Panasonic Ew 3153 Instructions Manual
Have a look at the manual Panasonic Ew 3153 Instructions Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 10737 Panasonic manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
9English30English Specifications Power sourceDC 6 V (Main unit: Four AA/LR6 size alkaline batteries) A C 10 V 60 Hz (when using AC adapter) DC 3 V (Wireless display: Two AA/LR6 size alkaline batteries) Display Digital LCD Method of measurement Oscillometric system Measurement range Blood pressure: 0 mmHg to 80 mmHg Pulse r ate: 30 to 160 beats per minute Accuracy Blood pressure: Within ±3 mmHg Pulse rate: Within ±5% Operation temperature/ humidity range 10 °C to 40 °C (50 °F to 104 °F), 15% to 90% RH Storage temperature/ humidity range – 0 °C to 60 °C (–4 °F to 140 °F), 10% to 95% RH Measurable arm circumference Approx. 7¾ to 13⅜ ( 0 cm to 34 cm) Weight Main unit: Approx. 1550 g (54.7 oz.) (not including batteries) Wireless displa y: Approx. 160 g (5.6 oz.) (not including batteries) Dimensions Main unit: 11⅛ × 6⅝ × 11⅛ ( 8.3 cm × 16.9 cm × 8.1 cm) Wireless display: 5¾ × 3¼ × ⅝ (14.6 cm × 8.4 cm × 6.8 cm) Protection against electric shock Internally powered equipment, Type BF applied part AC adapter Input AC 10 V 60 Hz Output DC 6 V .0 A Q & A Q Why are the blood pressure values measured at home different from those measured at the doctor’s office? A Blood pressure v aries in a 4-hour period. In addition, it can vary greatly depending on the weather, your mental state, amount of exercise, etc. In particular, there is the phenomenon of “White uniform hypertension” at doctor’s offices, where the blood pressure measured in a doctor’s office is higher because of stress or anxiety. In addition, if the blood pressure measured at home is sometimes higher and sometimes lower than the blood pressure measured at the doctor’s office, the following can be considered. Please pay attention to these points when taking measurements. 1 Measuring posture was poor. Did you take measurements while leaning forward with your body at an angle? Measurements w ere not taken at the upper arm. Please take measurements with your elbow in the arm support of the main unit. (See pages 15–17.) 3 Were you anxious or upset during measurements? Before taking measurements, take or 3 deep breaths and relax, so that your blood pressure stabilizes. Q Wh y do the main unit and the wireless display operate together even when they are separated? A The main unit and the wireless display communicate using infrared rays. Infrared rays are transmitted from the transmitter on the side panel of the main unit, and reflected on your body and then received on the bottom of the front panel of the wireless display. Q The mark does not light when an arm is moved. A The mark is a function to notify you that your arm is bent or to indicate extra loads on the pressure cuff. Therefore, it may not indicate motion (bending an arm slightly, etc.). If you are concerned about the result, retry the measurement. 29-302008/01/15 14:54:30
31English3English Q & A (continued) Q Since the mark was displayed, I retried the measurement by setting my arm on the unit properly. But the indicated value is still higher than usual. A The mark is just a guide for the proper measurement method. (See page 0.) If you correct your posture when the mark is displayed but the measured value is still higher or lower than usual, retry the measurement. The following causes are suspected. Did you measure properly (in proper posture)? Measure as indicated on pages 15–17. Measured immediately after smoking. Measured when irritated or emotional. Measured in cold environments For persons who have a physical constitution with less fluctuation in the pulse rate. If you are concerned about the results, retry the measurement after sta ying quiet for 4 to 5 minutes. • • • • • • Q When should I measure my blood pressure? A We recommend three measurements a day. First ........ Upon Waking (Because, with almost no influence from body movement, it’s the blood pressure is closest to that at night.) Second ... Bef ore Lunch (Because it provides blood pressure values close to the highest of the day.) Third ....... At Bedtime (So as to obtain data close to y our average blood pressure values during the day.) This way, by taking measurements morning, noon and night, you can use the unit to learn the patterns of and monitor your blood pressure. Use of the AM and PM average systolic blood pressure comparison function can aid in the detection of a blood pressure pattern known as “early morning surge” in which blood pressure rises rapidly in the mor ning. (See page 7.) • Care and Maintenance Do not attempt to disassemble, repair or modify the unit. Doing so may cause fire or cause the unit to malfunction. It may also lead to injury. Do not apply excessive force to the unit or drop the unit. Doing so may cause damage. Do not insert dust or foreign objects into the unit. Doing so may result in damage. If the unit is stored at temperatures below the freezing point, do not use it immediately. Leave it in a warm place for at least an hour before use. If the unit is not allowed to warm up, it may not pressurize. Do not touch the unit when measurement is in progress. Doing so may cause incorrect readings. If the unit becomes dirty, clean it with a soft cloth moistened with warm water or soapy water. (Do not use alcohol, benzine, or paint thinner.) Use of such chemicals may result in cracking or discoloration. When the unit will not be used f or a long time (30 days or more), be sure to remove the batteries. Otherwise, the batteries may leak and damage the unit. The pressure cuff cannot be washed. For storage, avoid high temperatures, high humidity, and direct sunlight. Shortage under such conditions may result in damage. • • • • • • • • 31-322008/01/15 14:54:31
31English3English Q & A (continued) Q Since the mark was displayed, I retried the measurement by setting my arm on the unit properly. But the indicated value is still higher than usual. A The mark is just a guide for the proper measurement method. (See page 0.) If you correct your posture when the mark is displayed but the measured value is still higher or lower than usual, retry the measurement. The following causes are suspected. Did you measure properly (in proper posture)? Measure as indicated on pages 15–17. Measured immediately after smoking. Measured when irritated or emotional. Measured in cold environments For persons who have a physical constitution with less fluctuation in the pulse rate. If you are concerned about the results, retry the measurement after sta ying quiet for 4 to 5 minutes. • • • • • • Q When should I measure my blood pressure? A We recommend three measurements a day. First ........ Upon Waking (Because, with almost no influence from body movement, it’s the blood pressure is closest to that at night.) Second ... Bef ore Lunch (Because it provides blood pressure values close to the highest of the day.) Third ....... At Bedtime (So as to obtain data close to y our average blood pressure values during the day.) This way, by taking measurements morning, noon and night, you can use the unit to learn the patterns of and monitor your blood pressure. Use of the AM and PM average systolic blood pressure comparison function can aid in the detection of a blood pressure pattern known as “early morning surge” in which blood pressure rises rapidly in the mor ning. (See page 7.) • Care and Maintenance Do not attempt to disassemble, repair or modify the unit. Doing so may cause fire or cause the unit to malfunction. It may also lead to injury. Do not apply excessive force to the unit or drop the unit. Doing so may cause damage. Do not insert dust or foreign objects into the unit. Doing so may result in damage. If the unit is stored at temperatures below the freezing point, do not use it immediately. Leave it in a warm place for at least an hour before use. If the unit is not allowed to warm up, it may not pressurize. Do not touch the unit when measurement is in progress. Doing so may cause incorrect readings. If the unit becomes dirty, clean it with a soft cloth moistened with warm water or soapy water. (Do not use alcohol, benzine, or paint thinner.) Use of such chemicals may result in cracking or discoloration. When the unit will not be used f or a long time (30 days or more), be sure to remove the batteries. Otherwise, the batteries may leak and damage the unit. The pressure cuff cannot be washed. For storage, avoid high temperatures, high humidity, and direct sunlight. Shortage under such conditions may result in damage. • • • • • • • • 31-322008/01/15 14:54:31
33English34English Troubleshooting Display Status before error Cause and solution is displayed. The indication disappears in the middle of the operation. (the or mark is displayed.) Batteries are exhausted. (See pages 11–13.) is displayed. Pressurization to more than 80 mmHg was performed. Were measurements taken according to the proper procedure and in the correct posture? (See pages 15–18.) The mark flashes for minutes or longer. Pressure decreased suddenly. Be sure your arm is properly inserted into the main unit. (See pages 15–18.) The heart mark blinked only a few times or not at all. Cuff does not inflate. Your arm or body moved. Has the unit detected movement? (See page 0.) is displayed. Failure of the main unit Take it to the store where it was purchased for testing and repairs. Symptom Possible cause Systolic or diastolic blood pressure value is high or low. The position of the main unit is higher or lower than your heart when measuring. Your arm was not inserted properly into the main unit. You moved or talked during measurements. (See pages 15–18.) • • • Value is different from that measured by the doctor. Blood pressure varies in response to minute changes in your mental state, such as your reaction to a doctor’s presence. Sit quietly f or 4 to 5 minutes and then take another measurement. (See pages 15–18, 30–31.) • • Measurement values are different with each measurement. Symptom Possible cause Pressure cuff comes off. Press the ring portion on the outside of the pressure cuff into the main unit. (See page 3.) • Indications on the wireless display do not change during operation. The wireless display is improperly positioned. Your hand or another object is in front of the wireless displa y. (See page 18.) • • No display appears even when the On/Off and Start button is pressed. Batteries are installed with the polarities wrong. (See pages 11–13.) • Batteries are exhausted. (See pages 11–13.) • The mark flashes when readings are being taken. There was a significant fluctuation in pulse when readings were being taken. (See page 1.) • The mark flashes when readings are being taken. Your arm or body moved. (See page 0.) • The mark flashes even though I did not move my arm or body. In some cases this will be displayed when you have tensed the muscles in your hand. Relax the muscles in your hand and take the reading again. (See pages 15–17, 0.) • Varying sounds and pressurization speeds occur during inflation of the cuff. Changes in operating sounds and pressurization speeds are normal and occur due to the fact that measurements are being made as the cuff is being inflated. • The wireless display is lost. Take it to the store where it was purchased for testing and repairs. Instrument is broken. If the unit still appears to provide unusual or erroneous readings, consult your physician. If the unit does not appear to be functioning properly, contact Panasonic at 1-800-338-0552. 33-342008/01/15 14:54:31
33English34English Troubleshooting Display Status before error Cause and solution is displayed. The indication disappears in the middle of the operation. (the or mark is displayed.) Batteries are exhausted. (See pages 11–13.) is displayed. Pressurization to more than 80 mmHg was performed. Were measurements taken according to the proper procedure and in the correct posture? (See pages 15–18.) The mark flashes for minutes or longer. Pressure decreased suddenly. Be sure your arm is properly inserted into the main unit. (See pages 15–18.) The heart mark blinked only a few times or not at all. Cuff does not inflate. Your arm or body moved. Has the unit detected movement? (See page 0.) is displayed. Failure of the main unit Take it to the store where it was purchased for testing and repairs. Symptom Possible cause Systolic or diastolic blood pressure value is high or low. The position of the main unit is higher or lower than your heart when measuring. Your arm was not inserted properly into the main unit. You moved or talked during measurements. (See pages 15–18.) • • • Value is different from that measured by the doctor. Blood pressure varies in response to minute changes in your mental state, such as your reaction to a doctor’s presence. Sit quietly f or 4 to 5 minutes and then take another measurement. (See pages 15–18, 30–31.) • • Measurement values are different with each measurement. Symptom Possible cause Pressure cuff comes off. Press the ring portion on the outside of the pressure cuff into the main unit. (See page 3.) • Indications on the wireless display do not change during operation. The wireless display is improperly positioned. Your hand or another object is in front of the wireless displa y. (See page 18.) • • No display appears even when the On/Off and Start button is pressed. Batteries are installed with the polarities wrong. (See pages 11–13.) • Batteries are exhausted. (See pages 11–13.) • The mark flashes when readings are being taken. There was a significant fluctuation in pulse when readings were being taken. (See page 1.) • The mark flashes when readings are being taken. Your arm or body moved. (See page 0.) • The mark flashes even though I did not move my arm or body. In some cases this will be displayed when you have tensed the muscles in your hand. Relax the muscles in your hand and take the reading again. (See pages 15–17, 0.) • Varying sounds and pressurization speeds occur during inflation of the cuff. Changes in operating sounds and pressurization speeds are normal and occur due to the fact that measurements are being made as the cuff is being inflated. • The wireless display is lost. Take it to the store where it was purchased for testing and repairs. Instrument is broken. If the unit still appears to provide unusual or erroneous readings, consult your physician. If the unit does not appear to be functioning properly, contact Panasonic at 1-800-338-0552. 33-342008/01/15 14:54:31
No.1 EN, SP Printed in China © 008 Matsushita Electric Works, Ltd. All Rights Reserved. Impreso en China For questions or assistance with your blood pressure monitor, call us at 1-800-338-0552. Panasonic Consumer Electronics Company Division of Panasonic Corporation of North America One Panasonic Way 1H-1 Secaucus, NJ 07094 352008/01/15 14:54:31 Before operating this device, please read these instructions completely and save this manual for future use. Antes de utilizar este dispositivo, lea completamente estas instrucciones y conserve este manual para futura referencia. English 1 Español S1 Operating Instructions Instrucciones de Uso Upper Arm Blood Pressure MonitorMonitor de presión arterial de brazo Model No./Nº de Modelo EW3153 12008/01/15 14:54:23