Page 11
© Adam Equipment Company 2 0 1 6 11 5.0 FUNCTIONS 5.1 PERCENTAGE WEIGHING The balance will allow a reference weight to be shown as 100%. Then any other weight placed on the balance will be displayed as a percentage of the original sample. 1) Place your weight on the balance . 2) Press the [%] key . The weight will be displayed as 100.00%. 3) Remov e the weight and place your next sample on the balance . The new reading will be shown as a percentage of the first value / reference...
Page 12
© Adam Equipment Company 2 0 1 6 12 4) The balance will initially show SP 10 asking for a sample size of 10 parts. Press [Mode] to cycle through the options: 10, 2 0, 50, 100, 200 and back to 10 t o select the sample you have placed on the balance. 5) Press [Smpl] again once y ou have selected the sample. The display will show you the number. As more parts are added the display will show the total number of parts (PCS – will be shown in the alpha -numeric display). 6) By p...
Page 13
© Adam Equipment Company 2 0 1 6 13 5) Put a second weight on. I f set to automatic it will accumulate the weight automatically . If in manual mode, press [Print] , the display will show ACC 2 foll owed by the new total in the memory for 2 seconds before returning to displaying the weight of the item on the balance. 6) Continue until all weights have been added. To view the totals in memory press the [Print] key when the balance is at zero. The display will show ACC xx (where “...
Page 14
© Adam Equipment Company 2 0 1 6 14 6.0 PARAMETERS Non -approved balances ha ve 8 parameters that can be set by the user. FUNCTION SECTION DESCRIPTION F1 UNT See section 6.1 Sets the units to be used g / ct / Lb / oz / d / GN / OZt / dWt / MM / TL.T / TL.C / TL.t /t / N /g2. F2 EL See section 6.2 Sets the backlight and audible alert beep (SW ver. 2.5 onwards) F3 SEr See section 6.3 Sets the print parameters F4 oFF See section 6.4 Sets the auto power-off parameter F5 IEC See section...
Page 16
© Adam Equipment Company 2 0 1 6 16 6.2 SETTING THE BACKLIGHT AND AUDIBLE ALERT BEEP The backlight and audible alert may be enabled or disabled by the user. If the backlight is disabled, the battery life will be greater. The followin g settings are available: ( beep settings available in SW ver. 2.5 onwards) bL AU Sets the backlight to operate automatically when a weight is placed on the balance or a key is pressed. bL Off Sets the backlight to be off. bL On Sets the backlight to be...
Page 18
© Adam Equipment Company 2 0 1 6 18 Press the [Mode] key to change the settin g. Press [->0/T0/T0/T
Page 19
© Adam Equipment Company 2 0 1 6 19 2) After a few seconds, the display will show the first function F1 UNT . 3) Keep pressing the [Mode] key until F4 oFF is displayed. 4) Press the [- >0/T0/T0/T0/T
Page 20
© Adam Equipment Company 2 0 1 6 20 6.6 ADJUST THE VALUE OF THE INTERNAL MASS (NOT POSSIBLE ON APPRO VED MODELS) The internal mass value stored in memory can be adjusted to more clos ely match the value of the user’s external mass. 1) To set this parameter press the [Mode] key during self -checking test when turning on the balance . 2) After a few seconds, the display will show the first function F1 UNT . 3) Keep press ing the [Mode] key until F6 CA is displayed. 4) Press the...