Hitachi V-209 Service Manual
Here you can view all the pages of manual Hitachi V-209 Service Manual. The Hitachi manuals for Oscilloscope are available online for free. You can easily download all the documents as PDF.
Page 21
Then connect the battery pack connector to the connector of the oscilloscope. d) Mount the plate with screws. e) Fix the upper enclosure on the oscilloscope with four screws. Fig. a Fig. b Upper enclosure , Bauery Ba ucry pack Battery pack fi,ing �crew AD-209 When mounting the battery pack AD-209, take care not to damage the connecting cables and components. 6. HOW TO PRODUCE THE BRIGHT LINE Before turning 0 the POWER switch. insure the power supply...
Page 22
If the instrument is not used with the power supply turned on. rotate the I TENsity counterclockwise to reduce the brightness and also blur the FOCUS. NOTE For usual observation, leave the following non-calibrating function section set to ''CAL" position. VARIABLE Rotate in the direction of arrow. In this case, the VOLTS/DIV is calibrated to its indicating value. SWP V AR Leave the knob in depressed state. In this case, the TIME/DIV is...
Page 23
MODE Switch of Vertical Axis TRIG OURCE (2) When observing two wavefonns CH2 CH2 Observation of two wavefom1S can be made easily by setting the MODE switch of vertical axis to ALT or CHOP. When observing two wavefonns of high repetition frequen cies, set the MODE switch to ALT and, in the case of low frequencies, set it to CHOP. Measu rement of the phase difference is per formed after effecting synchronization with leading phase signal. (3) When...
Page 24
Connection of earth lead wire 20 (a) A good example (b) A bad example For better measurement. it is required to use an earth attachment available at option. Multiply the reading ofVOLTS DIY by 10. For example. if the VOLTS/DIY ts SOmV/DIV. then read the wavefonn as SOmV/DIV x 10 = SOOmV/DIV o To avoid measurement error, put the probe in the follow ing correction state and check it before measurement without fail. Connect the tip of the probe to the...
Page 25
In general. tt tS safe to avoid measunng with non shielded connecting wire. When using a shieldmg wire, connect one end of the shield to the earth terminal of the oscilloscope and the other end to the grounding of the circuit to be measured. It is deirable to use a coaxtal cable with B C type connector. The following cautions must be observed when per forming a wide band measurement. It ts necessary to terminate with the characteristic impedance of the cable...
Page 26
8. MEASURING PROCEDURE The first things to do are as follows. o Bring the brightness and FOCUS at optimum positions for easy read out. � Display the waveform as large as possible to minimiLe the read error. Check the capacity correction when using a probe. (Refer to paragraph (I) "When using a probe" of Sec tion 7. "Method for connecting signals" for correcting capacity.) (I) DC voltage measurement Set input coupling to G D and decide the...
Page 27
r/f\ ----�ff\ -- 1\ I 1\ \17 1\ H--+--f--t-1-t-+---+-_+=1_ -- --- (3) Measurement of frequency and period This will be explained with the following figure. One period covers the time A and time B, which are sepa rated from each other by 2.0 DIY on the screen. When the sweep time is I ms/DIY, the period is given by 1 ms/DIV x 2.0 = 2.0 ms =2.0 X 10-3 S Accordingly, the frequency is 1/(2.0 X J0- 3)= 500Hz (However, when the knob MAG x!O is at pulled out...
Page 28
24 This means that CHI is used as the triggering signal when investigating the length of time by which the signal of CH2 is delayed from the signal of CH 1. C H2 is used in the reversed case. In other words, the signal leading in phase is selected as the triggering signal source. If this process is reversed, the portion to be measured may sometimes not appear on the screen. Thereafter, equalize the amplitudes of the two signals appearing on the screen...
Page 29
f c x tr = 0.35 Where, f c : Frequency band (Hz) tr: Rise time (s) The rise time and fall time are determined by the time elapsed between the I 0'1 to 9Qc,r \'a lues of pulse width. (6) Synchronization of complexed waveform In the case shown in the Fig. (a) below where two wave forms have difference in amplitude alternate. the wave form is doubled if the trigger level is not set properly. In the case where the trigger level is selected as Y...
Page 30
26 Q) Difference in the circuits Exclusive circuit for conventional oscillograph General circuit Video �ignal To trigger circuit Hard to synchronize, because video signal is applied directly as trigger signal. Simple synchronizing circuit To trigger circuit Synch ronization is more easily effected than in the circuit shown at left, because the signal is integrated to remove high frequency components. Exclusive circuit for this instrument (Principle...