29 © Adam Equipment Company 2013 1.11. ACCUMULATED TOTAL The indicator can be set to accumulate when a weigh t is added to the platform automatically or manually by pressing [Print]. See Section 13.2. NOTE: 1. The accumulation function is available only during weighing. It is disabled during parts counting or percent weighing. 2. The accumulated weights will be stored in either kg or lb, depending upon the weighing unit in use. 3. If at any time the weighing units are changed, the accumulated data will be lost. 1.11.1 MANUAL ACCUMULATION When the indicator is set to manual accumulation, t he weight displayed will be stored in the memory when the [Print] key is pressed and the weight is stable. · Remove the weight and press [Print] when the indicator is at zero. The display will show "ACC 1" and then the weight in memory for 2 seconds before returning to normal. The weight can be outp ut to a printer or PC using the RS-232 interface.
30 © Adam Equipment Company 2013 · When the indicator is at zero place a second weight on the platform. When stable press [Print] to accumulate the weight. The display will show "ACC 2" for 2 seconds and then show the new total. · Continue until all weights have been added. This ca n continue for up to 99 entries until the capacity of display is exceede d. · To view the total in memory press the [Print] key when the indicator is at zero. The display will show the total number of accumulation "ACC xx" and the total weight before returning to zero. · To print the total, press [Print] to recall and then immediately press [Print] the second time to print the results. · To erase the memory, press [Print] to view the total and then immediately press [CE] to clear the memory.
31 © Adam Equipment Company 2013 1.11.2 AUTOMATIC ACCUMULATION When the indicator has been set to Automatic Accumu lation the value is stored in memory automatically. · Place a weight on the platform. The beeper will sou nd when the display is stable indicating the value is accepted. Remove the weight. · The display will show "ACC 1" and then the total in the memory before it returns to zero. Adding a 2nd weight will repeat the process. · While the weight is on the platform, press the [Print] key to view the values- first the accumulation number "ACC x" and then the total will be shown. NOTE: 1. The indicator will not show the value when a wei ght is removed. 2. In all cases the display must return to zero or a negative number, before another sample can be added to the memory. 3. More products can be added and [Print] be pressed again for up to 99 entries until the capacity of display is exceede d.
32 © Adam Equipment Company 2013 RS-232 SPECIFICATION The GK indicator is supplied with bi-directional RS -232 interface as standard. The indicator when connected to a printer or computer o utputs the weight with the selected weighing unit through the RS-232 interface . Default Specifications: Connector: The indicator can be set to print text in English, French, German or Spanish. See the RS-232 parameters section for details. RS-232 output of weighing data ASCII code 9600 Baud (user selectable) 8 data bits No Parity 9 pin d-sub miniature socket Pin 3 Output Pin 2 Input Pin 5 Signal Ground
33 © Adam Equipment Company 2013 LABEL = On Label Formats are printed Data Format-Normal Output: Only weight value along with the weighing unit is p rinted. If Percent weighing is used then % is shown in place of weighing units. Data Format-Parts Counting Output: Weight, Unit weight and number of parts are printed . Date 12/09/2006 Time 14:56:27 Scale ID 123456 If ID is zero, it is left blank User ID 234567 Net Wt 1.234 kg Net Wt. (or Gross Wt.)
34 © Adam Equipment Company 2013 Date 12/09/2006 Time 14:56:27 Scale ID 123456 User ID 234567 Net Wt. 1.234 kg Net Wt. (or Gross Wt.) Unit Wt. 123 g g for metric and lb for pounds PCS 10 pcs
35 © Adam Equipment Company 2013 Data Format- Memory Recall Output: Data Format- Continuous Output- Normal weighing: Data Format- Continuous Output- Parts Counting: Date 12/09/2006 Time 14:56:27 Scale ID 123456 User ID 234567 ------------------ TOTAL No. 5 Wt. 1.234 kg PCS 10 pcs ------------------ Net 1.234 kg Net Weight (or Gross wt.) Net 1.234 kg Net Weight (or Gross wt.) U.W. 123 g kg and g for metric and Lb for pounds PCS 10 pcs
36 © Adam Equipment Company 2013 LABEL = Off single line output Fixed line width, leading zeros are spaces. Data Format-Normal Output: Data Format- Memory Recall Print: NOTE: 1. The accumulated total will not be sent to the RS-23 2 when the continuous print is turned on. 2. The continuous print will only be for the current w eight and the display data. 3. In other languages the format is the same but the t ext will be in the language selected. 4. When the scale is in the Lb:oz weighing unit the RS -232 output will only show pounds. 10lb:8oz will be printed as 10.5 lb. 1.234 Kg TOTAL No. 5, 1.234 Kg, 10 pcs
37 © Adam Equipment Company 2013 Description ENGLISH FRENCH GERMAN SPANISH Net weight Net Wt. Pds Net Net-Gew Pso Net Weight per unit counted Unit Wt. Pds unit Gew/Einh Pso/Unid Number of items counted Pcs Pcs Stck. Piezas Number of weighing added to subtotals No. Nb. Anzhl Num. Total weight and count printed Total Total Gesamt Total Print date Date Date Datum Fecha Print time Time Heure Zeit Hora Scale ID number Scale ID Bal ID Waagen ID Bal ID User ID Number User ID Util ID Nutzer ID Usuario ID
38 © Adam Equipment Company 2013 1.12. INPUT COMMANDS FORMAT The indicator can be controlled with the following commands. Press the [Enter] key of the PC after each command. T Tares the indicator to display the net weight. This is the same as pressing [Tare]. Z Sets the zero point for all subsequent weighing. Th e display shows zero. P Prints the results to a PC or printer using the RS- 232 interface. It also adds the value to the accumulation memory if t he accumulation function is not set to automatic.