    19                              © Adam Equipment Company 2013 
     
         
     
     
     
                            
     
        
                            
     
     
    ·   To  count  a  different  sample  quantity,  press  the   [Count]  key.  The  display 
    will  show  the  last  used  sample  size.    Either  use  th is  sample  size  with  a 
    different part or enter a new sample size as above.   
     
    ·  To return to weighing, press  [Unit] when  “XX pcs”  is displayed.     
    						
    							 
    20                              © Adam Equipment Company 2013 
    1.7.  CHECK-WEIGHING 
     
    Check-weighing is a procedure to display an indicat or or sound an alarm when 
    the  weight  on  the  platform  meets  or  exceeds  the  val ues  stored  in  the 
    memory.  The memory holds values for a high limit a nd a low limit.  Either or 
    both the limits can be set by the user.    
    NOTE:   
    1. The alarm and the LED bargraph can each be set t o OFF (See section 13.1). 
    The LCD display will indicate whenever the weight i s within or exceeds the 
    limits by showing  ‘OK’, ‘HI’ or ‘LO’ . 
       
           
    2. The limits can be locked by the manager. A Limit  Password must be used to 
    change the limits or recall other limits from memor y.   
    3.  If  Limit  Password  is  enabled  then  enter  the  pass word  which  will  allow  you 
    to change the limits or the operation of the beeper  or bargraph. 
     Mass on the platform is above the high limit
     
     Mass is between the limits
     
     Mass is below the low limit
         
    						
    							 
    21                              © Adam Equipment Company 2013 
     
    1.7.1   SETTING UP WHILE WEIGHING 
     
    ·  Press  the   [Low  Limit]  key.  It  will  show  the  current  low  limit.  The  “LO” 
    symbol will appear on the display. 
     
    ·  Press  the  [CE] key  to  clear  the  old  value  and  then  enter  the  new  l ow 
    limit  using  the  numeric  keys.  The  decimal  point  is  fixed  at  the  position 
    that  is  used  for  the  current  weighing  unit.  When  th e  desired  value  is 
    shown press  [Tare] to accept the value. If you want to reset the valu e to 
    zero, press  [CE] to clear the value. 
     
    ·  The limits are displayed in the weighing unit in us e.   
     
    ·  To  set  the  high  limit  press   [High  Limit],  the  display  will  show  the  high 
    limit,  the  “HI” symbol  will  be  on  to  the  left  side  of  the  display.  Set  the 
    high limit in the same way the low limit was set.  
     
    ·  Pressing  the  [Tare]  key  to  enter  the  value  will  return  the  indicator  t o 
    weighing, with the Check-weighing function enabled.   
    						
    							 
    22                              © Adam Equipment Company 2013 
     
    1.7.2  SETTING UP WHILE PARTS COUNTING OR % WEIGHING 
        
    During  parts  counting  and  percent  weighing  the  limi ts  are  set  in  the  same  way  as 
    above. The limits are displayed in pcs or %.    
    See  Section  10.4  for  the  description  of  parts  count ing  and  Section  10.7  for  percent 
    weighing.    
    NOTE:  
     
    1.  T he  weight  must  be  greater  than  20  scale  divisions  f or  the  check-
    weighing to operate.   
      
    2.   To disable the check weighing function, enter zero  into both the limits as 
    described above.  When the current limits are shown , press  [CE] to clear 
    the settings, then press  [Tare] to store the zero values. 
      
    1.8.   LIMITS STORING AND RECALLING  
     
    The indicator can store up to 10 sets of high and l ow limits in memory along with the 
    weighing  units  in  use  (including  pcs  for  parts  coun ting  and  %  for  percent  weighing) 
    as well as settings for the beeper and bar graph.    
    During  Check  weighing  the  current  limits  can  be  sto red  or  previously  stored  units 
    can be recalled.  
    						
    							 
    23                              © Adam Equipment Company 2013 
     
     
     
     
      Press  the
      [
    
    
    Lim]   key.      If  you  are  already  in 
    the check weighing mode the display will ask  
    if  you  wish  to  store  the  current  limits  by  
    showing  “StOrE” or recall another set of limits 
    by  showing  “rECALL”.  The [
    
    
    Lim]   key  can  be 
    used to toggle between  “StOrE” and “rECALL”.  
     
     
     
     
       
      I
    f  you  want  to  store  the  limits,  when  “StOrE” 
    is  displayed  press  the  [Tare]  key.  The  display 
    shows  “St  ”.    Enter  a  number  corresponding 
    to  the  desired  memory  location  (0  to  9).  “St 
    X”   will  be  displayed  for  2  seconds  indicating 
    the  location  X  where  the  current  limits,  
    weighing  units  and  settings  for  the  beeper  
    and  bar  graph  are  stored.  The  indicator  will  
    continue to work with the current settings as  
    active.  
     
       
     
       
      If  you  wish  to  recall  any  of  the  pre-stored  
    limits,  press
      [Tare]  when  “rECALL”  is 
    displayed. The display shows  “rEC ”. Enter the 
    number  corresponding  to  the  desired  
    memory location (0 to 9)  to be recalled.  “rEC  
    X”   will  be  displayed  for  2  seconds  indicating 
    the  values  stored  in  the  location  “X”  is  being  
    recalled.    The  indicator  will  change  to  the  
    recalled  limits,  weighing  units  and  settings  
    for the beeper and bar graph.   
     
              
    						
    							 
    24                              © Adam Equipment Company 2013 
    NOTE:    
    1. If the recalled limit is for parts counting, the di splay will show the last sample 
    value used, ready for a new sample to be counted.    
     
    2. If  the  recalled  unit  is  a  percent  weighing  limit,  t he  display  will  show  the  last 
    sample value used, ready for a new sample to be wei ghed.   
     
    3. If the memory location was empty the indicator will  return to weighing.  
     
     
    1.9.   PERCENT WEIGHING  
     
    The indicator can be set to perform percent weighin g.  See Section 13.1. 
      
    The  indicator  will  use  a  mass  on  the  platform  as  th e  100%  reference  weight.  If  the 
    platform  is  empty  (or  the  indicator  is  tared)  then  the  user  can  input  a  reference 
    weight using the keypad.    
    ·   If using a reference weight (or object) as your 100 % reference, add the weight 
    to the to the platform. 
      ·   Press  [Func] . The first option is  “FUnC 1”, press the  [Func] key 3 more times to 
    display  “FUnC 4” .  
     
    ·   Press the  [Tare] key. “F4 PCt”  will be displayed.  
     
     
        
    						
    							 
    25                              © Adam Equipment Company 2013 
    ·  Press  [Tare]  again to enter percent weighing. The indicator wil l set the sample 
    mass on the platform as 100% reference weight.   
     
    NOTE:  If  there  is  no  reference  weight  on  the  pan  and  perc ent  weighing 
    function  is  entered,  pressing  [Tare]  again  will  return  the  indicator  to  normal 
    weighing. 
         
     
     
     
    ·  Remove  the  sample  weight.    Then  any  other  weight  pl aced  on  the  platform 
    will  be  displayed  as  a  percentage  of  the  original  s ample.    For  example,  if 
    3500g is placed on the platform and percent weighin g is selected, the display 
    will show  100.00%. Remove  the  3500g  weight  and  plac e  a  3000g  weight.  The 
    display will show 85.7% as 3000g is 85.7% of 3500g.  
     
     
    ·   The  number  of  decimal  points  will  depend  on  the  wei ght  used.  A  smaller 
    weight will show only “100%” while a larger weight  might show “100.00%”.    
     
    ·  If  the  indicator  was  showing  zero  weight  before  ent ering  this  function,  then 
    the user must manually enter the weight to be set a s 100%. When “F4 PCT” is 
    displayed,  enter  the  weight  to  be  used  for  the  100%   reference,  then  press 
    [Tare]  to accept the reference weight. The display will s how “0.00   %”.     
    						
    							 
    26                              © Adam Equipment Company 2013 
    ·  If the indicator shows  “x x . x x %”, which is the last weight used as a reference, 
    press   [CE]  to  clear  and  use  the  numeric  keypad  to  enter  a  new   value.    Press 
    [Tare]  to accept the new reference weight.    
     
    ·  The weight entered must be greater than 50 scale di visions.   
       
    ·  Press  [Unit]  to return to normal weighing. 
      
    NOTE:   
    The  display  may  jump  by  large  numbers  unexpectedly  if  small  weights  are  used  to 
    set as 100% reference.   The indicator checks if th e weight is too small and will show 
    Error 7.     
    						
    							 
    27                              © Adam Equipment Company 2013 
    1.10.  ANIMAL (DYNAMIC) WEIGHING  
    The  indicator  can  be  set  to  animal  (dynamic)  weighi ng  for  weighing  any  items  that 
    are unstable or may move.    See Section 13.4.  
    The indicator  uses a  special filter to minimise the   effects of any   movement   on  the 
    platform.   
      
    ·  Press  [Func] . The first option is  “FUnC 1”, press the  [Func] key 3 more times to 
    display  “Func 4” .  
     
     
    ·   Press  the   [Tare] key. “F4  PCt”  will  be  displayed.  Press  the  [Func] key  to 
    advance to the second function,   “F4 AnL”,   Animal weighing. 
       
     
    ·   Press  [Tare]  to enter the animal weighing function.  
     
    ·  To  use  the  Animal  Weighing  function  it  is  necessary   to  set  the  amount  of 
    filtering  required  for  the  animals  to  be  weighed.      More  active  animals  will 
    require a higher level of filtering to give a stabl e result.    The display will show 
    “Filt x”   where  x  is  a  value  from  1  to  5.    The  higher  the  va lue  the  greater  the 
    amount  of  filter will be.    To  increment  the  value  s hown  press the   [Func] key 
    then press the  [Tare] key to accept it. 
     
    ·  The  display  will  flash  “Ani   “  2  times  then  show  the  current  weight,  0.00.    The 
    indicator is now ready to weigh.  
        
    						
    							 
    28                              © Adam Equipment Company 2013 
    1.10.1  A NIMAL WEIGHING PROCEDURE 
     
    ·  With the platform empty the display will show zero  weight..  Place containers 
    or  blankets  onto  the  platform  and  press  the   [Tare] key  to  zero  the  display.   
    The indicator may go into the animal weighing proce dure when the items are 
    placed on the platform but will return to showing z ero when the [Tare] key is  
     
    ·  Place the animal to be weighed on the platform.    
     
    ·  When  a  stable  reading  is  found,  the  display  will  sh ow  this  value,  and  the 
    display will be locked until the [Unit] key is pres sed.  The display will show the 
    “Hold” symbol while the display is locked. Remove t he animal, the display will 
    hold the weight value.  
     
    ·  Press  the  [Unit] key  to  unlock  the  display.    The  display  will  flash  “Ani ”  twice, 
    and be ready for the next animal.  
      ·   To weigh a second animal press the  [Tare] key if necessary to zero the display, 
    and  place  the  next  animal  on  the  platform.    It  is  a lso  possible  to  simply  place 
    the next animal on the scale without clearing the l ast value first.  The indicator 
    will detect the new weight and hold it as before. 
     
    ·  The  indicator  will  remain  in  the  animal  weighing  mo de  until  the [Zero] key  is 
    pressed. Then it will return to normal weighing.   
      
    						