Home > Panasonic > Blood Pressure Monitor > Panasonic Ew 3153 Instructions Manual

Panasonic Ew 3153 Instructions Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Panasonic Ew 3153 Instructions Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 10737 Panasonic manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							19English0English
    	Taking	a	Reading	(continued)
    Measurement
    The heart mark    
    will blink when pulse  is detected during  pressurization.
    The pressure cuff 
    will be tightened  automatically.
    The entire display 
    will light for approx.  1 second.
    1	 Press	the	On/Off	and	Start	button	on	the	main	
    unit.
    (1) After the entire display on the wireless display  has been lit, the pressure cuff will be tightened 
    automatically.
    (
    
    ) Pressurization will be performed automatically  and measurements will be taken.
    *
    
      It may take about 1 second for the wireless display 
    to turn on after pressing the On/Off and the Start 
    button. This is not abnormal.
    *
    
      The angle of the main unit may change slightly 
    while pressurizing. This will not affect accuracy.
    2	
    
    When	measurement	has	been	completed, 	the	
    blood	pressure	and	pulse	values	will	be	
    displayed.
    The	cuff	will	loosen	after	the	measurement	
    result	has	been	shown	for	about	5	seconds. 	
    The blood pressure indication (
     mark) will flash 
    for 6 seconds then stay lit. If the measurement is 
    high, it will also flash for 6 seconds to indicate high 
    blood pressure. (See page .)
    The 
     mark will light. (For storing measurement 
    values, see page 4.)
    The pulse rate value is the beats per minute 
    calculated based on the pulse rate during 
    measurement.
    The 
     mark is displayed when pulse rate could not 
    be measured.
    3	
    Remove	your	arm	from	the	main	unit.
    4	 To	switch	off	the	blood	pressure	monitor, 	press	
    the	On/Off	and	Start	button	on	the	main	unit. If you forget to switch it off, the blood pressure 
    monitor will switch off automatically after 
    appro
     ximately  minutes.
    When the Adjust button “” on the wireless display 
    is pressed, indications disappears but the power 
    of the main unit is still on f
     or about  minutes.
    •
    •
    •
    •
    •
    •
    Movement	detection	function
    If movement is detected while measurement is being taken and more 
    pressure is applied by the pressure cuff, the 
     mark will be displayed.
    Please redo blood pressure measurement again if the  mark flashes 
    or is lit.
    If the arm to which the pressure cuff is attached is moved while 
    measurement is being taken (e.g. the elbow is bent suddenly), in some 
    cases correct measurements may not be obtained.
    When movement has been detected during measurement, the 
     
    mark flashes when the blood pressure reading obtained during that 
    measurement is displayed. 
    (If this reading is saved, the 
     mark will light whenever it is recalled.)
    Movement Mark Explanation
    FlashingMovement has been detected.
    · The elbow has been bent suddenly.
    · The upper arm has been tensed.
    · The arm has been moved, etc.
    With correct posture, please take the measurement 
    again after measurement has been completed (see 
    page
     16).
    Lit Movement has been detected that will significantly 
    affect the reading.
    · Significant movement has been detected.
    · Movement has been detected repeatedly.
    · Significant arm movement has been detected, etc.
    Please take the measurement again (a  will 
    appear on the display).
    *
      This function is designed to help you to take measurements accurately 
    (see page 33). Even if the Movement mark is not displayed, we 
    recommend that measurements be taken  times, with 5 minutes 
    between readings, in order to obtain an accurate reading (see pages 15–
    18 for correct position).
    •
    •
    •
       19-202008/01/15   14:54:27 
    						
    							1EnglishEnglish
    	Taking	a	Reading	(continued)
    Irregular	pulse	function
    If the pulse r
     ate varies by more than ±5% from the average while a 
    measurement is being taken, when measurement is completed the 
     mark will be displayed.
    The  mark will be lit after approx. 6 seconds of blinking when 
    the measurement is completed.
    If the 
     mark is lit, please take the measurement again.
    In some cases, if the pulse rate fluctuates greatly while a measurement 
    is being taken, it may be impossible to obtain an accurate reading.
    (Flashes.) (Is lit.)
    If you save a reading where an irregular pulse was recorded, the  mark will be displayed when you recall that reading.
    If the  mark is displayed an irregular pulse rate is detected, 
    not necessarily due to atrial fibrillation. However, if this mark continues 
    to be displayed in subsequent readings, please consult with your 
    physician.
    *
       This function is designed to act as a guide to help you to obtain 
    accurate readings (Please see pages 15–18).
    *  If the 
     mark is frequently displayed, please consult a health 
    professional for your health condition.
    Don’t 	try	to	interpret	readings	or	attempt	to	treat	any	condition	
    yourself.
    Al
     ways	follow	the	guidance	of	a	health	professional.
    •
    •
    •
    •
    Alarm	function	signals	high	blood	pressure
    Blood pressure values for both systolic and diastolic readings are 
    determined and indicated via  mark in the display as being within normal 
    or hypertensive ranges based on definitions and classification of blood 
    pressure levels by the JNC 7. Moreover, when blood pressure is in the high 
    range, the reading displayed on screen flashes on and off to alert user.
    According to JNC 7 Classification, values consistently in excess of 
    140 mmHg (Sys.) and/or 90 mmHg (Dia.) are considered to constitute high 
    blood pressure.
    (mmH g
    )
    Systolic
    mark
    positions
    120140
    80 90
    (mmH
    g
    )
    Diastolic
    Blood pressure
    sign
    Orange
    YellowGreenHypertension
    (Stage 1, Stage 2)
    Prehypertension
    Normal
    Blood pressure readings will flash for approx. 6 seconds 
    when in the high blood pressure range.
    Systolic b lood pressure: 140 mmHg and over
    Diastolic b
    lood pressure: 90 mmHg and over
    (Only when measurement is complete)
    •
    When
    
    	an	Error	Occurs	During	Measurement
    A  will appear on the display to indicate that measurement 
    was unsuccessful and should be performed again.
    Before performing another measurement, always press the 
    On/Off and Start button to turn off the unit and make sure to 
    allo w a rest period of 4 to 5 minutes before starting again.
    Please consult your doctor without delay if the blood 
    pressure indicator repeatedly displays to elevated blood 
    pressure values.
       21-222008/01/15   14:54:28 
    						
    							1EnglishEnglish
    	Taking	a	Reading	(continued)
    Irregular	pulse	function
    If the pulse r
     ate varies by more than ±5% from the average while a 
    measurement is being taken, when measurement is completed the 
     mark will be displayed.
    The  mark will be lit after approx. 6 seconds of blinking when 
    the measurement is completed.
    If the 
     mark is lit, please take the measurement again.
    In some cases, if the pulse rate fluctuates greatly while a measurement 
    is being taken, it may be impossible to obtain an accurate reading.
    (Flashes.) (Is lit.)
    If you save a reading where an irregular pulse was recorded, the  mark will be displayed when you recall that reading.
    If the  mark is displayed an irregular pulse rate is detected, 
    not necessarily due to atrial fibrillation. However, if this mark continues 
    to be displayed in subsequent readings, please consult with your 
    physician.
    *
       This function is designed to act as a guide to help you to obtain 
    accurate readings (Please see pages 15–18).
    *  If the 
     mark is frequently displayed, please consult a health 
    professional for your health condition.
    Don’t 	try	to	interpret	readings	or	attempt	to	treat	any	condition	
    yourself.
    Al
     ways	follow	the	guidance	of	a	health	professional.
    •
    •
    •
    •
    Alarm	function	signals	high	blood	pressure
    Blood pressure values for both systolic and diastolic readings are 
    determined and indicated via  mark in the display as being within normal 
    or hypertensive ranges based on definitions and classification of blood 
    pressure levels by the JNC 7. Moreover, when blood pressure is in the high 
    range, the reading displayed on screen flashes on and off to alert user.
    According to JNC 7 Classification, values consistently in excess of 
    140 mmHg (Sys.) and/or 90 mmHg (Dia.) are considered to constitute high 
    blood pressure.
    (mmH g
    )
    Systolic
    mark
    positions
    120140
    80 90
    (mmH
    g
    )
    Diastolic
    Blood pressure
    sign
    Orange
    YellowGreenHypertension
    (Stage 1, Stage 2)
    Prehypertension
    Normal
    Blood pressure readings will flash for approx. 6 seconds 
    when in the high blood pressure range.
    Systolic b lood pressure: 140 mmHg and over
    Diastolic b
    lood pressure: 90 mmHg and over
    (Only when measurement is complete)
    •
    When
    
    	an	Error	Occurs	During	Measurement
    A  will appear on the display to indicate that measurement 
    was unsuccessful and should be performed again.
    Before performing another measurement, always press the 
    On/Off and Start button to turn off the unit and make sure to 
    allo w a rest period of 4 to 5 minutes before starting again.
    Please consult your doctor without delay if the blood 
    pressure indicator repeatedly displays to elevated blood 
    pressure values.
       21-222008/01/15   14:54:28 
    						
    							3English4English
    	Taking	a	Reading	(continued)
    To	interrupt	measurements	quickly
    Press	the	Quick	air	release	button.
    If pressure cuff will not deflate quickly enough 
    even if the On/Off and Start button is pressed, 
    press and hold the quick air release button on top 
    of the unit. The air will be released in 
    appro
     ximately 4 seconds.
    If	the	pressure	cuff	comes	off:
    1	 With	the	belt	and	seam	 downwards, 	slide	the	pressure	
    cuff	into	the	main	unit.
    Belt
    With the belt and seam 
    downwards, slide the 
    pressure cuff into the 
    main unit. Seam
    2	
     Press	the	pressure	cuff	into	
    the	groove	of	the	main	unit.
    Either the front or the back can 
    be attached first.
    Pressure cuff
    Groove
    Check that the  
    pressure cuff is  
    inserted in the groove  
    on the main unit.
    *When
     	re-attaching, 	be	sure	not	to	touch	the	inside	of	the	main	unit	
    where	the	pressure	cuff	has	been	removed. 
    Doing so may result in an accident or injury.
    	Storing	and	Calling	up	Blood	Pressure	Readings
    Up to 90 readings each can be stored in the memory for two people.
    Storing	blood	pressure	readingsAll data is retained even when the batteries go flat or are changed.
    When measurements are taken and no date has been set, the date and 
    time will be displayed as “
    ” on the wireless display.
    (After approx. 
     2 seconds.)
    (After approx. 
     2 seconds.)
    (After approx. 
     2 seconds.)
    1.	 After	measurement	has	been	completed, 	the		
    mark	will	be	lit	up.
    Press the user selection button on the wireless 
    display to select the desired user.
    To select User 
     … To Step .
    To select User  …   Press the user selection button.
    2.	 Press	the	Memory/Recall	button.
    3.	 Memory	storage	is	completed. A total for 
    	and 	of 90 readings can be stored 
    in memory. Readings over the memory capacity 
    will be recorded over the oldest measurement in 
    memory.
    Readings cannot be saved when a measurement 
    has been unsuccessful (display of the error mark 
     for blood pressure readings). Readings can 
    be saved, however, when the  mark is displayed 
    for the pulse rate.
    The memory number, measurement date, and 
    time are alternately displayed in approximately 
     
     seconds.
    •
    •
    •
    •
    •
       23-242008/01/15   14:54:29 
    						
    							3English4English
    	Taking	a	Reading	(continued)
    To	interrupt	measurements	quickly
    Press	the	Quick	air	release	button.
    If pressure cuff will not deflate quickly enough 
    even if the On/Off and Start button is pressed, 
    press and hold the quick air release button on top 
    of the unit. The air will be released in 
    appro
     ximately 4 seconds.
    If	the	pressure	cuff	comes	off:
    1	 With	the	belt	and	seam	 downwards, 	slide	the	pressure	
    cuff	into	the	main	unit.
    Belt
    With the belt and seam 
    downwards, slide the 
    pressure cuff into the 
    main unit. Seam
    2	
     Press	the	pressure	cuff	into	
    the	groove	of	the	main	unit.
    Either the front or the back can 
    be attached first.
    Pressure cuff
    Groove
    Check that the  
    pressure cuff is  
    inserted in the groove  
    on the main unit.
    *When
     	re-attaching, 	be	sure	not	to	touch	the	inside	of	the	main	unit	
    where	the	pressure	cuff	has	been	removed. 
    Doing so may result in an accident or injury.
    	Storing	and	Calling	up	Blood	Pressure	Readings
    Up to 90 readings each can be stored in the memory for two people.
    Storing	blood	pressure	readingsAll data is retained even when the batteries go flat or are changed.
    When measurements are taken and no date has been set, the date and 
    time will be displayed as “
    ” on the wireless display.
    (After approx. 
     2 seconds.)
    (After approx. 
     2 seconds.)
    (After approx. 
     2 seconds.)
    1.	 After	measurement	has	been	completed, 	the		
    mark	will	be	lit	up.
    Press the user selection button on the wireless 
    display to select the desired user.
    To select User 
     … To Step .
    To select User  …   Press the user selection button.
    2.	 Press	the	Memory/Recall	button.
    3.	 Memory	storage	is	completed. A total for 
    	and 	of 90 readings can be stored 
    in memory. Readings over the memory capacity 
    will be recorded over the oldest measurement in 
    memory.
    Readings cannot be saved when a measurement 
    has been unsuccessful (display of the error mark 
     for blood pressure readings). Readings can 
    be saved, however, when the  mark is displayed 
    for the pulse rate.
    The memory number, measurement date, and 
    time are alternately displayed in approximately 
     
     seconds.
    •
    •
    •
    •
    •
       23-242008/01/15   14:54:29 
    						
    							5English6English
    	Storing	and	Calling	up	Blood	Pressure	Readings	(continued)
    Calling	up	readings	from	memory
    Readings can be recalled after being saved without having to switch the 
    unit on.
    (After approx. 
     2 seconds.)
    (After approx. 
     2 seconds.)
    (After approx. 
     2 seconds.)
    1.	 Press	the	user	selection	button	on	the	wireless	 display	to	select	the	desired	user.
    To select User 
     … To Step .
    To select User  …   Press the user selection button.
    2.	 Press	the	Memory/Recall	button. The average readings of all recorded 
    measurements are displayed.
    When readings from 90 measurements have been 
    saved the average of the readings from all 
    90
     measurements will be displayed. (If there is 
    only 1 measurement recorded, mark 
     will not 
    be displayed.)
    The mark 
     will be displayed.
    Blood pressure indicator  will be lit up for 
    approximately 6 seconds.
    3.	 Press	the	Memory/Recall	button	to	display	 readings	starting	with	from	the	most	recent	
    measurement. Each time the Memory/Recall button is pressed 
    readings are displayed in order of the most recent 
    measurement.
    The memory number, measurement date, and 
    time are alternately displayed in approximately 
     
     seconds.
    4.	 Press	the	Adjust	button	 “2”	on	the	wireless	
    display	to	switch	off	the	unit. If you forget to switch off the unit it will do so 
    automatically after appro
     x. 30 seconds.
    •
    •
    •
    •
    •
    •
    • Deleting 	all	data	stored	in	memory
    (1)   Recall the user whom you want to delete using the user selection 
    button.
    (
    
    )   Recall stored data using the Memory/Recall button.
    (3)   Press the Memory/Recall button again and hold it (for approx. 
    3 seconds) until all readings are deleted.
    NOTE: Readings from individual measurements cannot be deleted.
    (4) 
       If everything has been deleted, the selected  or  mark and  
    will be displayed.
       25-262008/01/15   14:54:30 
    						
    							5English6English
    	Storing	and	Calling	up	Blood	Pressure	Readings	(continued)
    Calling	up	readings	from	memory
    Readings can be recalled after being saved without having to switch the 
    unit on.
    (After approx. 
     2 seconds.)
    (After approx. 
     2 seconds.)
    (After approx. 
     2 seconds.)
    1.	 Press	the	user	selection	button	on	the	wireless	 display	to	select	the	desired	user.
    To select User 
     … To Step .
    To select User  …   Press the user selection button.
    2.	 Press	the	Memory/Recall	button. The average readings of all recorded 
    measurements are displayed.
    When readings from 90 measurements have been 
    saved the average of the readings from all 
    90
     measurements will be displayed. (If there is 
    only 1 measurement recorded, mark 
     will not 
    be displayed.)
    The mark 
     will be displayed.
    Blood pressure indicator  will be lit up for 
    approximately 6 seconds.
    3.	 Press	the	Memory/Recall	button	to	display	 readings	starting	with	from	the	most	recent	
    measurement. Each time the Memory/Recall button is pressed 
    readings are displayed in order of the most recent 
    measurement.
    The memory number, measurement date, and 
    time are alternately displayed in approximately 
     
     seconds.
    4.	 Press	the	Adjust	button	 “2”	on	the	wireless	
    display	to	switch	off	the	unit. If you forget to switch off the unit it will do so 
    automatically after appro
     x. 30 seconds.
    •
    •
    •
    •
    •
    •
    • Deleting 	all	data	stored	in	memory
    (1)   Recall the user whom you want to delete using the user selection 
    button.
    (
    
    )   Recall stored data using the Memory/Recall button.
    (3)   Press the Memory/Recall button again and hold it (for approx. 
    3 seconds) until all readings are deleted.
    NOTE: Readings from individual measurements cannot be deleted.
    (4) 
       If everything has been deleted, the selected  or  mark and  
    will be displayed.
       25-262008/01/15   14:54:30 
    						
    							7English8English
    	Comparing	AM	and	PM	Average	Systolic	Blood	Pressure
    Average AM and PM systolic blood pressure is displayed in order to allow 
    you to compare readings from both time periods.
    Use of the AM and PM average systolic blood pressure comparison 
    function can aid in the detection of a blood pressure pattern known as 
    “early morning surge” in which blood pressure rises rapidly in the morning.
    AM/PM comparison can only be done when the date and time have been 
    set up.
    These can be compared after readings have been saved, and even when 
    the unit is turned off.
    1.	 Press	the	user	selection	button	to	select	the	user	for	a	comparison	of	average	AM	and	PM	
    systolic	blood	pressures.Each time the button is pressed the user switches 
    from one to the other.
    2.
    
    	 Press	the	
    	button.
    The average AM and PM systolic blood pressure 
    for the selected user will be displayed.
    AM… 3:00
      AM to 11:00 AM
    PM… 6:00 PM to :59 AM The average systolic blood pressure recorded 
    during the time periods shown above will be 
    displayed.
    When the time has not been recorded or a single 
    reading has been recorded, “
    ” will be 
    displayed.
    •
    •
    3.	 Press	the	Adjust	button	
    “2”	to	switch	off	the	
    unit. If you forget to switch off the unit it will do so 
    automatically after appro
     x. 30 seconds.
    •
    •
    •
    	After	Use
    Since the wireless display 
    cannot be secured in place, be 
    careful to avoid dropping the 
    unit.
    1	
     Tilt	the	main	unit	(using	the	
    handle)	in	the	direction	of	the	
    arrow	until	the	unit	is	locked.
    2	 Place	the	wireless	display	on	top	 of	the	unit	securely.
    To	transport	the	unit Carry the main unit and the 
    wireless display separately.
    •
       27-282008/01/15   14:54:30 
    						
    							7English8English
    	Comparing	AM	and	PM	Average	Systolic	Blood	Pressure
    Average AM and PM systolic blood pressure is displayed in order to allow 
    you to compare readings from both time periods.
    Use of the AM and PM average systolic blood pressure comparison 
    function can aid in the detection of a blood pressure pattern known as 
    “early morning surge” in which blood pressure rises rapidly in the morning.
    AM/PM comparison can only be done when the date and time have been 
    set up.
    These can be compared after readings have been saved, and even when 
    the unit is turned off.
    1.	 Press	the	user	selection	button	to	select	the	user	for	a	comparison	of	average	AM	and	PM	
    systolic	blood	pressures.Each time the button is pressed the user switches 
    from one to the other.
    2.
    
    	 Press	the	
    	button.
    The average AM and PM systolic blood pressure 
    for the selected user will be displayed.
    AM… 3:00
      AM to 11:00 AM
    PM… 6:00 PM to :59 AM The average systolic blood pressure recorded 
    during the time periods shown above will be 
    displayed.
    When the time has not been recorded or a single 
    reading has been recorded, “
    ” will be 
    displayed.
    •
    •
    3.	 Press	the	Adjust	button	
    “2”	to	switch	off	the	
    unit. If you forget to switch off the unit it will do so 
    automatically after appro
     x. 30 seconds.
    •
    •
    •
    	After	Use
    Since the wireless display 
    cannot be secured in place, be 
    careful to avoid dropping the 
    unit.
    1	
     Tilt	the	main	unit	(using	the	
    handle)	in	the	direction	of	the	
    arrow	until	the	unit	is	locked.
    2	 Place	the	wireless	display	on	top	 of	the	unit	securely.
    To	transport	the	unit Carry the main unit and the 
    wireless display separately.
    •
       27-282008/01/15   14:54:30 
    						
    							9English30English
    	Specifications
    Power sourceDC 6 V (Main unit: Four AA/LR6 size alkaline 
    batteries)
    A C 10  V 60 Hz (when using AC adapter)
    DC 3  V (Wireless display: Two AA/LR6 size 
    alkaline batteries)
    Display Digital LCD
    Method of measurement Oscillometric system
    Measurement range Blood pressure:
      0 mmHg to 80 mmHg
    Pulse r
    ate: 30 to 160 beats per minute
    Accuracy Blood pressure:
      Within ±3 mmHg
    Pulse rate: Within ±5%
    Operation temperature/
    humidity range 10
      °C to 40 °C (50 °F to 104 °F), 15% to 90% 
    RH
    Storage temperature/
    humidity range –
     0 °C to 60 °C (–4 °F to 140 °F), 
    10% to 95% RH
    Measurable arm 
    circumference Approx. 7¾ to 13⅜ (
    
    0 cm to 34 cm)
    Weight Main unit:
      Approx. 1550 g (54.7 oz.) (not 
    including batteries)
    Wireless displa y: Approx. 160 g (5.6 oz.) (not 
    including batteries)
    Dimensions Main unit: 11⅛ × 6⅝ × 11⅛ (
    8.3 cm × 
    16.9 cm × 8.1 cm)
    Wireless display: 5¾ × 3¼ × 
     ⅝ (14.6  cm × 
    8.4 cm × 6.8 cm)
    Protection against 
    electric shock Internally powered equipment, Type BF 
    applied part
    AC adapter
    Input AC 10 V 60 Hz
    Output DC 6
      V .0 A
    	Q	&	A
    Q Why	are	the	blood	pressure	values	measured	at	home	different	
    from	those	measured	at	the	doctor’s	office?
    A Blood pressure v
     aries in a 4-hour period.
    In addition, it can vary greatly depending on the weather, your 
    mental state, amount of exercise, etc.
    In particular, there is the phenomenon of “White uniform 
    hypertension” at doctor’s offices, where the blood pressure 
    measured in a doctor’s office is higher because of stress or anxiety.
    In
    
    	addition, 	if	the	blood	pressure	measured	at	home	is	
    sometimes	higher	and	sometimes	lower	than	the	blood	
    pressure	measured	at	the	doctor’s	office, 	the	following	can	be	
    considered. 	Please	pay	attention	to	these	points	when	taking	
    measurements.
    1
    
      Measuring posture was poor.
    Did you take measurements while leaning forward with your body 
    at an angle?
      Measurements w
    
    ere not taken at the upper arm.
    Please take measurements with your elbow in the arm support of 
    the main unit.
      (See pages 15–17.)
    3  Were you anxious or upset during measurements?
    Before taking measurements, take  or 3 deep breaths and relax, 
    so that your blood pressure stabilizes.
    Q Wh
    y	do	the	main	unit	and	the	wireless	display	operate	together	 even	when	they	are	separated?
    A The main unit and the wireless display communicate using infrared  rays. Infrared rays are transmitted from the transmitter on the side 
    panel of the main unit, and reflected on your body and then received 
    on the bottom of the front panel of the wireless display.
    Q The
     		mark	does	not	light	when	an	arm	is	moved.
    A The 
     mark is a function to notify you that your arm is bent or to 
    indicate extra loads on the pressure cuff.
    Therefore, it may not indicate motion (bending an arm slightly, etc.).
    If you are concerned about the result, retry the measurement.
       29-302008/01/15   14:54:30 
    						
    All Panasonic manuals Comments (0)

    Related Manuals for Panasonic Ew 3153 Instructions Manual