9English10English Preparations Prior to Taking Measurements To use this product, the power supplies are required for both the main unit and the wireless display. Main unit Power supply Exclusive AC adapter (EW31555) (included) or 4 AA-size Panasonic LR6 batteries Approximately 150 measurements can be taken. Depending on the conditions, battery life may be dramatically shorter. Use of the main unit with the dedicated AC adapter (EW31555) (included) is recommended.• • Wireless display Power supply AA-size LR6 batteries Appro ximately 000 measurements can be taken. • Main unit and wireless display are not sold separately and cannot be purchased as replacement parts. For Batteries Unit with batteries can be used up to 6 times a day when using Panasonic Alkaline batteries (AA-sized LR6 batteries) under the following conditions: Room temper ature: 73 °F (3 °C) Pressur ized to 170 mmHg Circumference of the arm: 11-51/64 (30 cm) If batteries other than alkaline batteries are used or batteries are used in a room colder than 73 °F (3 °C), and when measuring a person with thin arms or high blood pressure, battery life may be dramatically shortened. *Use alkaline batteries. (Panasonic Oxyride batteries can also be used.) • • Parts Identification Main unit View from Top 1 Measuring section Pressure cuff 2 Battery cover for main unit Bottom of main unit Battery cover for main unit 3 IR transmitter (on both sides) 4 Connector 5 Arm support The angle of this product can be adjusted so that measurements can be taken without having to be in a forced posture. Elbow rest 6 Quick air release button 7 Handle 8 On/Off and Start button Plug AC adaptor Power cord AC Adapter (EW31555) (included) 9-102008/01/15 14:54:24
11English1English When Using the Main Unit with Batteries *Always use alkaline batteries. (4 AA-size Panasonic LR6 alkaline batteries) When replacing batteries, always use 4 fresh alkaline batteries of the same type from the same manufacturer, and replace them simultaneously. 1. Turn the main unit over and remove the battery cover by sliding it in the direction of the arrow while pressing lightly on the mark. 2. Check the polarities of the batteries and install them correctly. 3. Replace the battery cover and slide it closed. Notes for using the main unit with batteries Approximately 150 measurements can be taken. Depending on the conditions, battery life may be dramatically shortened. Use of the main unit with the dedicated AC adapter (EW31555) (included) is recommended. Replace batteries when: The mark shown to the left appears during use. No display appears even if the On/Off and Start button is pressed. • • CAUTION: Always follow the cautions printed on the batteries. Check the polarities of the batteries and install them correctly. Remove exhausted batteries promptly. If the product will not be used for long periods of time, remove the batteries first. Batteries should be used before their expiration date. Use after the expiration date may result in injury or staining of the surrounding area due to generation of heat, bursting, or leakage. • • • • • • Parts Identification (continued) Wireless display View from Bottom 1 Blood pressure indicator6 Memory/Recall button 2 Display7 User selection button 3 Set button8 Battery cover for wireless display 4 Adjust button (Display off)9 IR receiver 5 AM/PM comparison button * Specifications of the product or parts may be changed without prior notice. When Using the Main Unit with the AC Adapter Be sure to use only the exclusive AC adapter (included). Do not use the power cord with any product other than this AC adaptor (EW31555). Use of other AC adapters could cause damage to the unit or physical harm to the user. AC 10 V1. Connect the power cord to the AC adaptor. 2. Insert the AC adapter plug into the connector on either side of the main unit. 3. Plug the power cord fully into the AC outlet. 11-122008/01/15 14:54:25
11English1English When Using the Main Unit with Batteries *Always use alkaline batteries. (4 AA-size Panasonic LR6 alkaline batteries) When replacing batteries, always use 4 fresh alkaline batteries of the same type from the same manufacturer, and replace them simultaneously. 1. Turn the main unit over and remove the battery cover by sliding it in the direction of the arrow while pressing lightly on the mark. 2. Check the polarities of the batteries and install them correctly. 3. Replace the battery cover and slide it closed. Notes for using the main unit with batteries Approximately 150 measurements can be taken. Depending on the conditions, battery life may be dramatically shortened. Use of the main unit with the dedicated AC adapter (EW31555) (included) is recommended. Replace batteries when: The mark shown to the left appears during use. No display appears even if the On/Off and Start button is pressed. • • CAUTION: Always follow the cautions printed on the batteries. Check the polarities of the batteries and install them correctly. Remove exhausted batteries promptly. If the product will not be used for long periods of time, remove the batteries first. Batteries should be used before their expiration date. Use after the expiration date may result in injury or staining of the surrounding area due to generation of heat, bursting, or leakage. • • • • • • Parts Identification (continued) Wireless display View from Bottom 1 Blood pressure indicator6 Memory/Recall button 2 Display7 User selection button 3 Set button8 Battery cover for wireless display 4 Adjust button (Display off)9 IR receiver 5 AM/PM comparison button * Specifications of the product or parts may be changed without prior notice. When Using the Main Unit with the AC Adapter Be sure to use only the exclusive AC adapter (included). Do not use the power cord with any product other than this AC adaptor (EW31555). Use of other AC adapters could cause damage to the unit or physical harm to the user. AC 10 V1. Connect the power cord to the AC adaptor. 2. Insert the AC adapter plug into the connector on either side of the main unit. 3. Plug the power cord fully into the AC outlet. 11-122008/01/15 14:54:25
13English14English Inserting the Batteries into the Wireless Display *Always use alkaline batteries. (2 AA-size LR6 alkaline batteries) When replacing batteries, always use fresh alkaline batteries of the same type from the same manufacturer, and replace them simultaneously. 1. Turn the wireless display over and remove the battery cover by sliding it in the direction of the arrow while pressing lightly on the mark. 2. Check the polarities of the batteries and install them correctly. 3. Replace the battery cover and slide it closed. Replace batteries when: The mark shown at left appears during use. In this case, press the On/Off and the Start button on the main unit to interrupt the measurement. • If both batteries for the main unit and the wireless display must be replaced at the same time, (main unit) and (wireless display) are alternately displayed. • Setting the time and date * Please be sure to set the date and time before taking the first measurement after purchasing the blood pressure monitor, or after replacing the batteries. (If you attempt to take measurements without first setting the time and date then “–” will be displayed.) 1. Press the Set button “1” on the wireless display for at least 3 seconds. 2. Adjust the month. Press the Adjust button “” to adjust the number of the month. 3. Adjust the date. (1) Press the Set button “1” and the number for the date will flash. ( ) Press the Adjust button “” to adjust the number of the date. 4. Adjust the hour. (1) Press the Set button “1” and the time will flash on and off. ( ) Press the Adjust button “” to adjust the number of the hour. 5. Adjust the minutes. (1) Press the Set button “1” and the number for minutes will flash. ( ) Press the Adjust button “” to adjust the number of minutes. 6. Finally, synchronizing with a time signal, press the Set button “1” to complete the procedure. The date and time are alternately displayed in appro ximately seconds. 7. Press the Adjust button “2” to turn off the unit. • During leap years, 3/1 will be displayed when it is February 9th. Please manually reset the date to /9. 13-142008/01/15 14:54:25
13English14English Inserting the Batteries into the Wireless Display *Always use alkaline batteries. (2 AA-size LR6 alkaline batteries) When replacing batteries, always use fresh alkaline batteries of the same type from the same manufacturer, and replace them simultaneously. 1. Turn the wireless display over and remove the battery cover by sliding it in the direction of the arrow while pressing lightly on the mark. 2. Check the polarities of the batteries and install them correctly. 3. Replace the battery cover and slide it closed. Replace batteries when: The mark shown at left appears during use. In this case, press the On/Off and the Start button on the main unit to interrupt the measurement. • If both batteries for the main unit and the wireless display must be replaced at the same time, (main unit) and (wireless display) are alternately displayed. • Setting the time and date * Please be sure to set the date and time before taking the first measurement after purchasing the blood pressure monitor, or after replacing the batteries. (If you attempt to take measurements without first setting the time and date then “–” will be displayed.) 1. Press the Set button “1” on the wireless display for at least 3 seconds. 2. Adjust the month. Press the Adjust button “” to adjust the number of the month. 3. Adjust the date. (1) Press the Set button “1” and the number for the date will flash. ( ) Press the Adjust button “” to adjust the number of the date. 4. Adjust the hour. (1) Press the Set button “1” and the time will flash on and off. ( ) Press the Adjust button “” to adjust the number of the hour. 5. Adjust the minutes. (1) Press the Set button “1” and the number for minutes will flash. ( ) Press the Adjust button “” to adjust the number of minutes. 6. Finally, synchronizing with a time signal, press the Set button “1” to complete the procedure. The date and time are alternately displayed in appro ximately seconds. 7. Press the Adjust button “2” to turn off the unit. • During leap years, 3/1 will be displayed when it is February 9th. Please manually reset the date to /9. 13-142008/01/15 14:54:25
15English16English Taking a Reading Advice: In general, blood pressure measurements are lower on the second attempt because you are calmer. For people who easily get tense, measuring twice each time is advised. When two measurements are taken, normally the smaller measurement is stored. When taking tw o measurements, always stay quiet for 4 to 5 minutes after the first measurement before taking the second measurement. Getting into the proper position for taking a measurement Place the main unit on a table so that it is flat and the center of the unit is at heart height. *Differences in height will cause differences in blood pressure values. If the heights of the chair and table are not suitable, you may have to lean forward. In this case, form a stable stack of books or similar objects under the main unit to adjust the height of the main unit. *You cannot use the unit while lying down. Use only in a sitting position. Elbow rest 1 Put the main unit on a table in front of you and slide your arm through the measurement section of the unit.Remove your shirt or roll up your sleeve so that the measuring section will be on your bare skin. If rolling up your sleeve causes pressure on your arm or underarm, remove your shirt for measurements. Be sure that the rolled-up sleeve is not inside the cuff area when taking measurements. • • 2 Place your elbow on the elbow rest with your palm up and relax. 3 Place the wireless display in front of your body where you can see it easily.The main unit and the wireless display transmit/receive the measured values through infr ared rays. (See page 18.) Do not put your hand or any other objects in front of the wireless display. Differences in blood pressure values between the left and right arms may be around 10 mmHg. The blood pressure can be measured in either the left or right arm, but all measurements should be taken using the same arm for comparative consistency. • • P osture to obtain accurate measurements Stretch your back. Place the main unit in front of you on a table at heart height. Measure on bare skin. Put your elbow on the elbow rest.Put your palm up and relax. Place the wireless display where you can see it easily. Do not put your hand or any other objects in front of the wireless display. Cautions to obtain accurate measurements Sit quietly and relax while taking measurements. Do not lean forward with your body at an angle during measurements. Take measurements in the proper posture. Do not let your rolled- up sleeve be inside the measuring section. Roll your sleeve far up your arm. Do not take measurements at the elbow. Place your elbow in the elbow rest. 15-162008/01/15 14:54:26
15English16English Taking a Reading Advice: In general, blood pressure measurements are lower on the second attempt because you are calmer. For people who easily get tense, measuring twice each time is advised. When two measurements are taken, normally the smaller measurement is stored. When taking tw o measurements, always stay quiet for 4 to 5 minutes after the first measurement before taking the second measurement. Getting into the proper position for taking a measurement Place the main unit on a table so that it is flat and the center of the unit is at heart height. *Differences in height will cause differences in blood pressure values. If the heights of the chair and table are not suitable, you may have to lean forward. In this case, form a stable stack of books or similar objects under the main unit to adjust the height of the main unit. *You cannot use the unit while lying down. Use only in a sitting position. Elbow rest 1 Put the main unit on a table in front of you and slide your arm through the measurement section of the unit.Remove your shirt or roll up your sleeve so that the measuring section will be on your bare skin. If rolling up your sleeve causes pressure on your arm or underarm, remove your shirt for measurements. Be sure that the rolled-up sleeve is not inside the cuff area when taking measurements. • • 2 Place your elbow on the elbow rest with your palm up and relax. 3 Place the wireless display in front of your body where you can see it easily.The main unit and the wireless display transmit/receive the measured values through infr ared rays. (See page 18.) Do not put your hand or any other objects in front of the wireless display. Differences in blood pressure values between the left and right arms may be around 10 mmHg. The blood pressure can be measured in either the left or right arm, but all measurements should be taken using the same arm for comparative consistency. • • P osture to obtain accurate measurements Stretch your back. Place the main unit in front of you on a table at heart height. Measure on bare skin. Put your elbow on the elbow rest.Put your palm up and relax. Place the wireless display where you can see it easily. Do not put your hand or any other objects in front of the wireless display. Cautions to obtain accurate measurements Sit quietly and relax while taking measurements. Do not lean forward with your body at an angle during measurements. Take measurements in the proper posture. Do not let your rolled- up sleeve be inside the measuring section. Roll your sleeve far up your arm. Do not take measurements at the elbow. Place your elbow in the elbow rest. 15-162008/01/15 14:54:26
17English18English Taking a Reading (continued) Do not take measurements. Within 1 hour of eating or drinking alcohol Immediately after drinking coffee or tea, or smoking In a moving vehicle Immediately after exercising or taking a bath (W ait at least 0 minutes and take a measurement while staying quiet.) In cold places [T ake measurements at room temperatures around 0 °C (68 °F).] When you need to urinate (Wait several minutes after urinating before taking measurements.) Be sure to obey the following points for accurate measurements: Do not move your body, arm, or fingertips. Do not touch the main unit. Do not talk. Do not use a cellular phone near the main unit. • • • • • • • • • • Positions of the main unit and the wireless display Your body reflects infrared rays. For the measurement on your left arm Face the wireless display towards you into position within reach of your hand [within approx. 23⅝ (60 cm)]. * Do not put your hand or any other objects in front of the wireless display. • The main unit and the wireless display communicate with each other using infrared rays. Infrared rays are transmitted from the transmitter on the side panel of the main unit, and reflected on your body and then received on the bottom of the front panel of the wireless display. The units may not communicate due to improper positioning of the wireless display or other environmental conditions (lighting, sunlight, etc.) In those cases, place the wireless display closer to your body. * If you place the wireless display where it cannot receive signals from the main unit, and then press the On/Off and the Start button, the main unit starts a measurement but nothing appears on the wireless display. * If the wireless display is placed where it can receive signals from the main unit while the measurement is executed, results are displayed. The same range (restrictions) as the measurement on your left arm is applied. For the measurement on your right arm If the wireless display cannot receive signals from the main unit, the mark flashes on the wireless display (on the display panel) as shown in the figure on the left. In this case, set the wireless display in the proper position. 17-182008/01/15 14:54:26
17English18English Taking a Reading (continued) Do not take measurements. Within 1 hour of eating or drinking alcohol Immediately after drinking coffee or tea, or smoking In a moving vehicle Immediately after exercising or taking a bath (W ait at least 0 minutes and take a measurement while staying quiet.) In cold places [T ake measurements at room temperatures around 0 °C (68 °F).] When you need to urinate (Wait several minutes after urinating before taking measurements.) Be sure to obey the following points for accurate measurements: Do not move your body, arm, or fingertips. Do not touch the main unit. Do not talk. Do not use a cellular phone near the main unit. • • • • • • • • • • Positions of the main unit and the wireless display Your body reflects infrared rays. For the measurement on your left arm Face the wireless display towards you into position within reach of your hand [within approx. 23⅝ (60 cm)]. * Do not put your hand or any other objects in front of the wireless display. • The main unit and the wireless display communicate with each other using infrared rays. Infrared rays are transmitted from the transmitter on the side panel of the main unit, and reflected on your body and then received on the bottom of the front panel of the wireless display. The units may not communicate due to improper positioning of the wireless display or other environmental conditions (lighting, sunlight, etc.) In those cases, place the wireless display closer to your body. * If you place the wireless display where it cannot receive signals from the main unit, and then press the On/Off and the Start button, the main unit starts a measurement but nothing appears on the wireless display. * If the wireless display is placed where it can receive signals from the main unit while the measurement is executed, results are displayed. The same range (restrictions) as the measurement on your left arm is applied. For the measurement on your right arm If the wireless display cannot receive signals from the main unit, the mark flashes on the wireless display (on the display panel) as shown in the figure on the left. In this case, set the wireless display in the proper position. 17-182008/01/15 14:54:26
19English0English Taking a Reading (continued) Measurement The heart mark will blink when pulse is detected during pressurization. The pressure cuff will be tightened automatically. The entire display will light for approx. 1 second. 1 Press the On/Off and Start button on the main unit. (1) After the entire display on the wireless display has been lit, the pressure cuff will be tightened automatically. ( ) Pressurization will be performed automatically and measurements will be taken. * It may take about 1 second for the wireless display to turn on after pressing the On/Off and the Start button. This is not abnormal. * The angle of the main unit may change slightly while pressurizing. This will not affect accuracy. 2 When measurement has been completed, the blood pressure and pulse values will be displayed. The cuff will loosen after the measurement result has been shown for about 5 seconds. The blood pressure indication ( mark) will flash for 6 seconds then stay lit. If the measurement is high, it will also flash for 6 seconds to indicate high blood pressure. (See page .) The mark will light. (For storing measurement values, see page 4.) The pulse rate value is the beats per minute calculated based on the pulse rate during measurement. The mark is displayed when pulse rate could not be measured. 3 Remove your arm from the main unit. 4 To switch off the blood pressure monitor, press the On/Off and Start button on the main unit. If you forget to switch it off, the blood pressure monitor will switch off automatically after appro ximately minutes. When the Adjust button “” on the wireless display is pressed, indications disappears but the power of the main unit is still on f or about minutes. • • • • • • Movement detection function If movement is detected while measurement is being taken and more pressure is applied by the pressure cuff, the mark will be displayed. Please redo blood pressure measurement again if the mark flashes or is lit. If the arm to which the pressure cuff is attached is moved while measurement is being taken (e.g. the elbow is bent suddenly), in some cases correct measurements may not be obtained. When movement has been detected during measurement, the mark flashes when the blood pressure reading obtained during that measurement is displayed. (If this reading is saved, the mark will light whenever it is recalled.) Movement Mark Explanation FlashingMovement has been detected. · The elbow has been bent suddenly. · The upper arm has been tensed. · The arm has been moved, etc. With correct posture, please take the measurement again after measurement has been completed (see page 16). Lit Movement has been detected that will significantly affect the reading. · Significant movement has been detected. · Movement has been detected repeatedly. · Significant arm movement has been detected, etc. Please take the measurement again (a will appear on the display). * This function is designed to help you to take measurements accurately (see page 33). Even if the Movement mark is not displayed, we recommend that measurements be taken times, with 5 minutes between readings, in order to obtain an accurate reading (see pages 15– 18 for correct position). • • • 19-202008/01/15 14:54:27