    							Process CalibratorTechChek™ 830 
    						
    							1
    Your New TechChek 830
    Lighten your load…take the TechChek 830 to every site.
    It’s like bringing a cartload of test equipment from the shop
    to the control room or the field. The TechChek 830 sources
    and reads DC like a milliamp or voltage calibrator, simulates
    and measures T/Cs & RTDs like a temperature calibrator,
    generates and counts frequency and Counts-Per-Minute
    like a frequency calibrator and displays pressure like a preci-
    sion test gauge. Troubleshooting? It checks continuity with a
    beeper and measures AC line voltage like a multimeter!Calibrate Milliamp Inputs
    Calibrate controllers, recorders and other devices in 4
    to 20 or 0 to 20 mA loops.  Source and read 0.00 to 24.00
    mA, or Simulate a 2-Wire Transmitter. Use the optional AC
    adaptor for continuous operation. Display to 0.01 mA , 0.1
    % of 4 to 20 and 0.1 % DP Flow.Calibrate 2-Wire Transmitters
    Easily calibrate 2-Wire Pressure and Electronic
    Transmitters by connecting the TechChek 830 to both the
    input and output of the transmitter. The TechChek 830 will
    simultaneously indicate the input and output of the transmit-
    ter on the graphical display.Calibrate Pressure Systems
    Read pressure with extreme accuracy using an external
    QuikCal pressure module in a companion ModPak module
    holder. Attach the module directly to the pressure connec-
    tion for the best accuracy or with optional tubing for tight
    spots. Display to five digits within 0.025% of reading in up
    to 20 engineering units including psi, pa, Kpa, Mpa, BAR,
    mBar, Atm, Kgf plus torr, inches and mm of mercury orGENERAL INFORMATION
    CONTENTS
    GENERAL OPERATING INSTRUCTIONS...........................3
    TURN ON ...................................................................3
    CONNECTIONS .........................................................3
    FIELD & BENCH USE (TILT STAND) ........................4
    CHANGING BATTERIES............................................4
    CONFIGURING TEMPERATURE SCALES ...............5
    ENABLING AUTO-OFF ..............................................5
    RESTORING DEFAULT SETTINGS ..........................5
    DISPLAY CONTRAST & BACKLIGHTING .................6
    SELECTING FUNCTIONS & RANGES......................7
    SOURCE MODE OPERATING INSTRUCTIONS.................8
    STORING & RECALLING QUIK-CHEK
    ®OUTPUTS .8
    SOURCE MODE CONNECTION INSTRUCTIONS
    CALIBRATE MILLIAMP INPUTS ................................9
    SIMULATE 2-WIRE TRANSMITTERS .....................10
    CALIBRATE VOLTAGE INPUTS...............................11
    CHECK 1-5 VOLT INPUTS WITHOUT
    DISCONNECTING WIRES .......................................12
    CALIBRATE THERMOCOUPLE INPUTS ................14
    CALIBRATE RESISTANCE INPUTS ........................15
    CALIBRATE RTD INPUTS .......................................16
    CALIBRATE FREQUENCY INPUTS ........................18
    READ MODE OPERATING INSTRUCTIONS....................19
    MIN/MAX ..................................................................19
    OUT OF RANGE SIGNALS......................................19
    READ MODE CONNECTION INSTRUCTIONS
    READ MILLIAMP OUTPUTS ....................................20
    READ VOLTAGE OUTPUTS ....................................21
    READ AC  VOLTAGES ..............................................22
    MEASURE THERMOCOUPLE SENSORS ..............23
    READ RESISTANCE ................................................24
    CHECK CONTINUITY ..............................................25
    MEASURE RTD SENSORS .....................................26
    COUNT FREQUENCIES ..........................................28
    CALIBRATE 2 & 4 WIRE TRANSMITTERS .............30
    PRESSURE (WITH COMPANION PRESSURE MODULES)
    READ PRESSURE ...................................................32
    CALIBRATE PRESSURE TRANSMITTERS ............34
    SPECIFICATIONS...............................................................36
    WARRANTY........................................................................44 
    						
    							3 2
    water at a variety of temperatures.Each module is fully
    characterized for temperature and linearity to give you labora-
    tory accuracy in the shop, control room or field.
    Voltage Calibration
    Calibrate all your DC millivolt and voltage instrumenta-
    tion. Source from 0.00 to 110.00 mV and 0.00 to 10.25 V.
    Read up to 110.00 mV, 11.00 V and 200.0 VDC.
    Temperature Calibration
    Source and Read directly in °C and °F for T/C types J,
    K, T, E, R, S & N and four Pt 100 Ohm, Ni 120 Ohm and Cu
    10 Ohm RTDs. Resolution to 0.1°. Cold junction compen-
    sation continuously tracks ambient temperature changes.
    Frequency Calibration
    Generate zero crossing square waves from 1 to 1000 Hz,
    0.01 to 10.00 kHz and from 1 to 1000 CPM (Counts-Per-
    Minute). Built-in frequency counter measures Hz, kHz & CPM.Measure AC Voltage
    Check line voltage or mains from 0.0 to 250.0 volts AC.
    Great for troubleshooting power problems.Check Continuity
    Locate pairs of wires, open connections and shorts with
    the built-in beeper.
    Rear Label - Condensed Operation Guide
    GENERAL
    TURN-ON
    Each time you turn on the TechChek 830  it runs through a self
    check then returns to the most recently selected settings.
    SOURCE - The three QUIK-CHEK outputs will be the same as
    previously stored. Each time a different function is selected, the
    associated three QUIK-CHEK outputs will be recalled. 
    READ - The 830 is ready to measure the same signal as the last
    time it was used and automatically updates the MAX & MIN
    readings for recall at any time.
    CONNECTIONS
    TechChek 830 has protected banana jacks compatible with
    standard and safety banana plugs.
    Included with your TechChek are:a pair of safety test leads with
    test probes, safety alligator clips, standard alligator clips and
    spade lugs for attachment to a wide variety of instruments. An
    additional test lead and spade lug are also included for 3-Wire RTD
    connections.
    A second pair of test leads with right angled safety banana plugs
    and alligator clips for mA Read and Power Transmitter functions.
    Thermocouple connections are made through a miniature thermo-
    couple socket. 
    Optional KIT-1 includes T/C wires for Type J, T, E & K with miniature
    T/C plugs (not included).
    CAUTION! To prevent accidently overloading the instrument
    being tested, correctly set up the outputs before connecting
    the TechChek 830 to any instruments to be calibrated.
    !
    !
    !
    ! 
    						
    							4
    5
    GENERAL
    CONFIGURING TEMPERATURE SCALES
    The thermocouple and RTD ranges may be configured for full time
    use of °C, full time use of °F or selectable °C and °F operation.
    This configuration is part of the DEFAULT SETTINGS below.
    AUTO-OFF
    TechChek 830 can be set up to turn itself off after 30 minutes of
    inactivity. The internal timer is reset to 30 minutes each time the
    digital pot is turned or a pushbutton is pressed.
    DEFAULT SETTINGS
    TechChek 830 may be restored to the factory default setting. This
    will reset the HI and LO “QUIK-CHEK” memories according to the
    table below and the SET memory to midrange between HI and LO.
    OPERATING INSTRUCTIONS
    1) Press and hold the STORE/RESET push-button while turning
    the 830 on.
    2) Keep pressing the push-button until SETTING UP DEFAULT
    appears on the display then release the push-button.
    3) °F/°C SETUP appears on the display.
    4) Turn the digital pot to select °F & °C, °F Only or °C Only. °C &
    °F is the default if no selection is made.
    5) Press STORE/RESET push-button to store your choice or wait
    five seconds for the 830 to automatically store your choice and
    AUTO-OFF SETUP appears on the display.
    6) Turn the digital pot to select AUTO-OFF ENABLE. AUTO-OFF
    causes the 830 to turn itself off after 30 minutes of inactivity to
    preserve the batteries. Select AUTO-OFF DISABLE for contin-
    uous operation.
    7) Press STORE/RESET push-button
    to store your choice or wait five
    seconds for the 830 to automatical-
    ly store your choice. 
    8) The firmware revision will be
    displayed, then the 830 will begin
    operation with the new settings.
    OPERATING INSTRUCTIONS
    GENERAL
    FIELD & BENCH USE
    TechChek 830 comes with a carrying case
    and a built-in tilt stand/hanger. The 830 is
    held securely in the case by VELCRO® for
    use with the carrying case open. The carry-
    ing case also has a snap-on belt loop which
    can be looped around a pipe or rail.
    The tilt stand is easily raised by pulling the
    stand until it locks into place. The stand can
    also be reversed for use as a hanger to
    suspend the 830.
    CHANGING BATTERIES 
    Low battery is indicated by a battery symbol
    on the display. Approximately four hours of
    operation remain before the LCD blanks
    and TechChek 830 shuts itself down.  Turn
    the 830 off, loosen the  captive screw
    securing the battery compartment and lift
    off the  cover from the bottom of the case.
    The six “AA” batteries are easily removed
    and replaced (alkaline supplied and recom-
    mended). Replace the battery compartment
    cover by inserting the tabs and tightening
    the screw.
    QUIK-CHEK DEFAULTSRANGE LO       SET HI
    mA 4.00    12.00   20.00
    mV 1.00     5.00   10.00
    V 1.00     5.00   10.00
    T/C J,T,E,K,N  All points 0°C/32°F
    T/C R & S  All points 538°C/1000°F
    Ohms      100.0   200.0  400.0
    RTD All points 0°C/32°F
    kHz 1.00    5.00   10.00
    Hz 100 500    1000
    CPM 100 500    1000 
    						
    							GENERAL
    SELECTING FUNCTIONS
    Turn the selector knob to choose among
    mA, V, T/C, Ω, RTD, FREQ and
    PRESSURE.
    SELECTING RANGES
    Press the TYPE/ENG UNIT pushbutton to
    select the desired range and scale. 
    SOURCE RANGES
    Milliamp:
    Source: mA, %mA (% 4-20), DP% (% Differential Pressure) 
    2-Wire Simulator: mA, %mA (% 4-20), DP% (% Differential Pressure)
    VDC: mV, V
    T/C: Types J, T, E, K, N, R & S in °C &°F
    Ohms: Ohms
    RTD: Four Pt 100Ω, Ni 120Ω& Cu 10Ωin °C &°F
    Frequency: KHz, Hz, CPM
    READ RANGESMilliamp: mA, %mA (% 4-20), DP% (% Differential Pressure) 
    VDC: mV, 10V, 200V
    T/C: Types J, T, E, K, N, R & S in °C &°F
    Ohms: Ohms
    RTD: Four Pt 100Ω, Ni 120Ω& Cu 10Ωin °C &°F
    Frequency: KHz, Hz, CPM
    AC Volts: VAC
    TRANSMITTER RANGESPwr Xmtr: Supplies nominal 24 VDC to power the transmitter while 
    measuring the transmitter milliamp signal . Can be simul-
    taneously  displayed with any Source Range or Pressure 
    Measurement
    Read mA: Measures transmitter output
    7
    6
    OPERATING INSTRUCTIONS
    GENERAL
    DISPLAY CONTRAST
    The contrast of the Liquid Crystal Display may adjusted for best
    readability.
    1) Press and hold the DISPLAY/SOURCE/READ push-button
    while turning the 830 on to adjust the contrast of the display.
    2) Turn the digital pot (knob) until the display is most legible.
    3) Press the STORE/RESET push-button or wait five seconds for
    your selection to be stored.
    DISPLAY BACKLIGHTING
    The LCD may have too little contrast where the lighting is dim or
    the 830 is in shadow. Turn on the LCD backlighting to make the
    display easy to read.
    1) If the TechChek 830 is off, turn it on and wait for the display to
    come up in normal mode.
    2) Press and hold the DISPLAY/SOURCE/READ push-button for
    three seconds and the backlight will turn on.
    To extend the life of the batteries, it is recommended that the
    backlighting be turned off when the LCD is in normal light. Press
    and hold the DISPLAY/SOURCE/READ push-button for  three
    seconds and the backlight will turn off.
    TURN OFF
    Press the POWER push-button to turn the
    830 off. If AUTO-OFF is enabled, the 830 will
    turn itself off after 30 minutes of inactivity.
    TYPE
    ENG UNITS
    POWER
    ΩRTD T/CVmA
    FREQPRESSUREOPERATING INSTRUCTIONS 
    						
    							9
    CALIBRATE MILLIAMP INPUTS
    mA, mA % (Percent of 4 to 20 mA), DP% (DP Flow)
    Choose this function to provide an output from 0.00 to 24.00
    milliamps. The compliance voltage is a nominal 24 VDC to provide
    the driving power to your milliamp receivers.
    1) Disconnect one or both input wires from the device to be
    calibrated.
    2) Turn the Selector Knob to mA
    3) Press the DISPLAY/SOURCE/READ push-button until SOURCE
    or 2-WIRE appear on the display
    4) If 2-WIRE is on the display, press the TYPE/ENG UNITS push-
    button once to indicate SOURCE on the display
    5) Press the mA/%/% DP FLOW push-button to display mA, % 4-20
    or % DP Flow.
    6) Connect the red SOURCE lead of the calibrator to the plus (+)
    input of the device and the black SOURCE lead to the minus (-).
    Output current is continuously adjustable with the QUIK-CHEK
    switch in the SET position. Zero & Span (or any other values) are
    available by using the LO and HI QUIK-CHEKs.OPERATING INSTRUCTIONS
    Milliamp Receiver Input
    Controller
    Transmitter
    Computer
    Logger
    I/P
    DCS
    REDBLACK
    SOURCE MODE
    Select source by pressing the DISPLAY/SOURCE/READ
    pushbutton until  the word SOURCE appears on the LCD display.
    To change the output value, turn the speed sensitive digital pot.
    Turning the knob slowly will cause a gradual change in the output.
    A faster rate of change will occur when the knob is turned faster.
    This function operates in all three output positions (HI, SET & LO). 
    STORING QUIK-CHEK OUTPUTS
    1) Switch to HI or LO 
    2) Turn the knob to desired value
    3) Press the STORE push-button
    The LCD will flash once to show that the
    value was saved
    If a value is in the SET position and you
    want that value stored in HI or LO, press
    and hold the STORE push-button while
    moving the switch to HI or LO. The display
    will flash once to indicate the value has
    been stored. Then release the STORE
    push-button.
    RECALLING QUIK-CHEK OUTPUTS
    When you need a stored value just flip the
    QUIK-CHEK switch. Any value for the
    selected range may be stored in HI & LO.
    The TechChek 830 remembers the HI, LO
    and SET values for each function with the
    power on or off. Each time a different
    function is selected, the last three QUIK-
    CHEK values for that function will be
    recalled.
    OPERATING INSTRUCTIONS
    SOURCEHI
    SET
    LOSOURCEHI
    SET
    LO
    8
    STORE
    RESET 
    						
    							11
    CALIBRATE VOLTAGE INPUTS
    V,  m V
    Choose this function to provide an output from 0.00 mV to 110.00
    mV and from 0.00 to 10.25 VDC. Current compliance up to 20 mA
    to provide the driving power to your voltage receivers.
    1) Disconnect one or both input wires from the device to be
    calibrated
    2) Turn the Selector Knob to V
    3) Press the DISPLAY/SOURCE/READ push-button until SOURCE
    and V or mV appear on the display
    4) Press the TYPE/ENG UNITS push-button once to switch
    between V and mV on the display
    5) Connect the red SOURCE lead of the calibrator to the plus (+)
    input of the device and the black SOURCE lead to the minus (-).
    Output voltage is continuously adjustable with the QUIK-CHEK
    switch in the SET position. Zero & Span (or any other values) are
    available by using the LO and HI QUIK-CHEKs.OPERATING INSTRUCTIONS
    REDBLACK
    Volta
    ge Receiver Input
    Controller
    Transmitter
    Computer
    Logger
    DCS
    10
    OPERATING INSTRUCTIONSSIMULATE 2-WIRE TRANSMITTERS
    2-WIRE mA, 2-WIRE % (Percent of 4 to 20 mA) 2-WIRE DP %
    Choose this function to simulate a 2-Wire Transmitter output from
    1.00 to 24.00 milliamps. Operates in loops with power supply
    voltages from 3 to 45 VDC.
    1) Disconnect existing 2-Wire Transmitter from the loop
    2) Turn the Selector Knob to mA
    3) Press the DISPLAY/SOURCE/READ push-button until SOURCE
    or 2-WIRE appear on the display
    4) If SOURCE is on the display, press the TYPE/ENG UNITS push-
    button once to indicate 2-WIRE on the display
    5) Press the mA/%/% DP FLOW push-button to display mA, % 4-20
    or % DP Flow.
    6) Connect the red SOURCE lead of the calibrator to the plus (+)
    input of the device and the black SOURCE lead to the minus (-).
    The simulated output of the 2-Wire Transmitter is continuously
    adjustable from 1.00 to 24.00 mA with the QUIK-CHEK switch in
    the SET position. Zero & Span (or any other values) are available
    by using the LO and HI QUIK-CHEKs.+IN-
    REF +OUT-
    Power Supply
    2 to 45 VDCReceiver
    (Powers external
    2-Wire Transmitter)
    To
    Sensor
    Typical
    2-Wire
    Transmitter
    (Disconnected)
    REDBLACK 
    						
    							13 12
    CHECK 1-5 VOLT INPUTS
    WITHOUT DISCONNECTING WIRES
    Most 1-5 Volt receivers in 4-20mA loops have a 250 Ohm resis-
    tor across the input of the receiver. This resistor may be mounted
    internally or externally. TechChek 830 is connected directly across
    the input of the 1-5 Volt receiver without disconnecting any field
    wiring. This saves a great deal of time when a large number of
    voltage receivers, such as chart recorders or computer systems,
    require calibration.
    Make certain that changing the signal input will not disturb the
    process or cause unexpected alarms when checking on-line
    instruments. It is important to remember the 830 drives only the
    device to which it is connected.It has no effect on other devices
    in the 4 to 20 mA loop. TechChek 830 will clamp the selected
    value in the mV and V Ranges to the maximum source or sink
    current of >16 mA.
    1) Turn the Selector Knob to V
    2) Press the DISPLAY/SOURCE/READ push-button until SOURCE
    and V or mV appear on the display
    3) Press the TYPE/ENG UNITS push-button once to switch
    between V and mV on the display
    4) Connect the red SOURCE lead of the calibrator to the plus (+)
    input of the device and the black SOURCE lead to the minus (-).
    Any associated 250 Ohm resistor must not be disconnected.OPERATING INSTRUCTIONS
    REDBLACK
    CONTROLLER
    (TYPICAL)
    ADDITIONAL
    4-20 mA OR 1-5V
    INSTRUMENTS RECORDER
    COMPUTER
    LOGGER
    ETC,
    1-5V DC 250Ω
    Typical
    2-Wire
    Transmitter
    +IN-
    REF +OUT-
    CHECK 1-5 VOLT INPUTS
    WITHOUT DISCONNECTING WIRESOPERATING INSTRUCTIONS 
    						
    							15 14
    CALIBRATE RESISTANCE INPUTS
    Choose this function to simulate a resistance into a variety of
    instruments. 
    1) Disconnect one or both input wires from the device to be
    calibrated
    2) Turn the Selector Knob to Ω
    3) Press the DISPLAY/SOURCE/READ push-button until SOURCE
    appears on the display
    4) Connect the red SOURCE lead of the calibrator to the plus (+)
    input of the device and the black SOURCE lead to the minus (-). OPERATING INSTRUCTIONS
    REDBLACK
    Receiver InputOhms or 2-Wire R
    TD
    Controller
    Transmitter
    Computer
    CALIBRATE THERMOCOUPLE INPUTS
    Choose this function to simulate a thermocouple signal into any
    instrument requiring a thermocouple input. The output of the 830
    is automatically cold junction compensated.
    1) Disconnect the thermocouple from the instrument being
    calibrated.
    2) Turn the Selector Knob to T/C
    3) Press the DISPLAY/SOURCE/READ push-button until SOURCE
    and any T/C TYPE appear on the display
    4) Use the proper thermocouple wire and corresponding miniature
    T/C connector to connect TechChek 830 to the thermocouple.
    Output temperature is continuously adjustable with the QUIK-
    CHEK switch in the SET position. Zero & Span (or any other
    values) are available by using the LO and HI QUIK-CHEKs
    To Change the Thermocouple type
    1) Press the TYPE/ENG UNITS push-button. The words SELECT
    T/C TYPE and the current selection appear on the display.
    until the desired T/C type and temperature scale appear.
    2) Turn the digital pot (knob) until the required type appears on
    the display.
    3) Press the STORE/RESET push-button or wait 5 seconds to
    store the selectionOPERATING INSTRUCTIONS+-
    Instrument withThermocouple Input
    Controller
    Temperature Indicator
    Temperature Trip or Alarm
    Temperature Transmitter 
    						
    							17 16
    OPERATING INSTRUCTIONS
    REDBLACK
    BLACK
    Instrument withRTD Input
    Controller
    Temperature Transmitter
    Temperature Indicator
    Temperature Trip or Alarm
    REDBLACK
    Receiver InputOhms or 2-Wire R
    TD
    Controller
    Transmitter
    Computer
    Three Wire RTD ConnectionTwo Wire RTD ConnectionCALIBRATE RTD INPUTSFour Wire RTD Connection
    SENSOR}
    + COMP
    - COMP
    Jumper
    Jumper
    REDBLACK
    CALIBRATE RTD INPUTS
    Choose this function to simulate a temperature signal into any
    instrument requiring an RTD input. 
    1) Disconnect the RTD from the instrument being calibrated.
    2) Turn the Selector Knob to RTD
    3) Press the DISPLAY/SOURCE/READ push-button until SOURCE
    and any RTD TYPE appear on the display
    4) Connect using 2 or 3 wires as in the diagrams on the opposite
    page. Spade lugs are recommended to minimize any contact
    resistance.
    Output temperature is continuously adjustable with the QUIK-
    CHEK switch in the SET position. Zero & Span (or any other
    values) are available by using the LO and HI QUIK-CHEKs
    To Change the RTD type
    1) Press the TYPE/ENG UNITS push-button. The words SELECT
    RTD TYPE and the current selection appear on the display.
    2) Turn the digital pot (knob) until the required type appears on
    the display.
    3) Press the STORE/RESET push-button or wait 5 seconds to
    store the selection 
    Note on 4-Wire Connections:
    The TechChek 830 may be used to calibrate instruments
    requiring a 4-Wire RTD connection. This connection is valid
    only when the lead wires included with the 830 are used to
    connect the 830 directly to the instrument. Use of longer wires
    can cause errors in the output setting.
    For long 4-Wire cable runs use the 3-Wire connection and add
    a stacking banana jack (not available from Altek) to connect a
    second wire to jack #1. Connect this extra wire to the fourth
    wire field connection.
    OPERATING INSTRUCTIONS 
    						