Philips Fm2 Service Manual
Here you can view all the pages of manual Philips Fm2 Service Manual. The Philips manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 91
Introduction Chapter 8 - Test and Inspection Matrix 79 Caution General-purpose personal computers and modems are not designed to meet the electrical safety requirements of medical devices. The RS-232 and parallel connectors on the monitor are electrically isolated to permit safe connections to non-medical devices, which should be connected with a cable of sufficient length to prevent the non-medical equipment from contacting the patient. If a non-medical device is connected to the serial port, a...
Page 92
When to Perform Tests Blocks 80Chapter 8 - Test and Inspection Matrix When to Perform Tests Blocks Table 20 shows which test blocks to perform after repair and preventive maintenance tasks. Table 20 FM-2: When to Perform Test Blocks Service EventTest Block(s) Required (See Table 21 on page 81) Installation The product is customer installed. Refer to the Instructions for Use.Visual and Power On Preventive Maintenance Preventive maintenance is limited to exchanging the batteryPo we r O n Safety Tests...
Page 93
When to Perform Tests Blocks Chapter 8 - Test and Inspection Matrix 81 Table 21 Test and Inspection Matrix Test Block NameTest Expected ResultsWhat to Record Visual Test (See 20)Inspect exterior of fetal monitor for damage.No Visual Damage V:P or V:F Pow e r - O n Se l f - Te s t (See 20)Power on the monitor Displays Normal Monitoring Screen Configuration and emits tonesPO:P or PO:F Safety Tests: • Safety Test 1 (See 86)Enclosure Leakage Current - S.F.C. Open SupplyAll Safety Test results are in...
Page 94
Battery Charging and Maintenance 82Chapter 8 - Test and Inspection Matrix Battery Charging and Maintenance Before attempting to perform any tests, it is important to verify that the battery is charged and functional. Using the Battery You can run the monitor for approximately six hours on battery power at room temperature of 25o C from a fully charged battery (with a minimum of two hours monitoring at 10o C from a fully charged battery). The age of the battery, operating temperature and number of...
Page 95
Preventive Maintenance Chapter 8 - Test and Inspection Matrix 83 Recharging the Battery Use the power supply to recharge the battery from an AC power source. This takes a maximum of 14 hours when monitoring simultaneously, or eight hours if no monitoring occurs. Typical times are 11 hours if you are monitoring simultaneously, or 4.75 hours if no monitoring occurs. Storing A stored battery loses capacity over time. To minimize loss of capacity, recharge the battery at least every six months. A battery...
Page 96
Safety Testing 84Chapter 8 - Test and Inspection Matrix Safety Testing This section defines the test and inspection procedures applicable to the FM-2 fetal monitors. Use the tables in the following section to determine what test and inspection results must be reported after a repair has been carried out. Test Blocks in Table 20 tells you when to carry out the tests blocks Test and Inspection Matrix in Table 21 tells you the expected results of the tests and what to record. Warning Apply the safety...
Page 97
Safety Testing Chapter 8 - Test and Inspection Matrix 85 Safety Test Procedures The test procedures outlined in this section are to be used only for verifying the safe installation or service of the product. “Safety checks at installation refer to safety aspects directly related to the installation and setup activities and not to intrinsic safety features that have already been checked during final acceptance testing at the factory.” [FDA 97-4179 Medical Quality Systems Manual: A Small Entity...
Page 98
Safety Testing 86Chapter 8 - Test and Inspection Matrix Ground Integrity The FM-2 is a Class 2, Type BF instrument. It has an external power supply, and is constructed with double insulation. The FM-2 does not require an isolated Earth Ground terminal, neither is one installed. No Protective Ground Continuity check is required. Electrical Leakage The following tests verify the electrical leakage of the monitor: Enclosure leakage Current Patient Leakage Current Safety Test 1: Enclosure Leakage...
Page 99
Safety Testing Chapter 8 - Test and Inspection Matrix 87 Te s t S t e p s1. Connect the external power supply output cord to the monitor. Then connect the appropriate external power supply input power cord to the analyzer as recommended by the analyzer operating instructions. 2. Using the appropriate test cable, connect the analyzer to a fixing nut of the RS-232 connector on the back of the monitor. 3. Turn the monitor on. 4. Perform the test as recommended by the analyzer operating instructions....
Page 100
Safety Testing 88Chapter 8 - Test and Inspection Matrix Safety Test 2: Patient Leakage Current Test Normal ConditionThis test is applicable to Class 1 and 2 equipment, type B, BF, and CF Applied Parts. The test measures patient leakage current from any individual patient connection to earth (power ground) in parallel. It tests both normal and reversed polarity. Safety Test according IEC 601-1 (Clause 19). Report the highest value. Single Fault Condition (S.F.C.) Open SupplyThis test is applicable to...
All Philips manuals