Hitachi V-209 Service Manual
Here you can view all the pages of manual Hitachi V-209 Service Manual. The Hitachi manuals for Oscilloscope are available online for free. You can easily download all the documents as PDF.
Page 11
The S\\ itches are set as shown in the table below to establi�h each power �upply mode Switche Po\.1.-er sourc e Power mode upply lnode select witch (J) select witch @ ---- t AC IXOP DC D OP DC t- 0 OP When charging DC BATT CI-IARGC (AD-209) *DC II to 14V IS supphed via EXT DC I PUT connector. After completing the charge of AD-209, be sure to set power mode select sw1tch to DC OPE. When the output voltage of DC power supply becomes less than II...
Page 14
CD POWER switch After checking the line voltage select switch, set POWER switch set to OFF and then connect the AC cord to AC wall socket. (}) POWER lamp This lamp goes on in red when the power supply is in ON state. G) CHARGE lamp This lamp goes on in red in the battery charge mode. When the battery is completely charged, this becomes dim. @) FOCUS control After obtaining an appropriate brightness by operating INTENsity,adjust FOCUS until the bright...
Page 15
@ CH2 INPUT connector Thr same as CHI, but when the instrument is used as an X- Y oscilloscope, the signal input to this terminal becomes the Y -axis signal. @@ Input coupling switches (AC-GND- DC) Used to select the coupling system between the input signal and vertical axis amplifier. AC At this setting. the signal is connected through a condenser. The DC component of the input signal is cut off and only the AC component is dis· played. GND At this...
Page 16
AL T Signals applied respectively to CHI and CH2 appear on the screen alternatively at each sweep. This setting is used when the sweep time is short in 2-channel observation. CHOP At this setting, the input signals applied respec tively to CHI and CH2 are switched at about 250 kHz independent of the sweep and at the same time appear on the screen. This setting is used when the sweep time is long in 2-channel observation. ADD The algebraic sum of the...
Page 17
It is indisp ensable to the measurement of the time of waveform. Bright line is moved toward right when the knob IS ro tated clockwise and toward left with counterclockwise rotation. The outer shaft is for coarse adjustment and the inner shaft for fine adjustment. @ x 10 MAG control A and B sweeps are magnified I 0 ume s by pulling out FINE knob (inner shaft) of POSITION. In this case, the sweep time is I /10 of the value indicated by TIME tDJV. Bring...
Page 18
LINE This setting is used when observing a signal trig gering with power supply line frequency. EXT External triggering signal applied to TRIG IN PUT becomes the triggering signal. This setting is used when triggering with a special independency of the vertical axis signal. @ TRIG INput connector Input terminal used for external triggering signal of A sweep. @ TRIG LEVEL control 14 Used to decide at which portion of the waveform the sweep should be...
Page 19
quency signal (30Hz or less). TV(V) This setting is used when observing the enure vertical picture of television signal. TV(H) Th1s setting is used when observing the entire horizontal picture of television signal. (NOTE) Both TV V and TV H synchronize only when the synchronizmg signal is negative. (5) Miscellaneous @ EXT BLANK connector Input terminal for brightness modulation. It is of the DC coupling. The brightness is reduced with a positive signal and...
Page 20
Charging of battery I) ( onne..:t the \C cord 10 an AC outlet and set the ,\(-DC swnch (J) to DC and the DC 01' B \ fT CIIARGL switch@ to BAIT CHARGE. :::! ) Turn on the pl)>N er swit..:h CD and the charging will stan. In this case. the POWER lamp is not illuminated. 3) T he ( 111\RG! la mp Q) is bemg J!luminated dunng the � h arg e. II will take about IS hours until the batte ry is charged completely. When the charging is completed, the CHARGf...