Adam Equipment PGW Precision Balances PGW2revA User Manual
Have a look at the manual Adam Equipment PGW Precision Balances PGW2revA User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 35 Adam Equipment manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
© Adam Equipment Company 2010 39 15.0 SAFETY AND MAINTENANCE CAUTION Use the AC adapter designed by the manufacturer for the balance. Other adapters may cause damage to the balance. Avoid overloading or dropping material onto the pla tform which could damage the balance. Do not spill liquids on the balance as it is not wa ter-resistant. Liquids may damage the case and if it gets inside the balance i t may cause damage to the electronics. Material that has a static electric charge could in fluence the weighing. Discharge the static electricity of the samples, if possible. Another solution to the problem is to wipe both sides of the pan and th e top of the case with an anti-static agent. 16.0 TROUBLE-SHOOTING Service of a PGW balance will generally be necessar y when the balance does not perform as expected. The balances are not user-serviceable. For Service Information, see section 18.0 and contact A dam equipment or your supplier. Problems usually fall into one of the following cat egories: User Problems: The user is asking the balance for something it can not do or is confused by the modes and functions of a balance. It is also p ossible the user has set a parameter that has affected the balance operation. Resetting the parameter to a normal value will restore operation. Mechanical Problems The balances consist of complicated and fragile mec hanical devices. They can be damaged by placing a weight on it which is t oo high for the balance or by dropping the balance or occasionally shipping it without taking care. The most fragile parts are the flexures. Dust, dirt, spills and other foreign objects in the balance can also cause problems. Electronic Problems:
© Adam Equipment Company 2010 40 These are the rarest of the problems affecting bala nces. If an electronic problem is suspected make sure the mechanical probl ems that can cause similar symptoms have been eliminated before attemp ting electronic repairs. With the exception of cables most electronic repair s are solved by board replacement. The table that follows is a guide of common problem s and their solutions. Note that many problems may have multiple solutions and there may be problems found that are not listed in the table. F or Service Information, contact Adam Equipment or your supplier. BALANCE DOES NOT FUNCTION Problems Possible causes Suggestions The balance is dead when power is applied Power supply failure Check adapter is working Check adapter is correct for the balance Normal adapter is 15VDC, 800mA. *Power supply circuit board failure *Short circuit on any circuit board The display does not turn on but the calibration motor moves when power is applied Power is getting to balance, display is not working *Display cables may be faulty *Display module failure The display stays on the initial test screen when power is applied. Calibration weight motor is on. Unstable balance Balance not working correct Power supply *Check if balance is stable by using service menu and view A/D values Put draught shield over pan Check power supplies BALANCE WORKS BUT IS NOT STABLE Balance is unstable by a few divisions Noise or vibration from environment Friction in mechanics Check the balance is positioned correctly to avoid vibration, wind or air movement, it is on a solid table, It is not near sources of heat or cool air, Check balance with weights if problem occurs when sample is used. Static electricity on the samples can cause drifting and instability. Check the area around the weighing pan for hair, dust, obstructions under the pan, *A complete inspection of the mechanics to look for sources of friction may be needed.
© Adam Equipment Company 2010 41 Balance is very unstable and does not weigh correctly Mechanical problems Balance programming Electronic problems *A complete inspection of the mechanics to look for sources of friction. *Verify the A/D is also unstable. If the A/D is OK then suspect the programming of the balance. Reset parameters, check linearity and redo the calibration. Some electronic problems can also cause this. But all mechanical problems must be resolved first. BALANCE IS NOT ACCURATE You must have accurate and trusted weights to test a balance. If you suspect that the balance is not accurate then you must know your wei ghts are accurate. A balance calibrated using a bag of flour is not accurate eve n if it works OK otherwise. Balance is not accurate Repeatability Eccentric loading Linearity Verify the balance shows the same value when the same mass is placed on the centre of the pan for a few tests. Verify the balance shows the same reading (within a tolerance depending upon the model) when a mass is placed at positions around the pan. Verify the balance is acceptable throughout the weighing range. The balance must give acceptable readings from low weights up to the capacity. Poor Repeatability Usually a mechanical problem. Inspect the area around the pan for hair, dust or other obstructions, *Inspection of the mechanics may be needed for any possible problems. Poor Eccentric loading A mechanical problem Inspect the area around the pan for hair, dust or other obstructions, *Inspection of the mechanics may be needed for any possible problems. *Readjusting of the Eccentric loading is recommended.
© Adam Equipment Company 2010 42 Poor Linearity Usually a mechanical problem Electronic Problems Re-check repeatability *Inspection of the flexures for damage or loose hardware may be required *Use the Linearity Function in the service menu to reset linearity *A problem in the analogue circuit board or power supplies can cause poor linearity. Make sure all mechanical problems have been eliminated first OTHER PROBLEMS: Cannot calibrate Zero shifted more than allowed Calibration timeout *Check all flexures for damage *Reset dealer calibration *Verify linearity and repeatability *The balance may be unstable. Verify stability as above. Calibration weight motor does not stop *Check the cables to the motor, try plugging the balance into the power again *Look for friction in the calibration weight movement *Check the opto-coupler that controls the motor position. RS-232 not working Doesn’t print Check parameters m atch the device connected Verify cable is correct *RS-232 circuits damaged Display dark, keys beep Display contrast poor Cable unplugged or damaged *Check the cables to the display *Replace the display which could be damaged *To be carried out by authorised technicians only.
© Adam Equipment Company 2010 43 17.0 REPLACEMENT PARTS AND ACCESSORIES If you need to order any spare parts and accessorie s, contact your supplier or Adam Equipment. A partial list of such items is as follows- Power Supply Module Stainless Steel top Pan Below Balance Hanger Density Determination Kit Anti-Vibration Table Security Lock and Cable Dust Cover Printers, etc. 18.0 SERVICE INFORMATION This manual covers the details of operation. If you have a problem with the balance that is not directly addressed by this manu al then contact your supplier for assistance. In order to provide furthe r assistance, the supplier will need the following information which should be kept ready: A. Details of your company -Name of your company: -Contact person’s name: -Contact telephone, e-mail, fax or any other methods: B. Details of the unit purchased (This part of information should always be availabl e for any future correspondence. We suggest you to fill in this form as soon as the unit is received and keep a print- out in your record for ready reference.) Model name of the balance: PGW________ Serial number of the unit: Software revision number (Displayed when power is first turned on): Date of Purchase: Name of the supplier and place: C. Brief description of the problem Include any recent history of the unit. F or example: -Has it been working since it’s delivered -Has it been in contact with water -Damaged from a fire -Electrical Storms in the area -Dropped on the floor, etc.
© Adam Equipment Company 2010 45 19.0 BALANCE MENU STRUCTURE
© Adam Equipment Company 2010 48 WARRANTY INFORMATION Adam Equipment offers Limited Warranty (Parts and Labor) for the components failed due to defects in materials or workmanship. Warranty starts from the date of delivery. During the warranty period, should any repairs be necessary, the customer must inform the supplier or Adam Equipment. The company or its authorised Technician reserves the right to repair or replace any components at its own discretion. Any shipping costs involved in sending the faulty units to a service centre is the customers responsibility. The warranty will cease to operate if the equipment is not returned in the original packaging and with correct documentation for a claim to be processed. All claims are at the sole discretion of Adam Equipment. This warranty does not cover equipment where defects or poor performance is due to misuse, accidental damage, exposure to radioactive or corrosive materials, negligence, faulty installation, unauthorised modifications or attempted repair or failure to observe the requirements and recommendations as given in this User Manual. Repairs carried out under the warranty does not extend the warranty period. Components removed during the warranty repairs become the company property. The statutory right of the purchaser is not affected by this warranty. The terms of this warranty is governed by the UK law. For complete details on Warranty Information, see the terms and conditions of sale available on our web-site