Philips Fm2 Service Manual
Have a look at the manual Philips Fm2 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 619 Philips manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
Introduction Chapter 8 - Test and Inspection Matrix 79 Caution General-purpose personal computers and modems are not designed to meet the electrical safety requirements of medical devices. The RS-232 and parallel connectors on the monitor are electrically isolated to permit safe connections to non-medical devices, which should be connected with a cable of sufficient length to prevent the non-medical equipment from contacting the patient. If a non-medical device is connected to the serial port, a recorder or printer connected at the same time to the parallel port should be located outside the patient vicinity, as the two interfaces are not isolated against each other. Introduction This chapter contains maintenance and safety information for the FM-2 antepartum fetal monitor and its accessories. All checks that require the instrument to be opened must be made by qualified service personnel. Please contact your local Agilent representative when you wish safety and maintenance checks to be carried out. To ensure that your monitor functions properly and safely, you must adhere to the standards stated in this manual.
When to Perform Tests Blocks 80Chapter 8 - Test and Inspection Matrix When to Perform Tests Blocks Table 20 shows which test blocks to perform after repair and preventive maintenance tasks. Table 20 FM-2: When to Perform Test Blocks Service EventTest Block(s) Required (See Table 21 on page 81) Installation The product is customer installed. Refer to the Instructions for Use.Visual and Power On Preventive Maintenance Preventive maintenance is limited to exchanging the batteryPo we r O n Safety Tests Repairs Unit Exchange Visual and Power On Unit opened Visual and Power On Front End Connector Board exchangedPo we r O n Performance Tests for Trans- ducers Safety Tests Rear Connector Board exchanged Power On Safety Tests Power Supply exchanged Power On Battery exchanged Power On Safety Tests
When to Perform Tests Blocks Chapter 8 - Test and Inspection Matrix 81 Table 21 Test and Inspection Matrix Test Block NameTest Expected ResultsWhat to Record Visual Test (See 20)Inspect exterior of fetal monitor for damage.No Visual Damage V:P or V:F Pow e r - O n Se l f - Te s t (See 20)Power on the monitor Displays Normal Monitoring Screen Configuration and emits tonesPO:P or PO:F Safety Tests: • Safety Test 1 (See 86)Enclosure Leakage Current - S.F.C. Open SupplyAll Safety Test results are in range of Table 22 on page 87S:P or S:F Safety Test 2 (See 88)Patient Leakage Cur- rent - S.F.C. Open SupplyAll Safety Test results are in range of Table 23 on page 89S:P or S:F Key to Table:P = Pass F = Fail
Battery Charging and Maintenance 82Chapter 8 - Test and Inspection Matrix Battery Charging and Maintenance Before attempting to perform any tests, it is important to verify that the battery is charged and functional. Using the Battery You can run the monitor for approximately six hours on battery power at room temperature of 25o C from a fully charged battery (with a minimum of two hours monitoring at 10o C from a fully charged battery). The age of the battery, operating temperature and number of recharge cycles affects this. A battery life cycle is approximately 200 charge/discharge cycles. Low Power Warning A low power warning occurs when the battery power is running low. The battery fuel gauge flashes, and a repetitive audible tone sounds. Remaining monitoring time is typically 30 minutes with a new battery at room temperature of 25 o C. In an older monitor, in which the battery has had a number of recharge cycles, it may occur when fewer minutes remain. The warning continues until the low power condition disappears (when you connect the monitor to an AC power supply) or until the system stops functioning because of the low power.You cannot silence it, and you cannot adjust its volume. The monitor retains any stored fetal traces, even if it shuts down because of lack of battery power.
Preventive Maintenance Chapter 8 - Test and Inspection Matrix 83 Recharging the Battery Use the power supply to recharge the battery from an AC power source. This takes a maximum of 14 hours when monitoring simultaneously, or eight hours if no monitoring occurs. Typical times are 11 hours if you are monitoring simultaneously, or 4.75 hours if no monitoring occurs. Storing A stored battery loses capacity over time. To minimize loss of capacity, recharge the battery at least every six months. A battery typically has three to five years shelf life. Preventive Maintenance The only preventive maintenance task required for the monitor is battery exchange. Life expectancy of a battery depends on the frequency and duration of use. The battery should be replaced every two years. See “Removing the Battery” on page 43, and “Replacing the Battery” on page 44.
Safety Testing 84Chapter 8 - Test and Inspection Matrix Safety Testing This section defines the test and inspection procedures applicable to the FM-2 fetal monitors. Use the tables in the following section to determine what test and inspection results must be reported after a repair has been carried out. Test Blocks in Table 20 tells you when to carry out the tests blocks Test and Inspection Matrix in Table 21 tells you the expected results of the tests and what to record. Warning Apply the safety test limits set by the local standards and statutes applicable to the country of installation, such as IEC 601-1, UL2601, CAN/CSA- C22.2 No. 601.1.M90. The safety tests defined in this chapter are derived from local and international standards but may not be equivalent. They are NOT a substitute for local safety testing. Caution Successful completion of the patient safety tests does not ensure the correct and accurate functioning of the equipment.
Safety Testing Chapter 8 - Test and Inspection Matrix 85 Safety Test Procedures The test procedures outlined in this section are to be used only for verifying the safe installation or service of the product. “Safety checks at installation refer to safety aspects directly related to the installation and setup activities and not to intrinsic safety features that have already been checked during final acceptance testing at the factory.” [FDA 97-4179 Medical Quality Systems Manual: A Small Entity Compliance Guide (QSR manual). First Edition December 1996. Chapter 14: Storage, Distribution and Installation. This document may be viewed at http:// www.fda.gov. When running safety tests on the peripheral components of the system, refer to the documentation included with those products. When performing safety tests, you must use a standard safety tester. You can use testers complying with IEC 601-1 internationally. You may also use testers complying with any local standards and statutes applicable to the country of the installation. For safety test procedures see the operation instructions of the safety tester used. If using the Metron safety tester use your local regulations to perform the test, for example in Europe: IEC60601-1/IEC60601-1-1 in the US: UL2601-1 The Metron Report should print results as detailed in this chapter, along with other data. Note—For information and ordering guides for Metron products contact: Metron AS, Travbaneveien 1, N-7044 Trondheim, Norway Tel. (+47) 73 82 85 00; Fax (+47) 73 91 70 09; E-mail: [email protected]/[email protected] www: http://www.metron-biomed.com There are two general categories of safety test: Ground integrity Electrical Leakage
Safety Testing 86Chapter 8 - Test and Inspection Matrix Ground Integrity The FM-2 is a Class 2, Type BF instrument. It has an external power supply, and is constructed with double insulation. The FM-2 does not require an isolated Earth Ground terminal, neither is one installed. No Protective Ground Continuity check is required. Electrical Leakage The following tests verify the electrical leakage of the monitor: Enclosure leakage Current Patient Leakage Current Safety Test 1: Enclosure Leakage Current Test Normal ConditionThis test is applicable to Class 1 and 2 equipment, type B, BF, and CF Applied Parts. The test measures leakage current of exposed metal parts of the Instrument Under Test. It tests both normal and reversed polarity. For type BF and CF Applied Parts the test measures AP/GND. Safety Test according IEC 601-1 (Clause 19). Report the highest value. Single Fault Condition (S.F.C.) Open SupplyThis test is applicable to Class 1 and 2 equipment, type B, BF, and CF Applied Parts. The test measures leakage current of exposed metal parts of the Instrument Under Test with one supply lead interrupted. It tests both normal and reversed polarity. For type BF and CF Applied Parts the test measures AP/GND. Safety Test according IEC 601-1 (Clause 19). Report the highest value.
Safety Testing Chapter 8 - Test and Inspection Matrix 87 Te s t S t e p s1. Connect the external power supply output cord to the monitor. Then connect the appropriate external power supply input power cord to the analyzer as recommended by the analyzer operating instructions. 2. Using the appropriate test cable, connect the analyzer to a fixing nut of the RS-232 connector on the back of the monitor. 3. Turn the monitor on. 4. Perform the test as recommended by the analyzer operating instructions. The analyzer leakage current indication must not exceed the values listed in Table 22. Table 22 Enclosure Leakage Current Test Condition PolarityMaximum Leakage Current (µA) IEC 601-1 ES1AAMI/ANSI Standard NormalNormal 100 µA 100 µA Reversed 100 µA 100 µA S.F.C.1 (Open Supply) 1. S.F.C. = Single Fault Condition Normal 500 µA 300 µA Reversed 500 µA 300 µA
Safety Testing 88Chapter 8 - Test and Inspection Matrix Safety Test 2: Patient Leakage Current Test Normal ConditionThis test is applicable to Class 1 and 2 equipment, type B, BF, and CF Applied Parts. The test measures patient leakage current from any individual patient connection to earth (power ground) in parallel. It tests both normal and reversed polarity. Safety Test according IEC 601-1 (Clause 19). Report the highest value. Single Fault Condition (S.F.C.) Open SupplyThis test is applicable to Class 1 and 2 equipment, type B, BF, and CF Applied Parts. The test measures patient leakage current from any individual patient connection to earth (power ground) in parallel, with one supply lead interrupted. It tests both normal and reversed polarity. Safety Test according IEC 601-1 (Clause 19). Report the highest value. Te s t S t e p sThis test requires a sample patient cable for each device parameter. These must be configured as recommended by the safety analyzer operating instructions. 1. Configure the electrical safety analyzer as recommended by the analyzer operating instructions. 2. Connect the appropriate external power supply input power cord to the analyzer as recommended by the analyzer operating instructions. Connect the external power supply output cord to the monitor. 3. Turn the monitor on. 4. Perform the test as recommended by the analyzer operating instructions. For each test condition, the measured leakage current must not exceed that indicated in Table 23.