Page 41
© Adam Equipment Company 2010 39 15.0 SAFETY AND MAINTENANCE CAUTION Use the AC adapter designed by the manufacturer for the balance. Other adapters may cause damage to the balance. Avoid overloading or dropping material onto the pla tform which could damage the balance. Do not spill liquids on the balance as it is not wa ter-resistant. Liquids may damage the case and if it gets inside the balance i t may cause damage to the...
Page 42
© Adam Equipment Company 2010 40 These are the rarest of the problems affecting bala nces. If an electronic problem is suspected make sure the mechanical probl ems that can cause similar symptoms have been eliminated before attemp ting electronic repairs. With the exception of cables most electronic repair s are solved by board replacement. The table that follows is a guide of common problem s and their solutions....
Page 43
© Adam Equipment Company 2010 41 Balance is very unstable and does not weigh correctly Mechanical problems Balance programming Electronic problems *A complete inspection of the mechanics to look for sources of friction. *Verify the A/D is also unstable. If the A/D is OK then suspect the programming of the balance. Reset parameters, check linearity and redo the calibration. Some electronic problems can also cause this. But all mechanical problems...
Page 44
© Adam Equipment Company 2010 42 Poor Linearity Usually a mechanical problem Electronic Problems Re-check repeatability *Inspection of the flexures for damage or loose hardware may be required *Use the Linearity Function in the service menu to reset linearity *A problem in the analogue circuit board or power supplies can cause poor linearity. Make sure all mechanical problems have been eliminated first OTHER PROBLEMS: Cannot calibrate Zero...
Page 45
© Adam Equipment Company 2010 43 17.0 REPLACEMENT PARTS AND ACCESSORIES If you need to order any spare parts and accessorie s, contact your supplier or Adam Equipment. A partial list of such items is as follows- Power Supply Module Stainless Steel top Pan Below Balance Hanger Density Determination Kit Anti-Vibration Table Security Lock and Cable Dust Cover Printers, etc. 18.0 SERVICE INFORMATION This manual covers the...
Page 50
© Adam Equipment Company 2010 48 WARRANTY INFORMATION Adam Equipment offers Limited Warranty (Parts and Labor) for the components failed due to defects in materials or workmanship. Warranty starts from the date of delivery. During the warranty period, should any repairs be necessary, the customer must inform the supplier or Adam Equipment. The company or its authorised Technician reserves the right to repair or replace any components at its own discretion. Any shipping costs involved in sending...