Miele G 7825 Manual
Here you can view all the pages of manual Miele G 7825 Manual. The Miele manuals for Washer Disinfector are available online for free. You can easily download all the documents as PDF.
Page 41
Operating levels There are four operating levels available on this machine: A = Fixed and free access programmes B = Free programme selection C = AWK – Automatic mobile unit recognition (programme selection via mobile unit coding) - optional - D = Programming / free programme selection / change code (see Programming Manual) To change operating level: Pressandat the same time. Operating levelsABCDareshown in the display. Select the operating level you want using theorbuttons. Pressto confirm...
Page 42
Starting a programme See the Programme charts in the Programming Manual supplied with the machine for detailed information and important notes on the standard Miele programmes. When processing medical devices, any changes made to programmes or dispensing systems must be documented in a log book kept with the machine in accordance with the Medical Devices directive . The machine's cleaning and disinfecting standards must also be re-validated. A. Fixed programme Set up fixed programme(s) once in...
Page 43
C. AWK – Automatic mobile unit recognition Select operating level C. Push the coded mobile unit into place. Make sure that there are no small metallic objects or instrument parts attracted to the magnetic strip, in particular to the underside of it. Any metallic objects on the strip can result in the coding being incorrectly read. Close the door. Before starting a programme it is absolutely essentialto check that the programme shown in the display is the correct on for the mobile unit being used....
Page 44
Switching off Press and release theI-0button. Turn off the stopcocks. Cancelling a programme A programme can only be cancelled in operating level B or D. When testing to EN ISO 15883-1 or HTM 2030 standards, the programme must not be interrupted or cancelled. In operating levels B or D Press the Stop button . The programme is interrupted. CANCEL or CONTINUE< will appear in the display. With established decontamination programmes any contaminated water must be decontaminated using a disinfecting agent...
Page 45
Interrupting a programme A programme can only be interrupted in operating level B or D. If you absolutely have to open the door, e.g. because the load is obviously unstable (conscious intervention): After interrupting a DESIN disinfection programme, if you wish to continue it please check the message in the display at the end of the programme. If the message PROCESS PARAMETERS NOT MET is displayed, it means the door was openedafterthe disinfection part of the programme and the DESIN parameters were not...
Page 46
Each machine is supplied witha5m long interface cable for data transfer between the Profitronic system and an external report printer or PC. The interface cable is rolled up inside the machine and must only be connected by an authorised Service Dealer or the Miele Service Dept. The serial interface is RS 232 compatible. For the interface configurations see "PC/Printer Function" in the Programming manual. Various printers can be used as external printers: – Epson-compatible (contact Miele for a...
Page 47
Periodic checks This washer-disinfector should be inspected in accordance with local and national safety regulations after every 1000 operating hours, or every 6 months by a Miele approved service technician. This maintenance will cover the following: –Electrical safety tests –Door mechanism and door seal –Any screw connections and connectors in the wash cabinet – Water inlet and drainage – Internal and external dispensing systems – Spray arms – Filter combination – Sump including drain pump and...
Page 48
Process validation The standard of cleaning and disinfection in the disinfection programmes must be confirmed by the user as a routine matter. Safety checks must be carried out to the machine every 3 months by a Miele approved service technician in accordance with EN ISO 15883 and HTM 2030. Routine checks Before the start of each working day the user must carry out a number of routine checks. A check list is supplied with the machine for this purpose. The following need to be inspected: –All filters in...
Page 49
Cleaning the filters in the wash cabinet The filters in the base of the wash cabinet are designed to prevent coarse soiling getting into the circulation system. A build-up of coarse soiling can cause the filters to clog up. They should, therefore, be checked daily and cleaned if necessary. The machine must not be used without all the filters in place. Watch out for glass splinters which could cause injury.Cleaning the fine filter Remove the filter and clean it. Cleaning the flat filter Remove the...
Page 50
Cleaning the filters for the circulation pump There are two filters located underneath the flat filter to the right of the heating element block to protect the circulation pump. Pull them up and out of their holders, and clean them. Replace the filters by carrying out the above steps in the reverse order. Maintenance 50