Page 71
Drainage –The drainage system is fitted with a non-return valve which prevents dirty water from flowing back into the machine via the drain hose. –The machine should preferably be connected to aseparatedrainage system on-site. If separate drainage is not available, contact your Miele application specialist for professional advice. The on-site drain connection point should be sited between 0.3 m and 1mabove the lower edge of the machine. If it is lower than 0.3 m, lay the hose in a curve at a height of at...
Page 72
Programme When to use AFreely programmable programme (programmed by a Miele authorised service technician). B INORGANICFor the removal of inorganic residues; general analysis and water analysis, suitable for low to medium soiling levels and for medium to high quality rinse requirements. C ORGANICFor the removal of organic residues such as oil, grease, wax and agar. Suitable for preparation and analytical applications with medium to high levels of soiling and medium quality rinse requirements....
Page 73
Programme sequence Pre-wash 12Main wash 12Rinse 12Rinse 34Final rinses* 12Dryingx 12 CW HW DOS 1 80°C 3 minHW DOS 3HW AD AD 75°C 1 min99°C 70°C 25 min 5 min CW DOS 3 50°C 1 minHW DOS 1 75°C 3 minHW DOS 3AD AD AD 70°C 1 min99°C 70°C 25 min 5 min HW HW DOS 1 DOS 1 65°C 85°C 3 min 3 minHW DOS 3HW AD AD 80°C 1 min99°C 70°C 25 min 5 min HW DOS 1 75°C 3 minHW DOS 3AD AD 75°C 1 min99°C 70°C 25 min 5 min CW HW DOS 1 80°C 3 minHW DOS 3HW AD AD 75°C 1 min99°C 70°C 25 min 5 min CW HW DOS 1 85°C 3 minHW DOS 3AD AD...
Page 74
Height with lid Height without lid85 cm 82 cm Width 60 cm Depth Depth with door open60 cm 120 cm Weight (net) 70 kg Voltage, connected load, fuse rating See data plate Mains cable Approx. 1.8 m Water temperature: Cold water Hot and AD waterMax. 20 °C max. 70 °C Static water pressure Max. 1000 kPa (10 bar) pressure Minimum flow pressure: Cold/hot water inlet connections AD water connection100 kPa pressure 60 kPa pressure Recommended flow pressure: Cold/hot water inlet connections AD water connection0250...