Home > Altek > Calibrator > Altek Model 942 Frequency Calibrator with Totalizer Manual

Altek Model 942 Frequency Calibrator with Totalizer Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Altek Model 942 Frequency Calibrator with Totalizer Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 Altek manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							Frequency Calibrator
    with Totalizer
    Model 942
    ¥ 0.001% accuracyLocked to high stability crystal
    ¥ Six real world ranges1 to 20000 counts-per-hour
    0.1 to 2000.0 counts-per-minute
    0.01 to 999.99 Hz
    0.1 to 9999.9 Hz
    0.001 to 99.999 kHz
    0.01 to 250.00 kHz
    ¥ Calibrate totalizers and batch systemsOutput an adjustable number of pulses
    Incremental counter totalizes
    ¥ ÒQuik-Chek¨
    Ó switchStore 18 values for instant recall
    Three memories for each range
    ¥ Three output waveformsSine waves, zero crossing
    Square waves, zero based 
    Square waves, zero crossing 
    ¥ Simulate and measure dry contactsOptional dry contact module for relay contacts Frequency calibrator
    Calibrate your turbine meters, frequency counters,
    vibration systems, tachometers, vortex shedders, inte-
    grators and any other frequency devices in the shop,
    plant or field. Measure from 0.01 Hz to 250.00 KHz
    with resolution down to 0.01Hz. For your slower
    signals, source or measure from 0.1 to 2000.0 counts
    per minute (CPM) or from 10 to 20000 counts per
    hour (CPH). Quickly indicate process signals, no more
    waiting around with stopwatch and calculator.
    Read a wide range of frequencies
    Use in the field to check the signals from your
    flowmeter pickups, velocity and motion detectors, Fre-
    quencies can be directly measured from 0.01 Hz to
    250.00 KHz in four frequency ranges.
    Measure extremely slow frequencies (
    						
    							General instructions
    Turn-onEach time the Model 942 is turned on, the LCD will
    display all segments for about 1 second. It then
    displays the currently selected waveform (Source
    Mode) for approximately 3 seconds. 
    1) Move the power switch to SOURCE or READ. Allsegments on the LCD are turned on during self
    test.
    2) (SOURCE MODE ONLY) The display will indicate the selected waveform for 3 seconds. Repeatedly
    press or press and hold the SCROLL/STORE
    pushbutton to change to the desired waveform.
    3) (SOURCE MODE ONLY) The three QUIK-CHEK frequency outputs will be the same as previously
    stored. Each time a different range is selected,
    the three QUIK-CHEK outputs for that range will
    be recalled. 
    4) Move the mode switch to RANGE and repeatedly press or press and hold the SCROLL pushbutton
    to change to the desired frequency range.
    5) Return the mode switch to FREQ/TRIG to source or read frequency signals.
    ConnectionsThe Model 942 has built-in test leads with alligator
    clips for attachment to instruments or sensors with
    terminal blocks or flying leads. An optional BNC
    connector attaches to instruments or sensors
    equipped with BNCs for fast connections.
    Troubleshooting 
    Changing the batteriesLow battery is indicated by BAT on the LCD Display.
    Approximately 10 Hours of operation remain before
    the LCD goes blank and the Model 942 shuts itself
    down.  Turn the 942 off, loosen the three captive
    screws securing the battery compartment cover. The
    six ÒAAÓ batteries are easily removed and replaced.
    Replace the battery compartment cover, tighten the
    screws and turn on when ready to use.
    ResetThe Model 942 may be reset from the front panel to
    factory default settings. This will reset all the ÒQUIK-
    CHEKÓ memories to display 1000 and will set the
    output and trigger levels to 1 V p-p.
    1) Press and hold the SCROLL/RESET pushbuttonwhile turning the Model 942 on to SOURCE or
    READ
    2) Keep pressing the STORE pushbutton for 10 seconds
    3)  All segments on the LCD will remain displayed until the Model 942 has been reset
    Field and bench useThe Model 942 comes with a carrying case and a
    built-in tilt stand/hanger. The 942 is held securely in
    the case by Velcro¨ even with the carrying case
    open. The carrying case also has a snap-on belt
    loop which can also be looped around a pipe or rail.
    The tilt stand is easily raised by pulling the stand until
    it locks into place. The stand can also be reversed for
    use as a hanger to suspend the Model 942.
               x10
    12
    10.0
    6.0
    3.0
    1.0
    0.6
    0.3
    0.1
    .06
    .03
    QUIK-CHEKªSTART/
    STOPRESETS CROL
    L
    ST ORE
    MAX
    READ
    MIN
    HI
    SETLOX1/X10
    TRIG
    RANGE
    LEVEL
    FREQRANGESET
    TIME
    SET
    COUNT
    TOTALSET
    TIME
    TIMELEFT
    TOTALTOTALIZE FUNCTIONS
    10.0
    6.0
    3.0
    1.0
    0.6
    0.3
    0.1
    .0610942
    FREQUENCY CALIBRATORSOURCEREAD
    STA RT/
    STOP
    RESET
    S CROL
    L
    S
    T ORE
    Reading frequencyIn order for the Model 942 to obtain the most accurate readings
    you must correctly set the ATTENUATOR, TRIGGER LEVEL and
    RANGE. Signals from 50 mV to 40 volts p-p, with or without
    DC offsets can be displayed.
    Symptom    Check     Solution
    GATE on LCD,
    Display shows
    0.0
    OVER/UNDER
    Range on LCD
    Unstable readingConnections
    Attenuator
    Input level
    DC Offset
    Range
    Trigger LevelMake sure all power and signals
    are properly connected.
    Set at x1 for signals from 50mV to
    12 Volts p-p,  x10 for signals over
    12 V p-p.
    Turn knob until GATE pulses and
    readings are displayed.
    Small signals with large DC offsets
    may require a series capacitor.
    Move the mode switch to RANGE
    and press the SCROLL/RESET
    pushbutton until the correct range
    appears in the LCD.
    Turn knob until GATE pulses and
    readings are displayed.
    Sourcing frequencySome receivers can only detect signals that go from positive to
    negative (Sine Wave or Zero Crossing Square Waves) while other
    receivers require only positive signals (Zero Based Square
    Waves). The Model 942 provides a choice of these outputs.
    Symptom    Check     Solution
    Lack of
    Response
    Wrong Range
    Lack of
    response or
    jittery signalConnections
    Waveform
    Range
    Peak VoltageMake sure all power and signals
    are properly connected.
    Turn Model 942 OFF and back on
    to SOURCE. Repeatedly press the
    SCROLL/STORE pushbutton until
    the correct waveform is displayed.
    Move the mode switch to RANGE
    and press the SCROLL/RESET
    pushbutton until the correct range
    appears in the LCD.
    Move the mode switch to LEVEL
    and turn the knob while observing
    the logarithmic bar graph to match
    the input level of the device being
    calibrated. Return the mode switch
    to FREQ.
    To Convert From:   To:          Divide By:
    CPM      Hz 60
    CPH      Hz 3600
    To Convert From:   To:         Multiply By: Hz     CPM 60
    Hz     CPH 3600
    CPM/CPH frequency conversions  
    						
    							Frequency output
    Source1) Move the POWER switch to SOURCE
    2) Move the mode switch to RANGE and repeated-ly press or press and hold the SCROLL/STORE
    pushbutton to change to the desired frequency
    range. Return the mode switch to FREQ.
    3) Move the mode switch to LEVEL (AMPLITUDE) and turn the Digipot (Knob) until the logrithmic
    bargraph on the display reaches the desired
    level.Return the mode switch to FREQ.
    4) Connect the Model 942 to the input terminals of the instrument or meter to be calibrated
    5) Adjust the digital pot to the desired output value or Quik-Chek with previously stored frequency
    outputs (see below)
    Whenever SOURCE mode is selected the word
    SOURCE will appear on the LCD display. To change
    the output value, turn the speed sensitive digital pot.
    Turning the pot slowly will cause a gradual change
    in the output. A faster change will occur when the
    pot is turned faster. This function operates in all
    three output positions (HI, SET & LO). Store1) Switch to HI or LO 
    2) Turn the digital pot to desired value
    3) Press the STORE/SCROLL pushbutton The LCD will flash once to show that the value
    was saved
    If a value is in the SET position and you want that
    value stored in HI or LO, press and hold the
    STORE/SCROLL pushbutton while moving the
    switch to HI or LO. The display will flash once to
    indicate the value has been stored. Then release
    the STORE/SCROLL button.
    Quik-ChekAny time you need a stored value just throw the
    Quik-Chek switch. Any value in the frequency range
    may be stored in HI & LO. The Model 942
    remembers the HI, LO and SET values for all
    ranges (18 memories) for you with the power on or
    off. Each time a different frequency range is select-
    ed, the last three Quik-Chek values for that type will
    be recalled.
    Read frequency
    Read1) Move the POWER switch to READ
    2) Move the mode switch to RANGE and repeatedly press or press and hold the SCROLL/RESET
    pushbutton to change to the desired frequency
    range. Return the mode switch to TRIG.
    3) Switch the MODE switch to LEVEL (AMPLITUDE) to toggle between x1 & x10
    attenuation (Use x1 for signals from 30 mV to 12
    V p-p, x10 for signals from 12 V to 40 V p-p).
    Return the mode switch to TRIG.
    4) Connect the Model 942 to the output of the in- strument or sensor to be measured. 
    5) Adjust the trigger level to obtain a stable frequency reading by turning the Digipot (knob).
    A bargraph on the display will show the approxi-
    mate trigger level. 
    6) Use the ÒQuik-ChekÓ switch to display present reading, maximum or minimum frequency.
    The word GATE will appear on the display whenev-
    er the Model 942 is measuring the frequency signal
    and will flash each time the displayed reading is
    updated.
    Period readingsSelect Counts-per-Minute (CPM) or Counts-per-
    Hour (CPH) to measure slow frequency signals.
    Frequencies as low as 0.1 CPM (0.001666 Hz) and
    10 CPH (0.002777 Hz) can be measured (See
    CPM/CPH frequency conversions for conversion
    factors).
    Min/maxTo read the Maximum or Minimum frequencies since
    READ mode was entered, simply switch to MAX or
    MIN. The value will appear on the LCD along with
    the word MAX or MIN. The MAX/MIN values are au-
    tomatically updated and may be viewed at any time
    without disturbing the other values. Pressing the
    RESET/SCROLL pushbutton will cause the 942 to
    stop counting frequencies and will display zeros.
    Upon releasing the RESET/SCROLL the Model 942
    will display GATE, resume counting frequencies and
    update the MAX & MIN values as the measured fre-
    quency changes.
    Trigger levelThe adjustable TRIGGER LEVEL is used in mea-
    surements of noisy signals, AC signals superim-
    posed on DC levels and to select Voltage threshold
    for all other signals. The bargraph on the display
    shows the approximate level from 0 to over 12 V
    positive peak with the attenuator set at x1. This
    bargraph should be read as 0 to over 120 V positive
    peak with the attenuator set at x10. For quickest
    readings, determine or estimate the voltage level to
    be detected and set the ATTENUATOR and TRIG-
    GER LEVEL to match.
    Out of range signalsFrequencies above or below those available for the
    currently selected range will be indicated by OVER
    and UNDER on the display (See Troubleshooting).
    SOURCE
    STA RT/
    STOP
    RESET
    S CROL
    L
    S
    T ORE
               x10
    12
    10.0
    6.0
    3.0
    1.0
    0.6
    0.3
    0.1
    .06
    .03
    CPM
    OVER
    Reading dry contacts
    Isolated dry contact, open collector transistor or
    opto-isolated frequencies may be measured with the
    Model 942 with the optional Dry contact module
    (Model C-41) or with the circuit shown. In order to
    detect contact opening or closing, an external bat-
    tery or power supply, in series with a 5000 Ohm
    resistor, may be used. Select connection polarity to
    provide desired signal upon contact transfer. Relay
    or switch contacts may require a resistor-capacitor
    filter in order to eliminate contact bounce errors.
    Typical filter values for mechanical contacts are 10
    K Ohms and 0.1 microfarads. 
    						
    							Calibrate totalizers (output pulses)1) Move the POWER switch to SOURCE
    2) Move the mode switch to RANGE and repeated-ly press or press and hold the SCROLL
    pushbutton until the word TOTAL appears on the
    LCD. The ouput waveform will automatically
    change to a zero based square wave. Return the
    mode switch to FREQ.
    3) Move the mode switch to LEVEL (AMPLITUDE) and turn the Digipot (Knob) until the logrithmic
    bargraph on the display reaches the desired
    level. Return the mode switch to FREQ.
    4) Move the TOTALIZE FUNCTIONS switch to SET TIME. Adjust the digital pot to the number of
    minutes (1 to 99) that you require the 942 to
    output pulses.
    5) Move the TOTALIZE FUNCTIONS switch to SET COUNT. Adjust the digital pot to the number of
    pulses (1 to 99999) that you require the 942 to
    output. Return the TOTALIZE FUNCTIONS
    switch to TOTAL.
    6) Connect the Model 942 to the input terminals of the instrument or meter to be calibrated.
    7) Reset to zero or record the value from the readout of the totalizer or device being
    calibrated.
    8) Press the START/STOP pushbutton to start out- putting pulses. The display of the 942 will
    increment with each output pulse. The Model
    942 will automatically stop outputting pulses
    when the selected time has elapsed.
    Whenever CALIBRATE TOTALIZER mode is
    selected the words SOURCE and TOTAL appear on
    the LCD display. You can recall the settings for
    TIME and COUNTS without interrupting the output
    pulses by moving the TOTALIZE FUNCTIONS
    switch to SET TIME and SET COUNTS.
    Restarting the totalizerThe Model 942 will automatically stop outputting
    pulses when the selected time has elapsed.To
    restart CALIBRATING TOTALIZERS press the
    START/STOP bushbutton while the Quik-Chek
    switch is in the TOTAL position. The display on the
    Model 942 will reset to 0 and will automatically
    begin outputting pulses.
    Changing totalizer settingsThe settings for TIME and COUNTS can be
    changed whenever the Model 942 isnt outputting
    pulses. Move the TOTALIZE FUNCTIONS switch to
    SET TIME and SET COUNTS and turn the digital
    pot to set the new values as in steps 4 and 5.
    Interrupting the totalizerPress the START/STOP pushbutton while the Quik-
    Chek switch is in the TOTAL position to interrupt the
    output pulses at any time. The Model 942 will stop
    sending the pulses and will display the number of
    pulses which have been put out. This can be done
    to verify the device being calibrated without waiting
    the entire selected number of minutes.
    Calibrate totalizers
    Totalize (count pulses1) Move the POWER switch to READ
    2) Move the mode switch to RANGE and repeated-ly press or press and hold the SCROLL
    pushbutton until the word TOTAL appears on the
    LCD. Return the mode switch to FREQ.
    3) Switch the MODE switch to LEVEL (AMPLI- TUDE) to toggle between x1 & x10 attenuation
    (Use x1 for signals from 30 mV to 12 Vp-p, x10
    for signals from 12 V to 40 Vp-p). Return the
    mode switch to TRIG.
    4) Adjust the trigger level by turning the Digital pot (knob). A bargraph on the display will show the
    approximate trigger level. 
    5) Move the TOTALIZE FUNCTIONS switch to SET TIME. Adjust the digital pot to the number of
    minutes (1 to 99) that you require the 942 to
    count pulses. Return the TOTALIZE
    FUNCTIONS switch to TOTAL.
    6) Connect the Model 942 to the output of the in- strument or sensor to be measured. 
    7) Press the START/STOP pushbutton to begin counting pulses.
    8) If the Model 942 doesnt increment the LCD display adjust the trigger level by turning the
    Digital pot until the display starts incrementing. A
    bargraph on the display will show the approxi-
    mate trigger level. Press the Start/Stop
    pushbutton twice to restart totalizing.
    Whenever TOTALIZER mode is selected the words
    READ and TOTAL appear on the LCD display. You
    can recall the settings for the SET TIME without
    interrupting the output pulses by moving the
    TOTALIZE FUNCTIONS switch to SET TIME. You
    can also see how much time is remaining by moving
    the TOTALIZE FUNCTIONS switch to TIME LEFT.
    Restarting totalizeThe Model 942 will automatically stop counting
    pulses when the selected time has elapsed.To restart
    TOTALIZE press the START/STOP bushbutton while
    the Quik-Chek switch is in the TOTAL position. The
    display on the Model 942 will reset to 0 and will
    automatically begin counting pulses.
    Changing totalize settingsThe setting for TIMEcan be changed whenever the
    Model 942 isntcounting pulses. Move the
    TOTALIZE FUNCTIONS switch to SET TIME and
    turn the digital pot to set the new value as in step 5.
    Interrupting totalizePress the START/STOP pushbutton while the Quik-
    Chek switch is in the TOTAL position to interrupt the
    pulse counting at any time. The Model 942 will stop
    counting the pulses and will display the number of
    pulses which have been counted. This can be done
    to verify the device being calibrated without waiting
    the entire selected number of minutes. To restart
    TOTALIZE press the START/STOP bushbutton
    while the QUIK-CHEK switch is in the TOTAL
    Totalize
    Totalizer connections
    Calibrate
    totalizersTotalize
    >
    Flowmeter
    RESET
    Totalizer
    START/
    STOP
    RESET
    S CROL
    L
    S
    T ORE
    STA RT/
    STOP
    RESET
    S CROL
    L
    S
    T ORE
    STA RT/
    STOP
    RESET
    S CROL
    L
    S
    T ORE
    STA RT/
    STOP
    RESET
    S CROL
    L
    S
    T ORE 
    						
    							PN 1887840 Rev B November 2002
    ©2002 Altek Industries, Inc. Specifications subject to change without notice. \
    All rights reserved. Printed in U.S.A.  
    Altek products are warranted to be free from defects
    in material and workmanship (excluding fuses,
    batteries and leads) for a period of three years from
    the date of shipment. Warranty repairs can be
    obtained by returning the equipment prepaid to our
    factory. Products will be replaced, repaired, or
    adjusted at our option. 
    Altek gives no other
    warranties, including any implied warranty of fitness for a particular purpose.
    Also, Altek shall not be liable
    for any special, indirect, incidental or consequential
    damages or losses arising from the sale or use of its
    products.
    Warranty
    (Unless otherwise indicated, specifications are in ±% reading @ 23¡\
     C)
    GeneralFrequency stability: 80 hours, nominal
    Source: >50 hours, nominal at 250 KHz at 12V p-p
    Low battery: ÒBATÓ indication on the display at 6.5 V nominal,
    approximate 10 hours left
    Attenuator: Logarithmic for smooth input/output signal level control
    Operating temperature range: -5 to +140¡F (-20 to +60¡C)
    Storage temperature range: -13 to +149¡F  (-25 to +65¡C)
    Relative humidity: 10 to 90%, non-condensing, for 24 hours
    Warmup time: 5 seconds to rated accuracy
    Overall size: 73/16 x 4 x 27/16 inches (183 x 102 x 62 mm)
    Weight: 1lb, 7oz (0.650 kg)
    942-BNC: A male BNC is added to the top of the calibrator in
    addition to the attached leads.
    Specifications
    ReadAccuracy: ±(0.001% of reading + 1 LSD)
    Sensitivity: Triggers on positive peaks down to 40 mV peak, DC
    coupled
    Maximum usable input voltage:  1 Meg Ohm + 60pF
    Trigger level adjustment: x1 & x10 Attenuator plus logarithmic
    control
    NOTE: High signal noise and low slew rate (Volts-per-second) may
    affect reading uncertainty (may not read sine waves below 10 Hz)
    SourceAccuracy: ±0.001% of reading
    Waveforms: Sine, zero based square & zero centered square waves
    Totalizer output: from 1 to 99999 pulses for a period from 1 to 99
    minutes up to a maximum rate of 1666.65 Hz
    Output amplitude: 50 mV to 12 V p-p, 50% ±1% duty cycle
    Risetime: 
    						
    All Altek manuals Comments (0)

    Related Manuals for Altek Model 942 Frequency Calibrator with Totalizer Manual