Hitachi V-209 Service Manual
Have a look at the manual Hitachi V-209 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
Then connect the battery pack connector to the connector of the oscilloscope. d) Mount the plate with screws. e) Fix the upper enclosure on the oscilloscope with four screws. Fig. a Fig. b Upper enclosure , Bauery Ba ucry pack Battery pack fi,ing �crew AD-209 When mounting the battery pack AD-209, take care not to damage the connecting cables and components. 6. HOW TO PRODUCE THE BRIGHT LINE Before turning 0 the POWER switch. insure the power supply voltage is within the range of 90-J30V for AC IOOV set. 180 260V for AC 200V set. Insert the plug of the power cord on the rear panel into the power supply wall socket and set the controls as follows. POWER OFF IN TEN Counterclockwise to the full FOCUS M1drange AC-GND-DC GND �I POSITION Midrange {the knob is left depressed ) V. MODE (Ill TRIG AUTO TRIG SOURCE CHI TIME/DIY 0.5 ms/div ::: POSITION I Midrange Set all the levers of the switches to the upper side. After ending all the setting mentioned above. turn 0 the POWER and. 15 second later, rotate the INTEN knob clock wise. Then the sweep bright line will appear. If the observation is to be started immediately. set the FOCUS control at a point where the bright line is sharpest. 17
If the instrument is not used with the power supply turned on. rotate the I TENsity counterclockwise to reduce the brightness and also blur the FOCUS. NOTE For usual observation, leave the following non-calibrating function section set to ''CAL" position. VARIABLE Rotate in the direction of arrow. In this case, the VOLTS/DIV is calibrated to its indicating value. SWP V AR Leave the knob in depressed state. In this case, the TIME/DIV is calibrated to its indicating value. Align the bright line with the horizontal scale line at the center of the screen by operating CH J POSIT ION. In some cases, the bright line may be oblique to the scale slightly by the effect of earth magnetism. In this case. bring the bright line until it lies on the horizontal scale line at the center of the screen by properly adjusting the semi- Gxed variable resistor TRACE ROTATION on the front panel. 18 GENERAL MEASUREMENT (I) In the case of observing a single waveform. Use CHI or CH2 when not observing the phase difference between two waveforms or when engaging in a operation other than X- Y operation. Make the following settings when using CHI . MODE Switch of Vertical defection system CHI MODE Switch of TRIG AUTO TRIG SOURCE CHI Under these settings, almost all the repetitive signals of about 30 Hz or more applied to CHI can be synchro nized and observed by adjusting TRIG LEVEL. Since the MODE of horizontal axis is at AUTO position, the bright line appears even when no signal is present or when input coupling switch is at GND position. This means that the measurement of DC voltage can be measured. The following switching is needed when observing low fre quency signals of about 30 Hz or less. MODE of TRIG ORM Synchronization can be effected by operating LEVER knob under this setting. When using only CH2, use the instrument after making the following settings.
MODE Switch of Vertical Axis TRIG OURCE (2) When observing two wavefonns CH2 CH2 Observation of two wavefom1S can be made easily by setting the MODE switch of vertical axis to ALT or CHOP. When observing two wavefonns of high repetition frequen cies, set the MODE switch to ALT and, in the case of low frequencies, set it to CHOP. Measu rement of the phase difference is per formed after effecting synchronization with leading phase signal. (3) When observing wavefonn with X-Y Set the MODE switch of vertical defection system to CH2 (X- Y) and TIM E/DIV switch to X-Y. Then the instrument works as an X-Y oscilloscope. Each input is applied to the instrument as follows. X-axis signal (horizontal axis signal) Y -axis signal (vertical axi� signal) CHI INPUT CH2 INPUT In this case.leave the horizontal axis magnification switch (PULL-MAG x 10 inner shaft knob) at depressed position. 7. METHOD FOR CONNECTING SIGNALS The first step of measurement is to introduce the signal desired to measure to the oscilloscope properly. Do it with utmost care. (I) When using a probe Use the attached probe, AT-I 0 AK I .5, when measuring a high frequency wave with high accura cy. It should be noted, however, that since the input signal is attenuated by this probe to 1/10 before it is input to the oscilloscope, the use of the probe is disadvantageous for low signals, and that at the same time, the measuring range is extended by that amount for high signals. o Do not apply a signal which exceeds 250V (DC+ peak AC at I kHz). o Bring the grounding point of the earth lead wire of the probe close to the point to be measured when measur ing a rapid rising signal or a high frequency signal. Long earth lead wire may cause wavefonn distortions such as ringing and overshoot. 19
Connection of earth lead wire 20 (a) A good example (b) A bad example For better measurement. it is required to use an earth attachment available at option. Multiply the reading ofVOLTS DIY by 10. For example. if the VOLTS/DIY ts SOmV/DIV. then read the wavefonn as SOmV/DIV x 10 = SOOmV/DIV o To avoid measurement error, put the probe in the follow ing correction state and check it before measurement without fail. Connect the tip of the probe to the output tenninal CAL O.SV of I kHz calibration square wave voltage. When this correction capacity value is at optimum, the wavefonn takes the shape as shown in Fig. (a) as follows. If the wavefonn is as shown in Fig. (b) or Fig. (c), rotate the semiftxed adjusting screw on the matching box of the probe by using a screwdriver until the opti mum state is obtained. L Capacitance correction trimmer JliUlJ1 MJUl MM (a) Optimum (b) Capacity too small (2) At time of direct connection (c) Capacity too large When connecting a signal directly to the oscilloscope not usmg the attached probe AT -10 AK 1.5 (I O· 1 ), pay atten tion to the following points in order to minimize the mea surement error. o When perfonning observation using a bare lead wire, no trouble occurs on the circuit to be measured at low impedance and high level. However. note that, in most cases. measurement error may be caused by static stray coupling with other circuit and power line. This measurement error cannot be ignored even in low frequency region.
In general. tt tS safe to avoid measunng with non shielded connecting wire. When using a shieldmg wire, connect one end of the shield to the earth terminal of the oscilloscope and the other end to the grounding of the circuit to be measured. It is deirable to use a coaxtal cable with B C type connector. The following cautions must be observed when per forming a wide band measurement. It ts necessary to terminate with the characteristic impedance of the cable when measuring a rapid nsing waveform or a high fre quency wave. Especially when using a long cable. the absence of a terminating resistor will necessarily lead to a measure ment error derived from ringing phenomenon. Some measunng circutls require a term mating reststor equal to the charactensttc unpedance of the cable also on the measurement terminal stde. B C type terminating reststor (50 Q) ts conveniently used for this purpose. c In order to perform measurement with the measuring circuit put in proper operating state, it is sometimes necessary to terminate the cable with an impedance which corresponds to the circuit to be measured. o The stray capacity of the shield wire must be taken in 10 account when performing measurement with a long shield wire. Since the shield wire normally in use has a capacity of about I 00 pF per meter. its effect on the circutt to be measured cannot be ignored. Use a probe to minimize the cffection the circuit. When the length of the shield wire is used or when the length of the non-terminated cable reaches I 4 wave length or its multiples within the band of V-:!09 type (I 4 wave length is about 3 meter when using a coaxial cable at }0 MHz), oscillation may be caused near 5 mV DIY range. This is caused by the resonance between the externally connected high-0 inductance and the input capacity and can be avoided by reducing the Q. Connect the cable or shield wire to the input connector by way of a senally connected I OOU to I n reststor, or perform measurement at other VOLT/DIY range. 21
8. MEASURING PROCEDURE The first things to do are as follows. o Bring the brightness and FOCUS at optimum positions for easy read out. � Display the waveform as large as possible to minimiLe the read error. Check the capacity correction when using a probe. (Refer to paragraph (I) "When using a probe" of Sec tion 7. "Method for connecting signals" for correcting capacity.) (I) DC voltage measurement Set input coupling to G D and decide the zero level prop erly. Set VOL TS/DIV appropriately and set AC-G D-DC to DC. Since the bright line shifts here by the amount of DC voltage, the DC voltage of the signal can be obtained by multiplying the shift width by the indicated value of VOLTS/DIV. When VOLTS/DIY is 50 mV/DIV, then SO mV/D IV x 4.2 = 210mV (However, if the probe AT- 10 AK I.S (10:1) is in use, the true value of the signal be comes 10 times of the value, or SO mV/DIV x 4.2 x 10 = 2.1V.) DC voltage (after shifting) Zero level (reference line) (2} AC voltage measurement The same as paragraph 8 (1). "DC voltage measurement", but here there is no need of align the zero level with the scale line. Set the zero level to the position that can be easily observed. In the drawing as follows, VOLTS/DIV is IV/DIY, IV/ DIY x S = S Vp-p (SOVp-p at time using the probe AT- 10 AK 1.5(10 : 1)). When magnifying and observing a small amplitude signal, superimposing on a high DV voltage, set the input coupling to AC. The DC voltage is cut off and AC voltage can be observed by increasing sensitivity.
r/f\ ----�ff\ -- 1\ I 1\ \17 1\ H--+--f--t-1-t-+---+-_+=1_ -- --- (3) Measurement of frequency and period This will be explained with the following figure. One period covers the time A and time B, which are sepa rated from each other by 2.0 DIY on the screen. When the sweep time is I ms/DIY, the period is given by 1 ms/DIV x 2.0 = 2.0 ms =2.0 X 10-3 S Accordingly, the frequency is 1/(2.0 X J0- 3)= 500Hz (However, when the knob MAG x!O is at pulled out posi tion, TIME/DIY must be converted to I /10 since the sweep is magnified.) I-+--1--I-�. * • -- - ---- --7 1\ f/ l I- T\ II\ 7 1\ 1/ I\ rr ., u -- - - Time A Time B II \ - - - 1- -lh__: v � I- (4) Measurement of time difference Triggering signal source ''SOURCE" is selected as an offer ing reference signal when measuring the time difference be tween two signals. Assume that pulse trains as shown in (a). Then (b) shows the case when CHI is taken as the triggering signal source and (c) the case where CH2 is taken. I I CHI� CH2��- (a) (b) I I -_I ,- +-.:::.CH:..:...:..I __ _ I CH2 (c) 23
24 This means that CHI is used as the triggering signal when investigating the length of time by which the signal of CH2 is delayed from the signal of CH 1. C H2 is used in the reversed case. In other words, the signal leading in phase is selected as the triggering signal source. If this process is reversed, the portion to be measured may sometimes not appear on the screen. Thereafter, equalize the amplitudes of the two signals appearing on the screen or superimpose one on another. Read the time difference between 50% amplitude points of the two signals. Sometimes the superimposing method is more convenient from the point of view of procedure. � it'----� Time => difference Equalize "' ��amplitud es - - --- J _-:;-by V AR ( a) Equal amplitude measuring method (b) Suerposition measuring method Since the pulsed wave contains many high-frequency wave components (higher harmonics) depending on its width or period, pay the same attention as given to high frequency signals when handling it. Accordingly, use a probe or coax ial cable and shorten the earth lead wire as much as possible. (5) Measurement of rise (fall) time To measure the rise time, pay attention not only to the abovementioned items but also to measurement error. The following equation is provided for the relation among the rise time Trx of the wavefom1 to be measured, the rise time Trs of oscilloscope, and the rise time Tro displayed on the screen. Tro =../ Trx2 + Trs2 When the rise time of the pulse going to be measured is sufficiently longer than the rise time of the oscilloscope (17.S ns in our case),the effectofthe rise time of the oscil loscope on the measurement can be neglected. However, if both are close to each other, measurement error may be caused. The true rise time is given by Trx = .../ Tro2 - Trs2 Moreover, in general, in a circuit free from waveform dis tortion such as overshoot and sag, the following relation ship is established between frequency band and rise time.
f c x tr = 0.35 Where, f c : Frequency band (Hz) tr: Rise time (s) The rise time and fall time are determined by the time elapsed between the I 0'1 to 9Qc,r \'a lues of pulse width. (6) Synchronization of complexed waveform In the case shown in the Fig. (a) below where two wave forms have difference in amplitude alternate. the wave form is doubled if the trigger level is not set properly. In the case where the trigger level is selected as Y line two waveforms. one starting with A and advancing to B. C. D, E. F .... and the other starting �ith E and advancing to F, G. H, I .... will appear alternately on the screen. They will be doubled as shown in Fig. (b), for which no synchro nization can be taken. In such a case, rotate LEVEL clockwise until the trigger level comes to Y' line. Then the waveform on the screen be comes the one shown in F1g. (c) above wh1ch starts with B and advances to C, E. r, ... and which allows synchro niLation. \/TC\ .. 7\ h/"'-� 0\ �� � v J.......,. L� p= (a) Signal waveform Y' Trigger level y �Lting line (b) When the uiggc r (c) When the uigger ...:ning level •� Y selling level i' y· Synchronization of complexed waveform (7) How to use TV exclusive synchronization CD On the image waveform of TV In the work concerned w1th TV. comple:-.ed signals con taining video signal. blanking pedestal signal. and syn chronizing s1gnal are often measured. llowever, since the waveform is complexed, a special circuit IS needed 10 effect a stable syncrhonization with vertical waveform. Synchronizing signal pulse (SYNC pulse) IV (vertical) ------� 25
26 Q) Difference in the circuits Exclusive circuit for conventional oscillograph General circuit Video �ignal To trigger circuit Hard to synchronize, because video signal is applied directly as trigger signal. Simple synchronizing circuit To trigger circuit Synch ronization is more easily effected than in the circuit shown at left, because the signal is integrated to remove high frequency components. Exclusive circuit for this instrument (Principle drawing) TV exclusive synchronizing separator circuit To trigger cucuit After picking up the SY C pulse, the vertical synchronization is separated. Then, the stable syn chro nization is obtained.