    Have a look at the manual Adam Equipment Highland Portable Precision Balances HIGHLANDHCB User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 35 Adam Equipment manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							© Adam Equipment  Company 2 0 1 6   11 
    5.0  FUNCTIONS  
    5.1 PERCENTAGE WEIGHING  
    The  balance  will allow a reference weight to be shown as 100%. Then any other 
    weight placed on the balance  will be displayed as a percentage of the original 
    sample.   
    1)  Place your weight on the balance . 
    2)  Press the [%]  key . The weight will be displayed as 100.00%.  
    3)  Remov e the weight and place your next sample on the balance . The new 
    reading will be shown as a percentage of the first value / reference used.   
    4)  Pressing the  [%]  key again will return the  balance to normal weighing. 
    Note:  The balance  may jump by large numbers unexpectedly if small weights are 
    used to set the 100% level. For example, if only 23.5g is on a balance  with 0.5g 
    increments and the  balance is set to 100%, the display will show 100.00%. Ho wever, 
    a small change of weight will cause the display to jump to 102.13%, as one balance  
    division (0.5g) increase to 24.0g will be equivalent to a 2.13% increase. 
     
    5.2  PARTS
     COU NTING  
    Parts counting lets you count small parts that are of equal weight,  quickly and easily. 
    To do this you must first set your sample by telling the balance  how many parts you 
    have.   
    1)  Place your container on the balance and press the [- >0/T
    						
    							© Adam Equipment  Company 2 0 1 6   12 
    4) The  balance  will initially show  SP   10 asking for a sample size of 10 parts.  
    Press  [Mode]  to cycle through the options: 10, 2 0, 50, 100, 200 and back to 10 
    t o select the sample you have placed on the balance.   
    5)  Press  [Smpl]  again once y ou have selected the sample. The display will show 
    you the number. As more parts are  added the display will show the total 
    number of parts (PCS  –  will be shown in the alpha -numeric display).   
    6)  By p ressing the  [Mode]  key you can view the unit weight (W/P ), total weight 
    ( g ) or the count (PCS ). The total and the unit weight are shown in the current 
    weighing  units. 
    7)  Press  [Smpl]  to return to normal weighing. 
     
    5.3  ACCUMULATION
      
    There are two types of accumulations –  automatic and manual.  
    Automatic Accumulation Manual Accumulation 
    When the balance is set for automatic 
    accumulation,  see Parameters 6.3  
    Setting the Printer Parameters / 
    Accumulation, the weight will be added 
    to the memory when the balance  
    becomes stable.  
    When the balance is set to manual 
    accumulation, see Parameters 6.3 
    Setting the Printer Parameters / 
    Accumulation ,  the weight displayed will 
    only be  stored in memory once  the  
    [Print]  
    key is pressed and the weight is 
    stable.   
    1) Place the weight on the balance . If automatic then the balance  will 
    automatica lly accumulate the weight . If set to manual then you will need to 
    press the [Print]  key.  
    2)  The display will show ACC  1  followed by the total value in the memory for 2 
    seconds before returning to displaying the weight of the item on the  balance.  
    3)  The weight will be transmitted  to a printer or PC. 
    4)  Remove the weight, allowing the balance  to return to zero.  
    COPY  
    						
    							© Adam Equipment  Company 2 0 1 6   13 
    5) Put a second weight on. I f set to automatic it will accumulate the weight 
    automatically . If in manual  mode,  press [Print] , the display will show ACC   2  
    foll owed by the new total in the memory for 2 seconds before returning to 
    displaying the weight of the item on the  balance. 
    6)  Continue until all weights have been added.  
    To view the totals in memory press the  [Print] key when the balance  is at zero.  The 
    display  will show ACC  xx  (where “ xx” is the total number of readings) and the total 
    weight, before returning to zero. At the same time the total will also be sent via the 
    RS -232 interface.  
    To clear the memory, press [Smpl ] once the total accumulation value is di splayed 
    after  pressing  [Print] .
     
    COPY  
    						
    							© Adam Equipment  Company 2 0 1 6   14 
    6.0  PARAMETERS  
    Non -approved  balances ha ve 8  parameters that can be set  by the user. 
    FUNCTION SECTION DESCRIPTION 
    F1 UNT See section 6.1 Sets the units to be used 
    g / ct / Lb / oz  / d / GN / OZt / dWt / MM / TL.T / TL.C / TL.t /t 
    / N /g2.  
    F2 EL See section 6.2 Sets the backlight and audible alert beep (SW ver. 2.5 onwards) 
    F3 SEr See section 6.3 Sets the print parameters  
    F4 oFF See section 6.4 Sets the auto power-off parameter  
    F5 IEC See section 6.5 Internal or external calibration select  
    F6 CA See section 6.6 Cal mass fine adjust  
    F7 SEt See section 6.7 User Parameters for autozero, filter and stability 
    tECH Service agents only Technical parameters setting mode / factory setting 
     
    Approved balances have 4  paramete rs that can be set by the user. 
    FUNCTION SECTION DESCRIPTION 
    F1 EL See section 6.2 Sets the backlight and audible alert beep (SW ver. 2.5 onwards) 
    F2 SEr See section 6.3 Sets the print parameters  
    F3 oFF See section 6.4 Sets the auto power-off parameter  
    F4 SEt See section 6.7 User Parameters for autozero, filter and stability 
    tECH Service agents only Technical parameters setting mode / factory setting 
     
    6.1 ENABLING WEIGHING UNITS (NOT APPROVED MODELS) 
    You can enable and disable the weighing units ava ilable to the user when they press 
    the  [Mode]  key as described in section  4.4 Weighing Units .     
    1)  To set this parameter  press  and hold  the [Mode]  key during self -checking test  
    when turning on the  balance. 
    2)  After a few seconds, the displa y will show the first function  F1   UNT.  
    3)  P ress the [- >0/T
    						
    							© Adam Equipment  Company 2 0 1 6   15 
    4) Pressing the  [- >0/T
    						
    							© Adam Equipment  Company 2 0 1 6   16 
    6.2  SETTING THE BACKLIGHT AND AUDIBLE ALERT BEEP  
    The backlight  and audible alert  may be enabled or disabled by the user. If the 
    backlight is disabled, the battery life will be greater.  The followin g settings are 
    available:  ( beep settings available in SW ver. 2.5 onwards)
     
    bL  AU Sets the backlight to operate automatically when a weight is 
    placed on the balance or a key is pressed. 
    bL  Off Sets the backlight to be off. 
    bL  On Sets the backlight to be permanently on. 
    bP on Sets audible alert beep on 
    bP oFF Sets audible alert beep off 
     
    1)  To set this parameter press the [Mode]  key during self -checking test when 
    turning on the balance . 
    2)  After a few seconds, the display will show the first function F1   U NT.  
    3)  Press the [Mode]  key to select F2  EL  . 
    4)  Press the [- >0/T0/T
    						
    							© Adam Equipment  Company 2 0 1 6   17 
    4) Press the [- >0/T0/T0/T
    						
    							© Adam Equipment  Company 2 0 1 6   18 
    Press the [Mode]  key to change the settin g. Press [->0/T0/T0/T
    						
    							© Adam Equipment  Company 2 0 1 6   19 
    2) After a few seconds, the display will show the first function F1   UNT .  
    3)  Keep pressing the [Mode]  key until F4  oFF  is displayed. 
    4)  Press the [- >0/T0/T0/T0/T
    						
    							© Adam Equipment  Company 2 0 1 6   20 
    6.6  ADJUST THE VALUE OF THE INTERNAL MASS (NOT POSSIBLE ON APPRO VED 
    MODELS) 
    The internal mass value stored in memory can be  adjusted to more clos ely match 
    the value of the user’s  external mass.     
    1)  To set this parameter press the  [Mode] key during self -checking test when 
    turning on the balance . 
    2)  After a few seconds, the display will show the first function F1   UNT .  
    3)  Keep press ing the [Mode]  key until F6  CA is displayed. 
    4)  Press the [- >0/T0/T
    						