible, the symptoms of the reported malfunction. Thereafter, it is important to categorize the problem as either one of two types, as illustrated in Figure 4. They are: Totally inoperative condition, or Degraded operation. The totally inoperative condition is indicated by the failure of the monitor to respond to the power ON front-panel action, or by complete loss of monitoring functions during operation. Contact your local Agilent Response Center. Degraded operation performance might be indicated by the appearance of artifacts in the display, or by gaps in the pixel display, by intermittent actions, or by obviously inaccurate measures of fetal activity, etc. Figure 4 Diagnosing Malfunction - Initial Diagnosis
![](/img/blank.gif)
Power Problem Diagnosis Chapter 5 - Troubleshooting 33 Power Problem Diagnosis Diagnostic Flow Chart The flow chart of Figure 5 shows a sequence of actions and decisions you can use after the initial diagnosis of a “total failure” as a power supply or display problem. Note—If the apparent “total failure” is caused by a malfunctioning display backlight, you may see a faintly lit display with the appropriate Power-On Self Test message. The troubleshooting under these circumstances is described in this section. Figure 5 Power Diagnosis - Power Supply or Display
![](/img/blank.gif)
Power Problem Diagnosis 34Chapter 5 - Troubleshooting Diagnostic/Repair Actions Examine Display IntenselyReduce background room illumination. If only the back light fluorescent tube or its high voltage driver has failed, the display may be faintly visible. Refer to page 50 for removal procedures. Replace BatteryRefer to page 43. Exchange the MonitorIf the display is not at all visible, and there is total failure, exchange the monitor.
![](/img/blank.gif)
Monitoring Degradation Problems Chapter 5 - Troubleshooting 35 Monitoring Degradation Problems Diagnostic Flow Chart Figure 6 illustrates a logical sequence of troubleshooting diagnostics when the monitor appears to be operative, but the displayed results are suspect for any number of reasons. The resulting diagnosis will result in replacing either the monitor or the transducers. Figure 6 Monitoring Degradation, Diagnostic Flow Chart
![](/img/blank.gif)
Monitoring Degradation Problems 36Chapter 5 - Troubleshooting Diagnostic Repair Actions Describe Degraded PerformanceExamine the display after an apparently successful power up sequence and self- test. Consider, first, the appearance of artifacts, gaps, or general quality of the display. Note the values being reported in the numerical frames (FHR(s)) and TOCO). Are they clearly unreasonable? Swap TransducersUse a spare transducer that is free of defects and meets factory specifications. At facilities where more than one monitor is installed, it may be convenient to borrow a transducer that is working satisfactorily with another monitor. Exchange the MonitorIf the monitoring problems persist, or there is total failure, exchange the monitor.
![](/img/blank.gif)
Chapter 6 - Disassembly/Reassembly Guide37 6 Disassembly/Reassembly Guide Warning Performance verification: do not place the monitor into operation after repair or maintenance has been performed, until all performance tests and safety tests listed in Chapter 8 of this service manual have been performed. Failure to perform all tests could result in erroneous monitor readings, or patient/operator injury. Warning Energized circuits are accessible with the covers open. Do not work on the monitor with the covers open. Only qualified service personnel should open or disassemble the monitor. Introduction Warning Before attempting to open or disassemble the monitor, disconnect the power supply from the monitor. Warning High voltage is generated by the LCD backlight driver. Do not operate the monitor with the covers open.
![](/img/blank.gif)
How to Use this Chapter 38Chapter 6 - Disassembly/Reassembly Guide Caution Observe ESD (electrostatic discharge) precautions when working within the unit. Caution Important for disassembly! When disassembling the monitor, make sure that the battery leads are disconnected. Otherwise, it is likely that circuit damage will occur. The monitor can be disassembled down to all major component parts, including: PCBs battery cables function buttons chassis enclosures Remember to store all screws, nuts, washers and other parts in a safe place for later refitting. How to Use this Chapter The step-by-step procedures that are used to access replaceable parts of the monitor are illustrated in the Disassembly sections which follow. These describe and photographically illustrate procedures for disassembling the monitor to enable removal and replacement of suspected defective assemblies and components. The monitor consists of two main assemblies: The front case assembly The rear case assembly The main PCB assembly is separable from the front case assembly.