Have a look at the manual Adam Equipment CBD Bench Counting Scales CBDREVBAE2 User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 35 Adam Equipment manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
![](/img/blank.gif)
© Adam Equipment Company 2011 Page 29 Sample of output: Using M+ function Using PRINT function DATE 01/27/2011 TIME 12:38:49 LOCAL SCALE SID: ABS123 UID: 123ABC NAME:Text 4 S. No. 1.234 kg TARE 12.456 kg NET 1.1234 g U.W. 11 PCS TOTAL - - - - - - - - - - - - 49.824 kg TW 44 TPC 4 No. DATE 01/27/2011 TIME 12:39:58 LOCAL SCALE SID: ABS123 UID: 123ABC NAME:Text 1.234 kg TARE 12.456 kg NET 1.1234 g U.W. 11 PCS 9.1 INPUT COMMANDS FORMAT The scale can be controlled with the following comm ands. Input Commands: · The scale has a number of commands to either cause an action or to enter data into memory. The commands are all uppe r case and are summarised below. · All commands are terminated by a carriage return ( Enter button on PC keyboard) with the line feed optional. · If an illegal command is received or a command can not be carried out then send the command back with the addition of ER in front of it. For example if the command is NN send back ER N N.
![](/img/blank.gif)
© Adam Equipment Company 2011 Page 30 Basic Commands: PLUxx Select PLU from scale memory to be used T Tare current weight value T123.456 preset tare value is 123.456 Z Zero the display P Print using selected format M+ Store current results into memory and print MR Recall memory values to scale display MC Clear memory U123.456 Store unit weight of 123.456 (grams if in kilograms or pounds if currently in po unds) S123 Enter sample size of 123 parts. Same as pressing [Smpl] key. SL Select local scale to be used SR Select remote scale to be used Immediate Printing Commands: Command Output from scale \L Scale: Local or Remote \I ID number same as PUID below \S Scale number same as PSID below \N Net weight \G Gross Weight \T Tare weight \U Unit weight \P Count \C Total Count \W Total Weight \M Number of items stored in memory \B A blank line printed 9.2 STORING DATA VIA RS232 To store data the commands are: SUIDxxxxxx Store user ID data SSIDxxxxxx Store scale ID data SPLUxx,xxxxxxxxxxxx Store text data for PLUxx When PLU text data is stored the Scale stores curre nt unit weight and current tare value along with the PLU number. For the SPLU command the data is: PLU number (2 cha racters), (Comma) description (max 12 characters). If the fields are less than the maximum all charact ers need not be used.
![](/img/blank.gif)
© Adam Equipment Company 2011 Page 31 9.3 PLU ENTRY USING RS-232 INTERFACE This method enables the user to send the data from a PC program as well as from the keypad. The most common PLUs can be stored and recalled from the scale memory. Other PLU data can be stored on a PC, then the Description (text data), Unit Weight and Tare Weigh t data can be sent from the PC for the PLUxx (where xx can be any desired n umber up to 99). This can then be used or over-written during operation. OPERATION: · Send Tare Weight data to set any tare value to be stored with PLU. i.e. “ T 1.45 ” . If no tare is needed then you may send “ T 0” to delete any present tare data. · Send the Unit Weight to be stored with PLU. i.e. “ U0.015” · Send PLU text data to be stored with current Tare and U/W values. i.e. “ SPLU27, M4 Nut Brass ” Any information stored for the Tare Weight, Descrip tion and the Unit weight will be stored in the scale and can be recalled for use in the future.
![](/img/blank.gif)
© Adam Equipment Company 2011 Page 32 10.0 CALIBRATION ACTION DISPLAYS Press the [Tare/Zero] key during the self- test at power on. The scale will ask to enter the password. “ Pi n” Default Password is 0000. Enter “0” four times. Password can be changed in technical parameters. Press [Print] . “ Pi n” “ - - - -” Select the scale to be set up by using the [Local/Remote] key. Press the [Print] key to enter the technical section. “tECH” “ LocAL” “ “ “tECH” “rEmo tE” “ “ Use the [U.Wt.] to select the weighing unit to be used for setting the calibration for the scale. The arrow in the “ Weight” window will indicate the unit selected. Press the [Print] key to continue. “tECH” “ Uni t” “ “ You will enter the first parameter - Calibration. Display will ask to unload any weight on the platform. Press the [Print] key to continue. “nO L o Ad” Load the calibration weight onto the scale If the mass is different than the value shown press the [CE] key then enter the calibration weight to be used . Then press [Print]. “LoAd ” “ 0010” “ KiLoS”
![](/img/blank.gif)
© Adam Equipment Company 2011 Page 33 The Calibration is complete, the scale will run the self-test during which the weight should be removed. If calibration is not possible an error message will be shown “- - - -” “ - - - -” “ - - - - ” “SPAn” “ PASS” “ ” “SPAn” “ FAiL” “ LoW ” or “HiGH If The Calibration is being done while in the TECH parameters (section 12) the linearity is set at the same time as calibration by performing t he calibration in 2 steps, Display will ask to unload any weight on the platform. Press the [Print] key to continue. “nO L o Ad” Load the first calibration weight onto the scale If the mass is different than the value shown press the [CE] key then enter the calibration weight to be used . Then press [Print] to continue or press [Zero] to complete with only one mass being used. “LoAd 1 ” “ 0010” “ KiLoS” Load the second calibration weight onto the scale If the mass is different than the value shown press the [CE] key then enter the calibration weight to be used . Then press [Print]. “LoAd 2” “ 0030” “ KiLoS” The Calibration is complete, the scale will run the self-test during which the weight should be removed. “- - - -” “ - - - -” “ - - - - ” “SPAn” “ PASS” “ ” If calibration is not possible an error message will be shown “SPAn” “ FAiL” “ LoW ” or “HiGH
![](/img/blank.gif)
© Adam Equipment Company 2011 Page 34 11.0 ERROR CODES During the initial power-on testing or during the o peration, the scale may show an error message. The meaning of the error me ssages are described below. If an error message is shown, repeat the procedure that caused the message, such as turning the balance on, calibratio n or any other functions. If the error message is still shown then contact yo ur dealer for further support. ERROR CODE DESCRIPTION POSSIBLE CAUSES Err 4 Initial Zero is greater than the permissible value (typically 4% of maximum capacity) when power is turned on or when the [Zero] key is pressed, Weight on the pan when turning the scale on. Excessive weight on the pan when zeroing the scale. Improper calibration of the scale. Damaged load cell. Damaged Electronics. Err 5 Keyboard error. Improper operation of the scale. Err 6 A/D count is not correct when turning the scale on. Platform is not installed. Load cell may be damaged. Electronics may be damaged. FAIL HiGH or FAIL LoW Calibration error Improper calibration. If the problem persists contact your dealer or Adam Equipment for assistance.
![](/img/blank.gif)
© Adam Equipment Company 2011 Page 35 12.0 TECHNICAL PARAMETERS The technical parameters accessed via the “ tECH” prompt at the end of user parameters and are password controlled to prevent u nauthorised access. These parameters set the metrology for the scales. Each scale is set independently. The parameters will set capacity, di vision, decimal point position, initial zero range, auto and manual zero range as well as factory calibration. ACTION DISPLAYS When “ tEch” is displayed, press the [Print] key. The scale will ask you to enter the password. “ Pi n” Default Password is 0000. Enter “0” four times. 2006 will override any other user password. Press [Print]. “ Pi n” “ ----” Select the scale to be set up by using the [Local/Rem] key. Press the [Print] to enter the technical section for that scale. “t ECH” “ Lo c A L” “ “ “t ECH” “r E mo t E” “ “ Use the [U.Wt./Units] to select the weighing unit to be used for setting up the parameters for the scale. The arrow in the “ Weight” window will indicate the unit selected. Press the [Print] key to continue. “t ECH” “ Uni t” “ “ Press the [U.Wt./Units] key to scroll through the menus for the chosen scale. User [Print] to select and set parameters and the [Tare/Zero ] to escape as without changing data as with the user parameter section. “Cn t” Both scales have the same menus except that the capacity of the local scale is fixed and cannot be entered.
![](/img/blank.gif)
© Adam Equipment Company 2011 Page 36 Press the [U.Wt/Unit] to go to next menu. Press [Print] to enter. Use [U.Wt/Unit] to select the decimal point position and press [Print]. To change the remote scale cap press CE to clear and then enter new value followed by [Print]. “ CAP ”. Scale capacity setting. “dESC” “ 0.00” “LOAd” “0060” Press [U.Wt/Unit] to go to next menu. Press [Print] to enter. Use [U.Wt/Unit] to select the division which you would like the display to increment in, followed by [Print]. “inC 5” Press the [U.Wt/Unit] to go to next menu, use [Print] key to enter. Use [U.Wt/Unit] to increment the values then [Print]. “AZt” Autozero tracking range. Select from 0.5d 0.5d 0.5d 0.5d, 1d, 2d, 4,d Press the [U.Wt/Unit] to go to next menu, use [Print] key to enter. Use [U.Wt/Unit] to increment the values then [Print]. “0 Auto” Zero auto range at power on. Select from 0%, to 10% Press the [U.Wt/Unit] to go to next menu, use [Print] key to enter. Use [U.Wt/Unit] to increment the values then [Print]. “0 manl ” Zero manual range. Select from 0%, to 10% Press the [U.Wt/Unit] to go to next menu, use [Print] key to enter. Enter the new PIN number and press [Print]. Confirm the new PIN number and press [Print]. “ Pin ” Password number for “ tEch”. “Pin1” “Pin2” “donE” Press [U.Wt./Units] to continue use [Print] key to enter. Escape the parameters with the [Zero/Tare] “ CAL ” Enter calibration section where the local or remote scale is selected. See section 10.
![](/img/blank.gif)
© Adam Equipment Company 2011 Page 38 13.0 REPLACEMENT PARTS AND ACCESSORIES If you need to order any spare parts and accessories, contact your supplier or Adam Equipment. A partial list of such items is as follows- ·Power Supply Module · Main Power cord · Replacement Battery · Stainless Steel Pan · In use cover · Printer, etc. 14.0 SERVICE INFORMATION This manual covers the details of operation. If you have a problem with the scale that is not directly addressed by this manual then contact your supplier for assistance. In order to provide further assistance, the supplier will need the following information which should be kept ready: A. Details of your company -Name of your company: -Contact person’s name: -Contact telephone, e-mail, fax or any other methods: B. Details of the unit purchased (This part of information should always be available for any future correspondence. We suggest you to fill in this form as soon as the unit is received and keep a print-out in your record for ready reference.) Model name of the scale: CBD____ Serial number of the unit: Software revision number (Displayed when power is first turned on): Date of Purchase: Name of the supplier and place: C . Brief description of the problem Include any recent history of the unit. For example: -Has it been working since it’s delivered -Has it been in contact with water -Damaged from a fire -Electrical Storms in the area -Dropped on the floor, etc.